A patient examination glove is a disposable device made of nitrile rubber intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
Disposable nitrile rubber examination glove; ambidextrous; non-sterile. Worn by clinicians or examiners to provide a physical barrier against contamination between patient and examiner. Manufactured with surface treatment for smooth donning without powder. Used in clinical or medical settings. Provides protection through material barrier properties. Benefits patient and examiner by reducing risk of cross-contamination during examinations.
Clinical Evidence
No clinical data. Bench testing only. Performance verified against ASTM D6319 (physical properties), ASTM D5151 (watertightness), ASTM D6124 (powder content), and ASTM D3767 (dimensions). Biocompatibility assessed per ISO 10993-5, ISO 10993-10, and ISO 10993-11.
Technological Characteristics
Material: Nitrile rubber. Form factor: Ambidextrous, powder-free, non-sterile. Standards: ASTM D6319 (physical properties/dimensions), ASTM D5151 (watertightness), ASTM D6124 (powder residue). Surface treatment: Chemical process for smooth donning. Biocompatibility: ISO 10993 compliant.
Indications for Use
Indicated for use as a disposable, non-sterile, ambidextrous nitrile rubber glove worn by medical examiners on hands or fingers to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Blue Nitrile Examination Gloves Powder Free (K192333)
Related Devices
K222720 — Blue Nitrile Powder Free Patient Examination Glove, Non Sterile · Safeskin Retailing (Hk) Limited · Dec 3, 2022
K212916 — Black Nitrile Powder Free Patient Examination Glove, Non Sterile · Rimba Glove Sdn Bhd · Mar 24, 2022
K212802 — powder free nitrile examination glove - white, non-sterile · Hong Seng Gloves Sdn Bhd · Dec 2, 2021
K222713 — White Nitrile Powder Free Patient Examination Glove, Non Sterile · Safeskin Retailing (Hk) Limited · Dec 3, 2022
K163553 — Powder Free Nitrile Patient Examination Gloves, Blue · Hebei Titans Medical Supply Technology Group., Ltd. · May 22, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 1, 2021
Hong Seng Gloves Sdn Bhd Michael Van Der Woude U S Agent Emergo Global Representative LLC, 2500 Bee Cave Road Building 1, Suite 300 Austin, Texas 78746
Re: K212804
Trade/Device Name: Powder free nitrile examination glove- blue, non-sterile Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: August 12, 2021 Received: September 2, 2021
Dear Michael Van Der Woude:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Clarence W. Murray III, Ph D Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212804
Device Name POWDER FREE NITRILE EXAMINATION GLOVE - BLUE, NON STERILE
Indications for Use (Describe)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
# K212804
### 1.0 Submitter:
| Name: | Ho Chia Yao |
|------------|------------------------------------------------------------------------------------------------------------------------|
| Address: | Hong Seng Gloves Sdn Bhd<br>Lot 97, Jalan 10, Kawasan Perusahaan Bakar Arang, 08000,<br>Sungai Petani, Kedah, Malaysia |
| Phone No.: | +604-4211555 |
| Fax No.: | +604-4211555 |
Date of Summary Prepared: 20th July 2021 (REVISED DATE: 30th NOVEMBER 2021)
### 2.0 Identification of the subject device:
| Trade Name | : Powder Free Nitrile Examination Glove - Blue, Non-Sterile |
|----------------------|-------------------------------------------------------------|
| Common Name | : Patient Examination Gloves |
| Classification Name: | Patient Examination Gloves Device |
| Classification | : 1 |
| Regulation Number | : 21 CFR 880.6250 |
| Product Code | : LZA. |
#### 3.0 Predicate Device:
### K192333
Blue Nitrile Examination Gloves Powder Free Company: JR Engineering & Medical Technologies (M) SDN. BHD.
#### 4.0 Description of The Device:
Powder Free Nitrile Examination Glove - Blue, Non-Sterile meet all requirements of ASTM standard D6319 and FDA 21 CFR 880.6250.
The powder free nitrile examination glove is manufactured from nitrile rubber. Inner surface of gloves undergoes surface treatment process to produce a smooth surface that assists the user in donning the gloves with ease without using any lubricant such as powder on the glove surface. The glove is ambidextrous, i.e., can be worn on right hand or left hand.
