K991897 · Catheter Innovations, Inc. · LJT · Oct 21, 1999 · General Hospital
Device Facts
Record ID
K991897
Device Name
CATHETER INNOVATIONS IMPLANTABLE PORT
Applicant
Catheter Innovations, Inc.
Product Code
LJT · General Hospital
Decision Date
Oct 21, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5965
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Catheter Innovations Implantable Port is designed for patients who require long-term access to the central venous system for administration of fluids including but not limited to hydration fluids, antibiotics, chemotherapy, analgesics, nutritional therapy, and blood products. It is also indicated for blood specimen withdrawal.
Device Story
Implantable port system for long-term central venous access; consists of titanium port body with single or dual independent reservoirs and radiopaque silicone catheter. Port implanted in subcutaneous pocket; accessed via non-coring needle through self-sealing silicone septum. Features integrated three-way valve within port-to-catheter adapter housing to control fluid flow; valve remains closed under normal venous pressure, opens with positive (infusion) or negative (aspiration) pressure. Used by healthcare professionals for repeatable infusion and blood withdrawal. Benefits include secure, long-term venous access for various therapies.
Clinical Evidence
Bench testing only. Performance characteristics compared to predicate devices (BardPort with Groshong and Hickman catheters) per FDA Guidance on 510(k) Submissions for Implanted Infusion Ports (October 1990). Results indicate performance equivalent or superior to predicates.
Technological Characteristics
Titanium port body; radiopaque silicone catheter; polysulfone adapter housing. Three-way valve mechanism located in adapter housing. Single or dual-lumen configurations. Implantable, non-powered, mechanical device.
Indications for Use
Indicated for patients requiring long-term central venous access for fluid administration (hydration, antibiotics, chemotherapy, analgesics, nutritional therapy, blood products) and blood specimen withdrawal.
Regulatory Classification
Identification
A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.
Predicate Devices
BardPort™ Implanted Ports with Groshong® catheters (K880571)
{1}------------------------------------------------
# OCT 21 1999
neter Innovations, Inc. Dual-Lumen Implantable Ports Premarket Notification
K99/897
## 510(k) SUMMARY Prepared May 24, 1999
### 1. Submitted Bv:
Catheter Innovations, Inc. 3598 West 1820 South Salt Lake City, UT 84104-4959 Tel: (801) 954-8444 Fax: (801) 954-8484
### 2. Contact Person:
Roger L. Richins V. P. Technology and Regulatory Affairs
### 3. Device Establishment Registration No.: 1723743 Owner/Operator No.: 9025151
#### 4. Device Identification:
| Trade Name: | Implantable Port |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Intravascular Implanted Port |
| Classification Name: | Port & Catheter, Implanted, Subcutaneous, Intravascular |
| Classification: | Unclassified |
| FDA Classification Advisory Committee: | 80 - General Hospital |
| Product Code: | LJT – The specific guidelines for Premarket Notification in the<br>LJT product classification are detailed in the "Guidance on 510(k)<br>Submissions for Implanted Infusion Ports" by the CDRH, ODE,<br>Division of Gastroenterology/Urology and General Use Devices<br>(October 1990)<br>Predicate Device(s): K880571 – Originally submitted as Catheter Technologies<br>Port Implantable Vascular Access System (Approved to market |
"BardPort™ Implanted Ports with Groshong® catheters" K873213 - Hickman® Subcutaneous Port currently marketed by
March 4, 1988) - Currently marketed by Bard Access Systems as
Bard Access Systems ﻴﺔ ﻣﺮ
### 6. Device Description:
5.
Catheter Innovations, Inc. Implantable Ports are constructed using a combination of traditional implanted port/catheter technology which is well known in the art (over fifteen years of successful use), Bard Access Systems BardPort™ Implanted Ports with Groshong® catheters-hereafter referred to as BardPorts™ (over eleven years in use) and its own previously approved 510(k) technology. The Catheter Innovations ports, like the BardPorts™ and other traditional ports, are implanted within the body and are used for access to the circulatory system for repeatable or periodic infusion of fluids and/or aspiration of blood. Like many currently marketed ports the Catheter Innovations ports have smooth, contoured titanium port bodies with either one (single-lumen) or two (duallumen) independent reservoirs.
{2}------------------------------------------------
# Catheter Innovations, Inc. Single and Dual-Lumen Implantable Ports Premarket Notification -
# 510(k) SUMMARY (Continued)
### Device Description Continued: 6.
A radiopaque silicone catheter with one or two lumens is provided with each port body. The catheter may come attached to the port body by the manufacturer or un-attached depending on user preference. If the catheter comes attached to the port, at the time of use the health practitioner measures and cuts off the distal end of the catheter to the desired length. If the catheter is not attached to the port when it is purchased, the catheter is placed, then cut off at the proximal end to the desired length and attached to the port by the practitioner. The port is then placed in a small pocket made just under the skin. The port bodies are designed to facilitate secure seating and anchoring in the port pocket.
The port bodies are accessed by inserting a non-coring needle through the skin and into the self-sealing silicone septum which caps these reservoirs.
Both BardPorts™ and Catheter Innovations ports contain a thin silicone membrane within their construction. In both cases these membranes have a small slit or cut through them that acts as a valve. In the BardPorts™ this valve is located at the distal end of the attached catheter. In the Catheter Innovations ports the valve is located in a small plastic polysulfone housing which comes attached to the port body. The catheter is connected to this port through this housing. In both examples the valve is used to control the flow of fluids both into and out of the blood stream. The valve remains closed when the ports are not in use and when subjected to normal central venous pressures. When positive fluid pressure is applied through the reservoir, the valve opens, allowing infusion through the catheter. When negative pressure (aspiration) is applied, the valve opens, allowing for withdrawal of blood into a syringe.
