K102010 · Medpro Safety Products, Inc. · FMI · Nov 19, 2010 · General Hospital
Device Facts
Record ID
K102010
Device Name
VACUETTE PREMIUM SAFETY BLOOD COLLECTION SET
Applicant
Medpro Safety Products, Inc.
Product Code
FMI · General Hospital
Decision Date
Nov 19, 2010
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.5570
Device Class
Class 2
Indications for Use
VACUETTE® PREMIUM Safety Blood Collection Set is used in routine venipuncture procedures. It is used for blood collection. The safety shield is activated automatically to cover the needle immediately following blood collection thus preventing accidental needlestick injury. The Vacuette Premium Safety Blood Collection Set is used in routine venipuncture procedures. It is used for blood sollection. The winged needle is designed with a safety shield, which is activated automatically to cover the needle immediately following blood eollection to aid in the protection against-accidental-needlestick injury-The product-is-to be-used by appropriately trained healtheare professionals only in accordance with these instructions.
Device Story
Single-use, sterile, winged blood collection needle with integrated safety shield and flexible tubing; female luer adapter allows connection to luer systems. Device used in clinical settings by trained healthcare professionals for routine venipuncture. During puncture, wings fall open, placing shield in staged position. Upon withdrawal, closing wings triggers spring-loaded mechanism to automatically deploy shield over needle, protecting user from needlestick injury. Device is PVC, DEHP, and latex-free. Benefits include reduced risk of occupational exposure to bloodborne pathogens via automated safety activation.
Clinical Evidence
Bench testing only. Simulated use testing performed by a panel of healthcare professionals to evaluate function, performance, and ease of use of the safety feature.
Technological Characteristics
Winged needle with integrated spring-loaded safety shield; flexible tubing; female luer adapter. Materials meet ISO 10993-1 for external communicating, blood path indirect, limited contact duration. PVC, DEHP, and latex-free. Mechanical activation via wing closure.
Indications for Use
Indicated for routine venipuncture and blood collection in patients requiring blood draws. Intended for use by trained healthcare professionals to prevent accidental needlestick injuries.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
BD VACUTAINER PUSH BUTTON BLOOD COLLECTION SET (K011984)
Submission Summary (Full Text)
{0}------------------------------------------------
# NOV 1 9 2010
# Kingalo
| K162016 | |
|----------------------------|----------|
| Section 5 - 510(k) Summary | Page 5-1 |
Date Prepared: July 13, 2009 Submitter / Contact: H. Carl Jenkins The Wood Burditt Group 10 E. Scranton Avenue, Suite 201 Lake Bluff, IL 60044 (ph) (847) 234-7500 x 205 (fax) (847) 574-0728 (email) hcjenkins@woodburditt.com Applicant / 510(k) Owner: MedPro Safety Products, Inc 817 Winchester Road, Suite 200 Lexington, KY 40505
Phone – (859) 225-5375 Fax - (859) 225-5347 Contact Person - Mrs. Tammy Nichols
#### Name of Device
| Trade Name: | VACUETTE® PREMIUM Safety Blood Collection Set |
|----------------------|-------------------------------------------------------|
| Common Name: | Safety device for blood collection |
| Classification Name: | Hypodermic Single Lumen Needle with antistick syringe |
| Regulations No: | 21 CFR 880.5570 |
| Product Code: | FMI |
#### Predicate Information
| Predicate Device: | VACUETTE SAFETY BLOOD COLLECTION SET |
|----------------------|------------------------------------------------|
| 510(k) Number: | K011786 |
| Date of Concurrence: | 07/12/2001 |
| Predicate Device: | BD VACUTAINER PUSH BUTTON BLOOD COLLECTION SET |
| 510(k) Number: | K011904 |
510(k) Number: K011984 Date of Concurrence: 08/29/2001
{1}------------------------------------------------
| Section 5 - 510(k) Summary | Page 5-2 |
|----------------------------|----------|
|----------------------------|----------|
# Substantial Equivalence Declaration:
The term "Substantial Equivalence" as used in this 510(k) Premarket Notification is limited to the definition of Substantial Equivalence found in the Federal Food, Drug, and Cosmetic Act, as amended and as applied under 21 CFR §807, Subpart E, under which a device can be marketed without premarket approval or reclassification. A determination of substantial equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to, or in support of, substantial equivalence herein shall be construed as an admission against interest under the US Patent Laws or their application by the courts.
