THERAPORT LOW-PROFILE VASCULAR ACCESS SYSTEM MODELS 1011 AND 1012
K974475 · Biocontrol Technology, Inc. · LJT · Mar 2, 1998 · General Hospital
Device Facts
Record ID
K974475
Device Name
THERAPORT LOW-PROFILE VASCULAR ACCESS SYSTEM MODELS 1011 AND 1012
Applicant
Biocontrol Technology, Inc.
Product Code
LJT · General Hospital
Decision Date
Mar 2, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5965
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The theraPORT® Low-Profile Vascular Access System is intended for use with patients that require repeated venous access for injection or infusion therapy and/or venous blood sampling. It is indicated for either traditional chest placement or placement in the upper arm, and is appropriate for use in pediatric patients.
Device Story
Totally implantable venous access system; consists of detached catheter and port. Used for repeated venous access; injection; infusion therapy; blood sampling. Implanted in chest or upper arm; used in clinical settings. Provides reliable access point for long-term therapy; reduces need for repeated venipuncture. Physician-operated; provides access for delivery of medications or withdrawal of blood samples.
Clinical Evidence
No clinical data. Bench testing only, conducted per FDA 1990 guidance for implanted infusion ports, including connection strength, septum durability, leakage, and clearance tests.
Technological Characteristics
Totally implantable venous access system; low-profile port and detached catheter. Materials, processing, and sterilization identical to predicate devices. Mechanical device; no energy source or software.
Indications for Use
Indicated for patients requiring repeated venous access for injection, infusion therapy, or blood sampling. Suitable for pediatric patients and small adults. Placement sites include chest or upper arm.
Regulatory Classification
Identification
A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.
theraPORT® Vascular Access System (Models 1001 and 1002)
Submission Summary (Full Text)
{0}
510(k) Summary for the
K974475
therePORT® Low-Profile Vascular Access System
MAR - 2 1998
## GENERAL INFORMATION:
**Common/Usual Names:**
Implanted Subcutaneous Intravascular Catheter; Implantable Vascular Access System; Implanted Infusion Port.
**Proprietary Name:**
*theraPORT®* Low-Profile Vascular Access System
**Applicant:**
Biocontrol Technology, Inc.
300 Indian Springs Road
Indiana, PA 15701
(412)349-1811
**Equivalence Device Comparison:** Bard Access Systems, BardPort® Titanium Low-Profile, Product Code 0606200; Cook Vascular, Inc., Vital-Port® Model 6113; SIMS Deltec, Inc., Port-A-Cath® Low Profile Unassembled, Kit No. 21-4034; Biocontrol Technology, Inc., *theraPORT®* Vascular Access System, Models 1001 and 1002.
## DEVICE DESCRIPTION:
The *theraPORT®* Low-Profile Vascular Access System is a totally implantable venous access system consisting of a detached catheter and port.
## INTENDED USE:
The *theraPORT®* Low-Profile Vascular Access System is intended for use with patients that require repeated venous access for injection or infusion therapy and/or venous blood sampling. It is indicated for either traditional chest placement or placement in the upper arm, and is appropriate for use in pediatric patients.
{1}
# ALTERNATIVE DEVICES:
Alternative devices to the *theraPORT®* Low-Profile Vascular Access System are other commercially available implantable vascular venous access systems such as the BardPort® Titanium Low-Profile, Product Code 0606200, by Bard Access Systems; the Vital-Port® Model 6113, by Cook Vascular, Inc.; the Port-A-Cath® Low Profile Unassembled, Kit No. 21-4034, by SIMS Deltec, Inc.; and the *theraPORT®* Vascular Access System, Model 1001, by Biocontrol Technology, Inc.
# POTENTIAL ADVERSE EFFECTS:
The following adverse effects, which are normally associated with the insertion or use of any implanted device or indwelling catheter, may occur when using the *theraPORT®* Low-Profile Vascular Access System: air embolism, bacteremia, catheter disconnection, catheter fragmentation, cardiac arrhythmia, cardiac puncture, cardiac tamponade, catheter occlusion, catheter rupture, catheter shearing, catheter/port erosion through blood vessel/skin, catheter/port migration, drug extravasation, hematoma, hemothorax, implant rejection, infection, laceration or puncture of vessels, pneumothorax, sepsis, thromboembolism, thrombophlebitis, thrombosis.
# SUMMARY OF STUDIES:
## Performance Testing:
Performance testing of the *theraPORT®* Low-Profile Vascular Access System was conducted in accordance with the “Guidance on 510(k) Submissions for Implanted Infusion Ports,” Center for Devices and Radiological Health, Office of Device Evaluation, Division of Gastroenterology/Urology and General Use Devices, Food and Drug Administration, October 1990. Catheter-to-port connection strength tests, septum puncture durability tests, port leakage integrity tests and port/catheter clearance tests were all performed.
Biocompatibility testing was not conducted since all materials, their processing, and their sterilization are identical to substantially equivalent devices.
## Clinical Studies:
Clinical studies were not conducted as they were determined to be not necessary due to the similarity in design, performance, materials, function and intended use of the *theraPORT®* Low-Profile Vascular Access System to other commercially available systems.
{2}
# CONCLUSIONS DETERMINED FROM TESTS:
The above described studies demonstrate that the *theraPORT®* Low-Profile Vascular Access System functions properly and is substantially equivalent to the aforementioned commercially available predicate devices. Therefore, the *theraPORT®* Low-Profile Vascular Access System is determined to be safe and effective for its intended use.
theraPORT® is a registered trademark of Biocontrol Technology, Inc.
BardPort® is a registered trademark of C. R. Bard., Inc.
Port-A-Cath® is a registered trademark of SIMS Deltec, Inc.
Vital-Port® is a registered trademark of Cook Pacemaker Corp.
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Patrick J. Cooper, Ph.D.
Manager, Sales and Marketing
Biocontrol Technology, Incorporated
300 Indian Springs Road
Indiana, Pennsylvania 15701
MAR - 2 1998
Re: K974475
Trade Name: theraPORT® Low-Profile Vascular Access System, Models 1011 and 1012
Regulatory Class: Unclassified
Product Code: LJT
Dated: February 5, 1998
Received: February 6, 1998
Dear Dr. Cooper:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the theraPORT® Low-Profile Vascular Access System have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of
{4}
Page 2 - Dr. Cooper
the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the *Federal Register*. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Director
Division of Dental, Infection Control, and General Hospital Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{5}
Page 1 of 1
510(k) Number: K974475
Device Name: *theraPORT Low-Profile Vascular Access System*
Indications for Use:
The *theraPORT* Low-Profile Vascular Access System is a totally implantable venous access system intended for use with patients that require repeated venous access for injection or infusion therapy and/or venous blood sampling. Because of its low-profile design, it is indicated for either traditional chest placement or placement in the upper arm, and is appropriate for use in pediatric patients and small adults.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
*Patricia Cucente*
(Division of Dental, Infection Control, and General Hospital Devices
510(k) Number *K974475*
Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over-the-Counter Use ☐ (Optional Format 1-2-96)
PROPRIETARY/CONFIDENTIAL
BIOCONTROL TECHNOLOGY INC
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.