K130130 · Selane Products, Inc. · LQZ · Aug 26, 2013 · Dental
Device Facts
Record ID
K130130
Device Name
ATG/SM-OSA APPLIANCES
Applicant
Selane Products, Inc.
Product Code
LQZ · Dental
Decision Date
Aug 26, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ATG/SM-OSA appliances are intended to reduce nighttime snoring and mild to moderate obstructive sleep apnea in adults. The ATG/SM-OSA appliances are worn while sleeping to support the lower jaw in a forward position prescribed by the dentist. The customized appliance is inserted and removed by the patient and adjusted by the prescribing dentist.
Device Story
ATG/SM-OSA appliances are custom-fabricated intraoral polymer trays; worn by patients during sleep to support mandible in forward position; intended to reduce snoring and mild-to-moderate obstructive sleep apnea. Three designs: Adjustable Herbst, Adjustable Dorsal, and MIRS. Herbst and Dorsal designs utilize stainless steel tubes, rods, or screws to control mandibular protrusion; MIRS design uses lingual rest. Adjustable Herbst may include metallic palate made of dental alloy. Devices fabricated by laboratory per dentist prescription; patient inserts/removes appliance; dentist performs adjustments. Mandibular advancement principle reduces airway obstruction. Clinical benefit: reduction of snoring and apnea symptoms. Risks managed by dentist-led fitting, monitoring, and patient instructions regarding TMJ and dental health.
Clinical Evidence
No clinical or non-clinical testing performed. Substantial equivalence based on design analysis, FMEA risk assessment, and evaluation of materials of construction.
Technological Characteristics
Materials: dental acrylic, dual-laminate polymers, stainless steel, dental alloy (for metallic palate). Principle: mandibular advancement via mechanical linkage (tubes/rods/screws) or lingual rest. Form factor: custom-fitted intraoral trays. Connectivity: none. Sterilization: not specified (custom dental device).
Indications for Use
Indicated for adults with mild to moderate obstructive sleep apnea and nighttime snoring. Contraindicated for patients where mandibular advancement may cause TMJ soreness, soft tissue soreness, or dentition complications.
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
Predicate Devices
DynaFlex Anti-snoring and Sleep Apnea Devices: LISA, DORSAL and HERBST ADJUSTABLE Appliances (K103076)
Related Devices
K130504 — ASA ORAL APPLIANCES · American Sleep Association · Oct 18, 2013
K181416 — Sapphire Dorsal Comfort Fit · Crane Dental Laboratory, Inc. · Sep 17, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
(21 CFR 801 Subpart C)
# 510(k) Summary- REVISED
per 21CFR807.92 CONTACT: Mr. John Christian Appliance Therapy Group/ Space Maintainers Lab 9129 Lurline Ave. Chatsworth, CA 91311 800-423-3270/818-998-7460 xtn 4578 Fax: 818-341-4684 http://ApplianceTherapy.com jc@appliancetherapy.com
AUG 2 6 2013
### DATE PREPARED: June 9, 2013
#### TRADE OR PROPRIETARY NAME: ATG/SM-OSA APPLIANCES
CLASSIFICATION NAME: Intraoral devices for snoring and obstructive sleep apnea: CFR 872.5570, Product Code LQZ & LRK
#### PREDICATE DEVICE: K103076
DEVICE DESCRIPTION: The ATG/SM-OSA APPLIANCES are primarily polymer trays that are used intraorally over the dentition to make the mandible protrude. Three designs are included: Adjustable Herbst, Adjustable Dorsal, and MIRS. The mandible protrusion is controlled by stainless steel tubes & rods or screws for the Herbst and Dorsal designs. The MIRS single-piece construction has a lingual rest to set the mandibular protrusion. Each device is fabricated to the prescription of a dentist.
INTENDED USE: The ATG/SM-OSA appliances are intended to reduce nighttime snoring and mild to moderate obstructive sleep apnea in adults. The ATG/SM-OSA appliances are worn while sleeping to support the lower jaw in a forward position prescribed by the dentist. The customized appliance is inserted and removed by the patient and adjusted by the prescribing dentist.
TECHNOLOGICAL CHARACTERISTICS vs. the predicate device: The ATG/SM-OSA APPLIANCES are essentially identical in indications for use to the predicate devices of DynaFlex® Anti-snoring and Sleep Apnea Devices: LISA, DORSAL and HERBST ADJUSTABLE Appliances, Both the present and predicate devices are customized for the patient using a prescription from the dentist. The scientific principle for the device is mandibular advancement determined by the design of the device.
