K230009 · Meta Biomed Co., Ltd. · KLE · Mar 3, 2023 · Dental
Device Facts
Record ID
K230009
Device Name
EZ Bond Universal
Applicant
Meta Biomed Co., Ltd.
Product Code
KLE · Dental
Decision Date
Mar 3, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3200
Device Class
Class 2
Indications for Use
All classes of fillings (according to Black) with light-curing composite or compomer filling materials Repair of composite resin or compomer fillings Intraoral repair of composite restorations, porcelain fused to metal, and all-ceramic restorations without extra primer Protective varnish for glass ionomer fillings
Device Story
EZ Bond Universal is a single-bottle, methacrylate-based resin tooth bonding agent. It functions as an adhesive layer to bond restorative materials (composites, compomers) to dentin, enamel, and indirect substrates (glass ceramics, alumina, zirconia). Used by dental professionals in clinical settings; applied to tooth surfaces or restorative substrates. Supports self-etching, total-etching, or selective-etching procedures. Eliminates need for extra primer steps. Enhances adhesion of polymerizable resins to dental structures; improves clinical outcomes for direct and indirect restorations.
Clinical Evidence
Bench testing only. Biocompatibility evaluated per ISO 10993-3, -5, -10, -11. Performance testing included film thickness, sensitivity to ambient light, bond strength (dentin/enamel), and fine leakage per ISO 4049:2019, ISO 29022:2013, and ISO/TC 11405:2015. Shelf life validated per ISO standards.
Technological Characteristics
Single-bottle methacrylate-based resin bonding agent. Composition: MDP phosphate monomer, HEMA, ethanol. Principle: Adhesive layer for polymerizable resin. Film thickness: 4 µm. Shelf life: 2 years. Biocompatibility: ISO 10993 compliant. Performance standards: ISO 4049, ISO 29022, ISO/TC 11405.
Indications for Use
Indicated for patients requiring dental restorations, including all classes of fillings (Black's classification) using light-curing composite or compomer materials, repair of composite/compomer fillings, intraoral repair of composite, porcelain-fused-to-metal, and all-ceramic restorations, and as a protective varnish for glass ionomer fillings.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 3, 2023
Meta Biomed Co., LTD. % April Lee Consultant Withus Group Inc 106 Superior Irvine, California 92620
Re: K230009
Trade/Device Name: EZ Bond Universal Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: Class II Product Code: KLE Dated: January 3, 2023 Received: January 3, 2023
Dear April Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael E. Adjodha -S
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230009
Device Name EZ Bond Universal
Indications for Use (Describe)
- · All classes of fillings (according to Black) with light-curing composite or compomer filling materials
- · Repair of composite resin or compomer fillings
- · Intraoral repair of composite restorations, porcelain fused to metal, and all-ceramic restorations without extra primer
- · Protective varnish for glass ionomer fillings
Type of Use (Select one or both, as applicable)
| | Prescription Use (Part 21 CFR 801 Subpart D) |
|--|----------------------------------------------|
| | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
#### Submitter
META BIOMED CO., LTD. Suk Song Oh 270, Osongsaengmyeong 1-ro, Osong-eup, Heungdeok-gu, Cheongju-si, Chungcheongbuk-do, South Korea Email: ef1459@metabiogw.bizmeka.com Phone: +82-43-230-8841 Fax: +82-43-217-1983
#### Device Information
- Trade Name: EZ Bond Universal ●
- . Classification Name: agent, tooth bonding, resin
- Common Name: Resin Tooth Bonding Agent
- . Product Code: KLE
- Panel: Dental
- Regulation Number: 21 CFR 872.3200
- Device Class: Class II
- Date prepared: 02/28/2023
#### Predicate Devices:
- . K110302, ADHESIVE EXL-759 by 3M ESPE AG
#### Device Description
EZ Bond Universal, manufactured by Meta Biomed, is classified as a Resin tooth bonding agent (21 C.F.R. §872.3200). EZ Bond universal is a single-bottle solution. Depending on the indication, the adhesive can be used for direct and indirect restorations with light cured composites in a "self-etching" procedure, "total-etching" procedure or" selective-etching" procedure. EZ Bond Universal is methacrylate-based restoratives, cement and sealant materials to dentin, enamel, glass ionomer and various indirect restorative substrates (glass ceramics, alumina and zirconia) without an extra primer step.
