White Nitrile Powder Free Patient Examination Glove, Non Sterile
K222713 · Safeskin Retailing (Hk) Limited · LZA · Dec 3, 2022 · General Hospital
Device Facts
Record ID
K222713
Device Name
White Nitrile Powder Free Patient Examination Glove, Non Sterile
Applicant
Safeskin Retailing (Hk) Limited
Product Code
LZA · General Hospital
Decision Date
Dec 3, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A nitrile patient examination glove is a disposable device made of nitrile rubber intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner
Device Story
Disposable nitrile rubber patient examination glove; ambidextrous; non-sterile; powder-free. Used by healthcare professionals in clinical settings to provide a barrier against contamination between patient and examiner. Device is donned manually; surface treatment ensures smooth donning without lubricants. Output is a physical barrier. Benefits include protection against cross-contamination during medical examinations.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via ASTM D412 (tensile/elongation), ASTM D3767 (dimensions), ASTM D5151 (watertightness/AQL 1.5), and ASTM D6124 (residual powder <2mg). Biocompatibility confirmed via ISO 10993-10 (irritation/sensitization) and ISO 10993-11 (acute systemic toxicity).
Technological Characteristics
Material: Nitrile rubber. Form factor: Ambidextrous, non-sterile, powder-free glove. Standards: ASTM D6319, ASTM D412, ASTM D3767, ASTM D5151, ASTM D6124. Biocompatibility: ISO 10993-10, ISO 10993-11. No energy source, software, or connectivity.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Xingyu Nitrile Powder Free Patient Examination Gloves, White Color (K171615)
Related Devices
K222720 — Blue Nitrile Powder Free Patient Examination Glove, Non Sterile · Safeskin Retailing (Hk) Limited · Dec 3, 2022
K212804 — Powder free nitrile examination glove- blue, non-sterile · Hong Seng Gloves Sdn Bhd · Dec 1, 2021
K211028 — Nitrile Patient Examination Gloves · Huayuan Medical Technology(Shangqiu) Co., Ltd. · Jun 4, 2021
K210587 — Violet Nitrile Powder Free Patient Examination Glove, Non Sterile · Onetexx Sdn Bhd · Jun 28, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 3, 2022
Safeskin Retailing (HK) Limited % Stephan Toupan President Dawa Medical LLC 7320 NW 12th Street Suite 103 Miami, Florida 33126
Re: K222713
Trade/Device Name: White Nitrile Powder Free Patient Examination Glove, Non Sterile Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: September 5, 2022 Received: September 8, 2022
Dear Stephan Toupan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combinationproducts/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Bifeng Qian -S
Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222713
Device Name WHITE NITRILE POWDER-FREE PATIENT EXAMINATION GLOVE, NON STERILE
Indications for Use (Describe)
A nitrile patient examination glove is a disposable device made of nitrile rubber intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner
| Type of Use (Select one or both, as applicable) |
|------------------------------------------------------------------------------------------------|
| |
| |
| <label><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
| <label><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Safeskin Retailing. The logo consists of a blue geometric shape on the left and the words "safeskin retailing" on the right. The geometric shape is a stylized letter "K" formed by two triangles, one in a darker shade of blue and the other in a lighter shade. The text "safeskin retailing" is in a bold, sans-serif font, with "safeskin" on the top line and "retailing" on the bottom line.
# K222713 510K Summary
Trade Name : White Nitrile Powder-free Patient
Examination Glove, Non-Sterile
{4}------------------------------------------------
#### 1.0 Submitter:
| Name | : | Alvin Ho |
|--------------------------|---|---------------------------------------------------------------------------------------------------------------|
| Address | : | Safeskin Retailing (HK) Limited<br>26th Floor, Beautiful Group Tower, 77 Connaught Road<br>Central, Hong Kong |
| Phone No. | : | +6012 826 5625 |
| Date of Summary Prepared | : | 8/1/2022 |
#### 2.0 ldentification of the subject device:
| Trade Name: | White Nitrile Powder-Free Patient Examination Glove, Non-<br>Sterile |
|------------------------|----------------------------------------------------------------------|
| Common Name: | Patient Examination Gloves |
| Classification Name : | Patient Examination Gloves |
| Device Classification: | 1 |
| Regulation Number : | 21 CFR 880.6250 |
| Product Code: | LZA |
#### 3.0 Predicate Device:
# K171615
Xingyu Nitrile Powder Free Patient Examination Gloves, White Color Company: Shandong Xingyu Gloves Co., Ltd
#### 4.0 Description of The Subject Device:
White Nitrile Powder-Free Patient Examination Glove, Non-Sterile is manufactured from nitrile rubber. Innersurface of gloves undergoes surface treatment process to produce a smooth surface that assists the user in donning the gloves without using any lubricant such as powder or silicone on the glove surface. The glove is ambidextrous, i.e., can be worn on right or left hand.
