K252636 · The Surgical Company International B.V. · LGZ · Nov 18, 2025 · General Hospital
Device Facts
Record ID
K252636
Device Name
Fluido Compact System
Applicant
The Surgical Company International B.V.
Product Code
LGZ · General Hospital
Decision Date
Nov 18, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The device can be used with the following products: • Plasma • Packed red blood cells • Crystalloid intra venous (IV) fluids • Synthetic Colloid intra venous (IV) fluids The system is indicated to be used in a healthcare environment to help prevent hypothermia. The devices can be used in the full clinical environment. The Fluido Compact System can be used in adult and pediatric patients (weighing 3.5 kg or more) who require fluid warming prior to the administration of blood products or IV fluids to help prevent hypothermia. The system is intended to be used by healthcare professionals, e.g., nurses, medical specialists, and doctors.
Device Story
Fluido Compact System is a blood and fluid warming system for low-to-moderate flow applications (5–100 ml/min). It consists of a reusable warming module, a control module, and a sterile, single-use disposable set (plastic housing with biocompatible coated aluminum heater plate). The system uses conduction heating; in-line sensors monitor fluid temperature to calculate required energy for safe warming to a 39°C set point. Operated by healthcare professionals (nurses, doctors) in clinical/hospital environments, the device warms infusates before patient administration. It includes closed-loop temperature control, over-temperature cut-off (49°C), and independent heater monitoring. The output is warmed fluid, which helps prevent hypothermia in patients. The system is used on physician order.
Clinical Evidence
No clinical data. Evidence consists of bench testing, toxicological risk assessments (aluminum leaching, EO/ECH residuals), and human factors analysis to support the pediatric indication (≥ 3.5 kg). Performance specifications (flow rates, temperature output) remain identical to the predicate.
Technological Characteristics
Conduction heating technology; reusable controller and warming module; sterile disposable set with biocompatible coated aluminum heater plate. Closed-loop temperature control; over-temperature cut-off at 49°C (ASTM F2172-02). Power: 100-240V AC. Dimensions: Warmer 16.5x7.5x5.0 cm; Controller 28.5x12.0x19.5 cm. Sterilization: Ethylene oxide. Biocompatibility: ISO 10993. Standards: IEC 60601-1, IEC 60601-1-2.
Indications for Use
Indicated for adult and pediatric patients (≥ 3.5 kg) requiring fluid warming of plasma, packed red blood cells, crystalloid IV fluids, or synthetic colloid IV fluids prior to administration to prevent hypothermia in healthcare environments.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
Fluido® Compact Blood and Fluid Warming System (K211618)
Related Devices
K211618 — Fluido Compact Blood and Fluid Warming System · The Surgical Company International B.V. · Jun 5, 2023
K112902 — ENFLOW IV FLUID WARMER · Vital Signs, Inc. · Mar 14, 2012
K012832 — COMFORTMATE FLUID WARMING SYSTEM, MODEL # FW-200; COMFORTMATE DISPORSABLE DISPOSABLE WARMING SET, MODEL #FWS-200 · Nxstage Medical, Inc. · Oct 24, 2001
K171215 — QiF Blood and Fluid Warmer · Quality IN Flow , Ltd. · Sep 27, 2017
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 18, 2025
The Surgical Company International B.V.
Gianpaolo Soares
RA Project Manager
Paalbergweg 3
Amsterdam, North Holland 1105 AG
Netherlands
Re: K252636
Trade/Device Name: Fluido Compact System
Regulation Number: 21 CFR 880.5725
Regulation Name: Infusion Pump
Regulatory Class: Class II
Product Code: LGZ
Dated: August 20, 2025
Received: August 20, 2025
Dear Gianpaolo Soares:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252636 - Gianpaolo Soares
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K252636 - Gianpaolo Soares
Page 3
Sincerely,
Jake K. Lindstrom -S
Jake Lindstrom, Ph.D.
