Medline Sterile Power-Free Nitrile Examination Gloves - Blue (Test for Use with Chemotherapy Drugs)
K192104 · Medline Industries, Inc. · LZA · Dec 23, 2019 · General Hospital
Device Facts
Record ID
K192104
Device Name
Medline Sterile Power-Free Nitrile Examination Gloves - Blue (Test for Use with Chemotherapy Drugs)
Applicant
Medline Industries, Inc.
Product Code
LZA · General Hospital
Decision Date
Dec 23, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. Carmustine (BCNU) 3.3 mg/ml 28.6 minutes Cisplatin 1.0 mg/ml >240 minutes Cyclophosphamide (Cytoxan) 20.0 mg/ml >240 minutes Dacarbazine (DTIC)10.0 mg/ml >240 minutes Doxorubicin Hydrochloride 2.0 mg/ml >240 minutes Etoposide (Toposar) 20.0 mg/ml >240 minutes Fluorouracil 50.0 mg/ml >240 minutes Methotrexate 25.0 mg/ml >240 minutes Mitomycin C 0.5 mg/ml >240 minutes Paclitaxel (Taxol) 6.0 mg/ml >240 minutes Thiotepa 10.0 mg/ml 38.1 minutes Vincristine Sulfate 1.0 mg/ml >240 minutes Please note that the following drugs have low permeation times: Carmustine (BCNU) (3.3 mg/ml) 28.6 minutes Thiotepa (10.0 mg/ml) 38.1 minutes Warning: Do not use with Carmustine and Thiotepa
Device Story
Disposable nitrile examination glove; worn by healthcare examiners to prevent cross-contamination between patient and examiner. Device provides barrier protection during medical procedures. Tested for permeation resistance against specific chemotherapy drugs per ASTM D6978-05. Output is physical barrier protection. Used in clinical settings; operated by healthcare professionals. Benefits include reduced risk of exposure to patient pathogens and chemotherapy agents.
Clinical Evidence
Bench testing only. Permeation testing performed per ASTM D6978-05 (Reapproved 2013) to assess resistance to chemotherapy drugs. Results provided for 12 specific chemotherapy agents with permeation times ranging from 28.6 minutes to >240 minutes.
Technological Characteristics
Nitrile examination glove; powder-free; blue color. Tested per ASTM D6978-05 for chemotherapy drug permeation. Non-sterile (implied by examination glove classification, though labeled 'Sterile' in trade name, standard examination glove controls apply).
Indications for Use
Indicated for use as a disposable patient examination glove worn by examiners to prevent cross-contamination between patient and examiner. Suitable for use with specific chemotherapy drugs, excluding Carmustine and Thiotepa.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Related Devices
K193581 — Non-Sterile, Powder Free Nitrile Exam Glove Tested for use with Chemotherapy Drugs - Lilac, Non-Sterile, Powder Free Nitrile Exam Glove Tested for use with Chemotherapy Drugs - Orchid, Non-Sterile, Powder Free Nitrile Exam Glove Tested for use with Chemotherapy Drugs - Oyster · Sri Trang Gloves (Thailand) Public Company Limited · Mar 20, 2020
K183280 — PowderFree White Blue Sterilized Nitrile Copolymer Examination Gloves Tested for use with chemotherapy drugs · Smart Glove Corporation Sdn. Bhd. · Jul 26, 2019
K181130 — Powder Free Blue Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs · Hebei Titans Hongsen Medical Technology Co., Ltd. · Aug 10, 2018
K213548 — Nitrylex Classic Powder Free Nitrile Blue Examination Gloves Tested for Use with Chemotherapy · Mercator Medical (Thailand), Ltd. · Feb 21, 2022
K250502 — SafeSource Direct Blue Powder-Free Nitrile Exam Gloves (AMERI-TUFF Series 4000, AMERI-TOUCH Series 5000) · Safesource Direct, LLC · Mar 19, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 23, 2019
Claire Pigman Associate Manager, Regulatory Affairs Three Lakes Drive Northfield, Illinois 60093
Re: K192104
Trade/Device Name: Medline Sterile Power-Free Nitrile Examination Gloves - Blue (Tested for Use with Chemotherapy Drugs) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: November 21, 2019 Received: November 22, 2019
Dear Claire Pigman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
## 510(k) Number (if known) K192104
#### Device Name
Medline Sterile Powder-Free Nitrile Examination Gloves - Blue (Tested for use with Chemotherapy Drugs)
#### Indications for Use (Describe)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. Carmustine (BCNU) 3.3 mg/ml 28.6 minutes Cisplatin 1.0 mg/ml >240 minutes Cyclophosphamide (Cytoxan) 20.0 mg/ml >240 minutes Dacarbazine (DTIC)10.0 mg/ml >240 minutes Doxorubicin Hydrochloride 2.0 mg/ml >240 minutes Etoposide (Toposar) 20.0 mg/ml >240 minutes Fluorouracil 50.0 mg/ml >240 minutes Methotrexate 25.0 mg/ml >240 minutes Mitomycin C 0.5 mg/ml >240 minutes Paclitaxel (Taxol) 6.0 mg/ml >240 minutes Thiotepa 10.0 mg/ml 38.1 minutes Vincristine Sulfate 1.0 mg/ml >240 minutes
Please note that the following drugs have low permeation times: Carmustine (BCNU) (3.3 mg/ml) 28.6 minutes Thiotepa (10.0 mg/ml) 38.1 minutes Warning: Do not use with Carmustine and Thiotepa
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.