#### 5.0 Indication for use:
A Nitrile patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
#### 6.0 Comparison of the Technological Characteristics of the Device:
The White Nitrile Powder-Free Patient Examination Gloves, Non-Sterile are summarized with the following technological characteristics compared to ASTM D6319 or equivalent standards as shown in Table 1
{5}------------------------------------------------
# Table 1
| CHARACTERISTICS | | STANDARDS | DEVICE PERFORMANCE | | COMPARISON ANALYSIS |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| | | | PREDICATE<br>WHITE | CURRENT<br>WHITE | |
| 510(k) Number | | - | K171615 | K222713 | |
| Manufacturer(s) | | - | Shandong Xingyu Gloves Co., Ltd | Safeskin Retailing (HK) Limited | Same |
| Material | | ASTM D6319 | Nitrile | Nitrile | Same |
| Color | | - | White | White | Same |
| Sterility | | - | Non-Sterile | Non-Sterile | Same |
| Handedness | | - | Ambidextrous | Ambidextrous | Same |
| Physical Properties | Before Aging<br>Tensile Strength:<br>Ultimate Elongation: | ASTM D6319 | 18-25Mpa<br>560-600% | 28.1 – 32.9 Mpa<br>530 - 600% | Different but<br>within the ASTM<br>standard |
| | After Aging Tensile<br>Strength:<br>Ultimate Elongation: | | 17-21Mpa<br>460-560% | 30.8 – 35.9 Mpa<br>460 - 560% | Different but<br>within the ASTM<br>standard |
| Thickness: | - Finger<br>- Palm | ASTM D6319 | 0.08mm<br>0.08mm | Min 0.09mm for (XS, S, M, L XL)<br>Min 0.06mm for (XS, S, M, L, XL) | Different but<br>within the ASTM<br>standard |
| | Powder Free | | | ASTM D6124 | Below 2mg of residual powder |
| CHARACTERISTICS | STANDARDS | DEVICE PERFORMANCE | | COMPARISON<br>ANALYSIS | |
| | | PREDICATE | CURRENT | | |
| | | WHITE | WHITE | | |
| Biocompatibility | Primary Skin Irritation<br>- ISO 10993-10:2010<br>(E) & Consumer<br>Product Safety<br>Commission Title 16,<br>Chapter II, Part 1500 | Under the conditions of this<br>study, the test article was a non-<br>irritant or non-sensitizer.<br><br>SKIN IRRITATION DERMAL and<br>SENSITIZATION STUDIES Meets<br>ISO 10993-10: Third Edition<br>2010-08-01 | Under conditions of this study, the<br>test material did not cause an<br>irritant response. The Primary<br>Irritant Response Category is<br>deemed 'Negligible' | Same | |
| | Dermal Sensitization-<br>ISO 10993-10: 2010<br>(E) & Consumer Product<br>Safety Commission,<br>Title 16,Chapter II, Part<br>1500.3 (c) (4) | Under the conditions of this<br>study, the test article was a non-<br>irritant or non-sensitizer.<br><br>SKIN IRRITATION DERMAL and<br>SENSITIZATION STUDIES<br>Meets ISO 10993-10: Third<br>Edition 2010-08-01 | Under conditions of this study,the<br>test material did not produce a skin<br>sensitization effect in the guinea<br>pigs. | Same | |
| | Acute Systemic<br>Toxicity, ISO 10993-<br>11:2017 (E) | Not Tested | Under conditions of this study,the<br>test item did not induce any<br>acute systemic toxicity in Swiss<br>albino mice. | Different. The<br>subject glove was<br>tested using<br>systemic toxicity test<br>and passed, but the<br>Predicate did not<br>have the test<br>performed | |
| CHARACTERISTICS | STANDARDS | DEVICE PERFORMANCE | | COMPARISON<br>ANALYSIS | |
| | | PREDICATE<br>WHITE | CURRENT<br>WHITE | | |
| Watertight (1000ml) | ASTM D5151:2019 | Meets 21 CFR 800.20 ASTM D6319-10<br>(Reapproved 2015) Tested in accordance with ASTM<br>D5151 (Reapproved 2015) with<br>acceptable results | Meets 21 CFR 800.20 ASTM D6319-10<br>(Reapproved 2015) Tested in accordance with ASTM<br>D5151 (Reapproved 2015) with<br>acceptable results<br>At an AQL 1.5 | Same | |
| Intended use | - | Powder Free Nitrile Patient<br>Examination Gloves, White Color is<br>a disposable device intended for<br>medical purposes that is worn on<br>the examiner'shand or finger to<br>prevent contamination between<br>patientand examiner. | A Nitrile patient examination glove<br>is a disposable device made of<br>nitrile rubber intended for medical<br>purposes that is worn on the<br>examiner's hand or finger to<br>prevent contamination between<br>patient and examiner. | Same | |
