SML-OSA2 Appliances

K162816 · Selane Products, Inc. · LQZ · Jul 27, 2017 · Dental

Device Facts

Record IDK162816
Device NameSML-OSA2 Appliances
ApplicantSelane Products, Inc.
Product CodeLQZ · Dental
Decision DateJul 27, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.5570
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The SML OSA2 Appliances are intended to reduce nighttime snoring and mild to moderate obstructive sleep apnea in adults. The SML-OSA2 appliances are worn while sleeping to support the lower jaw in a forward position prescribed by the dentist. The customized appliance is inserted and removed by the patient and adjusted by the prescribing dentist.

Device Story

SML-OSA2 Appliances are customized, prescription intraoral dental devices; used to reduce snoring and mild to moderate obstructive sleep apnea in adults. Device consists of two arch forms (maxillary and mandibular) that fit over dentition; mechanically repositions mandible forward during sleep. Forward mandibular position prevents tongue from obstructing airway; reduces snoring/apnea. Device is manufactured by dental laboratory; inserted/removed by patient; adjusted by prescribing dentist. Five models available (ASA, Clear Sleep Adjustable Dorsal, Clear Sleep Adjustable Anterior, QNA, V-force); designs based on Dorsal or Herbst principles. Materials include dental acrylic, stainless steel, POM, and orthodontic elastomers. No clinical testing performed; substantial equivalence based on design, material, and principle of operation similarity to predicate.

Clinical Evidence

No clinical data. Bench testing only; biocompatibility assessment based on established safety of materials in previously cleared intraoral devices per ISO 7405 and ISO 10993-1.

Technological Characteristics

Customized intraoral dental appliances. Materials: Polymethylmethacrylate (PMMA), SBS/EVA laminate, Polyurethane/copolyester, Stainless Steel 304/316, Polyoxymethylene (POM), and latex rubber elastomers. Principle: Mechanical mandibular advancement via two connected arch forms. Designs: Dorsal or Herbst-based. Non-sterile, reusable, prescription-only. Biocompatibility per ISO 7405 and ISO 10993-1.

Indications for Use

Indicated for adults with mild to moderate obstructive sleep apnea and nighttime snoring. Worn during sleep to support the lower jaw in a forward position.

Regulatory Classification

Identification

Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.