Assistant Director, Infusion Devices
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K252636 | ? |
| Please provide the device trade name(s). | | ? |
| Fluido Compact System | | |
| Please provide your Indications for Use below. | | ? |
| The device can be used with the following products:
• Plasma
• Packed red blood cells
• Crystalloid intra venous (IV) fluids
• Synthetic Colloid intra venous (IV) fluids
The system is indicated to be used in a healthcare environment to help prevent hypothermia. The devices
can be used in the full clinical environment. The Fluido Compact System can be used in adult and pediatric
patients
(weighing 3.5 kg or more) who require fluid warming prior to the administration of blood products or IV fluids
to help prevent hypothermia. The system is intended to be used by healthcare professionals, e.g., nurses,
medical specialists, and doctors. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
TSCLIFE
Innovation with purpose
# 510(k) Summary
Traditional 510(k) Premarket Notification
Fluido® Compact System
November 2025
The Surgical Company International B.V. | Paalbergweg 3 | 1105 AG Amsterdam | The Netherlands | T +31 20 7900 2100 | letsconnect@tsc-life.com
www.tsc-life.com | CC no. 31029240 | VAT no. NL 8111.49.948.B03 | BANK: ING Commercial Banking | Bijlmerplein 888 | 1102 MG Amsterdam | The Netherlands
IBAN NL15 INGB 0653 7522 02 | BIC: INGBNL2A
Page | 1
{5}
TSC LIFE
Innovation with purpose
Submitter Name and Address
The Surgical Company International B.V., doing business as TSC Life
Paalbergweg 3
Amsterdam, 1105 AG
The Netherlands
Contact Person
Name: Gianpaolo Soares
Title: Regulatory Affairs Project Manager, TSC Life
Company Telephone: +31 20 790 2100
Email: fda.tsclife-us@tsc-life.com
Device Information
Name of Device: Fluido® Compact System
Common or Usual Name: Infusion Pump
Classification Name: Infusion Pump, 21 CFR 880.5725
Regulatory Class: II
Product Code: LGZ
Predicate Information
Name of Predicate: Fluido® Compact Blood and Fluid Warming System
510(k) Number: K211618
This predicate has not been subject to a design-related recall.
No reference devices were used in this submission.
Device Description
The Fluido Compact System is a Blood and Fluid Warming System. The system is suitable for low to moderate flow applications. For high flow applications (higher than 100 ml/min for more than 5 minutes) the use of a high flow system is recommended.
The Fluido Compact System uses conduction heating technology to warm blood and fluids. Based on in-line sensors, the system calculates the energy required to safely warm the infusates.
The Fluido Compact System can be used with patients who require fluid warming prior to the administration of blood products or IV fluids, to help prevent hypothermia. The system is intended to be used in a healthcare environment by healthcare professionals, e.g. nurses, medical specialists, doctors. Only medical professionals shall interact with the system. The devices must only be used on the order of a physician.
Indications for Use
The device can be used with the following products:
- Plasma
- Packed red blood cells
- Crystalloid intra venous (IV) fluids
- Synthetic Colloid intra venous (IV) fluids
The system is indicated to be used in a healthcare environment to help prevent hypothermia. The devices can be used in the full clinical environment. The Fluido Compact System can be used in adult and pediatric patients (weighing 3.5 kg or more) who require fluid warming prior to the administration of blood products or IV fluids to help prevent hypothermia. The system is intended to be used by healthcare professionals, e.g., nurses, medical specialists, and doctors.
The Surgical Company International B.V. | Paalbergweg 3 | 1105 AG Amsterdam | The Netherlands | T +31 20 7900 2100 | letsconnect@tsc-life.com
www.tsc-life.com | CC no. 31029240 | VAT no. NL 8111.49.948.B03 | BANK: ING Commercial Banking | Bijlmerplein 888 | 1102 MG Amsterdam | The Netherlands
IBAN NL15 INGB 0653 7522 02 | BIC: INGBNL2A
{6}
TSCLIFE
Innovation with purpose
Table 1: Substantial Equivalence Comparison
| | Subject Device
Fluido® Compact Blood and Fluid Warming | Predicate Device
Fluido® Compact Blood and Fluid Warming | Substantial Equivalence |
| --- | --- | --- | --- |
| 510(k) Number | Not yet assigned | K211618 | N/A |
| Manufacturer | The Surgical Company International B.V. | The Surgical Company International B.V. | Same |
| Trade name | Fluido® Compact System | Fluido® Compact Blood and Fluid Warming System | Same |
| CFR Regulation Number | 21 CFR 880.5725 | 21 CFR 880.5725 | Same |
| Product Code | LGZ | LGZ | Same |
