The HANAROSTENT® Esophagus TTS (CCC) and HANAROSTENT® Esophagus TTS (NCN) are intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas.
Device Story
Self-expanding metal stent (SEMS) system for esophageal strictures and fistulas; consists of Nitinol wire stent, gold radiopaque markers, silicone membrane, and polyester repositioning lasso. Delivered via through-the-scope (TTS) delivery system compatible with 3.7mm endoscope working channels. Operated by physicians in clinical settings. Stent loaded into distal delivery device; deployed by pulling outer sheath; expands to exert outward radial force on esophageal lumen. Radiopaque markers facilitate placement accuracy. Stent maintains patency; occludes fistulas. Single-use device.
Clinical Evidence
No clinical performance data submitted. Bench testing only, including deployment force, deployment accuracy, expansion/compression force, dimensions, corrosion, tensile strength, foreshortening, trackability, repositioning force, and MR safety/compatibility.
Technological Characteristics
Self-expanding metal stent (SEMS) made of Nitinol wire; gold radiopaque markers; silicone membrane; polyester repositioning lasso. Delivery system made of polymeric materials. Compatible with 3.7mm endoscope working channel. Single-use. Non-clinical assessment per ISO 10993-1 and Nitinol guidance.
Indications for Use
Indicated for patients with esophageal strictures caused by intrinsic or extrinsic malignant tumors and concurrent esophageal fistulas requiring maintenance of esophageal luminal patency.
Regulatory Classification
Identification
An esophageal prosthesis is a rigid, flexible, or expandable tubular device made of a plastic, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the esophagus. The metal esophageal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.
Special Controls
The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”
Predicate Devices
Esophageal TTS Stent (currently commercialized as the Niti-STM TTS Esophageal Stent) (K123205)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
May 20, 2020
M.I. Tech Co., Ltd % Beryl St. Jeanne, MS, RAC Regulatory Consultant Namsa 400 Highway 169 South Suite 500 Minneapolis, MN 55426
Re: K200860
Trade/Device Name: HANAROSTENT Esophagus TTS (CCC), HANAROSTENT Esophagus TTS (NCN) Regulation Number: 21 CFR 878.3610 Regulation Name: Esophageal prosthesis Regulatory Class: II Product Code: ESW Dated: March 31, 2020 Received: April 1, 2020
## Dear Beryl St. Jeanne:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200860
Device Name
HANAROSTENT® Esophagus TTS (CCC); HANAROSTENT® Esophagus TTS (NCN)
Indications for Use (Describe)
The HANAROSTENT® Esophagus TTS (CCC) and HANAROSTENT® Esophagus TTS (NCN) are intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic malignant tumors, and occlusion of concurrent esophageal fistulas.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------|------------------------------------------------------------|
| <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for M.I.Tech. The logo consists of the letters "MIT" in a stylized blue font, with a red dot above the "I". To the right of the logo is the text "M.I.Tech" in a larger, bolder blue font. Below the logo is the website address "www.mitech.co.kr".
