KSEA THERMOFLATOR (MODEL 26 4320 20)
Device Facts
| Record ID | K955073 |
|---|---|
| Device Name | KSEA THERMOFLATOR (MODEL 26 4320 20) |
| Applicant | KARL STORZ Endoscopy-America, Inc. |
| Product Code | HIF · Obstetrics/Gynecology |
| Decision Date | Dec 6, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 884.1730 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
This device is designed to carefully deliver large flow rates for rapid insufflation, and monitor the amounts of CO₂ gas needed to establish and maintain proper distention of the peritoneal cavity during Ob/Gyn laparoscopic surgical and diagnostic procedures
Device Story
KSEA Thermoflator (model 26 4320 20) is a microprocessor-based laparoscopic insufflator; delivers CO₂ gas to peritoneal cavity; user-adjustable flow rates (0-30 L/min) and pressure (0-30 mmHg); used in clinical settings by surgeons; provides rapid insufflation and maintenance of distention; includes pressure and flow-sensing technology; incorporates safety features to monitor gas delivery; assists surgeons in maintaining surgical field visibility.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Microprocessor-based system; CO₂ gas delivery; pressure and flow-sensing technology; user-adjustable flow (0-30 L/min) and pressure (0-30 mmHg) controls.
Indications for Use
Indicated for patients undergoing Ob/Gyn laparoscopic surgical and diagnostic procedures requiring peritoneal cavity distention via CO₂ insufflation.
Regulatory Classification
Identification
A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
Related Devices
- K080852 — KSEA ELECTRONIC CO2 ENDOFLATOR LC, MODEL 26430320-1 · KARL STORZ Endoscopy-America, Inc. · Mar 30, 2009
- K962863 — KARL STORZ MODEL 264305-20 ELECTRONIC ENDOFLATOR · KARL STORZ Endoscopy-America, Inc. · Nov 19, 1996
- K955477 — HI-TEC CO2 INSUFFLATOR TYPE 1300 · Hi Tec Medical Vertriebs-GmbH Fur Innovative Mediz · Aug 6, 1997
- K991906 — LAPARO CO2-PNEU AUTOMATIC INSUFFLATOR WITH 30L FLOW RATE,AND VIDEO-SCREEN DISPLAY, HIGH FLOW INSUFFLATION TUBE WITH HEAT · Richard Wolf Medical Instruments Corp. · Nov 23, 1999
- K983326 — OMNIFLATOR 7640 MODEL 7-640-00 · Northgate Technologies, Inc. · Feb 5, 1999