K221482 · Taewoong Medical Co., Ltd. · ESW · Jun 22, 2022 · General, Plastic Surgery
Device Facts
Record ID
K221482
Device Name
Esophageal TTS Stent
Applicant
Taewoong Medical Co., Ltd.
Product Code
ESW · General, Plastic Surgery
Decision Date
Jun 22, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.3610
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Esophageal TTS Stent is intended for use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors and occlusion of concurrent esophageal fistulas.
Device Story
Esophageal TTS Stent is an implantable metallic stent with a disposable, flexible introducer system. Stent consists of Nitinol wire in a diamond shape; intended to restore esophageal structure/function by imparting outward radial force on the luminal surface to establish patency. Introducer is a co-axial tube system used for delivery and deployment at the target position via endoscopic placement. Device is used in clinical settings by physicians. Output is the physical restoration of esophageal patency, benefiting patients by relieving strictures and occluding fistulas. No electrical components or software.
Clinical Evidence
No clinical data provided. Substantial equivalence is established through identical design, materials, and technological characteristics to previously cleared predicate devices.
Technological Characteristics
Stent: Nitinol wire, Pt/Ir and STS 316L radiopaque markers, silicone covering. Introducer: Teflon, PE, ABS. Co-axial tube design. Dimensions: 60-150mm length; 18-22mm body diameter. Deployment: Endoscopic, release by pulling outer sheath. Sterilization: Ethylene Oxide (EO). No electrical components or software.
Indications for Use
Indicated for patients with esophageal strictures caused by intrinsic and/or extrinsic malignant tumors and occlusion of concurrent esophageal fistulas.
Regulatory Classification
Identification
An esophageal prosthesis is a rigid, flexible, or expandable tubular device made of a plastic, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the esophagus. The metal esophageal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.
Special Controls
The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 22, 2022
Taewoong Medical Co., Ltd. % Matthew Krueger Senior Consultant, Medical Devices Biologics Consulting Group, Inc. 1555 King Street, Suite 300 Alexandria, VA 22314
Re: K221482
Trade/Device Name: Esophageal TTS Stent Regulation Number: 21 CFR 878.3610 Regulation Name: Esophageal Prosthesis Regulatory Class: Class II Product Code: ESW Dated: May 19, 2022 Received: May 23, 2022
Dear Matthew Krueger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221482
Device Name Esophageal TTS Stent
Indications for Use (Describe)
The Esophageal TTS Stent is intended for use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors and occlusion of concurrent esophageal fistulas.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the Esophageal TTS Stent is provided below.
#### 1. SUBMITTER
| Applicant: | Taewoong Medical Co., Ltd.<br>14 Gojeong-ro, Wolgot-myeon, Gimpo-si, Gyeonggi<br>do, Korea, 10022<br>TEL: +82-31-996-0641 |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Yongjin Jeff Kim<br>International Regulatory Affairs Manager<br>Phone: +82 70 4649 1543<br>E-mail: jinjeff@stent.net |
| Submission Correspondent: | Matthew Krueger<br>Senior Consultant, Medical Devices<br>Biologics Consulting Group, Inc.<br>Phone: (667) 352-2578<br>Email: mkrueger@biologicsconsulting.com |
| Date Prepared: | June 22, 2022 |
#### 2. DEVICE
| Device Trade Name: | Esophageal TTS Stent |
|---------------------|----------------------------------|
| Device Common Name: | Esophageal Stent |
| Classification Name | 878.3610, Prosthesis, Esophageal |
| Regulatory Class: | Class II |
| Product Code: | ESW |
#### 3. PREDICATE DEVICE AND REFERENCE DEVICES
#### Predicate Devices 3.1.
K211706: Taewoong Esophageal TTS Stent
K123205: Taewoong Esophageal TTS Stent
K113551: Taewoong Esophageal TTS Stent
#### 3.2. Reference Device
K200860: HANAROSTENT Esophagus TTS (CCC) and HANAROSTENT Esophagus TTS (NCN)
{4}------------------------------------------------
#### DEVICE DESCRIPTION 4.
The Esophageal TTS Stents that are the subject of this 510(k) are identical to the devices cleared in K211706, K123205, and K113551, with the exception of a modification to the Indications for Use statement. The Esophageal TTS Stents consists of an implantable metallic stent and a disposable, flexible introducer system for placement of the stent is a flexible and expandable tubular device made of Nitinol wire that is intended to be implanted to restore the structure and/or function of the esophagus. The introducer is a disposable system for delivery and deployment of the stent at the target position. Upon deployment, the stent imparts an outward radial force on the luminal surface of the esophagus to establish patency.
#### 5. INTENDED USE/INDICATIONS FOR USE
The Esophageal TTS Stent is intended for use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors and occlusion of concurrent esophageal fistulas.
