K123205 · Taewoong Medical Co., Ltd. · ESW · Oct 9, 2013 · General, Plastic Surgery
Device Facts
Record ID
K123205
Device Name
ESOPHAGEAL TTS STENT
Applicant
Taewoong Medical Co., Ltd.
Product Code
ESW · General, Plastic Surgery
Decision Date
Oct 9, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3610
Device Class
Class 2
Attributes
Therapeutic
Intended Use
For use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors.
Device Story
Esophageal TTS Stent; implantable metallic stent with flexible introducer system. Stent composed of Nitinol wire; fully or partially silicone-covered. Device deployed in esophagus to establish patency via outward radial force against luminal surface. Used in clinical settings for management of malignant esophageal strictures. Introducer system facilitates delivery. Larger 20mm body diameter variant added to existing product line.
Clinical Evidence
Bench testing only. Conducted per FDA guidance for esophageal and tracheal prostheses. Tests included expansion force, compression force, dimensional, deployment, deployment force, tensile strength, packaging adhesive strength, and MR compatibility. No clinical data provided.
Technological Characteristics
Nitinol wire stent; fully or partially silicone-covered. Expandable tubular design. Includes disposable introducer system. 20mm body diameter. MR compatible.
Indications for Use
Indicated for patients with esophageal strictures resulting from intrinsic and/or extrinsic malignant tumors.
Regulatory Classification
Identification
An esophageal prosthesis is a rigid, flexible, or expandable tubular device made of a plastic, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the esophagus. The metal esophageal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.
Special Controls
The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”
K113551 — ESOPHAGEAL TTS STENT · Taewoong Medical Co., Ltd. · Feb 14, 2012
K041648 — NITI-S STENT & INTRODUCER, MODEL EOXXXX · Taewoong Medical Co., Ltd. · Apr 28, 2005
K211706 — Esophageal TTS Stent · Taewoong Medical Co., Ltd. · Oct 8, 2021
K080782 — NITI-S ESOPHAGEAL STENT · Taewoong Medical Co., Ltd. · Jan 30, 2009
K221482 — Esophageal TTS Stent · Taewoong Medical Co., Ltd. · Jun 22, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) Summary
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the Esophageal TTS Stent device is provided below.
| Device Common Name: | Esophageal Prosthesis |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Proprietary Name: | Esophageal TTS Stent |
| Submitter: | Y.J. Kim<br>Taewoong Medical Co., Ltd.<br>1-5, Gomak-ri, Wolgot-myeon, Gimpo-si, Gyeonggi-do, Korea<br>415871<br>TEL: +82-31-996-0641<br>FAX: +82-31-996-0645 |
| Contact: | Calley Herzog<br>Consultant, Medical Devices<br>Biologics Consulting Group, Inc.<br>Phone: 720-883-3633<br>Email: cherzog@bcg-usa.com<br>Fax: 720-293-0014 |
| | OCT 0 9 2013 |
| Date Summary Prepared: | October 10, 2012 |
| Classification | 21 CFR 878.3610 |
| Regulation: | |
| Panel: | General & Plastic Surgery Devices |
| Product Code: | ESW |
| Predicate Devices: | K113551 - Taewoong Esophageal TTS Stent<br>K080782 - Taewoong Niti-S Esophageal Stent |
#### Indication for Use:
For use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors.
### Device Description:
The Esophageal TTS Stent consists of an implantable metallic stent and a flexible introducer system. The stent is a flexible, expandable tubular device made of Nitinol wire and are either fully silicone covered or partially silicone covered (the body is covered but both heads remains uncovered). This device also includes a disposable introducer. Upon deployment, the stent imparts an outward radial force on the luminal surface esophagus to establish patency.
{1}------------------------------------------------
#### Comparison to the Predicates:
The Esophageal TTS Stent is substantially equivalent to the Esophageal TTS Stent cleared in K 113551 with the following exception: the submitted stent has a larger body diameter of 20mm. FDA has previously cleared a 20 mm diameter Esophageal stent with the same stent lengths as part of Taewoong's Niti-S stent system (K080782), so this new model does not raise any new types of safety and effectiveness questions.
The introducer used to deliver the Esophageal TTS Stent is the exact same introducer as the one cleared for use with the Esophageal TTS Stent (K113551). The matcrials, diameter, length and manufacturing are the same as previously cleared introducer.
#### Performance Data:
Per FDA "Guidance for Industry - Guidance for the Content of Premarket Notifications for Esophageal and Tracheal Prostheses," the following tests were conducted:
- Expansion Force Testing ●
- Compression Force Testing .
- Dimensional Testing 0
- . Deployment Testing
- Deployment Force Testing .
- Tensile Strength Testing .
The following testing was also conducted:
- . Packaging Adhesive Strength Testing
- MR Compatibility .
The performance testing showed that the Esophageal TTS Stent with a larger body diameter of 20mm is as safe, as effective, and performs as well as or better than the predicate device.
### Summary of Substantial Equivalence:
Based on the indications for use, technological characteristics and performance test results, the Esophageal TTS Stent is substantially equivalent to the predicate Esophageal TTS Stent.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, surrounded by a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is written around the border of the circle. The eagle is depicted in a simple, abstract style.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609. Silver Spring, MD 20993-0002
October 9, 2013
Taewoong Medical Co., Ltd. % Calley Herzog Consultant Biologics Consulting Group, Inc. 400 N. Washington Street, Suite 100 Alexandria, VA 22314
Re: K123205
> Trade/Device Name: Esophageal TTS Stent Regulation Number: 21 CFR\$ 878.3610 Regulation Name: Esophageal prosthesis Regulatory Class: II Product Code: ESW Dated: September 19, 2013 Received: September 23, 2013
Dear Calley Herzog,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRFI does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA`s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{3}------------------------------------------------
Page 2 - Calley Herzog
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda,gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Herbert P. Lerner -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### 4.0 Indications for Use Statement
510(k) Number (if known): Kl23205
Device Name: Esophageal TTS Stent
Indications For Use:
For use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Herbert P. Lerner -S 2013.10.09 15:51:17 -04'00'
Page 1 of __ 1__
Page 12
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.