VX130 Ophthalmic Diagnostic Device
K162067 · Luneau Sas · HKX · Apr 24, 2017 · Ophthalmic
Device Facts
| Record ID | K162067 |
| Device Name | VX130 Ophthalmic Diagnostic Device |
| Applicant | Luneau Sas |
| Product Code | HKX · Ophthalmic |
| Decision Date | Apr 24, 2017 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 886.1930 |
| Device Class | Class 2 |
Intended Use
The VX130 is a multi-function diagnostic device combining wavefront aberometer, corneal topographer, retro-illuminator, tonometer and pachymeter, indicated for: Measuring the refraction of the eye giving both lower and higher order aberrations Measurement of the shape of the cornea Retro-illumination imaging of the eye Measuring the intraocular pressure without contacting the eye for glaucoma evaluation. Photographing the eye and taking images of the eye to evaluate the thickness of the central cornea. Full corneal thickness map. Scheimpflug imaging. Anterior chamber imaging. Pupil image. Image of the cornea relative to the iris.
Device Story
VX130 is a multi-function ophthalmic diagnostic platform combining wavefront aberometer, corneal topographer, cataract screening, Scheimpflug pachymeter, and non-contact tonometer. Inputs include ocular optical signals and images; wavefront aberometer uses Shack-Hartmann principle for refraction; corneal topographer uses Placido disk; Scheimpflug pachymeter uses linear scanning slit-light photography; tonometer uses air-puff. Device is fully automated, performing alignment and focusing via single command. Used in clinical settings by eye care professionals. Output includes refraction maps, corneal shape/thickness maps, and anterior segment images displayed on 10.1" LCD. Data assists clinicians in glaucoma evaluation, cataract screening, and refractive assessment. Benefits include non-invasive, comprehensive anterior segment analysis in a single device.
Clinical Evidence
Bench testing and comparative study provided. Electrical safety and EMC testing per IEC60601-1 and IEC60601-1-2. Software validated per IEC62304. Risk management per ISO14971. Ophthalmic standards met (ISO15004-1/2). Optical hazard testing (IEC60825-1) classified device as Laser Class 1. Comparative study of anterior segment parameters (K1/K2 anterior/posterior, CCT) between VX130 and Galilei G4 showed small differences (e.g., CCT mean difference -0.43 μm, SD 7.45 μm). Repeatability testing (n=42 eyes) demonstrated consistent performance.
Technological Characteristics
Multifunctional diagnostic platform. Materials: non-invasive chin/forehead rest. Sensing: Shack-Hartmann wavefront sensor, Placido disk, linear scanning Scheimpflug camera, air-puff tonometer. Illumination: 880nm IR LED (pupil/corneal), 455nm blue LED (flash). Connectivity: 10.1" LCD display, digital data output. Power: 100-240 VAC. Software: automated alignment/focusing algorithms. Standards: IEC60601-1, IEC60601-1-2, IEC62304, ISO14971, ISO15004-1/2, IEC60825-1 (Laser Class 1).
Indications for Use
Indicated for patients requiring ophthalmic diagnostic evaluation, including refraction (lower/higher order aberrations), corneal topography, retro-illumination imaging, non-contact intraocular pressure measurement for glaucoma, and anterior segment imaging (corneal thickness, Scheimpflug, anterior chamber, pupil, iris).
Regulatory Classification
Identification
A tonometer and accessories is a manual device intended to measure intraocular pressure by applying a known force on the globe of the eye and measuring the amount of indentation produced (Schiotz type) or to measure intraocular tension by applanation (applying a small flat disk to the cornea). Accessories for the device may include a tonometer calibrator or a tonograph recording system. The device is intended for use in the diagnosis of glaucoma.
