The NightBlocks Appliance

K192581 · Advanced Facialdontics, LLC · LQZ · Jan 28, 2020 · Dental

Device Facts

Record IDK192581
Device NameThe NightBlocks Appliance
ApplicantAdvanced Facialdontics, LLC
Product CodeLQZ · Dental
Decision DateJan 28, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.5570
Device ClassClass 2
AttributesTherapeutic

Intended Use

The NightBlocks™ Appliance is intended to reduce nighttime snoring and mild to moderate obstructive sleep apnea in adults. The NightBlocks™ Appliance is worn while sleeping to support the lower jaw in a forward position prescribed by the dentist. The customized appliance is inserted and removed by the patient and adjusted by the prescribing dentist.

Device Story

NightBlocks™ is a custom-fabricated, removable intraoral mandibular advancement device. It consists of two acrylic trays (upper and lower) that fit over dental arches; stainless-steel ball clasps provide retention; buccal-mounted stainless-steel expansion screws allow for mandibular advancement. Used in home settings, the device is inserted/removed by the patient and adjusted by a prescribing dentist. During sleep, the device holds the mandible in a forward position to increase pharyngeal airway space, preventing tongue collapse and improving airflow. This reduces snoring and mild-to-moderate obstructive sleep apnea. The device is non-sterile and requires no external power or software.

Clinical Evidence

No clinical data. Substantial equivalence is based on bench testing, risk analysis per ISO 14971, and comparison of technological characteristics to the predicate device.

Technological Characteristics

Materials: Polymethylmethacrylate (Lang Orthodontic Acrylic), medical-grade stainless steel wire (AJ Wilcock), and orthodontic stainless-steel expansion screws (Leone). Principle: Mandibular advancement via buccal expansion screws. Form factor: Custom-fitted intraoral acrylic trays. Connectivity: None. Sterilization: Provided non-sterile.

Indications for Use

Indicated for adults with nighttime snoring and mild to moderate obstructive sleep apnea.

Regulatory Classification

Identification

Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.

