K211618 · The Surgical Company International B.V. · LGZ · Jun 5, 2023 · General Hospital
Device Facts
Record ID
K211618
Device Name
Fluido Compact Blood and Fluid Warming System
Applicant
The Surgical Company International B.V.
Product Code
LGZ · General Hospital
Decision Date
Jun 5, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Fluido Compact Blood and Fluid Warming System can be used in adult patients that need fluid warming pror to administration of blood, blood products and IV fluids to help prevent hypothermia. The system is intended to be used by healthcare professionals, e.g., nurses, medical specialists, doctors. Only medical professionals shall interact with the system. The devices must only be used on the order of a physician.
Device Story
Fluido Compact System warms blood, blood products, and IV fluids to 39°C; utilizes conduction warming technology via a disposable cassette with a biocompatible coated aluminum heater plate. System comprises a reusable warming module, control module, and sterile disposable set. Operated by healthcare professionals (nurses, doctors) in hospital settings; requires physician order. Control module powers warming module; closed-loop software monitors fluid temperature via sensors. Redundant safety module provides over-temperature shutoff. Output is warmed fluid delivered to patient line. Benefits include prevention of hypothermia during infusion/transfusion.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including temperature control, flow rate verification, alarm functionality, usability, heater safety (ASTM F2172-02), and electrical/EMC testing (IEC 60601-1, IEC 60601-1-2).
Technological Characteristics
Conduction warming technology; biocompatible coated aluminum heater plate; closed-loop temperature control; redundant over-temperature shutoff. Power: 100-240V AC. Standards: ASTM F2172-02, IEC 60601-1, IEC 60601-1-2, ISO 10993. Disposable set sterilized via ethylene oxide. Connectivity: Standalone system.
Indications for Use
Indicated for adult patients requiring fluid warming prior to administration of blood, blood products, and IV fluids to prevent hypothermia. Contraindications not specified.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
Vital Signs, Inc. enFlow Model IV Fluid Warmer (K112902)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
June 5, 2023
The Surgical Company International B.V. Inette Nieveen RAQA Director Beeldschermweg 6F Amersfoort, Utrecht 3821 AH Netherlands
Re: K211618
Trade/Device Name: Fluido Compact Blood and Fluid Warming System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion pump Regulatory Class: Class II Product Code: LGZ Dated: June 2, 2023 Received: June 2, 2023
Dear Inette Nieveen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Jake K. Jake K. Lindstrom -S Lindstrom -S Date: 2023.06.05 16:41:56 -04'00'
Jake Lindstrom, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211618
#### Device Name
Fluido Compact Blood and Fluid Warming System
#### Indications for Use (Describe)
The Fluido Compact Blood and Fluid Warming System can be used in adult patients that need fluid warming pror to administration of blood, blood products and IV fluids to help prevent hypothermia. The system is intended to be used by healthcare professionals, e.g., nurses, medical specialists, doctors. Only medical professionals shall interact with the system. The devices must only be used on the order of a physician.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|---------------------------------------------|
| <div></div> | |
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
The Surgical Company International B.V.
FLUIDO COMPACT SYSTEM TRADITIONAL 510(K) PREMARKET NOTIFICATION
## 510(k) Summary
This 510(k) summary is in accordance with the requirements of 21 CFR 807.92
| SUBMITTER 807.92(a)(1) | |
|----------------------------------|---------------------------------------------------------------------------------------|
| Submitter Name: | The Surgical Company International B.V. |
| Submitter Address: | Beeldschermweg 6F<br>3821 AH Amersfoort, The Netherlands |
| Phone Number: | +31 (0) 33 4507250 |
| Contact Person: | F (Federica) Federici |
| Date Prepared: | October 25, 2022 |
| DEVICE 807.92(a)(2) | |
| Device Trade Name: | Fluido® Compact System |
| Common Name | Infusion Pump |
| Classification | 21 CFR 880.5725 |
| Product Code: | LGZ |
| Review Panel: | General Hospital |
| Device Class: | Class II |
| PREDICATE DEVICE<br>807.92(a)(3) | Vital Signs, Inc. enFlow Model IV Fluid Warmer<br>(K112902) (a GE Healthcare Company) |
DEVICE DESCRIPTION 807.92(a)(4):
The Fluido® Compact System consists of Fluido Compact Warming Module, Fluido Compact Control Module and Fluido Compact Standard Set (disposable).
The Fluido Compact system is designed to supply warm fluids at 39°C set point with flow rates from 5 to 100 ml/min. The sterile disposable set consist of a plastic housing and biocompatible coated aluminum heater plate. Heat from the warming module is transferred through the heat transfer part into the fluids that is circulating through the inner part of the cassette. The disposable is connected to the blood or fluid source (infusate reservoir) and the patient line on the other end. The disposable has standard luer lock fittings on both ends. The Control Module serves as the power supply for the Warming module.