Predicate Devices
- VX120 Ophthalmic Diagnostic Device (K143086)
- CM3910 ROTATING DOUBLE SCHEIMPFLUG CAMERA (Galilei G4) (K051940)
Related Devices
- K143086 — VX120 Ophthalmic Diagnostic Device · Luneau Sas · Jun 1, 2015
- K202221 — VX650 · Luneau Technology Operations · Jan 29, 2021
- K201724 — Pentacam AXL Wave · Oculus Optikgerate GmbH · Oct 21, 2020
- K051940 — CM 3910 ROTATING DOUBLE SCHEIMPFLUG CAMERA · Sis Ltd., Surgical Instrument Systems · Sep 28, 2005
- K253039 — AUTO KERATO-REFRACTO TONOMETER TRK-3 OMNIA · Topcon Corporation · Nov 7, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing to the right.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 24, 2017
Luneau SAS Isabelle Durand Quality/ RA Manager 1 Ave de Malaguet Prunay-Le-Gillon, FR 28360
Re: K162067
Trade/Device Name: VX130 Ophthalmic Diagnostic Device Regulation Number: 21 CFR 886.1930 Regulation Name: Tonometer and Accessories Regulatory Class: Class II Product Code: HKX Dated: March 15, 2017 Received: March 17, 2017
Dear Isabelle Durand:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K162067
Device Name VX130
Indications for Use (Describe)
The VX130 is a multi-function diagnostic device combining wavefront aberometer, corneal topographer, retro-
illuminator, tonometer and pachymeter, indicated for:
Measuring the refraction of the eye giving both lower and higher order aberrations
Measurement of the shape of the cornea
Retro-illumination imaging of the eye
Measuring the intraocular pressure without contacting the eye for glaucoma evaluation.
Photographing the eye and taking images of the eye to evaluate the thickness of the central cornea.
Full corneal thickness map.
Scheimpflug imaging.
Anterior chamber imaging.
Pupil image.
Image of the cornea relative to the iris.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Visionix Luneau. The logo consists of two parts: a stylized eye symbol and the company name. The eye symbol is made up of three curved lines, and the company name is written in a sans-serif font.
# 510(k) Summary
# VX130 ophthalmic diagnostic device
April 24th, 2017
#### l. SUBMITTER
LUNEAU SAS 1 avenue de Malaguet 28360 Prunay le Gillon France Phone : +33 (0)2 37 25 25 25 Fax: +33(0)2 37 25 75 99
Contact person: Isabelle Durand
#### DEVICE II.
Name of Device: VX130 ophthalmic diagnostic device Common name: Ophthalmic diagnostic device Classification: tonometer , AC powered Regulatory class: II Product code: HKX Requlation number: 886.1930
#### III. PREDICATE DEVICES
The VX130 is claimed to be substantially equivalent to the following currently marketed devices:
#### . VX120 Ophthalmic Diagnostic Device
Manufacturer: LUNEAU SAS FDA K143086 issued June 1, 2015. Product code: HKX
### . CM3910 ROTATING DOUBLE SCHEIMPFLUG CAMERA (Galilei G4)
Manufacturer: SIS LTD. SURGICAL INSTRUMENT SYSTEMS Allmendstrasse 11 Port Bern, SWITZERLAND CH - 2562 FDA K051940 issued July 15, 2005. Product code: MXK
The VX130 is equivalent to VX120 for tonometry and pachymetry functions and other functions not subject to 510(k) : combined wave front aberrometer, corneal topographer and retroilluminator. It is equivalent to CM3910 for its pachymetry function.
{4}------------------------------------------------
#### DEVICE DESCRIPTION IV.
The VX130 is a multifunctional ophthalmic diagnostic device.
The VX130 combined wavefront aberrometer, corneal topographer, cataract screening device, Scheimpflug pachymeter, and non-contact tonometer is a single platform that contains five different measurement units.
The wavefront aberrometer works on the Shack-Hartmann principle and is used as an advanced autorefractometer that measures both lower and higher order aberrations of the eye. Retro illumination is used to image ocular opacities. The corneal topographer uses a Placido disk to measure keratometry and the detailed shape of the cornea. A linear scanning Scheimpflug pachymeter measures the thickness of the cornea by illuminating it with a slit of light and photographing it using the Scheimpflug technique. Anterior and posterior corneal shape are also measured from the Scheimpflug images. An air puff non-contact tonometer is included for measurement of the intraocular pressure. The device is fully automated and a number of different measurements can be performed by a single command including alignment and focusing.