| Classification | Class II | Class II | Same |
| Review Panel | General Hospital | General Hospital | Same |
| Intended Use | The Fluido Compact System can be used with patients who require fluid warming prior to the administration of blood products or IV fluids, to help prevent hypothermia. The system is intended to be used in a healthcare environment by healthcare professionals, e.g. nurses, medical specialists, doctors. Only medical professionals shall interact with the system. The devices must only be used on the order of a physician. | The Fluido® Compact System can be used with patients that need fluid warming prior to administration of blood products and IV fluids to help prevent hypothermia. The system is intended to be used in hospital by healthcare professionals, e.g., nurses, medical specialists, doctors. Only medical professionals shall interact with the system. The devices must only be used on the order of a physician. | Same, minor rewording to align intended use statement with use environment |
| Description of Device | The Fluido® Compact System consists of Fluido Compact Warming Module, Fluido Compact Control Module and Fluido Compact Standard Set & Fluido Compact Standard Set with Drip Chamber (disposable). The Fluido Compact system is designed to supply warm fluids at 39°C set point with flow rates from 5 to 100 ml/min. The sterile disposable set consist of a plastic housing and biocompatible coated aluminum heater plate. Heat from the warming module is transferred through the heat transfer part into the fluids that is circulating through the inner part of the cassette. The disposable is connected to the blood or fluid source (Infusate reservoir) and the patient line on the other end. The disposable has standard luer lock fittings on both ends. The Control Module serves as the power supply for the Warming module. | The Fluido® Compact System consists of Fluido Compact Warming Module, Fluido Compact Control Module and Fluido Compact Standard Set & Fluido Compact Standard Set with Drip Chamber (disposable). The Fluido Compact system is designed to supply warm fluids at 39°C set point with flow rates from 5 to 100 ml/min. The sterile disposable set consist of a plastic housing and biocompatible coated aluminum heater plate. Heat from the warming module is transferred through the heat transfer part into the fluids that is circulating through the inner part of the cassette. The disposable is connected to the blood or fluid source (Infusate reservoir) and the patient line on the other end. The disposable has standard luer | Same |
The Surgical Company International B.V. | Paalbergweg 3 | 1105 AG Amsterdam | The Netherlands | T +31 20 7900 2100 | letsconnect@tsc-life.com
www.tsc-life.com | CC no. 31029240 | VAT no. NL 8111.49.948.B03 | BANK: ING Commercial Banking | Bijlmerplein 888 | 1102 MG Amsterdam | The Netherlands
IBAN NL15 INGB 0653 7522 02 | BIC: INGBNL2A
{7}
TSC LIFE
Innovation with purpose
| | Subject Device
Fluido® Compact Blood and Fluid Warming | Predicate Device
Fluido® Compact Blood and Fluid Warming | Substantial Equivalence |
| --- | --- | --- | --- |
| | | lock fittings on both ends. The Control Module serves as the power supply for the Warming module. | |
| Patient Population | Adult and Pediatric (weighing 3.5 kg or more) | Adult | Different – subject device includes pediatric indication. Human factors, performance, biocompatibility, aluminum leaching, sterilization residues, and the risk file were thoroughly evaluated to confirm that none of these factors adversely affect the device. |
| User | Healthcare professionals e.g., nurses, medical specialists, doctors | Healthcare professionals e.g., nurses, medical specialists, doctors | Same |
| Use Environment | Hospital, clinical and field environments | Hospital, clinical and field environments | Same |
| Operating Principle | Reusable Conduction Warmer
Reusable Controller
Disposable sterile | Reusable Conduction Warmer
Reusable Controller
Disposable sterile | Same |
| Prescription Use/Over-The-Counter use | Prescription Use | Prescription Use | Same |
| Administered Fluids | IV Fluids, Blood Products | IV Fluids, Blood Products | Same |
| Fluid Path | Sterile Fluid Path | Sterile Fluid Path | Same |
| System Components | 1. Reusable Warmer
2. Reusable Controller
3. Single Use Disposable sterile set (Standard Set, Standard Set with drip chamber) | 1. Reusable Warmer
2. Reusable Controller
3. Single Use Disposable sterile set (Standard Set, Standard Set with drip chamber) | Same |
| Safety Features | • Closed-loop temperature control