Special Controls

*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA acronym is in blue, and the full name of the agency, "U.S. Food & Drug Administration," is in a smaller font size below the acronym. January 28, 2020 Advanced Facialdontics LLC Scott Simonetti President 325 Lake Ave. Unit 759 St James, New York 11780 Re: K192581 Trade/Device Name: The NightBlocks Appliance Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LOZ, LRK Dated: December 17, 2019 Received: December 19, 2019 Dear Scott Simonetti: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### Indications for Use 510(k) Number (if known) K192581 Device Name The NightBlocks™ Appliance #### Indications for Use (Describe) The NightBlocks™ Appliance is intended to reduce nighttime snoring and mild to moderate obstructive sleep apnea in adults. The NightBlocks™ Appliance is worn while sleeping to support the lower jaw in a forward position prescribed by the dentist. The customized appliance is inserted and removed by the patient and adjusted by the prescribing dentist. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # nced Facialdontics® # 510(k) Summary #### 1. Owner Information | Owner's Name: | Advanced Facialdontics LLC | |---------------------|------------------------------------------------| | Owner's Address: | 325 Lake Ave<br>Unit 759<br>St James, NY 11780 | | FDA Registration #: | 3015733038 | | Telephone: | (631) 379-3902 | | Fax: | (631) 234-7985 | | Contact Person: | Dr Scott Simonetti | | Email: | ssimonettidds@yahoo.com | | Date Prepared: | 1/10/2020 | #### 2. Device Name | Trade Name: | The NightBlocks™ Appliance | |-----------------|----------------------------| | 510K Number: | K192581 | | Common Name: | Device, Jaw Repositioning | | Classification: | Class II | | Product Code: | LQZ, LRK | | Review Panel: | Dental | | Regulation: | CFR 872.5570 | # 3. Predicate Devices | Primary Predicate: | ATG/SM-OSA Appliances (K130130) | |--------------------|----------------------------------------------------------------------------------| | Reference Devices: | SML-OSA2 Appliances (K162816)<br>Lamberg Sleep Well-Smartrusion, LSW-S (K101709) | #### 4. Device Description The NightBlocks™ appliance consist of two removable, custom fabricated acrylic trays that fit separately over the upper and lower dental arches and engage each other in order to reposition the mandible in a slightly anterior position. They contain stainlesssteel ball clasps for extra retention and stainless-steel orthodontic expansion screws on {4}------------------------------------------------ the buccal portion of the upper tray that engage with the dorsal wings of the lower tray. The trays occlude on the left or right molar and premolar region. Every device is custom prescribed and fabricated for each patient from dental models and is adjustable at the time of delivery and anytime thereafter as it is intended to be used in the home setting. The principal of operation is jaw repositioning. Worn during sleep, the NightBlocks™ function as a mandibular advancement device which acts to increase the patient's pharyngeal airway space, improving their ability to exchange air during sleep. The anterior repositioning of the mandible prevents the tongue from sliding backward or downward for a prone sleeping individual. Therefore, the airway will be larger, contributing to the reduction in snoring and mild to moderate sleep apnea. Materials: Polymethylmethacrylate Polymer and Monomer (Lang Orthodontic Acrylic K141439), medical grade stainless steel wire (AJ Wilcock), and orthodontic stainlesssteel expansion screws (Leone). #### 5. Indications for Use: The NightBlocks™ Appliance is intended to reduce nighttime snoring and mild to moderate obstructive sleep apnea in adults. The NightBlocks™ Appliance is worn while sleeping to support the lower jaw in a forward position prescribed by the dentist. The customized appliance is inserted and removed by the patient and adjusted by the prescribing dentist. #### 6. Technological Characteristics Comparison to Predicate Device: Intended Use: The intended use and the Indications of Use for the proposed device and predicate are identical. Technological Characteristics: The predicate and the subject device have the exact same following technological characteristics: Principle of operation, Indications of Use, Prescription device custom fabricated, intraoral jaw repositioning device for sleep apnea, environment of use: home, removable by patient and reusable, location of advancement mechanism, provided nonsterile, and identical raw materials, manufacturing, finishing and cleaning process. The only minor design differences between the proposed device and predicate is the location of the occluding surface of the upper and lower trays, and the inclusion of additional braided stainless-steel supporting wire within the dorsal wings. {5}------------------------------------------------ | | Subject Device | Predicate Device | Reference Device | Reference Device | |------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device | NightBlocks™<br>Appliance | ATG/SM-OSA,<br>Adjustable Dorsal<br>Appliance<br>K130130 | SML-OSA2<br>Appliances<br>K162816 | Lamberg Sleep<br>Well Smartrusion<br>K101709 | | Product Codes | LQZ, LRK | LQZ, LRK | LQZ, LRK | LQZ, LRK | | Indications<br>for Use | The NightBlocks™<br>Appliance is intended<br>to reduce nighttime<br>snoring and mild to<br>moderate obstructive<br>sleep apnea in adults.