HEAD OFFICE 174, HABUK2-GIL, JINWI-MYEON, PYEONGTAEK-SI, GYEONGGI-DO, 17706, REPUBLIC OF KOREA TEL +82 31 662 5645 FAX +82 31 662 5648
#### 6 510(k) Summary
| Preparation Date | March 31, 2020 | |
|------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| Submitter | M.I. Tech Co., Ltd.<br>174 Habuk 2-gil, Jinwi-myeon,<br>Pyeongtaek-si, Gyeonggi-do<br>17706, Republic of Korea<br>Phone: 82-31-662-5645<br>Fax: 82-31-662-5648 | |
| Primary Contact | Inae Kim<br>Medical Affairs Team Manager<br>M.I. Tech Co., Ltd.<br>174 Habuk 2-gil, Jinwi-myeon,<br>Pyeongtaek-si, Gyeonggi-do<br>17706, Republic of Korea<br>Email: inae116@mitech.co.kr<br>Phone: 82-70-4304-7450<br>Fax: 82-2-3463-4703 | |
| Subject Devices | Trade Name: | HANAROSTENT® Esophagus TTS (CCC) |
| | Device: | Prosthesis, Esophageal |
| | Regulation Description: | Esophageal prosthesis |
| | Review Panel: | Gastroenterology/Urology |
| | Regulation Number: | 21 CFR 878.3610 |
| | Device Class: | Class II |
| | Product Code: | ESW |
| | Regulation Medical Specialty: | General & Plastic Surgery |
| | Trade Name: | HANAROSTENT® Esophagus TTS (NCN) |
| | Device: | Prosthesis, Esophageal |
| | Regulation Description: | Esophageal prosthesis |
| | Review Panel: | Gastroenterology/Urology |
| | Regulation Number: | 21 CFR 878.3610 |
| | Device Class: | Class II |
| | Product Code: | ESW |
| | Regulation Medical Specialty: | General & Plastic Surgery |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for M.I.Tech. The logo consists of the company name in blue letters, with a red dot above the "i" in "MIT". Below the logo is the company's website address, www.mitech.co.kr.
HEAD OFFICE
174, HABUK2-GIL, JINWHMYEON, PYEONGTAEK-SI, GYEONGGI-DO,
17706, REPUBLIC OF KOREA
TEL +82 31 662 5645 FAX +82 31 662 5648
| Intended Use /<br>Indications for Use | The HANAROSTENT® Esophagus TTS (CCC) and HANAROSTENT®<br>Esophagus TTS (NCN) are intended for maintaining esophageal luminal patency<br>in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and<br>occlusion of concurrent esophageal fistulas. | Technological | The subject devices and predicate device have substantially equivalent |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Description | This self-expanding tubular prosthesis is designed to maintain esophageal luminal<br>patency in esophageal strictures caused by intrinsic and/or extrinsic malignant<br>tumors, and occlusion of concurrent esophageal fistulas. It consists of a self-<br>expandable metal stent and a through-the scope delivery system. The self-expandable<br>metal stent is made of nickel titanium alloy (Nitinol) wire, radiopaque markers<br>made of gold wire, fully or partially covered silicone membrane, and repositioning<br>lasso made of polyester. The delivery device is made of polymeric materials. The<br>stent is loaded into the distal part of the delivery device, and expanded in the body<br>by pulling the outer sheath of the delivery device. The delivery system is<br>compatible with a minimum 3.7mm working channel of a therapeutic endoscope.<br>The HANAROSTENT® Esophagus TTS (CCC) and HANAROSTENT®<br>Esophagus TTS (NCN) are intended for single use only. | Characteristics | technological characteristics with only minor differences regarding: |
| Predicate Device | Device Classification Name: Prosthesis, Esophageal<br>510(k) Number: K123205<br>Device Name: Esophageal TTS Stent<br>(currently commercialized as the<br>Niti-STM TTS Esophageal Stent)<br>Applicant: Taewoong Medical Co., Ltd.<br>Regulation Number: 21 CFR 878.3610<br>Classification Product Code: ESW<br>Decision Date: 10/09/2013<br>Regulation Medical Specialty: General & Plastic Surgery<br>510(k) Review Panel: Gastroenterology/Urology<br>Device Class: Class II | | Packaging: The subject devices do not have a Tyvek pouch. The predicate device has a Tyvek pouch. Radiopaque marker material and quantity: The subject devices have 12 gold radiopaque markers. The predicate device has 10 platinum-iridium radiopaque markers. Stent lengths: The subject device is offered in 60, 70, 80, 90, 100, 110, 120, 130, 140, and 150mm stent lengths. The predicate device is offered in 60, 80, 100, 120, 140, and 150mm stent lengths. Lasso materials: The subject devices use a polyester lasso material. The predicate device uses a nylon lasso material. Delivery device usable lengths: The subject devices offer an 1800mm or 2300mm delivery device. The predicate device offers an 1800mm delivery device. Performance – Bench: The subject and predicate devices have equivalent expansion forces. The subject and predicate devices have equivalent compression forces. The subject device's 1800mm delivery device and the predicate device's 1800mm delivery device have equivalent deployment forces. The subject device's 2300mm delivery device has greater deployment force than the predicate device's 1800mm delivery device. |
| Mechanism of Action | The stent is loaded by the delivery device. Upon deployment of the stent, it<br>imparts an outward radial force on the luminal surface of the esophagus to<br>establish patency. The stent is constrained and loaded between the two sheaths.<br>The delivery device and stent are introduced to the intended target location<br>through the use of a 0.035 inch guidewire. Radiopaque markers allow visualizing<br>and measuring placement accuracy. The delivery device is removed and discarded<br>after deployment of the stent. | | |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for M.I.Tech. The logo is blue and features a stylized "MIT" with a red dot above the "I". Below the logo is the company's website address, www.mitech.co.kr.