### SUBSTANTIAL EQUIVALENCE 6.
## Comparison of Indications
| Table 1: | Indications for Use Comparison | |
|----------------|--------------------------------|---------------------------|
| Subject Device | | Predicate Device (K21170) |
| Subject Device | Predicate Device (K211706) | Explanation of Differences |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The Esophageal TTS Stent is<br>intended for use in esophageal<br>strictures caused by intrinsic<br>and/or extrinsic malignant<br>tumors and occlusion of<br>concurrent esophageal fistulas. | The Esophageal TTS Stent is<br>intended for use in esophageal<br>strictures caused by intrinsic<br>and/or extrinsic malignant<br>tumors. | Similar to predicate and is<br>nearly identical to the reference<br>device K200860 which used the<br>Taewoong TTS Stent<br>(K123205) as its predicate.<br>This addition to the Indications<br>for Use does not affect the<br>safety and efficacy of the<br>device because the device<br>design and intended use is<br>identical to the predicate. |
| | Subject Device | Predicate |
| | | K211706 |
| | | (K123205 and K113551) |
| Device Name | Esophageal TTS Stent | Esophageal TTS Stent |
| Common<br>Name | Esophageal Stent | Esophageal Stent |
| Manufacturer | Taewoong Medical Co., Ltd | Taewoong Medical Co., Ltd |
| Classification<br>Regulation | 878.3610 | 878.3610 |
| Product Code | ESW | ESW |
| Indications<br>for use | For use in esophageal strictures caused by<br>intrinsic and/or extrinsic malignant tumors<br>and occlusion of concurrent esophageal<br>fistulas. | For use in esophageal strictures caused by<br>intrinsic and/or extrinsic malignant tumors. |
| Design<br>(Stents) | | |
| | Image: Full Covered Stent | Image: Full Covered Stent |
| | | |
| | Image: Both Bare Stent | Image: Both Bare Stent |
| | Nitinol wire in diamond shape | Nitinol wire in diamond shape |
| | Lengths; 60, 80, 100, 120, 140, 150mm | Lengths; 60, 80, 100, 120, 140, 150mm |
| | Body Diameter/Head Diameter: | Body Diameter/Head Diameter: |
| | ・ 18mm/26mm | ・ 18mm/26mm |
| | ・20mm/26mm | ・ 20mm/26mm |
| | ・ 22mm/28mm | ・ 22mm/28mm |
| | Coverage: | Coverage: |
| | · Silicone Full Covered | · Silicone Full Covered |
| | • Silicone Both Bare | • Silicone Both Bare |
| | Radiopaque Markers: 8 Pt/Ir, 2 STS 316L | Radiopaque Markers: 8 Pt/Ir, 2 STS 316L |
| | Subject Device | Predicate |
| | | K211706 |
| | | (K123205 and K113551) |
| Design<br>(Introducer) | Image: Subject Device | Image: Predicate K211706 |
| | Co-axial tube type | Co-axial tube type |
| | Usable Length: | Usable Length; |
| | · 180cm (for diameter 22mm stent) | · 180cm (for diameter 22mm stent) |
| | · 220cm (for diameter 18, 20 & 22 mm stent) | · 220cm (for diameter 18, 20 & 22 mm stent) |
| | Diameter; 10.5 Fr (3.5mm) | Diameter; 10.5 Fr (3.5mm) |
| Single Use | Yes | Yes |
| Sterile | EO Sterilization | EO Sterilization |
| Method of<br>Placement | Endoscopic | Endoscopic |
| Method of<br>Deployment | Release by pulling outer sheath | Release by pulling outer sheath |
| Materials | Stent - Nitinol, Pt/Ir, STS316L | Stent - Nitinol, Pt/Ir, STS316L |
| | Cover – Silicone | Cover – Silicone |
| | Introducer – Teflon, PE, ABS | Introducer – Teflon, PE, ABS |
# Technological Comparisons
The table below compares the key technological feature of the subject devices to the predicate device (Taewoong Esophageal TTS Stent, K211706).
{5}------------------------------------------------
#### Table 2: Technological Comparison
{6}------------------------------------------------
### 7. PERFORMANCE DATA
# Biocompatibility Testing
The subject device in its final finished form is identical to the Esophageal TTS Stent cleared in K211706, K123205, and K113551, in formulation, processing, sterilization, and geometry and no other chemicals have been added (e.g., plasticizers, fillers, additives, cleaning agents, mold release agents).
Therefore, no additional biocompatibility information is required to establish substantial equivalence.
# Electrical safety and electromagnetic compatibility (EMC)
Not applicable. The device contains no electric components, generates no electrical emissions, and uses no electrical energy of any type.
# Software Verification and Validation Testing
Not applicable. The device contains no software.
{7}------------------------------------------------
# Bench Testing
No design modifications have been made to the cleared Esophageal TTS Stent. The only change is to the Indications for Use. An assessment of the change was performed, and it was determined that the modified Indications for Use do not introduce any new risks. Thus, testing previously performed for the Esophageal TTS Stent is appropriate and sufficient for the modified Indications for Use. The testing on the Esophageal TTS Stent is consistent with the testing performed on the reference device which uses this identical Indications for Use statement. Therefore, no additional performance data is required.
# Animal Testing
Not applicable. Animal studies are not necessary to establish the substantial equivalence of this device.
# Clinical Data
Not applicable. Clinical studies are not necessary to establish the substantial equivalence of this device.
#### CONCLUSION 8.
Based on the modified Indications for Use, the technological characteristics and materials of construction, the Esophageal TTS Stent can be found substantially equivalent to the identified predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.