PRIME & BOND 2.0 MULTIPURPOSE DENTIN/ENAMEL BONDING AGENT

K960823 · Dentsply Intl. · KLE · Apr 26, 1996 · Dental

Device Facts

Record IDK960823
Device NamePRIME & BOND 2.0 MULTIPURPOSE DENTIN/ENAMEL BONDING AGENT
ApplicantDentsply Intl.
Product CodeKLE · Dental
Decision DateApr 26, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3200
Device ClassClass 2

Indications for Use

PRIME & BOND™ 2.0 MULTIPURPOSE DENTIN/ENAMEL BONDING AGENT is a self-priming bonding agent for use in direct and indirect restorative procedures, for adhesive repairs of amalgams, porcelain/metal and composite, and for an adhesive cavity varnish.

Device Story

One-component, no-mix, visible light-curable dental bonding agent; consists of monomers, adhesion promoter, acetone solvent, photo-initiators, accelerators, and polymerization inhibitors. Applied by clinician to prepared tooth surface; left for 20-30 seconds; air-dried; light-cured for 10 seconds. Used in dental clinics for restorative procedures and repairs. Provides adhesive interface between tooth structure/restorative materials and restorative composites/amalgams/porcelain/metal.

Clinical Evidence

Bench testing only; no clinical data presented. Safety supported by toxicity testing (cytotoxicity and irritation) performed on predicate device K931812.

Technological Characteristics

One-component, light-curable resin bonding agent. Composition: monomers, adhesion promoter, acetone, photo-initiators, accelerators, polymerization inhibitors. Polymerization via visible light energy.

Indications for Use

Indicated for use in direct and indirect dental restorative procedures, adhesive repairs of amalgams, porcelain/metal and composite restorations, and as an adhesive cavity varnish.

Regulatory Classification

Identification

A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.

Predicate Devices

Submission Summary (Full Text)

{1} 510(k) SUMMARY K960823 APR 26 1996 # DENTSPLY NAME & ADDRESS: DENTSPLY International 570 West College Avenue P.O. Box 872 York, PA 17405-0872 (717) 845-7511 Fax (717) 854-2343 P. J. Lehn Telefax (717) 849-4343 CONTACT: P. Jeffery Lehn DATE PREPARED: February 27, 1996 TRADE OR PROPRIETARY NAME: PRIME & BOND™ 2.0 MULTIPURPOSE DENTIN/ENAMEL BONDING AGENT COMMON OR USUAL NAME: Bonding agent CLASSIFICATION NAME: Resin tooth bonding agent 21 CFR, 872.3200 LEGALLY MARKETED DEVICE: Prime & Bond™ Direct Composite Bonding Agent K931812 Prime & Bond - New Uses K951864 DEVICE DESCRIPTION: PRIME & BOND™ 2.0 MULTIPURPOSE DENTIN/ENAMEL BONDING AGENT is a one-component, no-mix visible light curable dental bonding agent. The new bonding agent consists of a mixture of monomers and an adhesion promoter in an acetone solution. Additionally, the mixture contains photo-initiators, accelerators and polymerization inhibitors which allow for a shelf stable material that can be polymerized using visible light energy. Once the tooth to be restored has been prepared, PRIME & BOND™ 2.0 MULTIPURPOSE DENTIN/ENAMEL BONDING AGENT is applied, left on the surface for 20-30 seconds, air dried, and then light cured for 10 seconds. A second application may be applied for direct restorative procedures, but this is not required for indirect procedures. INTENDED USE: PRIME & BOND™ 2.0 MULTIPURPOSE DENTIN/ENAMEL BONDING AGENT is a self-priming bonding agent for use in direct and indirect restorative procedures, for adhesive repairs of amalgams, porcelain/metal and composite, and for an adhesive cavity varnish. TECHNOLOGICAL CHARACTERISTICS: All of the components of PRIME & BOND™ 2.0 SELF-PRIMING DENTIN/ENAMEL BONDING AGENT have been used in Dentsply legally marketed predicate devices. In K931812, Prime & Bond™ Direct Composite Bonding Agent was evaluated for cytotoxicity and for irritation and found to be non-cytotoxic and a non-irritant. Due to the similarity of the new formulation to this predicate device, we believe that additional toxicity testing for PRIME & BOND™ 2.0 MULTIPURPOSE DENTIN/ENAMEL BONDING AGENT is unnecessary. We believe that the prior use of the PRIME & BOND™ 2.0 MULTIPURPOSE DENTIN/ENAMEL BONDING AGENT components in predicate devices, the performance data presented, and the results of toxicity testing on predicate devices, support the safety and effectiveness of PRIME & BOND™ 2.0 MULTIPURPOSE DENTIN/ENAMEL BONDING AGENT for the intended uses. 000014
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...