Special Controls

*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it. The symbol is placed to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 27, 2017 Selane Products, Inc. % Carolyn Primus Consultant Primus Consulting 7046 Owl's Nest Terrace Bradenton, Florida 34203 Re: K162816 Trade/Device Name: Sml-osa2 Appliances Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LQZ, LRK Dated: June 25, 2017 Received: June 29, 2017 Dear Ms. Carolyn Primus: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {1}------------------------------------------------ or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Mary S. Runner -S for Lori Wiggins Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K162816 Device Name SML-OSA2 Appliances #### Indications for Use (Describe) The SML OSA2 Appliances are intended to reduce nighttime snoring and mild to moderate obstructive sleep apnea in adults. The SML-OSA2 appliances are worn while sleeping to support the lower jaw in a forward position prescribed by the dentist. The customized appliance is inserted and removed by the patient and adjusted by the prescribing dentist. | Type of Use (Select one or both, as applicable) | | |---------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------| | <div> <span> <b>Prescription Use (Part 21 CFR 801 Subpart D)</b> </span> </div> | | | | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW * The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # K162816 510(k) Summary | Submitted by: | Space Maintainer Laboratory, Division of Selane products<br>9129 Lurline Ave.<br>Chatsworth, CA 91311 | |-----------------------|-------------------------------------------------------------------------------------------------------| | Contact Person: | Mr. John Christian | | Date Prepared: | July 25, 2017 | | Device Trade Name: | SML-OSA2 Appliances | | Device Description: | Device, Jaw Repositioning | | Device Class: | Class II | | Regulation: | 872.5570 | | Classification Panel: | Dental | | Product Code: | LQZ, LRK | | Predicate Device: | ATG/SM-OSA Appliances (K130130). | Device Description: The SML-OSA2 APPLIANCES include five device models: Adjustable Sleep Appliance (ASA), Clear Sleep Adjustable Dorsal, Clear Sleep Adjustable Anterior, Quiet Night Appliance (QNA) and V-force. The SML-OSA2 Appliances are customized, intraoral dental devices designed to reduce snoring and mild to moderate obstructive sleep apnea in adults. The devices fit over the dentition and reposition the mandible in a slightly protrusive position. The SML-OSA2 APPLIANCES consist of 2 arch forms that are linked to advance the mandible, and are worn while sleeping to support the lower jaw in a forward position prescribed by the dentist. The customized appliance is inserted and removed by the patient and adjusted by the prescribing dentist. The SML-OSA2 APPLIANCES are made by Appliance Therapy Group/Space Maintainer Laboratory, which is owned/operated by Selane Products, Inc. Appliance Therapy Group/Space Maintainer Laboratory will manufacture these customized, prescription devices. These reusable (repeated use by one patient) devices will be provided to the patient non-sterile. Indications for Use: The SML-OSA2 Appliances are intended to reduce nighttime snoring and mild to moderate obstructive sleep apnea in adults. The SML-OSA2 appliances are worn while sleeping to support the lower jaw in a forward position prescribed by the dentist. The customized {4}------------------------------------------------ appliance is inserted and removed by the patient and adjusted by the prescribing dentist. ## TECHNOLOGICAL CHARACTERISTICS: ## Intended Use The Indications for Use for the SML-OSA2A Appliances are identical to the ATG/SM-OSA Appliances' Indications for Use. Both the proposed and predicate devices are intended to reduce nighttime snoring and mild to moderate obstructive sleep apnea in adults A more in-depth comparison of the materials of construction is provided in the Substantial Equivalence section. ## Design The SML-OSA2 APPLIANCES are made primarily of dental acrylic. Three modified designs (ASA, Clear Sleep Adjustable Dorsal, and Clear Sleep Adjustable Anterior) are a variation of the Dorsal design. Two modified designs (QNA and V-force) are a variation of the Herbst design. Stainless steel or elastomeric parts are incorporated to allow adjustment of the appliances. These designs may have dual-laminate dental polymers in place of the acrylic. Table S-1 summarizes the design characteristics of the five models of the proposed device. The function, scientific concept and principle of operation are the same for each device design. The principle of operation is that two connected arch forms maneuver the mandible forward. The position of the mandible prevents the tongue from sliding backward or downward for a prone sleeping individual. Therefore, the airway will be more open, contributing to the reduction in snoring and mild to moderate sleep apnea. | SML-OSA2<br>Appliance | Similarities to<br>K130130 | Differences from K130130 | |-------------------------------------|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Adjustable