#### V. INDICATIONS FOR USE:
The VX130 is a multi-function diagnostic device combining wavefront aberometer, corneal topographer, retro-illuminator, tonometer and pachymeter, indicated for:
Measuring the refraction of the eye giving both lower and higher order aberrations Measurement of the shape of the cornea
- Retro-illumination imaging of the eye
Measuring the intraocular pressure without contacting the eye for glaucoma evaluation Photographing the eye and taking images of the eye to evaluate the thickness of the central cornea.
- Full cornea thickness map
- Scheimpflug imaging
- Anterior chamber imaging
- Pupil image
- Image of the cornea relative to the iris
{5}------------------------------------------------
- VI. COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
### Comparison with VX120
The VX130 has same hardware as VX120. The difference is that VX130 provides multi-slits of the Scheimpflug camera resulting of multiple results of the same indication in different place in the cornea as well as additional maps, while VX120 provides only one Scheimpflug image. That doesn't affect indications for use below:
"Indications for use:
The VX120 is a multi-function diagnostic device combining wavefront aberometer, corneal topographer, retro-illuminator, tonometer and pachymeter, indicated for: Measuring the refraction of the eye giving both lower and higher order aberrations Measuring the shape of the cornea Retro-illumination imaging of the eye Measuring the intraocular pressure without contacting the eye for glaucoma evaluation Photographing the eye and taking images of the eye to evaluate the thickness of the cornea."
This variant doesn't impact on performances and safety. VX130 is not different in safety or efficacy, and is substantially equivalent to VX120
### Comparison with Galilei
Indications for use:
The pachymeter function of the VX130 has the same intended use as the Galilei G4 for photographing the eye and taking images of the eye to evaluate the thickness of the central cornea.
Full cornea thickness map Scheimpflug imaging Anterior chamber imaging Pupil image Image of the cornea relative to the iris
| Characteristics | VX130 | Galilei |
|-----------------------------|------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| References | User manual<br>UM30200013 | 510(k) K051940 |
| Measuring Principle | Linear scanning<br>Scheimpflug principle for<br>slit image photography | Rotational scan of<br>Double-Scheimpflug slit<br>images merged with<br>Placido disk images. |
| Viewing Means | Display on built in<br>LCD screen 10.1" | LCD on Measurement<br>Unit, 17" monitor on<br>table |
| Observation<br>Illumination | Infrared LED 880nm for<br>pupil and corneal<br>illumination | Infrared LED 810nm |
{6}------------------------------------------------
| Flash Output<br>Illumination | Blue LED light (UV free)<br>455nm, max 50μW | Blue LED light (UV<br>free) 470nm max,<br>15mW sec |
|------------------------------|---------------------------------------------|----------------------------------------------------|
| Photography<br>Camera | CMOS camera | CCD camera |
| Display | Data digital | Data digital, displayed<br>on a CPU |
| Image Resolution | 1600 × 1000 pixels | 1004 × 1004 pixels |
| Image Size | 9 × 6mm | 7.4 × 7.4 mm |
| Photographic Range | Linear 6mm | Eligible 0 to 180° |
| Photographic Series | 30 | 1 to 60 images |
| Slit Length | 8mm | 15mm |
| Power Requirement | 100-240 VAC, 50/60Hz | 110/220 VAC 50/60Hz |
| Weight | 25 kg | 10Kg (Measurement<br>Unit) |
### Rationale for Substantial Equivalence
The pachymeter function of the VX130 has the same intended use as the Galilei G4 for photographing the eye and taking images of the anterior segment of the eye to evaluate the thickness of the cornea.