#### 5.0 Intended Use/Indication for use:
A patient examination glove is a disposable device made of nitrile rubberintended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
### 6.0 Comparison of the Technological Characteristics of the Device:
The Powder Free Nitrile Examination Glove - Blue, Non-Sterile are summarized with the following technological characteristics compared to ASTM D6319 or equivalent standards as shown in Table 1.
{4}------------------------------------------------
## Table 1
| CHARACTERISTICS<br>STANDARDS | | DEVICE PERFORMANCE | | COMPARISON<br>ANALYSIS |
|----------------------------------------------------------------------------------|--------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| | | PREDICATE<br>BLUE | CURRENT<br>BLUE | |
| 510(k) Number | - | K192333 | K212804 | |
| Manufacturer(s) | - | JR Engineering & Medical<br>Technologies (M) SDN. BHD. | Hong Seng Gloves Sdn Bhd | Same |
| Intended<br>use/Indications for Use | - | JR MEDIC Blue Nitrile Examination<br>Gloves Powder Free is disposable<br>devices intended for medical purpose<br>that are worn on the examiner's<br>hand to prevent contamination<br>between patient and examiner. | A patient examination glove is a<br>disposable device made of nitrile<br>rubber intended for medical<br>purposes that is worn on the<br>examiner's hand or finger to prevent<br>contamination between patient and<br>examiner. | Same |
| Material | ASTM D6319 | Nitrile | Nitrile | Same |
| Color | - | Blue | Blue | Same |
| Texture | - | Finger texture | Finger Texture | Same |
| Physical Properties<br>Before Aging<br>Tensile Strength:<br>Ultimate Elongation: | ASTM D6319 | 25.6Mpa<br>868% | 28.6Mpa<br>579% | Different but<br>within the ASTM<br>standard |
| After Aging<br>Tensile Strength:<br>Ultimate Elongation: | ASTM D6319 | 22.0Mpa<br>828% | 33.1Mpa<br>535% | Different but<br>within the ASTM<br>standard |
| Thickness:<br>- Finger<br>- Palm | ASTM D6319 | 0.22mm<br>0.20mm | 0.10mm<br>0.07mm | Different but<br>within the ASTM<br>standard |
| Powder Free | ASTM D6124 | 0.21 mg/glove | 0.26 mg/glove | Different but<br>within the ASTM<br>standard |
| Watertight (1000ml) | ASTM<br>D5151:2019 | Gloves passes AQL 1.5 | Gloves passed AQL 1.5 | Same |
| Size | Medical Glove<br>Guidance Manual<br>– Labeling | Extra Small Small Medium Large<br>Extra Large | Extra Small Small Medium Large<br>Extra Large | Same |
| Single use | Medical Glove<br>Guidance Manual<br>– Labeling | Single Use | Single Use | Same |
| | Primary Skin<br>Irritation<br>– ISO 10993-<br>10:2010<br>(E) | Under the condition of study not an<br>irritant | The test material did not cause an<br>irritant response. The Primary<br>Irritant Response Category is<br>deemed `Negligible' | Similar |
| | Dermal<br>Sensitization- ISO<br>10993-10: 2010<br>(E) | Under the conditions of the study not<br>a sensitizer | The test material did not produce a<br>skin sensitization effect in the guinea<br>pigs. | Similar |
| Biocompatibility | Cytotoxicity -<br>MEM Elution, ISO<br>10993-5:<br>2009 (E) | Under the conditions of the study,<br>cytotoxic. Additional testing was<br>performed to determine if this was a<br>systemic toxicity concern. | The test material demonstrated a<br>cytotoxic effect under the condition<br>of this study. Additional test i.e.<br>Acute Systemic Toxicity was<br>tested. | Similar |
| | Acute Systemic<br>Toxicity, ISO<br>10993-<br>11:2017 (E) | Under the conditions of study, the<br>device extracts do not pose a<br>systemic toxicity concern | The test item did not induce any<br>systemic toxicity in Swiss albino<br>mice. | Similar |
{5}------------------------------------------------
{6}------------------------------------------------
There are no significant differences between the two products and they are the same or similar in terms of intended use, materials design, physical properties, thickness and biocompatibility test.
