Xingyu Nitrile Powder Free Patient Examination Gloves, White color

K171615 · Shandong Xingyu Gloves Co., Ltd. · LZA · Jul 31, 2017 · General Hospital

Device Facts

Record IDK171615
Device NameXingyu Nitrile Powder Free Patient Examination Gloves, White color
ApplicantShandong Xingyu Gloves Co., Ltd.
Product CodeLZA · General Hospital
Decision DateJul 31, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Intended Use

Xingyu Powder Free Nitrile Patient Examination Gloves, White Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between the patient and examiner

Device Story

Nitrile examination glove; acts as physical barrier to prevent contamination between patient and examiner. Device worn on hand/finger; conforms to hand via tensile properties. Used in medical settings; intended for single-patient use. Provides protection during medical procedures. No clinical data required; substantial equivalence established via bench testing against ASTM standards.

Clinical Evidence

No clinical data provided. Substantial equivalence demonstrated through bench testing, including dimension, tensile strength, elongation, water leak (pinhole) testing, powder residual, and biocompatibility (skin irritation and sensitization) per ISO 10993-10.

Technological Characteristics

Nitrile rubber material; non-sterile; powder-free. Dimensions and physical properties comply with ASTM D6319-10 (Reapproved 2015). Freedom from pinholes tested per ASTM D5151-06. Biocompatibility tested per ISO 10993-10. No software or electronic components.

