K193339 · Guangzhou T.K Medical Instrument Co., Ltd. · HIF · Jan 30, 2020 · Obstetrics/Gynecology
Device Facts
Record ID
K193339
Device Name
GTK Veress Needles
Applicant
Guangzhou T.K Medical Instrument Co., Ltd.
Product Code
HIF · Obstetrics/Gynecology
Decision Date
Jan 30, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1730
Device Class
Class 2
Indications for Use
The GTK Veress Needles are intended for percutaneous insertion into the peritoneal cavity for the purpose of insufflation with carbon dioxide to establish pneumoperitoneum prior to the placement of trocars during laparoscopic procedures.
Device Story
GTK Veress Needles are single-use, sterile, spring-loaded instruments used to establish pneumoperitoneum for laparoscopic surgery. The device consists of an inner needle, outer needle, warning block, spring, handle, and valve. Operation involves percutaneous insertion into the peritoneal cavity; a spring-loaded blunt stylet mechanism facilitates safe entry. The device connects to an external insufflator via a luer connector to deliver carbon dioxide. Used in hospital settings by trained clinicians. The device provides a pathway for gas insufflation, enabling surgeons to create the necessary space for subsequent trocar and cannula placement. Benefits include facilitating minimally invasive surgical access.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including appearance, size, tip pull, switch operation, spring obturator operation, and needle puncture force tests. Biocompatibility testing per ISO 10993 (cytotoxicity, sensitization, reactivity, systemic toxicity, pyrogenicity) and sterilization validation per ISO 11135 were performed. Shelf-life and package integrity were validated via accelerated aging (ASTM F1980) and associated standards (ASTM D3078, ASTM F88, ASTM F1929, DIN 58953-6).
Technological Characteristics
Materials: SUS 304 (needles), PC (handle). Principle: Spring-loaded blunt stylet mechanism. Energy: None (mechanical). Form factor: 120mm and 150mm lengths. Connectivity: Luer connector for insufflation tubing. Sterilization: Ethylene oxide (EO).
Indications for Use
Indicated for percutaneous insertion into the peritoneal cavity to establish pneumoperitoneum via carbon dioxide insufflation prior to trocar placement in laparoscopic procedures.
Regulatory Classification
Identification
A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 30, 2020
Guangzhou T.K Medical Instrument Co., Ltd. % Christy Young Consultant Shenzhen Joyantech Consulting Co., Ltd 1713A, Block A, Zhongguan Times Square, Liuxian Avenue Xili Town, Nanshan District, Shenzhen Guangdong Province, 518000 CHINA
K193339 Trade/Device Name: GTK Veress Needles Regulation Number: 21 CFR 884.1730 Regulation Name: Laparoscopic Insufflator Regulatory Class: II Product Code: HIF, FHO Dated: December 2, 2019 Received: December 2, 2019
#### Dear Christy Young:
Re:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Sharon M. Andrews Acting Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K193339
Device Name GTK Veress Needles
Indications for Use (Describe)
The GTK Veress Needles are intended for percutaneous insertion into the peritoneal cavity for the purpose of insufflation with carbon dioxide to establish pneumoperitoneum prior to the placement of trocars during laparoscopic procedures.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 1. Submission Sponsor | |
|-----------------------|-----------------------------------------------------------------------------------------------------------------|
| Applicant Name | Guangzhou T.K Medical Instrument Co., Ltd. |
| Address | A601, Guangzhou International Business Incubator<br>, Guangzhou Science Park, Guangzhou, Guangdong, P. R. China |
| Post Code | 510663 |
| Phone No. | +86-20-32290169 |
| Fax No. | +86-20-32290395 |
| Contact Person | Ms. Tracy Weng |
| Date Prepared | 1-28-2020 |
# 510(k) Summary
# 2. Submission correspondent
| Name | Shenzhen Joyantech Consulting Co., Ltd |
|----------------|---------------------------------------------------------------------------------------------------------------------|
| Address | 1713A, Block A, Zhongguan Times Square, Liuxian Avenue, Xili Town, Nanshan District, Shenzhen, Guangdong, P.R.China |
| | Image: [logo] |
| Post Code | 518000 |
| Phone No. | 86-755-86069197 |
| Contact Person | Ms. Christy Young; Field Fu |
| Email | christy@cefda.com; |
### 3. Devices Identification
| Trade Name | GTK Veress Needles |
|-----------------------|-------------------------------------------------------|
| Common Name | Veress Needle |
| Model Names | VN-120 and VN-150 |
| Regulatory Class | II |
| Classification Name | Laparoscopic insufflator |
| Classification Number | 21 CFR 884.1730 |
| Product Code | HIF; FHO |
| Product Code Name | Insufflator, Laparoscopic; Pneumoperitoneum<br>Needle |
| 510(k) review panel | Obstetrics/Gynecology; Gastroenterology/Urology |
{5}------------------------------------------------
#### 4. Leqally Marketed Predicate Devices
| Trade Name | Veress Needle |
|-----------------------|--------------------------------------------------------------|
| Classification Number | 21 CFR 884.1730 |
| Classification Name | Laparoscopic insufflator |
| Regulatory Class | II |
| 510 (k) number | K172120 |
| Product Code | HIF; FHO |
| Product Code Name | Insufflator, Laparoscopic; Pneumoperitoneum Needle |
| Manufacturer | WickiMed (Huizhou) Medical Equipment Manufacturing Co., Ltd. |
The predicate device was not subject to a design-related recall.
