IBOND SELF ETCH BOTTLE ASSORTMENT, IBOND SELF ETCH BOTTLE VALUE PACK, IBOND SELF ETCH SINGLE DOSE ASSORTMENT, IBOND SELF
Applicant
Heraeus Kulzer, LLC
Product Code
KLE · Dental
Decision Date
Dec 7, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.3200
Device Class
Class 2
Indications for Use
iBond Self Etch (new formulation) it intended for use by dental professionals, for the bonding of direct light-cured composite restorations (including Polyglas® and copomers), bonding of indirect restorations in combination with a light curing luting cement; porcelain, polyglas®, and composite restorations (inlays, onlays, veneers crowns) and sealing hypersensitive areas of teeth.
Device Story
iBond Self Etch is a light-activated methylacrylate resin bonding agent used by dental professionals. It is applied to tooth surfaces to facilitate the adhesion of direct light-cured composite restorations or indirect restorations (e.g., crowns, veneers) when used with light-curing luting cement; it also seals hypersensitive tooth areas. The device is applied in a single layer with a 20-second dwell time and 20-second light-curing cycle. The new formulation includes a stabilizer and a modified rheological additive, allowing for room-temperature storage and eliminating the need for shaking before use. The clinician applies the liquid to the prepared site; the material cures upon exposure to a light-curing unit. This process ensures secure restoration retention and reduces patient sensitivity. The device is intended for professional clinical use.
Clinical Evidence
Bench testing only. Biocompatibility was verified per EN ISO 10993-12 using hydrophilic and lipophilic extraction media. Leachable compounds were evaluated and deemed covered by existing toxicological data. Shear bond strength was tested, demonstrating an improvement to ≥20 MPa compared to the predicate's >15 MPa. No in vivo toxicity or clinical human studies were performed due to low patient exposure and animal welfare considerations.
Technological Characteristics
Acetone/water-based light-activated methylacrylate resin. Single-layer application. pH <2.00. Refractive index 1.421-1.426. Dwell time 20s, curing time 20s. Room temperature storage. Biocompatibility compliant with EN ISO 10993. Risk management compliant with ISO 14971.
Indications for Use
Indicated for dental professionals to bond direct light-cured composite restorations (including Polyglas and compomers), bond indirect restorations (inlays, onlays, veneers, crowns) using light-curing luting cement, and seal hypersensitive tooth areas. Contraindicated in persons with known allergies to acrylates or methacrylates.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{0}------------------------------------------------
K1232-78
Taeus
# 510(k) Summary according to 21 CFR 807.92(c)
510(k) summary of safety and effectiveness information for iBond Seff Etch (new formulation)
| Submitter Information: | |
|-----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Heraeus Kulzer, LLC |
| Address | 300 Heraeus Way<br>South Bend, IN 46614 |
| Phone Number | (574) 299-5421 |
| Fax Number | (574) 291-0080 |
| Establishment Registration Number | 1925223 |
| Name of Contact Person | Jamie Mearna |
| Date Prepared | December 3, 2012 |
| Name of Device: | iBond Self Etch |
| Trade or Proprietary Name | iBond Self Etch |
| Common or Usual Name | Resin Tooth Bonding Agent |
| Classification Name | Dental KLE |
| Device Classification | Class II |
| Classification Panel: | 76 Dental |
| Regulation: | 21 CFR 872.3200 |
| Product code (s): | KLE |
| Legally marketed devices(s) to which<br>equivalence is claimed: | HK Bond K063062 |
| Reason for 510(k) Submission: | This submission reports a modification to HK<br>Bond K063062 regarding the addition of a<br>stabilizer, change in rheological additive and<br>storage requirements. |
| Device Description: | iBond Self Etch(new formulation is an<br>acetone/water-based formulation of light<br>activated methylacrylate resins. |
| Intended use of the device: | See indications for use below. |
| Indications for use: | iBond Self Etch (new formulation) it intended<br>for use by dental professionals, for the<br>bonding of direct light-cured composite<br>restorations (including Polyglas® and<br>copomers), bonding of indirect restorations in<br>combination with a light curing luting cement;<br>porcelain, polyglas®, and composite<br>restorations (inlays, onlays, veneers crowns)<br>and sealing hypersensitive areas of teeth. |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "Heraeus" in a stylized font. The letters are outlined with a double border, giving them a hollow appearance. The text is presented in a horizontal arrangement, with each letter clearly distinguishable despite the decorative outline.