| Size | Medical Glove<br>Guidance Manual -<br>Labeling | Small<br>Medium<br>Large<br>X Large | Extra Small<br>Small<br>Medium<br>Large<br>Extra Large | Similar<br>The subject glove<br>has an additional<br>size Extra Small,<br>which the Predicate<br>does not offer | |
| Single use | Medical Glove<br>Guidance Manual –<br>Labeling | Single Patient Use | Single Use | Same | |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
There are no significant differences between the two devices. They are the same or similar in terms of intended use, primary material (Nitrile), design, physical properties, thickness, powder-free compliance, color, and biocompatibility assessments.
#### Summary of Non-Clinical Testing 7.0
The performance test data of the non-clinical tests for this powder free nitrile examination glove is summarized as per below.
{9}------------------------------------------------
| Test<br>Method | Standard | Purpose of Testing | Acceptance Criteria | | | Results | | Status |
|------------------------|---------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|------------------------|-----------------|-----------------|-----------------|-------------|--------|
| Physical<br>Properties | ASTM D412<br>(Standard Test Method<br>for Vulcanized Rubber<br>and Thermoplastic<br>Elastomers-Tension) | To evaluate the<br>tensile (tension)<br>properties of glove. | Tensile<br>strength | Before<br>aging | After<br>aging | Before<br>aging | After aging | |
| | | | | Min 14.0<br>MPa | Min 14.0<br>MPa | XS – 28.8 | XS – 30.8 | Pass |
| | | | | | | S – 28.6 | S – 31.1 | |
| | | | | | | M – 28.9 | M – 31.8 | |
| | | | | | | L – 28.1 | L – 31.9 | |
| | | | | | | XL – 28.7 | XL – 30.8 | |
| | | | Ultimate<br>elongation | Min<br>500% | Min<br>400% | XS – 540 | XS – 460 | Pass |
| | | | | | | S – 540 | S – 480 | |
| | | | | | | M – 530 | M – 480 | |
| | | | | | | L – 540 | L – 480 | |
| | | | | | | XL – 540 | XL – 480 | |
{10}------------------------------------------------
| Test<br>Method | Standard | Purpose of<br>Testing | Glove<br>Size | Acceptance Criteria | Results | Status | |
|----------------|----------------------------------------------------------------------------------|-----------------------------------------------------------|---------------|------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|------------------------------|------|
| Dimension | ASTM D3767<br>Standard<br>Practice for<br>Rubber—<br>Measurementof<br>Dimensions | To measure the<br>length, width and<br>thickness of glove | X-Small | Length<br>Min 240 mm<br><br>Width<br>$70 \pm 10$ mm<br><br>Thickness<br>Finger – min 0.05mm<br>Palm - min 0.05mm | Length<br>248 mm<br><br>Width<br>75 mm<br><br>Thickness<br>0.09 mm<br>0.06 mm | Pass<br><br>Pass<br><br>Pass | |
| | | | Small | Length<br>Min 240 mm<br><br>Width<br>$80 \pm 10$ mm<br><br>Thickness<br>Finger – min 0.05mm<br>Palm - min 0.05mm | Length<br>245 mm<br><br>Width<br>85 mm<br><br>Thickness<br>0.09 mm<br>0.06 mm | Pass<br><br>Pass<br><br>Pass | |
| | | | Medium | Length<br>Min 240 mm<br><br>Width<br>$95 \pm 10$ mm<br><br>Thickness<br>Finger – min 0.05mm<br>Palm - min 0.05mm | Length<br>244 mm<br><br>Width<br>95 mm<br><br>Thickness<br>0.09 mm<br>0.06 mm | Pass<br><br>Pass<br><br>Pass | |
| | | | Large | Length<br>Min 240 mm<br><br>Width<br>$110 \pm 10$ mm | Length<br>248 mm<br><br>Width<br>106 mm | Pass<br><br>Pass | |
| | | | Thickness | Finger – min 0.05mm | Thickness | 0.09 mm | Pass |
| | | | | Palm – min 0.05mm | | 0.06 mm | |
| | | | Length | Min 240 mm | Length | 244 mm | Pass |
| | | X-Large | Width | $120 \pm 10$ mm | Width | 116 mm | Pass |