### 5. Legally Marketed Reference Devices
| Trade Name | Medline Pneumoperitoneum Needle |
|-----------------------|----------------------------------------------------|
| Classification Number | 21 CFR 884.1730 |
| Classification Name | Laparoscopic insufflator |
| Regulatory Class | II |
| 510 (k) number | K111955 |
| Product Code | HIF; FHO; |
| Product Code Name | Insufflator, Laparoscopic; Pneumoperitoneum Needle |
| Manufacturer | Medline Industries, Inc |
The reference device was not subject to a design-related recall
#### 6. Device Description
The structure of the GTK Veress Needles contain inner needle, outer needle, warning block, spring, handle and valve. It contains a spring-loaded blunt stylet mechanism, which is also the major principle mechanism of GTK Veress Needles to achieve the intended use. It is used to establish pneumoperitoneum prior to trocar and cannula insertion in laparoscopic procedures. The veress needle is connected to the insufflators with a luer connector. There are two models of veress needles: VN-120 and VN-150, with different lengths of needle bodies (120mm and 150mm).
The GTK Veress Needles are for single use and are supplied sterile. The contact time is no more than 24 hours. The product is sterilized by Ethylene oxide. The manufacture date and expiry date are shown on the labeling. The GTK Veress Needles are only intended for use in a hospital setting/environment.
### 7. Indications for Use Statement
The GTK Veress Needles are intended for percutaneous insertion into the peritoneal cavity for the purpose of insufflation with carbon dioxide to establish
{6}------------------------------------------------
pneumoperitoneum prior to the placement of trocars during laparoscopic procedures.
## 8. Substantial Equivalence Discussion
| Items | Proposed device | Predicate device | Comments | | |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|------------|------|
| Trade name | GTK Veress Needles | Veress Needle | / | | |
| 510(K)<br>submitter | Guangzhou T.K<br>Medical Instrument<br>Co., Ltd. | WickiMed (Huizhou)<br>Medical Equipment<br>Manufacturing Co., Ltd. | / | | |
| 510(K) number | / | K172120 | / | | |
| Classification<br>regulation | 21CFR 884.1730 | 21CFR 884.1730 | Same | | |
| Classification<br>and product<br>code | Class II ,<br>HIF,FHO | Class II ,<br>HIF,FHO | Same | | |
| Common name | Veress Needle | Veress Needle | Same | | |
| Intended use | The GTK Veress<br>Needles are intended<br>for percutaneous<br>insertion into the<br>peritoneal cavity for the<br>purpose of insufflation<br>with carbon dioxide to<br>establish<br>pneumoperitoneum<br>prior to the placement<br>of trocars during<br>laparoscopic<br>procedures. | The Veress Needle is<br>intended for<br>percutaneous insertion<br>into the peritoneal cavity<br>for the purpose of<br>insufflation with carbon<br>dioxide to establish<br>pneumoperitoneum prior<br>to the placement of<br>trocars during<br>laparoscopic<br>procedures. | Same | | |
| Components | Veress Needle<br>Obturator | Veress Needle<br>Obturator | Same | | |
| Model | VN-120 and VN-150 | WVN0112T, WVN0115T | / | | |
| Length | 120mm,150mm | 120mm,150mm | Same | | |
| Sterilization | EO Sterilized | EO Sterilized | Same | | |
| Disposable | Yes | Yes | Same | | |
| Materials | Patient-<br>contacting<br>Material<br>(Outer tube<br>inner tube) | SUS<br>304 | Patient-<br>contacting<br>Material<br>(Outer tube<br>inner tube) | SUS<br>304 | Same |
| | Handle material | PC | Handle material | PC | Same |
| Principles of operation | Connect the device to the insufflators with insufflation tubing, insufflating with carbon dioxide to establish pneumoperitoneum. | | Connect the device to the insufflators with insufflation tubing, insufflating with carbon dioxide to establish pneumoperitoneum. | | Same |
### Table 1: Substantial equivalence comparison
{7}------------------------------------------------
### 9. Non-Clinical Performance Data
#### 1) Biocompatibility test
The needle (inner and outer) of the GTK Veress Needles are contacting with abdominal skin and abdominal wall tissue of human for less than 24 hours. The following tests were conducted:
ISO 10993-5: 2009-In Vitro Cytotoxicity test;
ISO 10993-10: 2010-Skin Sensitization test and Intracutaneous reactivity test; ISO 10993-11: 2006- Acute systemic toxicity and Pyrogen test;
USP 39 <85>-Bacterial endotoxin test.
#### 2) Sterilization validation
The sterilization validation was conducted according to ISO 11135:2014. The EO and ECH residual of the proposed device was evaluated by ISO 10993-7:2008.
#### Shelf life validation test 3)
The package verification test on proposed devices were conducted after accelerated aging for 3 years (ASTM F 1980-07 (2011)):
ISO 11737-2:2009-sterility test;
ASTM D 3078-02:2013-vacuum leak test;
ASTM F 88-seal strength test;
ASTM F 1929-12-leakage (dye penetration test);
Din 58953-6:2010-microbial barrier properties (agar contact-attack test); And physical performance of the proposed device was conducted after natural aging. All the results meet the acceptance criteria.
#### 4) Performance test-bench
The performance tests of GTK Veress Needles contain appearance, size, tip pull test, switch operation test, spring obturator operation and needle puncture force test. All the results meet the acceptance criteria. they also demonstrate that the GTK Veress Needles meet the performance characteristics.
### 10. Statement of Substantial Equivalence
The Indications for Use and technological characteristics for GTK Veress Needles are same to the predicate devices (K172120). The non-clinical performance testing demonstrates that the proposed device is as safe and effective as the predicate devices. Therefore, the results show that it is Substantially Equivalent (SE) between
{8}------------------------------------------------
products and predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.