HEAD OFFICE
174, HABUK2-GIL, JINWI-MYEON, PYEONGTAEK-SI, GYEONGGI-DO,
17706, REPUBLIC OF KOREA TEL +82 31 662 5645 FAX +82 31 662 5648
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for M.I.Tech. The logo is blue and features a stylized "M" with a red dot above it. Below the logo is the website address: www.mitech.co.kr.
HEAD OFFICE
174, HABUK2-GIL, JINWHMYEON, PYEONGTAEK-SI, GYEONGGI-DO,
17706, REPUBLIC OF KOREA
TEL +82 31 662 5645 FAX +82 31 662 5648
| FDA Guidance<br>Documents | The following FDA guidance documents were consulted in preparing this<br>premarket submission:<br>Shelf Life of Medical Devices, issued April 1991 Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile, issued January 2016 Guidance for Industry: Pyrogen and Endotoxins Testing: Questions and Answers, issued June 2012 Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", issued June 16, 2016 Guidance for The Content of Premarket Notifications for Esophageal and Tracheal Prostheses, issued April 28, 1998 Technical Considerations for Non-Clinical Assessment of Medical Devices containing Nitinol, draft issued April 2019 Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment, issued December 2014 Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment, issued August 2019 |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance - Bench | Bench testing was performed to confirm the safety and effectiveness of the<br>proposed subject devices as compared to the predicate devices. Performance<br>testing was performed as per the design control system. The following tests were<br>conducted:<br>Deployment force Deployment accuracy Deployment in a simulated environment Expansion force Compression force Dimensions Corrosion Tensile strength Foreshortening Trackability Repositioning force MR safety and compatibility |
| Performance - Animal | No animal performance data is submitted in this 510(k). |
| Performance - Clinical | No clinical performance data is submitted in this 510(k). |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for M.I.Tech, a company with a website at www.mitech.co.kr. The logo features the company name in a stylized font, with the letters "MIT" connected and a red dot above the "I". The website address is printed in a smaller font below the logo.
HEAD OFFICE
174, HABUK2-GIL, JINWHMYEON, PYEONGTAEK-SI, GYEONGGI-DO,
17706, REPUBLIC OF KOREA
TEL +82 31 662 5645 FAX +82 31 662 5648
| Substantial<br>Equivalence | The subject devices are substantially equivalent to the predicate device when<br>evaluating intended use and technological characteristics. The subject devices have the identical intended use as the predicate device. There are no differences between the subject device and predicate device with respect to intended use. The subject devices have different indications for use than the predicate device. The subject devices are additionally indicated for occlusion of concurrent esophageal fistulas. The subject devices and predicate device are substantially equivalent with only minor technological differences. These differences do not raise new questions of safety and effectiveness. |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion | This comparison demonstrates the subject devices are substantially equivalent to the predicate device. The subject devices are as safe and effective as the predicate device and will perform as intended. Therefore, M.I. Tech respectfully requests market clearance for the subject devices. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.