K955791 · W.O.M. World of Medicine GmbH · HIF · Dec 17, 1996 · Obstetrics/Gynecology
Device Facts
Record ID
K955791
Device Name
SURGIFLATOR-20 PIM
Applicant
W.O.M. World of Medicine GmbH
Product Code
HIF · Obstetrics/Gynecology
Decision Date
Dec 17, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1730
Device Class
Class 2
Indications for Use
The Surgiflator-20 PIM is a laparoscopic high flow insufflator intended to fill the abdominal cavity with CO2 to enable laparoscopic procedures.
Device Story
Laparoscopic high flow insufflator; fills abdominal cavity with CO2 to create pneumoperitoneum for minimally invasive surgery. Input: CO2 gas source. Operation: electronic flow control; features low-pressure insufflation mode for gentle filling; fluid sensor detects backflow contamination; simultaneous pressure monitoring via direct abdominal measurement. Output: regulated CO2 flow and pressure data. Used in OR by surgeons/staff. Provides stable pneumoperitoneum for visualization and instrument access during gynecologic and other laparoscopic surgeries.
Clinical Evidence
No clinical data. Safety and utility of high-flow insufflators for laparoscopic techniques are established in referenced medical literature (Bruhat, 1992; Hunter & Sackier, 1993).
Technological Characteristics
Electronic high-flow insufflator; CO2 gas source; integrated fluid sensor for backflow detection; pressure monitoring system. Design incorporates electronic flow regulation and pressure sensing circuitry.
Indications for Use
Indicated for patients undergoing laparoscopic procedures requiring abdominal CO2 insufflation.
Regulatory Classification
Identification
A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
Snowden-Pencer High Flow Insufflator Model 89-8600
Snowden-Pencer High Flow Insufflator Model SP 88-9700
Related Devices
K120151 — NEBULAE I · Northgate Technologies, Inc. · Aug 17, 2012
K962863 — KARL STORZ MODEL 264305-20 ELECTRONIC ENDOFLATOR · KARL STORZ Endoscopy-America, Inc. · Nov 19, 1996
K030837 — 40 L HIGH FLOW INSUFFLATOR F108 · W.O.M. World of Medicine AG · Feb 5, 2004
K153513 — Insufflator 50L FM134 · W.O.M. World of Medicine GmbH · Mar 4, 2016
K080852 — KSEA ELECTRONIC CO2 ENDOFLATOR LC, MODEL 26430320-1 · KARL STORZ Endoscopy-America, Inc. · Mar 30, 2009
Submission Summary (Full Text)
{0}
K955791
K955791
Page 1/2
# G. Summary of Safety and Effectiveness - 510(k) Summary
W.O.M. GmbH
Pascalstr. 11
D-10587 Berlin
Germany.
DEC 17 1996
The Surgiflator-20 PIM is a laparoscopic high flow insufflator intended to fill the abdominal cavity with $\mathrm{CO}_{2}$ to enable laparoscopic procedures.
The Surgiflator-20 PIM described in this notification is similar in design and construction to the W.O.M. Surgiflator 20, K950035 and the Snowden-Pencer High Flow Insufflator, Model 89-8600 and Model SP 88-9700, distributed in the United States by Snowden-Pencer, Inc.
- The Surgiflator-20 PIM incorporates the same design features as the W.O.M. Surgiflator 20. The differences lie in the following features:
Low pressure insufflation mode - fills the abdomen more gently
Fluid sensor: shows the user when the device is contaminated by backflow
Simultaneous pressure monitoring: monitoring of the pressure directly in the abdomen
- The utility and safety of laparoscopic techniques using modern electronic high flow insufflators is thoroughly reported in the literature with the advantages and risks well articulated.
- A comprehensive discussion of the use of insufflation methods is presented in the book "Operative Laparoscopy" (1): 9-15, by M.-A. Bruhat, 1992, which observes that Laparoscopy, along with the entire concept of minimally invasive surgery through endoscopically guided intra-abdominal surgery, has become a mainstay in gynecologic surgery. The review of instrumentation in this field includes comments on the use of modern high flow insufflators, the establishment of the pneumoperitoneum, use of instrumentation and use of $\mathrm{CO}_{2}$-laser.
- Background information and experiences with the use of laparoscopic techniques including videoendoscopy are presented in the book "Minimal Invasive Surgery" (3): 57ff, 216-218, 291-295 by John G. Hunter, M.D. and Jonathan M. Sackier, M.D., McGraw-Hill, Inc., 3-6 and 216, 291. New technologies are discussed as well as advantages and disadvantages of minimally invasive surgery. The importance of effective, well-maintained instrumentation, i.e. insufflators and other instrumentation like light sources, television screens, and energy sources, is discussed.
DC953540.005
{1}
K955791
Page 2/2
# REFERENCES
1. Maurice-Antoine Bruhat. "Operative Laparoscopy". New York: McGraw-Hill, 1992, 226 pages.
2. John G. Hunter. "Minimally Invasive Surgery". New York: McGraw Hill, 1993, 358 pages.
DC953540.005
TOTAL P.04
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.