Sleep<br>Appliance (ASA) | Similar to the K130130<br>Dorsal. | Adjustable arch width with midline screw on<br>maxillary arch; partial anterior coverage. | | Clear Sleep<br>Adjustable Dorsal | Similar to the K130130<br>Dorsal. | Acrylic wings are replaced with acrylic<br>interlocking tabs. | | Clear Sleep<br>Adjustable Anterior | Similar to the K130130<br>Dorsal. | Adjustments in the anterior instead of the buccal<br>corridor (area between the lateral margin of the<br>lips and the outside of the posterior dental arches). | | Quiet Night<br>Appliance (QNA) | Similar to the K130130<br>Herbst. | Stainless steel connectors are replaced with polymer<br>(POM) connectors. | | V-force | Similar to the K130130<br>Herbst. | Stainless steel connecting hinges are replaced with<br>orthodontic elastomers and acrylic tabs. Acrylic<br>interlocking tabs are included to advance the<br>mandible. | Table S-1: Five Design Models of the Subject SML-OSA2 {5}------------------------------------------------ ## APPLIANCES' Materials SML-OSA2 Appliances are manufactured of Stainless Steel 304, Stainless Steel 316, Polymethylmethacrylate, SBS/EVA laminate or Polyurethane/copolyester, Polyoxymethylene (POM), and intraoral latex rubber elastomers. Table S-2.1 through Table S-2.5 list the materials of each of the SML-OSA2 Appliances, and indicate that these materials have been used in the predicate device, in previously cleared dental devices, or have otherwise been proven safe for intraoral use. | MATERIAL | DEVICE PART | Predicate Device<br>ATG/SM-OSA | Subject Device<br>SML-OSA2 | Acceptable Material<br>for Medical Device Use | |--------------------------------------------------|-----------------------------------------------------------|--------------------------------|----------------------------|-----------------------------------------------| | Stainless Steel 304/316 | Connectors | Yes | Yes | K130130 | | Polymethyl Methacrylate<br>(PMMA Acrylic) | Maxillary & Mandibular<br>Arch Forms | Yes | Yes | K130130 | | SBS/EVA laminate or<br>Polyurethane/ copolyester | Maxillary & Mandibular<br>Arch Forms | Yes | Yes | K130130 | | Vitallium 2000 partial<br>denture alloy | Maxillary | Yes | No | K130130 | | | Table S-2.2: Clear Sleep Adjustable Dorsal Construction | | | | | MATERIAL | DEVICE PART | Predicate Device<br>ATG/SM-OSA | Subject Device<br>SML-OSA2 | Acceptable Material<br>for Medical Device Use | | Stainless Steel 304/316 | Connectors | Yes | No | K130130 | | Polymethyl Methacrylate<br>(PMMA Acrylic) | Maxillary & Mandibular<br>Arch Forms | Yes | Yes | K130130 | | | Connectors | No | Yes | K130130 | | SBS/EVA laminate or<br>Polyurethane/ copolyester | Maxillary & Mandibular<br>Arch Forms | Yes | Yes | K130130 | | | Table S-2.3: Clear Sleep Adjustable Anterior Construction | | | | | MATERIAL | DEVICE PART | Predicate Device<br>ATG/SM-OSA | Subject Device<br>SML-OSA2 | Acceptable Material<br>for Medical Device Use | | Stainless Steel 304/316 | Connectors | Yes | No | K130130 | | Polymethyl Methacrylate<br>(PMMA Acrylic) | Maxillary & Mandibular<br>Arch Forms | Yes | Yes | K130130 | | SBS/EVA laminate or<br>Polyurethane/ copolyester | Maxillary & Mandibular<br>Arch Forms | Yes | Yes | K130130 | | | Table S-2.4: Quiet Night Appliance (QNA) Construction | | | | | MATERIAL | DEVICE PART | Predicate Device<br>ATG/SM-OSA | Subject Device<br>SML-OSA2 | Acceptable Material<br>for Medical Device Use | | Stainless Steel 304/316 | Connectors | Yes | No | K130130 | | Polyoxymethylene (POM) | Connectors | No | Yes | FDA Biomaterials<br>Compendium (1) | | Polymethyl Methacrylate<br>(PMMA Acrylic) | Maxillary & Mandibular<br>Arch Forms | Yes | Yes | K130130 | | SBS/EVA laminate or<br>Polyurethane/ copolvester | Maxillary & Mandibular<br>Arch Forms | Yes | Yes | K130130 | 1 Polyoxymethylene (POM) is a recognized polymeric thermoplastic in the FDA Biomaterials Compendium. #### 510(k) Summary of Traditional Premarket Notification for SML-OSA2 APPLIANCES Space Maintainers Laboratory; Division of Selane Products, Inc. K162816 – CONFIDENTIAL TRADE SECRET {6}------------------------------------------------ #### Table S-2.5: V-force Construction | MATERIAL | DEVICE PART | Predicate<br>Device<br>ATG/SM-OSA | Subject<br>Device SML-<br>OSA2 | Acceptable Material<br>for Medical Device<br>Use | |--------------------------------------------------|--------------------------------------|-----------------------------------|--------------------------------|---------------------------------------------------| | Stainless Steel 304/316 | Connectors | Yes | No | K130130 | | Latex Rubber Elastomers | Connectors | No | Yes | FDA Biomaterials<br>Compendium(1),<br>K935144 (2) | | Polymethyl Methacrylate<br>(PMMA Acrylic) | Maxillary & Mandibular<br>Arch Forms | Yes | Yes | K130130 | | SBS/EVA laminate or<br>Polyurethane/ copolyester | Maxillary & Mandibular<br>Arch Forms | Yes | Yes | K130130 | ı Natural latex is a recognized polymeric thermoset/elastomer in the FDA Biomaterials Compendium. 