## Summary of the Technological Characteristics of the New Device Compared to the predicate Device HK Bond K063062:
The formula of the proven all-in-one adhesive has been refined. The modifications include the addition of a stabilizer the exchange of a rheological additive. The enhanced iBOND Self Etch new formulation is now conveniently useable and storable at room temperature (up to 25°C). The adhesive no longer requires shaking. The product fulfills the internal specification. The basic fundamental scientific technology remains the same as well as the indications for use.
| | CLEARED iBOND Self Etch | NEW iBOND Self Etch |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| | (BHT)- K063062 | (ISEA-EC) |
| Content of BHT | 750 ppm | Same |
| Indications for Use | iBOND Self Etch is used for<br>bonding of direct light-cured<br>composite restorations<br>(including Polyglas® and<br>copomers), bonding of<br>indirect restorations in<br>combination with a light-<br>curing luting cement;<br>porcelain, Polyglas®, and<br>composite restorations<br>(inlays, onlays, veneers,<br>crowns) and sealing<br>hypersensitive areas of<br>teeth. | Same |
| Visual Appearance | Yellowish liquid,<br>homogeneous, without<br>sediment | Same |
| pH | <2,00 | Same |
| Refractive index | 1,421-1,426 | Same |
| Application | Application in 1 layer | Same |
| Dwell time | 20s | Same |
| Curing time | 20s | Same |
| Compatibility | Compatible with direct light<br>cured composite, light cured<br>luting cement, porcelain | Same |
### Similarities as Compared to Predicate:
{2}------------------------------------------------
|--|--|--|
| CHARLERS CREEN LE LE LE SE MERICLE CARRIES CLAND SERVICE | A S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S<br>Polvelas®<br>and composite | COLLECTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTIO |
|----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | restorations<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | |
Note: For formulation similarities as compared to predicate, please see section 11 of this submission.
#### Differences as Compared to Predicate:
| | CLEARED iBOND Self Etch | NEW iBOND Self Etch |
|---------------------|-------------------------|---------------------------|
| | <b>(BHT)-K063062</b> | <b>(ISEA-EC)</b> |
| Shear Bond Strength | >15 MPa | ≥20 MPa |
| Storage condition | Refrigeration required | No refrigeration required |
| Handling | Shaking necessary | No shaking necessary |
Note: For formulation differences as compared to predicate, please see section 11 of this submission.
#### Results Summary:
The biological compatibility of iBOND Self Etch was verified in accordance with the international standard EN ISO 10993.
The medical devices iBOND Self Etch and new formulation were extracted with hydrophilic and lipophilic extraction medium according to EN ISO 10993-12. Leachable compounds detected in n-hexane, isopropanole and saline extracts are considered covered by the toxicological tests performed with iBOND Self Etch.
Considering the low exposure of the patient and animal welfare requirements, no in vivo toxicity studies were performed with the uncured or cured material.
The biocompatibility of iBOND Self Etch _new formulation in the aforementioned indication is documented and the benefit/risk-relation has to be judged as positive.
An allergenic potential in predisposed persons cannot be excluded completely due to the presence of monomers such as acrylates and methacrylates.
#### Conclusions Drawn From Non-Clinical and Clinical Data:
The refined formula of the proven all-in-one adhesive is substantially equivalent to cleared iBOND Self Etch.
A biocompatibility evaluation has been performed by a toxicologist for iBOND Self Etch _ new formulation. It was confirmed that the product meets the requirements of ISO 10993 and it is concluded that the safety of iBOND Self Etch __new formulation is equivalent to that of the predicate device.
The risk analysis acc. ISO 14791 was carried out for iBOND Self Etch _new formulation. It is concluded that safety of iBOND Self Etch _new formulation for the intended use is substantially equivalent to the predicate device. iBOND Self Etch_new formulation and the predicate
Page 3 of 4
{3}------------------------------------------------
HERAEUS
formula have the same indication for use, warnings and contraindications. When used in accordance with the instruction for use, by qualified personnel, iBOND Self Etch _new formulation is safe and effective, as indicated for the intended use.
. . . .
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized emblem with three curved lines, resembling a person with outstretched arms, positioned to the right. Encircling the emblem is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
#### December 7, 2012
Ms. Jamie Mearna Associate Quality Assurance & Regulatory Manager Heraeus Kulzer, Limited Liability Company 300 Heraeus Way SOUTH BEND IN 46614
Re: K123278
Trade/Device Name: iBond Self Etch Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: II Product Code: KLE Dated: September 25, 2012 Received: November 19, 2012
Dear Ms. Mearna:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I vature the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical Or ice-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please n you desire specific advisiboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kwame O. Ulmer
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): _ K (23279
Device Name: iBond Self Etch
Indications for Use:
iBond Self Etch is an acetone/water-based formulation of light-activated methylacrylate resins.
iBond Self Etch is used for bonding of direct light-cured composite restorations (including Polyglas and copomers), bonding of indirect restorations in combination with a light-curing luting cement; porcelain, Polyglas®, and composite restorations (inlays, onlays, veneers, crowns) and sealing hypersensitive areas of teeth.
Over-The-Counter Use Prescription Use AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
i 2012.12.07
Susan Runner DDS, MA 15:06:19
-05'00'
(Division Sign-Off) (UNIsion Sign-On)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page _1_ of .
510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.