LAPARO CO2-PNEU AUTOMATIC INSUFFLATOR WITH 30L FLOW RATE,AND VIDEO-SCREEN DISPLAY, HIGH FLOW INSUFFLATION TUBE WITH HEAT

K991906 · Richard Wolf Medical Instruments Corp. · HIF · Nov 23, 1999 · Obstetrics/Gynecology

Device Facts

Record IDK991906
Device NameLAPARO CO2-PNEU AUTOMATIC INSUFFLATOR WITH 30L FLOW RATE,AND VIDEO-SCREEN DISPLAY, HIGH FLOW INSUFFLATION TUBE WITH HEAT
ApplicantRichard Wolf Medical Instruments Corp.
Product CodeHIF · Obstetrics/Gynecology
Decision DateNov 23, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.1730
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Laparo CO2-Pneu automatic insufflator generates and maintains pneumoperitoneum under CO2 gas and can be used in diagnostic and operative laparoscopy.

Device Story

The Laparo CO2-Pneu is an automated insufflator used in diagnostic and operative laparoscopy to create and maintain pneumoperitoneum. It connects to CO2 gas sources (cylinders or central supply) and delivers gas to the patient's abdomen. The device features microprocessor-controlled pressure and flow regulation, allowing users to preselect intra-abdominal pressure (5-25 mmHg) and flow rates (1-30 L/min). An integrated electronic circuit in the insufflation tube warms the gas to body temperature. The device displays real-time pressure, flow, and gas consumption data on video equipment for clinician monitoring. It is intended for use by physicians in clinical settings. The device includes safety mechanisms to manage pressure and prevent over-inflation. It is designed to be used in conjunction with other surgical equipment, requiring careful monitoring when used with argon plasma coagulators to avoid pressure-related risks.

Clinical Evidence

No clinical tests performed. Bench testing only.

Technological Characteristics

Microprocessor-controlled CO2 insufflator. Preselectable pressure (5-25 mmHg) and flow (1-30 L/min). Integrated electronic gas warming circuit. Video output for data display. Standalone unit.

Indications for Use

Indicated for patients undergoing diagnostic or operative laparoscopy requiring pneumoperitoneum. Contraindicated if laparoscopy is contraindicated. Not suitable for hysteroscopy or inflation of the cavum uteri.