Indications for Use

Indicated for use as a non-sterile, disposable, patient examination glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ July 31, 2017 Public Health Service Image /page/0/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three faces in profile, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Shandong Xingyu Gloves Co., Ltd % Chu Xiaoan Official Correspondent Beijing Easylink CO., LTD Rm. F302 Bldg., 41. Jing Cheng Ya Ju. Courtyard 6 of Southern Dou Ge Zhuang, Beijing, 100121, China Re: K171615 Trade/Device Name: Xingyu Nitrile Powder Free Patient Examination Gloves, White color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA Dated: July 9, 2017 Received: July 19, 2017 Dear Chu Xiaoan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply {1}------------------------------------------------ with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Mark S. Fellman -S for Lori Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) K171615 Device Name Xingyu Powder Free Nitrile Patient Examination Gloves, White Color Indications for Use (Describe) Xingyu Powder Free Nittile Patient Examination Gloves, White Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between the patient and examiner Type of Use (Select one or both, as applicable) ] Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(K) Summary "The assigned 510(k) number is: K171615 " ## Premarket Notification [510(k)] Summary #### 1.0 Submitter: | Submitter's name : | Shandong Xingyu Gloves Co.,Ltd | |-------------------------|-----------------------------------------------------------------------------------------| | Submitter's address : | No.2158, Yaoqian Road,Economic<br>Development Zone, Gaomi<br>City,Shandong,261502,China | | Phone number : | 0086-536-2588123 | | Fax number : | 0086-536-2586328 | | Name of contact person: | Xu Qiang | | Date of preparation : | 2017-07-07 | #### 2.0 Name of the Device | Device Name: | Powder Free Nitrile Patient Examination<br>Gloves,White Color | |-------------------------|----------------------------------------------------------------------| | Proprietary/Trade name: | Xingyu Powder Free Nitrile Patient<br>Examination Gloves,White Color | | Common Name: | Exam gloves | | Classification Name: | Patient examination glove | | Device Classification: | I | | Regulation Number: | 21 CFR 880.6250 | | Panel: | General Hospital (80) | | Product Code: | LZA | ## 3.0 Predicate device | Device Name: | Powder Free Nitrile Patient Examination<br>Gloves, White Color | |----------------|----------------------------------------------------------------| | Company name: | JiangSu DongLing Plastic & Rubber Co., Ltd. | | 510(K) Number: | K110248 | ## 4.0 Device Description: #### 4.1 How the device functions: Nitrile films form a barrier to prevent contamination between patient and examiner. #### 4.2 Scientific concepts that form the basis for the device The nitrile rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure. {4}------------------------------------------------ #### 4.3 Physical and performance characteristics such as design, materials and physical properties: The Nitrile glove acts as a barrier to prevent contamination between patient and examiner. The glove is manufactured in accordance with the requirements of ASTM D6319-10(Reapproved 2015) and ASTM D5151-06(Reapproved 2015). ## 5.0 Indication for use: Xingyu Powder Free Nitrile Patient Examination Gloves, White Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. ## 6.0 Summary of the Technological Characteristics of the Device: Xingyu Powder Free Nitrile Patient Examination Gloves,White Color, non-sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard. | Characteristics | Standard | | |--------------------------|---------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------| | Dimension | ASTM standard D 6319-10(Reapproved 2015) | | | | Length | ≥230mm | | | Width | Small 70-90 mm | | | | Medium 85-105mm | | | | Large 100-120mm | | | | X large 110-130 mm | | | Thickness | Fingertip ≥0.05mm | | | | Palm ≥0.05mm | | Physical<br>Properties | ASTM standard D 6319-10(Reapproved 2015) | | | | Tensile strength(Before aging) | ≥14MPa | | | Tensile strength(After aging) | | | | Elongated rate(Before aging) ≥500%<br>Elongated rate(After aging) ≥400% | | | Freedom from<br>pinholes | • 21 CFR 800.20<br>• ASTM standard D<br>6319-10(Reapproved 2015).<br>Test method in accordance with<br>ASTM D5151-06(Reapproved 2015) | Passed Standard<br>Acceptance Criteria | | | | | | | | | | Powder Residual | ASTM standard D<br>6319-10(Reapproved 2015)<br>Test method in accordance with<br>D6124-06(Reaffirmation 2011) | Meets<br>0.1mg/glove | | | | | | | | | | Biocompatibility | Primary Skin Irritation in rabbits | Passes | | | ISO 10993-10: Third Edition<br>2010-08-01 | Not a Primary Skin<br>Irritation | | | Dermal sensitization in the guinea<br>pig ISO 10993-10: Third Edition<br>2010-08-01 | Passes<br>Not a Primary Skin<br>Irritation | ## 7.1 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data: The following bench testing was conducted for design elements and performance characteristics deemed appropriate to demonstrate equivalence to the predicate device. The Powder Free Nitrile Patient Examination Gloves,White Color made by Shandong {5}------------------------------------------------ Xingyu Gloves Co.,Ltd met the predetermined acceptance criteria ensuring substantial equivalence to the predicate device. No new safety or performance issues were raised during testing. - . Dimension per ASTM