| | | | Thickness | Finger – min 0.05mm | Thickness | 0.09 mm | Pass |
| | | | | Palm – min 0.05mm | | 0.06 mm | |
{11}------------------------------------------------
{12}------------------------------------------------
| Test<br>Method | Standard | Purpose of<br>Testing | Acceptance Criteria | Results | Status |
|--------------------|---------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Watertight | ASTM D5151<br>(Standard Test<br>Method for<br>Detection of Holesin<br>Medical Gloves) | To detect holes<br>that leak water and<br>thereby<br>compromise the<br>usefulness of the<br>glove. | Sample size: 315 pcs<br>Inspection level: G1<br>AQL: 1.5, Acceptance No. 10 | The batch size for this<br>sampling is 150,001 to<br>500,000. Hence,<br>according to the single<br>sampling plan GI, the<br>sample to be drawn is<br>under code M equivalent<br>to 315 pieces with accept<br>10 and reject 11 to be<br>accepted under AQL<br>1.5.<br>For Size XS During the<br>test, 1piece was found<br>with leaks. Hence it falls<br>within the acceptance<br>criteria.<br>For Size S During the<br>test, 2piece was found<br>with leaks. Hence it falls<br>within the acceptance<br>criteria.<br>For Size M During the<br>test, 2piece was found<br>with leaks. Hence it falls<br>within the acceptance<br>criteria.<br>For Size L During the<br>test, 4piece was found<br>with leaks. Hence it falls<br>within the acceptance<br>criteria.<br>For Size XL During the<br>test, 3piece was found<br>with leaks. Hence it falls<br>within the acceptance<br>criteria. | Pass |
| Test<br>Method | Standard | Purpose of<br>Testing | Acceptance Criteria | Results | Status |
| Residual<br>Powder | ASTM D6124<br>(Standard Test Method<br>for Residual Powder on<br>Medical Gloves) | To determine the<br>amount of residual<br>and non-powder<br>solids found on<br>gloves | Less than 2 mg per glove<br>Requirement: <2mg/glove | Sample size : 5 pcs<br>Result XS :0.24mg/glove<br>Result S :0.28mg/glove<br>Result M :0.34mg/glove<br>Result L :0.32mg/glove<br>Result XL :0.38mg/glove | Pass |
{13}------------------------------------------------
{14}------------------------------------------------
#### 8.0 Non-clinical performance testing methods full titles:
- ASTM D412 Test Methods for Vulcanized Rubber and Thermoplastic Elastomers— O Tension
- ASTM D573 Test Method for Rubber-Deterioration in an Air Oven x ASTM D3578 o Specification for Rubber Examination Gloves
- ASTM D6319 Standard Specification for Nitrile Examination Gloves for Medical o Application
- ASTM D5151 Test Method for Detection of Holes in Medical Gloves o
- ASTM D6124 Test Method for Residual Powder on Medical Gloves O
- ISO 2859 Sampling Procedures and Tables for Inspection by Attributes Test results show O that under the conditions of the testing, there is no difference inphysical attributes between the proposed device and the predicate device.
#### Biocompatibility Testing utilizing: ISO 10993 Biological Evaluation ofMedical 9.0 Devices:
- ISO 10993 Part 10: Tests for Irritation and Sensitization. Both Skin Irritation andDermal O Magnuson/Kligman Sensitization performed.
- ISO 10993 Part 11: Tests for assessment of Systemic Toxicity O
#### 10.0 Summary of Clinical Testing:
No clinical studies are included in this submission.
# 11.0 Differences:
There are no significant differences between the current glove and the predicate. They are identical in terms of their intended use, base materials, design, color, and manufacturing process.
# Conclusion
The conclusion drawn from the non-clinical tests demonstrate that the subject "White Nitrile Powder-Free Patient Examination Glove, Non-Sterile," is as safe, as effective, and performs as well as, or better than, the legally marketed predicate device "Xinqyu Nitrile Powder Free Patient Examination Gloves, White Color" (K171615)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.