The Galilei G4 and the VX130 systems are based on the Scheimpflug Principle for slit image photography. The measurement systems use blue light (UV-free) through a slit to illuminate the eye, and a Camera for photography. The devices take a series of images of the anterior segment of the eye and analyse the images.
- The devices have the same intended use
- The devices utilize the same measuring principles (Slit Scan)
- The devices utilize the same photographic medium
- . Similar measurement wavelengths are used
The devices use the same features like a
- Head stabilizing device
- . Fixation target
- All devices are considered "non-invasive" as defined in 21 CFR §812.3(k)
The VX130 for all functions other than pachymetry is identical to the VX120
#### Safety
The VX130 is a non-invasive diagnostic system, which contacts the patient only on his/her chin and forehead. The VX130 does not present or pose any new or additional risks for the prescribed intended uses. The light output is of an eye safe intensity and wavelength. The electrical safety requirements for medical devices are met. The VX130 is proven effective for its intended uses through internal performance tests.
The following performance data were provided in support of the substantial equivalence determination:
- 1. Electrical safety and EMC testing were conducted on the VX130. The device complies with IEC60601-1:2006, and the IEC60601-1-2 for EMC.
- 2. Software verification and validation testing was done according to IEC62304.
{7}------------------------------------------------
- 3. Risk management: VX130 was evaluated according to ISO14971: 2012. All risks have been reduced to safe levels thus there is no conflict between risk and benefit.
- 4. Tests for ophthalmic products: VX130 was evaluated in accordance with ISO15004-1:2009 and ISO15004-2:2007 standards and was found to meet all requirements of the standards.
- 5. For optical hazards, VX130 was evaluated in accordance with IEC60825-1: 2008: the result is VX130 is laser class 1.
### Tests for pachymetry:
- 1. A comparison study of anterior segment parameters done with VX130 and Galilei G4 shows that there are only small differences between the devices. Table 1 shows the differences between the two devices for a range of anterior segment parameters. Table 2 shows a repeatability measurement of the VX130. There is not a significant difference between the repeatability measurement and the comparison measurement. The two devices can therefore be considered equivalent.
| | | | 95% CL | |
|-------------------|---------|---------|--------|-------|
| Parameter | Average | Std dev | min | max |
| K1 Anterior (D) | -0.27 | 0.17 | -0.59 | 0.06 |
| K2 Anterior (D) | -0.29 | 0.16 | -0.62 | 0.03 |
| Cyl Anterior (D) | -0.03 | 0.24 | -0.51 | 0.45 |
| K1 Posterior (D) | -0.09 | 0.06 | -0.21 | 0.03 |
| K2 Posterior (D) | -0.01 | 0.14 | -0.29 | 0.28 |
| Cyl Posterior (D) | 0.08 | 0.12 | -0.15 | 0.30 |
| CCT (\$\mu\$m) | -0.43 | 7.45 | -15.03 | 14.16 |
Table 1: Summary of the comparison of anterior segment parameters between the VX130 and Galilei G4 showing the average difference between the devices, the standard deviation of the difference and the 95% confidence levels.
| Parameter | Std dev 1 | Std dev 2 |
|-------------------|-----------|-----------|
| K1 Anterior (D) | 0.173 | 0.132 |
| K2 Anterior (D) | 0.095 | 0.237 |
| Cyl Anterior (D) | 0.166 | 0.229 |
| K1 Posterior (D) | 0.031 | 0.075 |
| K2 Posterior (D) | 0.093 | 0.107 |
| Cyl Posterior (D) | 0.103 | 0.112 |
| CCT (μm) | 2.89 | 3.68 |
Table 2: Repeatability of VX130,
1) a single eye was measured ten times and the std dev was calculated.
- 2) 42 eyes were measured 3 times each the average std dev is shown.
#### VII. CONCLUSIONS
The VX130 has the same intended use, technological characteristics, and principles of operation as its predicate devices.
The technological differences between the VX130 and its predicates raise no new issues of safety and effectiveness.
Performance data demonstrates that the VX130 is as safe and effective as the predicate devices.