<br>The NightBlocks™<br>Appliance is worn<br>while sleeping to<br>support the lower jaw<br>in a forward position<br>prescribed by the<br>dentist. The<br>customized appliance<br>is inserted and<br>removed by the<br>patient and adjusted<br>by the prescribing<br>dentist. | The ATG/SM-OSA<br>Appliances are<br>intended to reduce<br>nighttime snoring and<br>mild to moderate<br>obstructive sleep<br>apnea in adults. The<br>ATG/SM-OSA<br>appliances are worn<br>while sleeping to<br>support the lower jaw<br>in a forward position<br>prescribed by the<br>dentist. The<br>customized appliance<br>is inserted and<br>removed by the<br>patient and adjusted<br>by the prescribing<br>dentist. | The SML-OSA2<br>Appliances are<br>intended to reduce<br>nighttime snoring<br>and mild to<br>moderate<br>obstructive sleep<br>apnea in adults. The<br>SML-OSA2<br>appliances are worn<br>while sleeping to<br>support the lower<br>jaw in a forward<br>position prescribed<br>by the dentist. The<br>customized<br>appliance is<br>inserted and<br>removed by the<br>patient and<br>adjusted by the<br>prescribing dentist. | For the reduction<br>of nighttime<br>snoring and mild<br>to moderate<br>obstructive sleep<br>apnea in adults 18<br>years of age or<br>older. Prescription<br>use only. | | Target<br>Population | Adults who snore or<br>have obstructive<br>sleep apnea. | Same | Same | Same | | Design<br>Features | Prescription device<br>custom fit | Same | Same | Same | | | Intraoral device fitted<br>over upper and lower<br>dentition | Same | Same | Same | | Sterility | Provided Non-sterile | Same | Same | Same | | Human<br>Factors | Removable by patient<br>and reusable | Same | Same | Same | | Environment<br>of Use | Home | Same | Same | Same | | Principle of<br>Operation | Mandibular<br>Repositioning with<br>custom fitted acrylic<br>upper and lower<br>components. This<br>advances the<br>mandible anteriorly<br>to enlarge the airway. | Same | Same | Same | | | Expansion mechanism<br>placed on buccal<br>portions for<br>unobstructed airway<br>passage. | Same | Same | Expansion<br>mechanism<br>located in anterior<br>portion of mouth. | | | Mandible can be<br>advanced with two<br>buccal expansion<br>screws up to 6mm. | Same | Varying<br>advancement<br>mechanisms. | Advancement by<br>dentist adding<br>acrylic material to<br>the anterior<br>protrusive element | | Materials | Methyl methacrylate,<br>medical grade<br>stainless steel,<br>Orthodontic<br>Expansion screws | Same, and additional<br>colorant | Same | Methyl<br>methacrylate,<br>Stainless steel, and<br>copolyester. | | Biocompatibility | Uses all FDA cleared<br>Class I materials that<br>are commonly used in<br>dentistry. | Same | Same | Same | | Compatibility<br>with the<br>environment<br>and other<br>devices | Does not release any<br>compounds harmful<br>to the environment<br>and pother devices. | Same | Same | Same | | Electrical,<br>radiation,<br>chemical, and<br>thermal<br>safety | The device does not<br>use electricity,<br>radiation, heat or<br>chemicals to function. | Same | Same | Same | # Table 1: Comparison of Intraoral Jaw Repositioning Devices {6}------------------------------------------------ {7}------------------------------------------------ # 7. Non-Clinical Tests Performed The materials used in the predicate device are identical to those in the subject device, but without the colorant. The three raw materials are obtained from the same manufacturers (Lang Dental Orthodontic Acrylic K141439, medical grade stainless steel orthodontic wire from AJ Wilcock, and stainless-steel orthodontic expansion screws from Leone). Both the predicate device and the subject device are manufactured in the same manufacturing facility (Space Maintainers Laboratory). The manufacturing process and sterilization and cleaning process of the predicate device and the subject device are identical. Biocompatibility and physical properties have been leveraged from the predicate. A risk analysis was performed as per ISO 14971 which considered intraoral soreness, obstruction of oral breathing, tooth movement or loosening, TMD concerns, excessive saliva and breakage. The subject device was compared to the predicate in each area to show the risks were equivalent or less to the predicate device. # 8. Clinical Tests Performed Clinical testing has not been performed. #### 9. Conclusion The subject device has the identical Indications of Use, intended use, materials of construction, sterilization and processing, principle of operation and technological characteristics as the predicate device. The devices are both manufactured, finished and shipped from the same FDA registered facility. Based upon the comparative analysis of features, materials and design, the minor design difference between the subject device and the predicate device does not affect the substantial equivalence of the subject device since: it does not change the Indication of Use of the subject device, it does not change the principle of operation, it does not change the manufacturing or sterilization process, and it does not elevate the risk. It is concluded that the proposed NightBlocks™ Appliance is substantially equivalent to the previously cleared device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...