D6319-10(Reapproved 2015), - . Tensile strength (Before aging/ After aging) and Elongated rate (Before aging/ After aging) per ASTM D6319-10(Reapproved 2015), - . Water leak test on pinhole per ASTM D6319-10(Reapproved 2015) and per 21 CFR 800.20, - . Powder Residual tests per ASTM D6319-10(Reapproved 2015) - Biocompatibility test per ISO 10993-10: Third Edition 2010-08-01. #### 8.0 Substantial Equivalent Based on Assessment of Clinical Performance Data: Clinical data was not needed to demonstrate that the subject glove is substantially equivalent to the predicate glove. So determination of substantial equivalence is not based on an assessment of clinical performance data. | Features<br>& | Predicate Device | | Subject Device | | Result<br>of | |-----------------------------------------------|-------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|--------------------------|------------------------|--------------| | Description | | | | | Comparison | | Company | | JiangSu DongLing Plastic | | Shandong Xingyu Gloves | -- | | | & Rubber Co., Ltd. | | Co.,Ltd | | | | 510(K) Number | K110248 | | K171615 | | | | Product name | | Powder Free Nitrile Patient | | Powder Free Nitrile | Same | | | | Examination Gloves, White | Patient Examination | | | | | Color | | | Gloves, White Color | | | Product Code | LZA | | LZA | | Same | | | | | | | | | Size | Small/ Medium/ | | Small/ Medium/ | | Same | | | Large/X large | | Large/X large | | | | Intended use | | Powder Free Nitrile Patient | Powder | Free<br>Nitrile | Same | | | | Examination Gloves, White | Patient | Examination | | | | | Color is a disposable device | Gloves,White Color is a | | | | | intended for medical | | device<br>disposable | | | | | purposes that is worn on the | | intended | for medical | | | | examiner's hand or finger to | | purposes that is worn on | | | | | prevent contamination | | the examiner's hand or | | | | | between patient and | | finger<br>prevent<br>to | | | | | examiner. | | contamination<br>between | | | | | | | patient and examiner. | | | | Device | Meets ASTM D6319-10 | | Meets ASTM | | Same | | Description and | | | D6319-10(Reapproved | | | | Specifications | | | 2015) | | | | Dimensions | Meets ASTM | | 232 mm min for all sizes | | Same | | --Length | D6319-10 | | | | | | ILS-2 | ≥230mm min | | | | | | AQL4.0 | | | | | | | Dimensions | Meets ASTM D6319-10 | | | | Same | | -- Width | Small | 70-90 mm | Small | 80-88 mm | | | ILS-2 | Medium | 85-105mm | Medium | 89-102 mm | | | AQL4.0 | Large | 100-120mm | Large | 104-118mm | | | | X large | 110-130 mm | X large | 115-129 mm | | | Dimensions<br>--Thickness | Meets ASTM D6319-10 | | | | Same | | IL S-2<br>AQL4.0 | Finger 0.05mm min.<br>Palm 0.05mm min. | Thickness (mm) min.<br>Finger 0.08<br>Palm 0.08 | | | | | Physical<br>Properties | Meets ASTM D D6319-10 | Aging | Before | After | Same | | IL S-2<br>AQL4.0 | Before aging/after aging<br>Tensile Strength≥ 14MPa | Elongation<br>(%) | 560-600 | 460-560 | | | | Before aging<br>Elongation ≥500%<br>After aging<br>Elongation ≥400% | | 18-25 | 17-21 | | | | | Tensile<br>Strength<br>(MPa) | | | | | Freedom from<br>Pinholes | Meets<br>• 21 CFR 800.20<br>• ASTM D6319-10 | Meets<br>• 21 CFR 800.20<br>• ASTM D<br>6319-10(Reapproved<br>2015) | | | Same | | Inspection<br>Level I AQL2.5 | Test method in accordance<br>with ASTM D5151-06<br>(Reapproved 2015) | Tested in accordance<br>with ASTM D5151<br>(Reapproved 2015)<br>with acceptable results | | | | | Residual<br>Powder | Meets ASTM<br>D 6124-06<br>(Reaffirmation 2011) | Meets ASTM<br>D 6124-06<br>(Reaffirmation 2011) | | | Same | | | below 2mg of residual<br>powder | Results generated values<br>below 2mg of residual<br>powder | | | | | Materials used<br>to fabricate the<br>devices | Nitrile | Nitrile | | | Same | | Dusting or<br>Donning<br>Powder: name | Surface Coating Agent | Surface Coating Agent | | | Same | | Standards | Meets<br>ASTM D5151-06<br>(Reapproved 2015)<br>ASTM D6319-10<br>ASTM D6124-06<br>(Reaffirmation 2011) | Meets<br>ASTM D5151-06<br>(Reapproved 2015)<br>ASTM D6319-10<br>ASTM D6124-06<br>(Reaffirmation 2011) | | | Same | | Single Patient<br>Use | Single Patient Use | Single Patient Use | | | Same | #### 9.0 Substantial Equivalence Comparison: {6}------------------------------------------------ {7}------------------------------------------------ | Biocompatibility | Under the conditions of this study, the test article<br>was a non- irritant or non-<br>sensitizer.<br><br>SKIN IRRITATION<br>DERMAL and<br>SENSITIZATION<br>STUDIES Meets ISO<br>10993-10:2002/Amd.1:20<br>06 | Under the conditions of this study, the test<br>article was a non-<br>irritant or non-<br>sensitizer.<br><br>SKIN IRRITATION<br>DERMAL and<br>SENSITIZATION<br>STUDIES Meets ISO<br>10993-10: Third Edition<br>2010-08-01. | Same | |------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------| | Labeling | -Powder Free<br>-Patient Examination Glove<br>-Single Use Only<br>- Manufactured For:<br>- Lot<br>-White color<br>-Non sterile | -Powder Free<br>-Patient<br>Examination Glove<br>-Single Use Only<br>- Manufactured For:<br>- Lot<br>-White color<br>-Non sterile | Same | #### 10.0 Conclusion: Based on the nonclinical tests data, it can be concluded that the Xingyu Powder Free Nitrile Patient Examination Gloves,White Color is as safe, as effective, and performs as well as the predicate device, Powder Free Nitrile Patient Examination Gloves, White Color, JiangSu DongLing Plastic & Rubber Co., Ltd., K110248.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...