2 Latex rubber elastomers will be purchased from Dentaurum GmbH & Co. KG, Registration Number 9611458 or another FDA registered and listed establishment with FDA clearance for intraoral latex elastics. #### Substantial Equivalence Table S-3 has a comparison of the predicate and the subject devices with regard to their substantial equivalence. | | Predicate Device | Subject Device SML-OSA2 | |--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | ATG/SM-OSA Appliances | Appliances | | 510(k) # | K130130 | K162816 | | Class | II | II | | Pro Code | LQZ, LRK | LQZ, LRK | | Regulation # | 872.5570 | 872.5570 | | Regulation Name | Intraoral Devices for Snoring and<br>Obstructive Sleep Apnea | Intraoral Devices for Snoring and<br>Obstructive Sleep Apnea | | Indications for Use | The ATG/SM-OSA appliances are<br>intended to reduce nighttime snoring and<br>mild to moderate obstructive sleep apnea<br>in adults. The ATG/SM-OSA appliances<br>are worn while sleeping to support the | The SML-OSA2 appliances are intended to<br>reduce nighttime snoring and mild to<br>moderate obstructive sleep apnea in adults.<br>The SML-OSA2 appliances are worn while<br>sleeping to support the lower jaw in a | | | lower jaw in a forward position prescribed<br>by the dentist. The customized appliance is | forward position prescribed by the dentist.<br>The customized appliance is inserted and | | | inserted and removed by the patient and | removed by the patient and adjusted by the | | | adjusted by the prescribing dentist. | prescribing dentist. | | | Models | Dorsal Herbst | | Patient Contact | Acrylic, polymers, and stainless steel | Acrylic, polymers, elastomers, and stainless steel | | Biocompatibility per<br>ISO 7405 & 10993-1 | External communicating (>30 days)<br>Surface-contacting (mucosa) | External communicating (>30 days)<br>Surface-contacting (mucosa) | | Sterility | Non-sterile | Non-sterile | | Type of Device | Customized, prescription | Customized, prescription | | Scientific Principle | Moves lower mandible forward during sleep | Moves lower mandible forward during sleep | | Design | Upper and lower arches to support jaw | Upper and lower arches to support jaw | | Adjustable or Fixed | Adjustable | Adjustable | | Connection Parts | Stainless steel connectors | Polyoxymethylene (POM) connectors, acrylic<br>spacers, and orthodontic elastomers | Table S-3: Substantial Equivalence Comparison of Devices 510(k) Summary of Traditional Premarket Notification for SML-OSA2 APPLIANCES Space Maintainers Laboratory; Division of Selane Products, Inc. K162816-CONFIDENTIAL TRADE SECRET {7}------------------------------------------------ The indications for use are identical for the predicate K130130 and subject K162816 devices. No significant differences exist in technology, intended use, or design between the subject devices and the predicates selected. ## Performance Standards: - ISO 7405 Second edition 2008-12-15, dentistry evaluation of biocompatibility of medical . devices used in dentistry as published in the Federal Register (Food and Drug Administration Modernization Act of 1997: Modifications to the List of Recognized Standards, Recognition List Number: 036; 07/09/2014). - The special control for these devices is the FDA guidance document entitled "Class II . Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea: Guidance for Industry and FDA." [Nov. 12, 2002] ## Biocompatibility: The predicate and subject devices are considered to be external communicating devices of permanent exposure (>30 days) per ISO 7405 and surface-contacting devices (mucosa) of permanent contact (>30 days) per ISO 10993-1. No changes have been made in the primary materials of construction from the sponsor's own predicate device. In some models, stainless steel connecting parts have been substituted with Polyoxymethylene (POM) or intraoral latex rubber elastomers. No biocompatibility tests have been performed for the modified devices; the new materials have als o been used in other cleared intraoral devices and implantable devices. ## Clinical Testing: No clinical testing is required for these devices. ### Risk Analysis: A risk analysis including an evaluation of the materials of construction and design was performed. The function of mandibular advancement devices requires that the prescribing dentist be cognizant of the potential for TMJ soreness, soft tissue soreness, and dentition complications (soreness, motion, loosening) by mandibular advancement. The contraindications, warnings, precautions, storage directions, prescription preparation instructions, fitting and adjustment directions are provided to avoid potential problems from arising or persisting with the dentition, tissue, or joints. A risk related to the polymer spacers was addressed. No significant new risks were identified. ### Conclusions: The subject devices have identical indications for use to the predicate devices. The intended use, materials of construction, technology, principle of operation, and designs of the SML-OSA2 Appliances are comparable to the legally marketed predicate. Therefore, the proposed SML-OSA2 Appliances are substantially equivalent to the predicate.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...