Fluid and blood are warmed during infusion or transfusion. The delivery of heat is done by heat exchange. A warming module exchanges heat to a disposable set, which is connected to the fluid reservoir and the patient line via conduction warming technology The fluid temperature is controlled by closed loop software and guarded by sensors and has a redundant safety module for fluid overtemperature shutoff.
| INTENDED USE /<br>INDICATION FOR USE<br>807.92(a)(5): | The Fluido Compact System can be used in adult patients<br>that need fluid warming prior to administration of blood,<br>blood products and IV fluids to help prevent hypothermia. |
|-------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{4}------------------------------------------------
| | The system is intended to be used in hospital by healthcare professionals, e.g., nurses, medical specialists, doctors. Only medical professionals shall interact with the system. The devices must only be used on the order of a physician. |
|--|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
## SUMMARY OF TECHNOLOGICAL CHARCHTERISTICS COMPARISON 807.92(a)(6):
Fluido® Compact Blood And Fluid Warming System device has the same fundamental technological characteristics as the predicate device. Similarities and differences are listed in the table below.
| Use<br>Characteristics | Fluido® Compact Blood<br>And Fluid Warming System | Predicate enFlow IV<br>Fluid Warmer | Equivalence |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| Patient<br>Population | Adult | Not defined | Similar |
| User | Healthcare professionals | Healthcare<br>professionals | Same |
| Use<br>Environment | Hospital | Hospital, clinical and<br>field environments | Similar |
| Operating<br>Principle | Reusable Warmer<br>Reusable Controller<br>Disposable sterile | Reusable Warmer<br>Reusable Controller<br>Disposable sterile | Same |
| Prescription<br>Use/<br>Over-The-<br>Counter use | Prescription Use | Prescription Use | Same |
| Technological Characteristics | | | Same/different<br>technological<br>characteristics |
| Administered<br>Fluids | IV Fluids,<br>Blood Products | IV Fluids,<br>Blood,<br>Blood Products | Similar |
| Fluid Path | Sterile Fluid Path | Sterile Fluid Path | Same |
| System<br>Components | Reusable Warmer<br>Reusable Controller<br>Disposable sterile Set | Reusable Warmer<br>Reusable Controller<br>Disposable sterile<br>Cartridge | Same |
| Safety Features | Reusable Warmer<br>Reusable Controller<br>Disposable sterile Set<br>Closed-loop<br>temperature controlOver-temperature<br>control and cut off<br>(ASTM F2172-02)Audible and Visual<br>alarms | Reusable Warmer<br>Reusable Controller<br>Disposable sterile<br>Cartridge<br>Closed-loop<br>temperature controlOver temperature<br>control and cut off<br>(ASTM F2172-02)Audible and Visual<br>alarms | Same |
| | heater temperature<br>monitoring circuit | • Independent heater<br>temperature<br>monitoring circuit | |
| Warming<br>technology | Conduction | Conduction | Same |
| Heat exchange | Inline disposable | Inline disposable | Same |
| Temperature<br>control | Closed Loop temperature<br>control | Closed Loop<br>temperature control | Same |
| Fluid<br>Temperature<br>Output | 39°C ± 2°C | 40°C ± 2°C | Similar,<br>within the<br>physiological<br>range |
| Safety cut off<br>Temperature | 49°C | 45°C | Similar, still<br>within the<br>safety margins |
| Flow Rate<br>Range | Standard Set:<br>5 - 100 ml/min | Keep Vein Open (KVO)<br>to 200 ml/min | Similar |
| Storage<br>Conditions/<br>Specifications | Warmer and Controller:<br>-40°C - 50°C at 10% to<br>90% relative humidity<br>Disposable Set:<br>-20°C - 40°C at 10% to<br>90% relative humidity | -30°C – 70°C at 10% to<br>90% relative humidity | Similar |
| Power Source | AC power supply<br>100-240 V ~ 50/60Hz)<br>1.6A | AC power supply<br>110-120 or 220-240 V<br>5A | Similar |
| Dimensions | Warmer:<br>16.5 cm x 7.5 cm x 5.0 cm<br>Controller:<br>28.5 cm x 12.0 cm x 19.5<br>cm<br>Disposable Set:<br>14.6 cm x 3.5 cm x 1.1 cm | Warmer:<br>12.7 cm x 6.6 cm x 3.0<br>cm<br>Controller:<br>23.6 cm x 16.8 cm x 3.8<br>cm<br>Disposable cartridge:<br>11.4 cm x 3.8 cm x 1.0<br>cm | Similar |
| Weight | Warmer (without<br>disposable): 450g<br>Controller: 1.7kg<br>Disposable Set: 24g | Warmer (without<br>disposable): 279g<br>Controller: 1.8kg<br>Disposable Cartridge:<br>33g | Similar |
| Performance<br>standards | IEC 60601-1<br>IEC 60601-1-2 | IEC 60601-1<br>IEC 60601-1-2 | Same |
| Sterilization | Disposable Set:<br>Ethylene oxide sterilized | Disposable Cartridge:<br>Gamma sterilized | Similar |
| Biocompatibility | ISO 10993 | ISO 10993 | Same |
{5}------------------------------------------------
{6}------------------------------------------------
### FLUIDO COMPACT SYSTEM TRADITIONAL 510(K) PREMARKET NOTIFICATION
## Non-clinical testing 807.92(b)(1):
Performance testing according to applicable testing met all acceptance criteria.
Comparative bench performance testing:
- . Temperature: PASS
- . Flow rates: PASS
- Alarms: PASS ●
- . Useability: PASS
Performance testing heater safety
- . ASTM F2172-02 (Reapproved 2011): PASS
Electrical/EMC testing: IEC 60601-1: PASS IEC 60601-1-2: PASS
Clinical testing 807.92(b)(2): Clinical testing was not necessary to demonstrate substantial equivalence.
## CONCLUSION OF SUBSTANTIAL EQUIVALENCE 807.92(b)(3):
Performance testing to FDA recognized standards and FDA guidance demonstrates substantial equivalence to the predicate device. The predicate comparison table and performance testing provided in this 510(k) is sufficient to demonstrate that the Fluido® Compact Blood and Fluid Warming System is substantially equivalent to the legally marketed predicate device, Vital Signs, Inc. enFlow Model IV Fluid Warmer cleared under 510(k) K112902.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.