Regulatory Classification

Identification

A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K991906 353 Corporate Woods Parkway Vernon Hills, Illinois 60061 Phone: 847.913.1113 Fax: 847.913.1488 NOV 23 1999 트 | RICHARD WOLF | |--------------| |--------------| MEDICAL INSTRUMENTS CORPORATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Image /page/0/Picture/5 description: The image contains a black and white symbol that resembles a stylized letter or character. The symbol is composed of angular lines and curves, creating a unique and abstract design. The symbol is set against a white background, which provides contrast and makes the symbol stand out. The symbol is small and compact, and it appears to be a part of a larger text or design. # 510(k) Summary of Safety and Effectiveness | Date of Preparation | May 14, 1999 | |--------------------------------------|----------------------------------------| | Company / Institution name: | Richard Wolf Medical Instruments Corp. | | FDA establishment regulation number: | 14 184 79 | | Division name (if applicable): | N.A. | | Phone number (include area code): | (847) 913-1113 | | Street address: | 353 Corporate Woods Parkway | | FAX number (include area code). | (847) 913-0924 | | City: | Vernon Hills | | State/Province: | Illinois | | Country: | USA | | ZIP/Postal Code: | 60061 | | Contact name: | Mr. Robert L. Casarsa | | Contact title: | Quality Assurance Manager | | Trade name: | Laparo CO₂-Pneu | | Model number: | 2232.621, 631 | | Common name: | Automated Laparoscopic CO₂ Insufflator | | Classification Name: | Laparoscopic Insufflator | Information on devices to which substantial equivalence is claimed: | | 510(k) Number | Trade or proprietary or model name | Manufacturer | |---|---------------|------------------------------------|--------------| | 1 | K981334 | Laparo CO₂ Pneu 2232 (20 L) | Richard Wolf | | 2 | K955073 | Thermoflator 264320 20 | Karl Storz | | 3 | | | | | 4 | | | | #### 1.0 Description The Laparo CO2-Pneu 2232 supplies CO2 gas from cylinders or central gas supplies to dilate the abdomen by the CO2 gas insufflation. The gas pressure and the flow rate are adjusted by the user and electronically controlled. {1}------------------------------------------------ #### 2.0 Intended Use The Laparo CO2-Pneu automatic insufflator generates and maintains pneumoperitoneum under CO2 gas and can be used in diagnostic and operative laparoscopy. ### 3.0 Technological Characteristics The pressure and flow values are checked by microprocessor control. A number of independent safety devices provide safe and problem-free insufflation. The intra-abdominal pressure between 5 mmHg and 25 mmHg can be preselected. The pressure is reduced if the intra-abdominal pressure is higher than the preselected. THC The flow rate can be preselected between 1 L/minute and 30 L/minute at a pace of 1 V/min. The measured values can be displayed on-screen in order to document the pressure, flow, and gas consumption values during intervention via video equipment. An electronic circuit, integrated into the insufflation tube, warms the gas to body temperature. ### 4.0 Substantial Equivalence The Laparo CO2-Pneu is substantially equivalent to existing 510(k) devices sold by Richard Wolf and Karl Storz. - 5.0 Performance Data No performance data generated. - 6.0 Clinical Tests No clinical tests performed. #### 7.0 Conclusions Drawn Laparo CO2-Pneu 2232 was designed and tested to guarantee the safety and effectiveness during the expected life time of the device when used according to the instruction manual. By: Robert J. Casassa Robert L. Casarsa Quality Assurance Manager Date: May 18, 1999 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle or other bird, with its wings spread and head facing to the left. The logo is black and white. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 NOV 2 3 1999 Mr. Robert L. Casarsa Quality Assurance Manager Richard Wolf Medical Instruments Corp. 353 Corporate Woods Parkway Vernon Hills, IL 60061 Re: K991906 Laparo CO2 Pneu 2232 Automatic Insufflator Dated: September 1, 1999 Received: September 2, 1999 Regulatory Class: II 21 CFR 884.1730/Procode: 85 HIF Dear Mr. Casarsa: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act, You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. lf you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ K991906 Device Name: _ ## Intended Use: The Laparo CO2-Pneu automatic insufflator generates and maintains pneumoperitoneum under CO2 gas and can be used in diagnostic and operative laparoscopy. ## Contraindications: The use of this device is contraindicated if laparoscopy is contraindicated. Warning: The device has a high flow rate. The device is not suitable for hysteroscopy. Do not use this device to inflate the cavum uteri. Contractindications for the patient resulting from the general findings, which are described in the relevant literature must be observed. ## Possible Combinations: Important! In addition to the product instruction manual, be sure to observe the instruction manuals of the products used in combination with this product. Electromagnetic interference or other influences which may occur between the product and other products may lead to defects or malfunctions. Warning! Danger of potentially fatal gas or air embolism. Purge the connection tubes thoroughly with CO, before every use. The use of devices using additional gaseous media in combination with the Laparo CO-Pncu automatic insufflator lies exclusively in the responsibility of the user. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH Office of Device Evaluation (ODE) (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices | 510(k) Number | K991906 | |---------------|---------| |---------------|---------| | Prescription Use | <div style="text-align:center;">OR</div> | Over- | |------------------|------------------------------------------|-------| |------------------|------------------------------------------|-------| {4}------------------------------------------------ Warning! There is danger of intra-abdominal pressure when a second gas source is used. When argon plasma coagulations are used, it is vital for the user to also have additional visual or manual pressure monitoring available, as there is a potential risk that the breasure monitoring or ventilation of the Laparo CO2-Pneu is deactivated, e.g. by a kinked insufflation tube or a closed instrument port (stopcock). The preselected gas flow rate of argon plasma coagulations should not excecd 2 Vmin. Activate the argon plasma coagulator only for shot periods of time. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH Office of Device Evaluation (ODE) (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number K991906 Prescription Use X Per 21 CRF 801.109 OR
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...