CLOSED SWABABLE STOPCOCK AND MINIMAL RESIDUAL VOLUME LUER-ACTIVATED SWABABLE-STOPCOCK
K060231 · Elcam Medical Acal · FMG · Apr 13, 2006 · General Hospital
Device Facts
Record ID
K060231
Device Name
CLOSED SWABABLE STOPCOCK AND MINIMAL RESIDUAL VOLUME LUER-ACTIVATED SWABABLE-STOPCOCK
Applicant
Elcam Medical Acal
Product Code
FMG · General Hospital
Decision Date
Apr 13, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
Elcam Closed Swabable Stopcock (or MRVLS) is indicated for fluid flow directional control and for providing access port(s) for administration of solutions. Typical uses include pressure monitoring, intravenous fluid administration and transfusion.
Device Story
Closed Swabable Stopcock and MRVLS (Minimal Residual Volume Luer-activated Swabable-stopcock) functions as a fluid control device for IV sets. Device integrates a needleless luer-activated valve into a stopcock; valve automatically opens upon male luer insertion and closes upon removal, eliminating need for caps and reducing contamination risk. MRVLS variant features modified handle design and flow guide to enable continuous fluid flow through side ports, significantly reducing residual volume. Used in clinical settings by healthcare providers for pressure monitoring, fluid administration, and blood sampling. Healthcare providers manually operate stopcock handle and swab valve with aseptic fluid prior to use. Benefits include reduced residual volume, improved fluid path cleanliness, and simplified access for injections/sampling.
Clinical Evidence
Bench testing only. Performance testing conducted in accordance with device specifications; all acceptance criteria met. Biocompatibility and chemical testing performed per ISO 10993-1:2003 and FDA Memorandum #G95-1 with satisfactory results.
Technological Characteristics
Stopcock with integrated luer-activated valve. MRVLS variant includes modified handle and flow guide for reduced residual volume. Materials evaluated for biocompatibility per ISO 10993-1:2003. Mechanical fluid control device; no software or energy source.
Indications for Use
Indicated for patients requiring fluid flow directional control, pressure monitoring, IV fluid administration, or transfusion. No specific age or gender restrictions stated.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Elcam Medical. The logo consists of the words "Elcam Medical" stacked on top of each other, with the letters "l" and "c" stacked on top of each other to the right of the words. There is a black circle above the "l" and "c". The logo is black and white.
## APR 1 3 2006
## SPECIAL 510(K) SUMMARY FOR ELCAM'S CLOSED SWABABLE STOPCOCK (OR MRVLS)
## DATE PREPARED: APRIL 7, 2006
#### 510(K) OWNER NAME 1.
Elcam Medical ACAL Kibbutz BarAm, Merom HaGalil 13860, Israel
Submitter person name: Ms. Tali Hazan - R.A Specialist Telephone: 972-4-6988875, Fax: 972-4-6980777, E-mail: tali@elcam.co.il
ELCAM MEDICAL'S U.S AGENT: Elcam Medical, Inc. 2 University Plaza, Suite 620, Hackensack, NJ 07601, USA
Contact Person: Mr. Ehud Raivitz - CEO Telephone: 201-457-1120, Fax: 201-457-1125, E-mail: ehud@elcam-medical.com
#### 2. DEVICE NAME
Common/Usual Name: Closed Swabable Stopcock and *MRVLS Proprietary/Trade name: Closed Swabable Stopcock and MRVLS MRVLS = Minimal Residual Volume Luer-activated Swabable-stopcock
Classification: Elcam's Closed Swabable Stopcock (or MRVLS) has been classified as Class II devices under the following classification names:
| Classification Name | Product Code | 21 CFR Ref. | Panel |
|---------------------|--------------|-------------|------------------|
| Stopcock, I.V. Set | FMG | 880.5440 | General Hospital |
#### 3. PREDICATE DEVICES
Elcam's Closed Swabable Stopcock (or MRVLS) is substantially equivalent to Elcam's Stopcocks and Manifolds cleared under 510(k) number K022895.
{2}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for Elcam Medical. The logo consists of the words "Elcam Medical" stacked on top of each other on the left side of the image. To the right of the words is a stylized graphic that appears to be the letters "I" and "C" with a circle above them. The logo is black and white.
#### 4. DEVICE DESCRIPTION
Modified Device and change Description: Elcam Medical's Closed Swabable Stopcock and MRVLS is a similar version of Elcam legally marketed stopcock. cleared under 510(k) number K022895.
The Closed Swabable Stopcock includes an addition of a component that functions as a closed luer-activated valve.
The valve enables the female port (luer) to be closed when it is not in use and saves the need to close it with a cap, in order to avoid leakage and/or port's contamination. Once a male luer is introduced into the closed port the fluid path automatically opens to allow injection or blood sampling. Once the male luer is taken out, the female port is automatically closed.
The valve actually serves as a needleless injection site integrated in the stopcock. The only change from the end user point of view is the need to swab the valve's top (injection site) with aseptic fluid such as alcohol prior to male luer insertion.
The MRVLS (Minimal Residual Volume Luer-activated Swabable-stopcock) feature, is an additional option that can be added to the Closed Swabable Stopcock, described above, and provides a stopcock with a minimal residual volume.