• Over-temperature control and cut off (ASTM F2172-02)
• Independent shut-off of heater and notification
• Independent heater temperature monitoring circuit | • Closed-loop temperature control
• Over-temperature control and cut off (ASTM F2172-02)
• Independent shut-off of heater and notification
• Independent heater temperature monitoring circuit | Same |
| Warming technology | Conduction | Conduction | Same |
| Heat exchange | Inline disposable | Inline disposable | Same |
| Temperature control | Closed Loop temperature control | Closed Loop temperature control | Same |
| Fluid Temperature Output | 39°C ± 2°C | 39°C ± 2°C | Same |
The Surgical Company International B.V. | Paalbergweg 3 | 1105 AG Amsterdam | The Netherlands | T +31 20 7900 2100 | letsconnect@tsc-life.com
www.tsc-life.com | CC no. 31029240 | VAT no. NL 8111.49.948.B03 | BANK: ING Commercial Banking | Bijlmerplein 888 | 1102 MG Amsterdam | The Netherlands
IBAN NL15 INGB 0653 7522 02 | BIC: INGBNL2A
{8}
TSC LIFE
Innovation with purpose
| | Subject Device
Fluido® Compact Blood and Fluid Warming | Predicate Device
Fluido® Compact Blood and Fluid Warming | Substantial Equivalence |
| --- | --- | --- | --- |
| Safety cut off
Temperature | 49°C | 49°C | Same |
| Flow Rate Range | Standard Set: 5 – 100 ml/min | Standard Set: 5 – 100 ml/min | Same |
| Storage Conditions/
Specifications | Warmer and Controller:
-40°C – 50°C at 10% to 90% relative humidity
Disposable Set:
-20°C – 40°C at 10% to 90% relative humidity | Warmer and Controller:
-40°C – 50°C at 10% to 90% relative humidity
Disposable Set:
-20°C – 40°C at 10% to 90% relative humidity | Same |
| Power Source | AC power supply
100-240 V ~ 50/60Hz)
1.6A | AC power supply
100-240 V ~ 50/60Hz)
1.6A | Same |
| Dimensions | Warmer:
16.5 cm x 7.5 cm x 5.0 cm
Controller:
28.5 cm x 12.0 cm x 19.5 cm
Disposable Set:
14.6 cm x 3.5 cm x 1.1 cm | Warmer:
16.5 cm x 7.5 cm x 5.0 cm
Controller:
28.5 cm x 12.0 cm x 19.5 cm
Disposable Set:
14.6 cm x 3.5 cm x 1.1 cm | Same |
| Weight | Warmer (without disposable): 450g
Controller: 1.7kg
Disposable Set: 24g (cassette) | Warmer (without disposable):
450g Controller: 1.7kg
Disposable Set: 24g (cassette) | Same |
| Performance | • IEC 60601-1
• IEC 60601-1-2
• Acceptable aluminum leaching levels for intended population | • IEC 60601-1
• IEC 60601-1-2
• Acceptable aluminum leaching levels for intended population | Same |
| Sterilization | Disposable Set: Ethylene oxide sterilized | Disposable Set: Ethylene oxide sterilized | Same |
| Biocompatibility | ISO 10993 | ISO 10993 | Same |
The Surgical Company International B.V. | Paalbergweg 3 | 1105 AG Amsterdam | The Netherlands | T +31 20 7900 2100 | letsconnect@tsc-life.com www.tsc-life.com | CC no. 31029240 | VAT no. NL 8111.49.948.B03 | BANK: ING Commercial Banking | Bijlmerplein 888 | 1102 MG Amsterdam | The Netherlands IBAN NL15 INGB 0653 7522 02 | BIC: INGBNL2A
{9}
TSCLIFE
Innovation with purpose
## Performance Data
- Human factors – Determined no impact for added pediatric patient population.
- Performance – Same 5–100 mL/min flow and temperature specifications. No impact for added pediatric patient population.
## Safety Data
- Biocompatibility – Determined no new testing required for added pediatric patient population.
- Aluminum leaching – Toxicologic risk assessment shows aluminum leaching levels within safe limits for pediatric patient population ≥ 3.5 kg.
- Sterilization residuals (EO/ECH) – Toxicologic risk assessment shows EO/ECH levels within safe limits for pediatric patient population ≥ 3.5 kg.
- Cybersecurity – Data provided as required by FDA guidance “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions” issued June 27th, 2025.
- Risk assessment – No new risks related to added pediatric patient population identified.
## Conclusion
The sole purpose of this Traditional 510(k) submission is to expand the indicated patient population to include pediatric patients weighing 3.5 kg or more. The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The Fluido® Compact System is substantially equivalent to the Fluido® Compact Blood and Fluid Warming System cleared under K211618 with respect to the indications for use, target populations, treatment method, and technological characteristics.
The Surgical Company International B.V. | Paalbergweg 3 | 1105 AG Amsterdam | The Netherlands | T +31 20 7900 2100 | letsconnect@tsc-life.com www.tsc-life.com | CC no. 31029240 | VAT no. NL 8111.49.948.B03 | BANK: ING Commercial Banking | Bijlmerplein 888 | 1102 MG Amsterdam | The Netherlands IBAN NL15 INGB 0653 7522 02 | BIC: INGBNL2A
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.