## Device Description:
Trade or propriety name or model of this device: VACUETTE® PREMIUM Safety Blood Collection Set;
| Gauge | Tubing | Accessory | Accessory | Accessory |
|-------|--------|-------------|-----------|---------------------------------|
| 21 | 10 | Female Luer | | Male Luer |
| 23 | 10 | Female Luer | | Male Luer |
| 25 | 10 | Female Luer | | Male Luer |
| 21 | 19 | Female Luer | | Male Luer |
| 23 | 19 | Female Luer | | Male Luer |
| 25 | 19 | Female Luer | | Male Luer |
| 21 | 30 | Female Luer | | Male Luer |
| 23 | 30 | Female Luer | | Male Luer |
| 25 | 30 | Female Luer | | Male Luer |
| 21 | 10 | Female Luer | Male Luer | Greiner Bio-One Standard Holder |
| 23 | 10 | Female Luer | Male Luer | Greiner Bio-One Standard Holder |
{2}------------------------------------------------
#### Section 5 - 510(k) Summary
| Gauge | Tubing | Accessory | Accessory | Accessory |
|-------|--------|-------------|-----------|---------------------------------|
| 25 | 10 | Female Luer | Male Luer | Greiner Bio-One Standard Holder |
| 21 | 19 | Female Luer | Male Luer | Greiner Bio-One Standard Holder |
| 23 | 19 | Female Luer | Male Luer | Greiner Bio-One Standard Holder |
| 25 | 19 | Female Luer | Male Luer | Greiner Bio-One Standard Holder |
| 21 | 30 | Female Luer | Male Luer | Greiner Bio-One Standard Holder |
| 23 | 30 | Female Luer | Male Luer | Greiner Bio-One Standard Holder |
| 25 | 30 | Female Luer | Male Luer | Greiner Bio-One Standard Holder |
| 21 | 10 | Female Luer | | NA |
| 23 | 10 | Female Luer | | NA |
| 25 | 10 | Female Luer | | NA |
| 21 | 19 | Female Luer | | NA |
| 23 | 19 | Female Luer | | NA |
| 25 | 19 | Female Luer | | NA |
| 21 | 30 | Female Luer | | NA |
| 23 | 30 | Female Luer | | NA |
| 25 | 30 | Female Luer | | NA |
#### Product Description:
The Vacuette Premium Safety Blood Collection Set is a single use, individually wrapped, sterile winged blood collection needle with an integrated needle and safety shield bonded to a flexible tubing with a female luer adapter allowing the set to be used with a luer system (e.g. Holdex®). It is available with optional Luer Adapter and / or Luer Adapter + Holder. The device is free of PVC, DEHP, Latex.
{3}------------------------------------------------
# Indications for Use:
"VACUETTE® PREMIUM Safety Blood Collection Set is used in routine venipuncture procedures. It is used for blood collection. The safety shield is activated automatically to cover the needle immediately following blood collection thus preventing accidental needlestick injury."
## Intended Use:
The Vacuette Premium Safety Blood Collection Set is used in routine venipuncture procedures. It is used for blood sollection. The winged needle is designed with a safety shield, which is activated automatically to cover the needle immediately following blood
eollection to aid in the protection against-accidental-needlestick injury-The product-is-to be-used by appropriately trained healtheare professionals only in accordance with these instructions.
# Substantial Equivalence:
The VACUETTE® PREMIUM Safety Blood Collection Set is substantially equivalent to the Vacuette Safety Blood Collection Set (K011786) and the BD Pushbutton Butterfly (K011984).
## Simulated Use Testing:
Simulated use testing was performed by a panel of health care professionals to evaluate function and performance, ease of use and other functions of the safety feature.
# Material Testing:
The VACUETTE® PREMIUM Safety Blood Collection Set will meet ISO 10993-1 requirements for material safety and biocompatibility, per product category:
- . External Communicating
- . Blood Path Indirect
- . Contact Duration "A" (Limited)
# Conclusions:
Summary of Safety and Effectiveness -
The device has a release mechanism molded into the needle shield. During the blood collection process, as the user performs the initial puncture, release of the wings causes them to fall open and brings the needle shield into a staged position. When the user is ready to remove the device, they grasp the wings as they would during routine use, and upon closing the needle shield is deployed and covers the needle upon exit.
The device is substantially equivalent to the predicate devices (VACUETTE® Safety Blood Collection Set and BD Vacutainer Push Button Set) because it has a plastic safety shield that covers the needle
{4}------------------------------------------------
| Section 5 - 510(k) Summary | Page 5-5 | after<br>it<br>has<br>been<br>n<br>cont |
|----------------------------|----------|-----------------------------------------|
|----------------------------|----------|-----------------------------------------|
aminated, protecting the healthcare worker from a potential needlestick. The predicate VACUETTE® Safety Blood Collection Set's safety shield
is activated when the user manually slides the shield over the needle after use. The predicate BD Vacutainer Push Button Blood Collection Set's safety shield is activated when the user manually presses the button located on the top of the device, activating the spring mechanism to deploy the safety shield. The VACUETTE® PREMIUM Safety Blood Collection Set's safety shield is automatically activated when the user closes the wings to withdraw the device from the patient. The spring loaded safety shield covers the needle after it has been contaminated, protecting the healthcare worker from a potential needlestick.
{5}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features the department's emblem, which is a stylized representation of a human figure embracing a globe. The emblem is positioned to the right of the text, which is arranged in a circular pattern around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MedPro Safety Products, Incorporated C/O Mr. H. Carl Jenkins Wood Burditt Group 10 E. Scranton Avenue, Suite 201 · Lake Bluff, Illinois 60044
NOV 1 9 2010
Re: K102010
Trade/Device Name: VACUETTE® PREMIUM Safety Blood Collection Set Regulation Number: 21 CFR 880.5570 Regulation Name: Regulatory Class: II Product Code: JKA, FMI Dated: November 5, 2010 Received: November 8, 2010
Dear Mr. Jenkins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
#### Page 2- Mr. Jenkins
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
| Section 4 - Indications for Use Statement | Page 4-1 |
|-------------------------------------------|----------|
|-------------------------------------------|----------|
#### Indications for Use
510(k) number (if Known): K102010 NOV 1 9 2010
Device Name: VACUETTE® PREMIUM Safety Blood Collection Set
Indications for Use:
"VACUETTE® PREMIUM Safety Blood Collection Set is used in routine venipuncture procedures. It is used for blood collection. The safety shield is activated automatically to cover the needle immediately following blood collection thus preventing accidental needlestick injury."
Prescription Use __ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
An Str for RZC
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page_1_of_1
K102010 510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.