Conventional stopcocks are made of a body and a handle with a fluid path bore that allows the fluid to flow only through the handle. Therefore, it leaves other areas in the stopcock without a continuous flow through the stopcock fluid path and especially through the side port female luer. The MRVLS feature enables fluid flow around the handle so the fluid constant flow accesses the entire stopcock's internal volume and enables a continuous flow through the stopcock fluid path and through the side female luer during the medical procedure. This feature significantly reduces the "residual volume" in the stopcock and helps to keep the cleanliness of the stopcock's fluid path when it is in use.
The feature of minimal residual volume is achieved thanks to a modification in the stopcock handle design, which changes the fluid path and a flow guide formed therein.
The minimal residual volume design, together with the luer-activated valve creates the Minimal Residual Volume Luer-activated Swabable-stopcock (MRVLS).
Elcam Medical BarAm 13860, Israel Tel: 972-4-698-8120/1/2, Fax: 972-4-698-0777 sales@elcam.co.il
..............................................................................................................................................................................
www.elcam-medical.com
Elcam Medical Special 510(k) revised Summary Elcam Closed Swabable Stopcock and MRVLS
April 7, 2006
CONFIDENTIAL Page 2 of 4
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a logo for Elcam Medical. The logo consists of the words "Elcam Medical" stacked on top of each other, with the letters "IC" in a stylized font to the right of the words. There is a black circle above the "IC" letters. The logo is in black and white.
The colorant and raw materials changed and/or added are all validated to their intended use and evaluated for biocompatibility.
All body/fluid contact materials that compose the modified device: Closed Swabable Stopcock and MRVLS were tested for biocompatibility in accordance to FDA's Memorandum – #G95 1, May 1, 1995 and ISO 10993-1:2003 - Biological evaluation of medical devices – Part 1: Evaluation and testing with acceptable results.
#### ડ. INTENDED USE
Elcam Closed Swabable Stopcock (or MRVLS) is indicated for fluid flow directional control and for providing access port(s) for administration of solutions. Typical uses include pressure monitoring, intravenous fluid administration and transfusion.
#### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE 6.
Elcam's Closed Swabable Stopcock (or MRVLS) is substantially equivalent to Elcam's conventional legally marketed Stopcock cleared by 510(k) number K022895. Elcam's new product and the predicate devices have the same indication for use, same basic shape, characteristics, materials, manufacturing technology and design.
The modified device differs from the predicate device in some characteristics such as the addition of the valve assembly and the design change in the MRVLS handle. These differences are not affecting the device's intended use or alter the device's fundamental scientific technology. The device is, therefore, as safe and as effective as the predicate device.
#### 7. NONE CLINICAL PERFORMANCE DATA
Tests results are supporting all labeling claims and substantial equivalency. The modified device was tested with accordance to Elcam's legally marketed device specification and all acceptance criteria were met.
Biocompatibility and chemical tests, material characterization and risk assessment were performed on the patient-contact and fluid path materials of Elcam's modified device with satisfactory results.
Elcam Medical BarAm 13860, Israel Tel: 972-4-698-8120/1/2, Fax: 972-4-698-0777 sales@elcam.co.il
www.elcam-medical.com
.............................................................................................................................................................................. Elcam Medical Special 510(k) revised Summary Elcam Closed Swabable Stopcock and MRVLS
April 7, 2006
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Elcam Medical. The logo consists of the words "Elcam Medical" in a bold, sans-serif font, with the word "Elcam" stacked above the word "Medical". To the right of the words is a stylized graphic that appears to be the letters "i" and "c" with a circle above them.
#### 8. CONCLUSIONS
The evaluation of Elcam's Closed Swabuble Stopcock (or MRVLS) non-clinical tests demonstrates that the device is as safe, as effective, and performs as well as or better than the predicate device. Therefore, we believe it is substantially equivalent to the Elcam's legally marketed device.
……………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread. The eagle is facing to the right.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 1 3 2006
Ms. Tali Hazan Elcam Medical A.C.A.L. Kibbutz BarAm D.N. Merom Hagalil, 13860 Israel
Re: K060231
Trade/Device Name: Closed Swabable Stopcock and Minimal Residual Volume Luer - activated Swabable-stopcock Regulation Number: 880.5440 Regulation Name: Intravascular administration set Regulatory Class: II Product Code: FMG Dated: March 13, 2006 Received: March 16, 2006
Dear Ms. Hazan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent [{for the indications for use stated in the enclosure)] to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page-2 Ms. Tali Hazan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at their toll free number (800) 638-2041 or at (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Clive
Chiu Lin, Ph.D Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health.
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Closed Swabable Stopcock and MRVLS)
MRVLS - Minimal Residual Volume Luer-activated Swabable-Stopcock
Indications for Use: Elcam Closed Swabable Stopcock and MRVLS is indicated for fluid flow directional control and for providing access port(s) for administration of solutions. Typical uses include pressure monitoring, intravenous fluid administration and transfusion.
> Prescription Use Prescription Ose ------------------------------------------------------------------------------------------------------------------------------------------------------------- Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page _1_of_1_
(Posted November 13, 2003)
Anthony D. Watson
,
, General Hospital
Control Dental Devices
K4E0231
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.