Three designs of the ATG-/SM-OSA designs are included which are denoted as MIRS, Adjustable Dorsal and Adjustable Herbst, which we believe are the same as the predicate DynaFlex® Anti-snoring and Sleep Apnea Devices: LISA, DORSAL and
{1}------------------------------------------------
HERBST ADIUSTABLE Appliances, respectively, Both the ATG/SM-OSA APPLIANCES and the predicate DynaFlex® Anti-Snoring & Sleep Apnea Devices are composed primarily of dental polymers. Dental acrylic and dual-laminate polymers are used in the predicate and new designs.
The Herbst and Dorsal designs of both the new and predicate devices include stainless steel fixation attachments that unite the upper and lower mandible trays, to allow for adjustments in the mandibular advancement. Stainless steel attachments are used in the Herbst and Dorsal designs of both the predicate and new devices. The linkage of the upper and lower trays and the trays themselves seem identical for both the predicate and new devices.
A metallic palate design is included for the adjustable Herbst design of the new OSA appliance. The material of the metal palate is a dental allov that is used for partial dentures, which is unlike the predicate Herbst device. However, the metal palate is identical to partial denture designs for the palate, and uses the same materials as partial dentures, with the addition of acrylic and stainless steel attachments on the maxillary posterior buccal area. The designs are completed in the laboratory such that the metal palate does not contact the opposing dentition or polymer tray.
The Lisa design of the predicate and the new MIRS design are essentially the same, using a lingual rest protruding from a polymer tray formed to cover the maxillary dentition. This lingual rest creates mandibular protrusion, designed to meet the prescription of the dentist.
RISKS: Selane Products Inc. performed no clinical or non-clinical testing. However, an FMEA risk analysis, and evaluation of the materials of construction and design were performed. The function of mandibular advancement devices requires that the prescribing dentist be cognizant of the potential for TM| soreness, soft tissue soreness, and dentition complications (soreness, motion, loosening) by mandibular advancement. Management of these risks is achieved by advising the patient and dentist in the directions for use that early and repeated examination of the fit of the device, and its performance, must be performed in the dental office by the prescribing dentist. The contraindications, warnings, precautions, storage directions, prescription preparation instructions, fitting and adjustment directions are written to avoid potential problems from arising or persisting with the dentition, tissue, or joints, caused by the OSA devices. No new materials are being used in the devices; all materials are already used in dental laboratories for other devices of a related nature. No new risks are introduced with the new devices that are not present in the predicate devices.
CONCLUSIONS: By comparison and analysis of the new and predicate devices, we note herein that the intended use, the designs, the functions, and the polymeric and stainless steel materials for the acrylic and the dual-laminate polymer models of the devices are the same as the predicate. The metal palate design of the adjustable Herbst device requires the fabrication technology, material, and palate design that
{2}------------------------------------------------
are used for removable dentures; these characteristics are suitable for the ATG/SM-OSA devices, because they do not change the function of the Herbst adjustable device, and do not affect safety or efficacy for mandibular advancement and treatment of mild to moderate OSA.
We believe that the analysis provided herein supports the substantial equivalence in safety and effectiveness to the predicate device, and we believe the new devices are as safe, as effective, and perform as well as the legally marketed devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles a stylized caduceus or a representation of human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 26. 2013
Selane Products, Incorporated C/O Carolyn M. Primus, PhD Consultant Primus Consulting 7046 Owl's Nest Terrace BRADENTON FL 34203
Re: K130130
Trade/Device Name: ATG/SM-OSA APPLIANCES Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices for Snoring and Intraoral Devices for Snoring and Obstructive Sleep Apnea Regulatory Class: II
Product Code: LOZ, LRK Dated: August 2, 2013 Received: August 7, 2013
Dear Dr. Primus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Dr. Primus
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Mary S. Runner -S
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):K130130
Device Name: ATG/SM-OSA APPLIANCES
Indications For Use: The ATG/SM-OSA appliances are intended to reduce nighttime snoring and mild to moderate obstructive sleep apnea in adults. The ATG/SM-OSA appliances are worn while sleeping to support the lower jaw in a forward position prescribed by the dentist. The customized appliance is inserted and removed by the patient and adjusted by the prescribing dentist.
Mary S. Runner -S
Susan Runns DDS, MA 2013.08.26
12:35:53-04'00'
(Division Sign-Off) Division of Anesthesiology, General Hospital Respiratory, Infection Control and Dental Devices 510(k) Number: K130130
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
:
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
> Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Premarket Notification
Selane Products Inc. K130130 ATG/SM-OSA APPLIANCES Page 7
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.