#### Indication for Use
- . All classes of fillings (according to Black) with light-curing composite or compomer filling materials
- . Repair of composite resin or compomer fillings
- Intraoral repair of composite restorations, porcelain fused to metal, and all-ceramic restorations without extra primer
- . Protective varnish for glass ionomer fillings
#### Official Correspondent
Withus Group Inc April Lee 106 Superior, Irvine, CA 92620 USA Email: withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122
{5}------------------------------------------------
## Non-clinical Testing
The subject device was tested to evaluation its safety and effectiveness according to the following standards:
- Biocompatibility Tests according to ISO 10993-3:2014. ISO 10993-5:2009, ISO 10993-10:2010, -ISO 10993-11:2017.
- Performance tests such as Film thickness, Sensitivity to ambient Light, Bond strength-Dentine, -Bond strength-enamel, Fine leakage according to ISO 4049:2019, ISO 29022:2013, ISO/TC 11405:2015
- Shelf Life tests according to ISO 4049:2019, ISO 29022:2013. -
### Summary of Technological Characteristics:
The subject device and predicate device have same intended use and principle of operation and similar technological characteristics such as capacity, film thickness, Sensitivity to Ambient Light, depth of cure, bond strength and shelf life.
The difference between the subject and predicate device is solubility. As the solubility value of the subject device is within the range that ISO 4049:2009, it doesn't affect safety and effectiveness.
| | Subject Device | Predicate Device |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Meta Biomed Co., Ltd. | 3M ESPE AG |
| Device Name | EZ Bond Universal | ADHESIVE EXL-759 |
| 510(k) Number | NA | K110302 |
| Common Name | Resin Tooth Bonding Agent | Resin Tooth Bonding Agent |
| Product Code | KLE | KLE |
| Regulation<br>Number | 21 CFR 872.3200 | 21 CFR 872.3200 |
| Indications for Use | All classes of fillings<br>(according to Black) with<br>light-curing composite or<br>compomer filling materialsRepair of composite resin or<br>compomer fillingsIntraoral repair of composite<br>restorations, porcelain fused<br>to metal, and all-ceramic<br>restorations without extra<br>primerProtective varnish for glass<br>ionomer fillings | All classes of fillings (according to<br>Black) with light-curing composite or<br>compomer filling materialsCementation of indirect restorations<br>when combined with RelyX Ultimate<br>Adhesive Resin CementCementation of veneers when combined<br>with RelyX Veneer CementBonding of core build-ups made of light-<br>curing composite or core build-up<br>materialsBonding of dual-cure cements and core<br>build-up materials and self-cure<br>composites when combined with Single<br>Bond Universal DCARepair of composite or compomer<br>fillingsIntraoral repair of composite<br>restorations, porcelain fused to metal, and |
{6}------------------------------------------------
| | | all-ceramic restorations without extra<br>primer |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | • Root surface desensitization |
| | | • Sealing of cavities prior to cementation<br>of amalgam restorations |
| | | • Sealing of cavities and preparation of<br>tooth stumps prior to temporary<br>cementation of indirect restorations |
| | | • Bonding of fissure sealants |
| | | • Protective varnish for glass ionomer<br>fillings |
| | | • Surface treatment of porcelain, ceramics<br>(including glass ceramics, zirconia and<br>alumina), metal and composite. |
| Principle of<br>operation | The device functions as an<br>adhesive layer for<br>enhancing an adhesiveness<br>of polymerizable resin to<br>adherends, and is used by<br>applying on the surface of<br>adherends. | The device functions as an<br>adhesive layer for<br>enhancing an adhesiveness<br>of polymerizable resin to<br>adherends, and is used by<br>applying on the surface of<br>adherends. |
| Raw Material | -MDP Phosphate Monomer<br>-HEMA<br>-Ethanol | -MDP Phosphate Monomer<br>-HEMA<br>-Ethanol |
| Film thickness | $4 \mu m$ | $10.6 \mu m$ |
| Sensitivity to<br>Ambient | Material remained physically<br>homogeneous. | Material remained physically<br>homogeneous. |
| Bond strength,<br>Dentine | 23.2 Mpa | 35.8 Mpa |
| Bond strength,<br>enamel | 23.7 Mpa | 26.4 Mpa |
| Shelf Life | 2 Years | 2 Years |
#### Conclusion:
Metabiomed Co., Ltd believes that EZ Bond Universal is substantially equivalent to currently legally marketed product. Product based on comparison of similar intended use and technologies tougher with the non-clinical performance testing. Any differences do not raise different questions of safety and effectiveness than the predicate, nor do they affect the safety or effectiveness of the subject device. These differences therefore, do not render the new device NSE in comparison to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.