40 L HIGH FLOW INSUFFLATOR F108

K030837 · W.O.M. World of Medicine AG · HIF · Feb 5, 2004 · Obstetrics/Gynecology

Device Facts

Record IDK030837
Device Name40 L HIGH FLOW INSUFFLATOR F108
ApplicantW.O.M. World of Medicine AG
Product CodeHIF · Obstetrics/Gynecology
Decision DateFeb 5, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.1730
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The 40 L High Flow Insufflator F108 is a device intended to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.

Device Story

Microprocessor-controlled CO2 insufflator for diagnostic/therapeutic laparoscopic procedures; distends peritoneal cavity with gas. Components: housing, power supply, pressure reducers, venting system, fluid sensor, gas heater, user interface. Continuous redundant pressure measurement ensures actual abdominal pressure matches nominal setpoint. Features Low Flow Mode for small abdominal cavities (pediatric use). Operated by clinicians in surgical settings. Provides accurate flow measurements (0.1 L/min to 1 L/min). Benefits include safe, effective insufflation for varied patient sizes.

Clinical Evidence

No clinical tests performed. Bench testing demonstrated safety and effectiveness of Low Flow Mode for small abdominal cavities. Safety and effectiveness of CO2 insufflators in pediatric laparoscopy supported by literature review.

Technological Characteristics

Microprocessor-controlled CO2 insufflator. Components: pressure reducers, venting system, fluid sensor, gas heater. Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and UL 2601-1. CE marked per 93/42/EEC.

Indications for Use

Indicated for pediatric patients requiring peritoneal cavity distension during laparoscopic procedures via low flow mode.

Regulatory Classification

Identification

A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K030837 ## 510(k) SUMMARY 40L High Flow Insufflator F108 ### I. Applicant: W.O.M. WORLD OF MEDICINE AG Alte Poststraße 11 96337 Ludwigsstadt Germany Phone number: 011 49 9263 877 0 FAX number: 011 49 9263 877 128 Establishment registration number: 8043980 | Contact name: | Marc Engelhardt | |-------------------------|-----------------------------| | Contact title: | Regulatory Affairs Manager | | Phone number: | 011 49 30 399 81 610 | | FAX number: | 011 49 30 399 81 593 | | Contact e-mail address: | marc.engelhardt@womcorp.com | #### II. Submission Correspondent: Susanne Raab 368 St. North Asaph Street Alexandria, VA 22314 | Contact title: | Regulatory Counsultant | |-------------------------|--------------------------------------| | Phone number: | 703-299-0523 | | FAX number: | 703-299-0523 | | Contact e-mail address: | sbraab@comcast.net<br>soraab@aol.com | #### III. Date Summary was Prepared: March 13, 2003 #### IV. Product Information: | Classification name: | Laparoscopic Insufflator | |----------------------|--------------------------| |----------------------|--------------------------| Common name: Laparoscopic Insufflator Trade name: 40 L High Flow Insufflator F108 #### V. Product Classification: | Product code: | 85 HIF | |-----------------|----------------------| | C.F.R. Section: | 21 C.F.R. § 884.1730 | | Device Class : | II | {1}------------------------------------------------ #### VI. Predicate Devices: - Surgiflator 40 (K001533) manufactured by W.O.M. WORLD OF . MEDICINE AG #### VII. Intended Use: The 40 L High Flow Insufflator F108 is a device intended to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it. #### VIII. Device Description: The 40 L High Flow Insufflator F108 is a microprocessor controlled device designed to insufflate medical CO2 gas into peritoneal cavities during diagnostic and therapeutic laparoscopic procedures. The 40 L High Flow Insufflator F108 consists of the following main components: a housing, power supply, pressure reducers, a venting system, a fluid sensor, a gas heater, various setting kevs and display elements. A continuous redundant pressure measurement controls the conformity of the actual pressure in the abdomen with the pre-set nominal pressure. The device is to be used with a specially designed heating tube and with an optional remote control. #### IX. Substantial Equivalence: The device described in this notification is similar in intended use. design and technological characteristics to the Sugiflator 40 (K001533) manufactured by W.O.M. WORLD OF MEDICINE AG. Both the 40 L High Flow Insufflator F108 and the Surgiflator 40 (K001533) are intended to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it. In addition, both devices are similar in design and technical characteristics. In particular, the pressure reduction concept, the venting system and the maximum flow and pressure ranges are identical. Unlike the predicate device, the 40 L High Flow Insufflator F108 is designed with a new software logarithm which allows for accurate flow measurements in the range of 0.1 Vmin and 1 Vmin and for the integration of a Low Flow Modc to ensure the safe and effective performance of the device when insufflating small abdominal cavities. The differences between the 40 L High Flow Insufflator F108 and the predicate device are minor and raise no new questions of safety and effectiveness. Accordingly, W.O.M. WORLD OF MEDICINE AG believes that the 40 L High Flow Insufflator F108 is substantially equivalent to the predicate device. {2}------------------------------------------------ #### X. Performance Data: Bench tests demonstrate the safety and effectiveness of the 40 L High Flow Insufflator F108 when used in the Low Flow Mode to insufflate small abdominal cavities. #### XI. Clinical Data: No clinical tests were performed. A review of professional literature demonstrates the safety and effectiveness of CO2 insufflators in pediatric laparoscopy. #### XII. Voluntary Standards: The device complies with the International Standard IEC 60601-1 (Electrical Safety) and IEC 60601-1-2 (Electromagnetic Compatibility). In addition, the device meets the requirements of the Underwriter Laboratories Standard UL 2601-1 and bears the CE mark in accordance with the Medical Device Directive 93/42/EEC. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 FEB - 5 2004 W.O.M. World of Medicinc AG % Ms. Susanne Raab Regulatory Consultant 368 St. North Asaph Street ALEXANDRIA VA 22314 Re: K030837 Trade/Device Name: 40 L High Flow Insufflator F108 Regulation Number: 21 CFR 884.1730 Regulation Name: Laparoscopic insufflator Regulatory Class: II Product Code: 85 HIF Dated: November 7, 2003 Received: November 7, 2003 Dear Ms. Raab: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050, {4}------------------------------------------------ Page 2 This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter: | 8xx. 1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Brydon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): K030837 40L High Flow Insufflator F108 Device Name: _________________________________________________________________________________________________________________________________________________________________ Indications For Use: "The 40L High Flow Insufflator F108 with low flow mode is a device intended to facilitate the use of the laparosoope by filling the peritoneal cavity with gas to distend it. The low flow application of this device is indicated for pediatric use." ﮐﯽ ﺍ Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Nancy C. Brogdon (Division Sign-Off) Division of Reproductive, A and Radiological Devices 510(k) Number _
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...