### 7. Statement of Indications for Use:
The Catheter Innovations ports are designed for patients who require long-term access to the central venous system for administration of fluids including but not limited to hydration fluids, antibiotics, chemotherapy, analqesics, nutritional therapy, and blood products. Thev are also indicated for blood specimen withdrawal.
The Catheter Innovations ports, the BardPorts™ (both with and without valve), and the numerous other approved intravascular port and catheter combinations have the same indications for use: long-term access to the central venous system for administration of fluids and for blood specimen withdrawal.
### 8. Summary of Technological Characteristics of Device in Relation to Predicate Device(s):
Physical characteristics of Catheter Innovations ports, compared to BardPorts™ with Groshong® and Hickman® catheters, demonstrate that the physical characteristics of the Catheter Innovations ports are identical to the predicate devices for all items listed in the "FDA Guidance on 510(k) Submissions for Implanted Infusion Ports." (October 1990)
Physically, the Catheter Innovations ports differ from the BardPorts™ with Groshong® catheters, in that the valve in the Catheter Innovations port is located in the catheter-toport adapter housing instead of close to the distal tip of the catheter as in the Groshong® catheter.
{3}------------------------------------------------
# Catheter Innovations, Inc. Single and Dual-Lumen Implantable Ports Premarket Notification ~
# 510(k) SUMMARY (Continued)
### 8. Summary of Technological Characteristics of Device in Relation to Predicate Device(s) Continued:
Additionally, in the BardPorts™ with Groshong® catheters are supplied in shorter lengths. This length difference is only apparent when the product is removed from the package. In actual use, the catheters are trimmed to length so that their distal tips are located in the distal 1/3 of the Superior Vena Cava (SVC). Thus, the difference in length of catheters, as supplied, is insignificant in relation to end-use of the product.
Both BardPorts™ with Hickman® catheters and Catheter Innovations ports are openended, and the distal end is trimmed by the health professional in order to obtain the desired length. Because the Groshong® catheter is close-ended and removal of a distal segment would remove the valve, these catheters are trimmed at the proximal end and subsequently fitted over their port-connecting adapter. Placement and trimming of Catheter Innovations catheters, because they are open-ended, is identical to that of BardPort™ with Hickman® catheters.
Catheter Innovations three-way valve is located in the port-to-catheter adapter housing at the proximal end of the catheter, immediately adjacent to the port body, while the Groshong® three-way valve is located in the distal tip of the catheter in the distal third of the SVC, and directly in the fluid stream path. There are no new safety or effectiveness issues posed by the difference in location of these valves within the body.
Based on these physical characteristic comparisons, we consider the Catheter Innovations Implantable Port substantially equivalent to the BardPort™ with Groshong® and Hickman® catheter predicate device ports.
### 9. Assessment of Performance Data:
Performance test results indicate Catheter Innovations ports are substantially equivalent to predicate device port/catheter combinations for all performance characteristics itemized in the "FDA Guidance on 510(k) Submissions for Implanted Infusion Ports." (October 1990)
Catheter Innovations performance characteristics related to port and valve function were compared to the predicate BardPort™ with Groshong® catheters. Comparison of all other Catheter Innovations port catheter performance characteristics were made to the range of values found for the performance of the predicate BardPort™ with Hickman® and Groshong® catheters. These comparisons indicate that the Catheter Innovations port with catheters performed equivalent to, or better than, the predicate devices for all performance characteristics studied.
## Conclusion:
Data indicates that the performance characteristics of the Catheter Innovations Port with valved catheters are equivalent or superior to the currently marketed BardPort™ Implanted Ports with Groshong® and Hickman® catheters while posing no new safety or efficacy issues.
{4}------------------------------------------------
# Catheter Innovations, Inc. Single and Dual-Lumen Implantable Ports Premarket Notification
# 510(k) SUMMARY (Continued)
### 10. Risk Analysis:
A risk analysis was completed on the products covered by this submission. This analysis showed that these ports have the same inherent risks of use as predicate devices and all other central venous access port/catheter products. Their use should be carefully considered before placement. Their placement and care should only be performed by persons knowledgeable of the risks involved and qualified in the procedures.
, k
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines suggesting movement or connection. The profiles are arranged vertically, one above the other, creating a sense of depth. Encircling the image is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 21 1999
Mr. Roger L. Richins V.P. Technology and Regulatory Affairs Catheter Innovations Inc. 3598 West 1820 South Salt Lake City, Utah 84104-4959
Re : K991897 Trade Name: Catheter Innovations Implantable Port Regulatory Class: Unclassified Product Code: LJT Dated: August 17, 1999 August 18, 1999 Received:
Dear Mr. Richins:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ಕ್ಕ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any
{6}------------------------------------------------
Page 2 - Mr. Richins
obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# 510(k) Number (if known):
Catheter Innovations Impantable Ports Device Name:
Indications For Use:
The Catheter Innovations Implantable Port is designed for patients who require long-term access to the central venous system for administration of fluids including but not limited to hydration fluids, antibiotics, chemotherapy, analgesics, nutritional therapy, and blood products. It is also indicated for blood specimen withdrawal.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Patricia Cuscente
(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 5 10(k) Number
Image /page/6/Picture/8 description: The image shows the words "Prescription Use" in bold font. Below this text is the text "(Per 21 CFR 801.109)". There is a line under the words "Prescription Use" with a check mark above the line.
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.