### 7.0 Summary of Non-Clinical Testing
The performance test data of the non-clinical test for this powder free nitrile examination glove is summarized as per below.
| Test<br>Method | Standard | Purpose of Testing | | Acceptance Criteria | | Results | | Status |
|------------------------|------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|---------------------|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|--------|
| | | | | Before<br>aging | After<br>aging | Before<br>aging | After aging | |
| Physical<br>Properties | ASTM D412<br>(Standard Test<br>Method for Vulcanized<br>Rubber and<br>Thermoplastic<br>Elastomers-Tension) | To evaluate the<br>tensile (tension)<br>properties of glove. | Tensile<br>strength | Min<br>14.0<br>MPa | Min 14.0<br>MPa | 28.6Mpa | 33.1Mpa | Pass |
| | | | Ultimate<br>elongation | Min<br>500% | Min<br>400% | 579% | 535% | Pass |
| Watertight | ASTM D5151<br>(Standard Test<br>Method for<br>Detection of Holes in<br>Medical Gloves) | To detect holes that<br>leak water and<br>thereby compromise<br>the usefulness of the<br>glove. | Sample size: 500 pcs Inspection<br>level: G1<br>AQL: 1.5, Acceptance No. 10 | | Less than 2 mg per glove | The batch size for this sampling<br>is 150,001 to 500,000. Hence,<br>according to the single<br>sampling plan GI, the sample<br>to be drawn is under code M<br>equivalent to 315 pieces with<br>accept 10 and reject 11 to be<br>accepted under AQL<br>1.5. 500 pcs were tested, 1<br>piece was found with leaks.<br>Hence it falls within the<br>acceptance criteria. | | Pass |
| | | | | | | Sample size : 5 pcs | Requirement : <2mg/glove<br>Result :0.26mg/glove | Pass |
{7}------------------------------------------------
| Test<br>Method | Standard | Purpose of<br>Testing | Glove<br>Size | Acceptance Criteria | Results | Status | |
|----------------|------------------------------------------------------------------------------------|-----------------------------------------------------------|---------------|-------------------------------------------------------|---------------------------------|--------------------|------|
| Dimension | ASTM D3767<br>Standard<br>Practice for<br>Rubber--<br>Measurement<br>of Dimensions | To measure the<br>length, width and<br>thickness of glove | X-Small | Length<br>Min 240 mm | Length<br>249 mm | Pass | |
| | | | | Width<br>$70 \pm 10$ mm | Width<br>78.0 mm | Pass | |
| | | | | Thickness<br>Finger - min 0.05mm<br>Palm - min 0.05mm | Thickness<br>0.10 mm<br>0.07 mm | Pass | |
| | | | Small | Length<br>Min 240 mm | Length<br>249 mm | Pass | |
| | | | | Width<br>$80 \pm 10$ mm | Width<br>88.0 mm | Pass | |
| | | | | Thickness<br>Finger - min 0.05mm<br>Palm - min 0.05mm | Thickness<br>0.10 mm<br>0.07 mm | Pass | |
| | | | Medium | Length<br>Min 240 mm | Length<br>249 mm | Pass | |
| | | | | Width<br>$95 \pm 10$ mm | Width<br>98.0 mm | Pass | |
| | | | | Thickness<br>Finger - min 0.05mm<br>Palm - min 0.05mm | Thickness<br>0.10 mm<br>0.07 mm | Pass | |
| | | | Large | Length<br>Min 240 mm | Length<br>249 mm | Pass | |
| | | | | Width<br>$110 \pm 10$ mm | Width<br>107 mm | Pass | |
| | | | Thickness | Finger – min 0.05mm<br>Palm – min 0.05mm | Thickness | 0.10 mm<br>0.07 mm | Pass |
| | | | Length | Min 240 mm | Length | 249 mm | Pass |
| | | X-Large | Width | $120 \pm 10$ mm | Width | 118 mm | Pass |
| | | | Thickness | Finger - min 0.05mm<br>Palm - min 0.05mm | Thickness | 0.10 mm<br>0.07 mm | Pass |
{8}------------------------------------------------
### Summary of Clinical Testing: 8.0
No clinical study is included in this submission.
### 9.0 Conclusion
The conclusion drawn from the non-clinical tests demonstrate that the subject Powder Free Nitrile Examination Glove - Blue, Non-Sterile is as safe, as effective, and performs as well as or better than the legally marketed predicate devices K19233.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.