NITI-S ESOPHAGEAL STENT

K080782 · Taewoong Medical Co., Ltd. · ESW · Jan 30, 2009 · General, Plastic Surgery

Device Facts

Record IDK080782
Device NameNITI-S ESOPHAGEAL STENT
ApplicantTaewoong Medical Co., Ltd.
Product CodeESW · General, Plastic Surgery
Decision DateJan 30, 2009
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 878.3610
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Niti-S Esophageal Stent is intended for use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors.

Device Story

Niti-S Esophageal Stent is a self-expanding, metallic tubular prosthesis designed to restore esophageal structure and function. The device consists of a Nitinol (Nickel-Titanium alloy) wire stent and a flexible introducer system. During clinical use, a physician deploys the stent into the esophagus; the stent exerts outward radial force on the luminal surface to establish patency in the presence of malignant strictures. The device is provided in various diameters (16-20 mm body; 24-28 mm head) and lengths (30-120 mm) to accommodate individual patient anatomy. The introducer system utilizes a co-axial tube design where the stent is constrained by an outer sheath; pulling the sheath releases the stent for deployment. The device is intended for prescription use.

Clinical Evidence

Bench testing only. Biocompatibility assessment performed on patient-contact materials. No clinical data provided.

Technological Characteristics

Self-expanding stent constructed of Nickel-Titanium alloy (Nitinol). Available in body diameters of 16, 18, or 20 mm and lengths of 30, 50, 70, 90, or 120 mm. Head diameters are 24, 26, or 28 mm with 15 mm length. Deployment via co-axial introducer system with outer sheath. Non-powered, mechanical device.

Indications for Use

Indicated for patients with esophageal strictures resulting from intrinsic or extrinsic malignant tumors.

Regulatory Classification

Identification

An esophageal prosthesis is a rigid, flexible, or expandable tubular device made of a plastic, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the esophagus. The metal esophageal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.

Special Controls

The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary This summary of 510(k)-safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92. Date: March 12, 2008 #### 1. Submitter: ## JAN 30 2009 0807 82 | Name: | Taewoong Medical Co., Ltd.<br>1-5 Gomak-ri, Wolgot-myeon,<br>Gimpo-si, Gyeonggi-do,<br>Republic of Korea 415871 | |----------|------------------------------------------------------------------------------------------------------------------| | Contact: | J.H. Nam /Director<br>Phone +82 31 996-0641 | #### 2. Device: | Proprietary Name: | Niti-S Esophageal Stent | |-----------------------|-------------------------| | Common Name: | Prosthesis, Esophageal | | Classification Name: | Esophageal Prosthesis | | Classification: | 21 CFR 878.3610 | | Product Code: | ESW | | Third Party Reviewed: | NO | #### 3. Predicate Device: Microvasive Ultraflex Esophageal Stent System Boston Scientific Corporation K940838 Niti-S Stent & Introducer, Model Esophageal, Taewoong Medical Co., Ltd K041648 Cook Esophageal Z Stent with DUA Anti-Reflux Wilson-Cook Medical K011591 #### 4. Description: The proposed Niti-S Esophageal Stent consists of an implantable metallic stent and a flexible introducer system. The stent is a rigid, flexible, and expandable tubular device Taewoong Medical Co., Ltd. 11 {1}------------------------------------------------ produ made of a Nitinol wire that is intended to be implanted to restore the structure and/or function of the esophagus. This device also includes the introducer. Upon deployment, the stent imparts an outward radial force on the luminal surface of the duct to establish patency. #### 5. Indications for use: Niti-S Esophageal Stent is intended for use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors. #### Technological Characteristics: 6. The proposed Niti-S Esophageal Stent is a self-expanding Nickel Titanium alloy (Nitinol) stent mounted on an introducer. The stent is available in multiple sizes. The proposed stent's body diameter is 16, 18 or 20 mm with a 30, 50, 70, 90, or 120 mm length and each head's diameter is 24, 26 or 28 mm and 15 mm in length. Physician preference and individual patient condition and/or anatomy will determine the appropriate size. Niti-S Esophageal Stent is substantially equivalent to Microvasive Ultraflex Esophageal Stent System, Boston scientific Corporation K940838, Cook Esophageal Z Stent with DUA Anti-Reflux, Wilson-Cook K011591 and Niti-S Stent & Introducer, Model Esophageal, Taewoong Medical Co., Ltd K041648. The devices have the same intended use and self-expanding stents constructed of Nitinol. The stents are mounted on an introducer system; however the designs are different. The predicate Microvasive Ultraflex Esophageal Stent System is constrained onto the introducer system with knitted loop by pulling string. The Niti-S Stent & Introducer, Model Esophageal and the proposed Niti-S Esophageal Stent utilized a co-axial tube and the stents are constrained onto the introducer system shaft by the outer sheath by pulling outer sheath. #### 7. Performance Data: Laboratory testing regarding characteristics was performed on the Niti-S Esophageal Stent to verify its safety and performance. A biocompatibility assessment was performed on the patient contact materials of the Niti-S Esophageal Stent. #### 8. Conclusions: Taewoong Medical Co., Ltd. concludes that the Niti-S Esophageal Stent is safe and effective. It is substantially equivalent to the predicate devices, Microvasive Ultraflex Esophageal Stent System, Cook Esophageal Z Stent with DUA Anti-Reflux and Niti-S Stent & Introducer, Model Esophageal. END K080982 {2}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. ### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Taewoong Medical Co., Ltd. c/o Cathryn N. Cambria, RAC Consultant Arkin Consulting Group, LLC 5536 Trowbridge Drive DUNWOODY GEORGIA 30338 ## JAN 3 0 2009 Re: K080782 > Trade/Device Name: Taewoong Niti-S Esophageal Stent System Regulation Number: 21 CFR §878.3610 Regulation Name: Esophageal prosthesis Regulatory Class: II Product Code: FGE Dated: December 23, 2008 Received: December 30, 2008 Dear Ms. Cambria: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter. | 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 | |----------------|----------------------------------|----------------| | 21 CFR 884.xxx | (Obstetrics/Gynecology) | (240) 276-0115 | | 21 CFR 892.xxx | (Radiology) | (240) 276-0120 | | Other | | (240) 276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppor/index.html. Sincerely yours. Laura B. Merrill anine M. M. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K080782 10 510(k) Submission - Niti-S Esophageal Stent 510(k) Number K 080782 Device Name: Niti-S Esophageal Stent Indication for use: Niti-S Esophageal Stent is intended for use in esophageal w strictures caused by intrinsic and/or extrinsic malignant tumors. Prescription Use (Per 21CFR801 Subpart D) Over-The-Counter Use (Per 21CFR807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) OR Concurrence of CDRH, Office of Device Evaluation (ODE) Arvito Whay Division Sign-Off Division of Reproductive, Abdominal, and Radiological Devices 8079 510(k) Number Taewoong Medical Co., Ltd.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...