Nitrylex Classic Powder Free Nitrile Blue Examination Gloves Tested for Use with Chemotherapy

K213548 · Mercator Medical (Thailand), Ltd. · LZA · Feb 21, 2022 · General Hospital

Device Facts

Record IDK213548
Device NameNitrylex Classic Powder Free Nitrile Blue Examination Gloves Tested for Use with Chemotherapy
ApplicantMercator Medical (Thailand), Ltd.
Product CodeLZA · General Hospital
Decision DateFeb 21, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Intended Use

Nitrylex® Classic Powder Free Nitrile Blue Examination Gloves Tested for use with Chemotherapy Drugs are disposable devices intended for medical propose that are worn on the examiner's hands to prevent contamination between patent and examiner.

Device Story

Disposable nitrile examination gloves; worn by healthcare examiners to prevent cross-contamination between patient and examiner. Device provides barrier protection during medical procedures. Tested for permeation resistance against various chemotherapy drugs per ASTM D678-05. Output is physical barrier protection; used in clinical settings. Benefits include reduced risk of exposure to patient fluids and specific hazardous chemotherapy agents. Not for use with Carmustine or Thiotepa.

Clinical Evidence

Bench testing only. Permeation testing performed per ASTM D678-05 for 31 chemotherapy drugs and Fentanyl Citrate. Breakthrough times exceeded 240 minutes for all tested substances except Carmustine (34.8 min) and Thiotepa (47.4 min).

Technological Characteristics

Nitrile examination glove; powder-free; blue color. Tested per ASTM D678-05 for chemotherapy drug permeation resistance.

Indications for Use

Indicated for use as a disposable medical glove worn on the examiner's hands to prevent cross-contamination between patient and examiner. Tested for resistance to permeation by specific chemotherapy drugs. Contraindicated for use with Carmustine and Thiotepa.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. February 21, 2022 Mercator Medical (Thailand) LTD % Kevin Walls Principal Consultant Regulatory Insight, Inc. 33 Golden Eagle Lane Littleton. Colorado 80127 Re: K213548 Trade/Device Name: Nitrylex Classic Powder Free Nitrile Blue Examination Gloves Tested for Use with Chemotherapy Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: January 31, 2022 Received: February 1, 2022 Dear Kevin Walls: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K213548 #### Device Name Nitrylex® Classic Powder Free Nitrile Blue Examination Gloves Tested for use with Chemotherapy Drugs #### Indications for Use (Describe) Nitrylex® Classic Powder Free Nitrile Blue Examination Gloves Tested for use with Chemotherapy Drugs are disposable devices intended for medical propose that are worn on the examiner's hands to prevent contamination between patent and examiner. The proposed device was tested for use with Chemotherapy Drugs as per ASTM D678-05 Standard Practice for Assessment Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. Chemotherapy Drugs Permeation. The following chemicals have been tested with the proposed device. | Test Chemotherapy Drugs | Concentration | Breakthrough Detection<br>(Time In Minutes) | |----------------------------|-------------------------|---------------------------------------------| | Bleomycin Sulfate | 15 mg/ml (15,000 ppm) | >240 min. | | Busulfan | 6 mg/ml (6,000 ppm) | >240 min. | | Carboplatin | 10 mg/ml (10,000 ppm) | >240 min. | | Cisplatin | 1 mg/ml (1,000 ppm) | >240 min. | | Cyclophosphamide (Cytoxan) | 20.0 mg/ml (20,000 ppm) | >240 min. | | Cytarabine HCl | 100 mg/ml (100,000 ppm) | >240 min. | | Cytovene | 10 mg/ml (10,000 ppm) | >240 min. | | Dacarbazine | 10 mg/ml (10,000 ppm) | >240 min. | | Daunorubicin HCl | 5 mg/ml (5,000 ppm) | >240 min. | | Docetaxel | 10 mg/ml (10,000 ppm) | >240 min. | | Doxorubicin HCl | 2.0 mg/ml (2,000 ppm) | >240 min. | | Epirubicin HCl | 2 mg/ml (2,000 ppm) | >240 min. | | Etoposide | 20.0 mg/ml (20,000 ppm) | >240 min. | | Fludarabine | 25 mg/ml (25,000 ppm) | >240 min. | | Fluorouracil | 50.0 mg/ml (50,000 ppm) | >240 min. | | Gemcitabine | 38 mg/ml (38,000 ppm) | >240 min. | | Idarubicin HCl | 1 mg/ml (1,000 ppm) | >240 min. | | Ifosfamide | 50 mg/ml (50,000 ppm) | >240 min. | | Irinotecan | 20 mg/ml (20,000 ppm) | >240 min. | | Mechlorethamine HCl | 1 mg/ml (1,000 ppm) | >240 min. | | Melphalan | 5 mg/ml (5,000 ppm) | >240 min. | | Methotrexate | 25 mg/ml (25,000 ppm) | >240 min. | | Mitomycin C | 0.5 mg/ml (500 ppm) | >240 min. | | Mitoxantrone | 2 mg/ml (2,000 ppm) | >240 min. | | Oxaliplatin | 2 mg/ml (2,000 ppm) | >240 min. | | Paclitaxel | 6.0 mg/ml (6,000 ppm) | >240 min. | | Rituximab | 10 mg/ml (10,000 ppm) | >240 min. | | Trisenox | 1 mg/ml (1,000 ppm) | >240 min. | | Vincristine Sulfate | 1 mg/ml (1,000 ppm) | >240 min. | | Vinorelbine | 10 mg/ml (10,000 ppm) | >240 min. | | Fentanyl Citrate Injection | 100 mcg/2 ml | >240 min. | {3}------------------------------------------------ Please note that the following drugs showed break through detected in less than 240 minutes: - Carmustine (BCNU), 3.3 mg/ml: 34.8 minutes - Thiotepa (THT) 10.0 mg/ml: 47.4 minutes Warning-Not for Use with Carmustine and Thiotepa Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...