K232730 · Inquis Medical · QEW · Nov 1, 2023 · Cardiovascular
Device Facts
Record ID
K232730
Device Name
Aventus Thrombectomy System
Applicant
Inquis Medical
Product Code
QEW · Cardiovascular
Decision Date
Nov 1, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Aventus Thrombectomy System is indicated for: · The non-surgical removal of emboli and thrombi from blood vessels. · Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel. The Aventus Thrombectomy System is intended for use in the peripheral vasculature.
Device Story
Aventus Thrombectomy System is a manual aspiration catheter system for minimally invasive removal of emboli/thrombi from peripheral vasculature. System components include an aspiration catheter with radiopaque distal tip, navigation catheter, infusion lumen, handle, 60-cc dual-action manual syringe, and clot canister. Physician operates device under fluoroscopic guidance; catheter is navigated to target site using 0.035" guidewires. Manual syringe provides vacuum for aspiration, drawing blood and clot into canister; infusion lumen allows contrast/fluid delivery. Device facilitates clot removal, potentially restoring blood flow and improving patient outcomes. Used in clinical settings; single-use; sterile.
Clinical Evidence
No human clinical data. Evidence includes successful bench testing (tensile strength, pressure/leak integrity, clot removal validation, torque, kink testing, simulated-use) and a GLP chronic large animal study demonstrating safety and performance meeting all pre-defined endpoints.
Technological Characteristics
Catheter-based manual aspiration system. Materials: stainless steel reinforcement, polymeric jacket with variable stiffness. Features: radiopaque distal tip, 60-cc dual-action manual syringe, clot canister. Compatible with 24F introducer sheaths, 4F support catheters, 0.035" guidewires. Sterilization: ISO 14937:2009 compliant (SAL 10^-6). Biocompatibility: ISO 10993-1:2018. Non-software device.
Indications for Use
Indicated for non-surgical removal of emboli and thrombi from peripheral vasculature and injection, infusion, or aspiration of contrast media and fluids in patients requiring such procedures.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
AngioDynamics: AlphaVac Multipurpose Mechanical Aspiration (MMA) F1885 System (K213388)
Related Devices
K250668 — Aventus Thrombectomy System (IM-2100) · Inquis Medical · Mar 25, 2025
K250202 — Aventus Thrombectomy System · Inquis Medical · Feb 26, 2025
K240117 — Aventus Thrombectomy System (IM-2001/IM-0002) · Inquis Medical · Mar 15, 2024
K251189 — Aventus Thrombectomy System · Inquis Medical · Jun 13, 2025
K253925 — Aventus Thrombectomy System · Inquis Medical · Jan 15, 2026
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 1, 2023
Inquis Medical Moigan Saadat, President 1755 East Bayshore Road, Suite 26 Redwood City, California 94063
Re: K232730
Trade/Device Name: Aventus Thrombectomy System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: QEW, KRA Dated: September 6, 2023 Received: September 7, 2023
Dear Mojgan Saadat:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Gregory W. Digitally signed by O'connell -S 10:39:32 -04'00'
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232730
Device Name Aventus Thrombectomy System
### Indications for Use (Describe) The Aventus Thrombectomy System is indicated for:
- · The non-surgical removal of emboli and thrombi from blood vessels.
- · Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.
The Aventus Thrombectomy System is intended for use in the peripheral vasculature.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Inquis Medical. The word "INQUIS" is in large, bold, dark blue font, with a green circle in the middle of the "Q". Below that, the word "MEDICAL" is in a smaller, lighter blue font. The logo is simple and modern.
# 510(k) Summary
#### I. SUBMITTER
Inquis Medical 127 Independence Drive Menlo Park, CA 94025 408-209-8326
Contact Person: Mojgan Saadat, President Inquis Medical
Date Prepared: September 6, 2023
#### II. DEVICE
Name of Device: Common or Usual Name: Classification Name: Regulatory Class: Product Code: Regulation Number:
Aventus Thrombectomy System Aspiration Thrombectomy Catheter Embolectomy Catheter Class II QEW, KRA 21 CFR 878.5150
#### III. PREDICATE/REFERENCE DEVICE
| Predicate Device: | Malibu Aspiration Catheter (K223929) |
|-------------------|--------------------------------------------------------------------------------------------|
| Reference Device: | AngioDynamics: AlphaVac Multipurpose Mechanical Aspiration<br>(MMA) F1885 System (K213388) |
#### IV. DEVICE DESCRIPTION
The Aventus Thrombectomy System is a catheter-based manual aspiration system designed for minimally invasive removal of emboli and thrombi from the peripheral vasculature and/or infusing fluids into the peripheral vasculature. The System is comprised of the following major components:
- Aspiration Catheter •
- . Clot Management System consisting of:
- Aspiration Syringe
- Clot Canister .
The System is compatible with a standard 24-French (24F) introducer sheath, 4-French (4F) Support Catheters and is to be used with 0.035" guidewires. The System is provided sterile and is intended for single use only.
The Aspiration Catheter includes an atraumatic radiopaque Distal Tip with angled aspiration orifice for directional aspiration and navigation without a dilator, a dedicated navigation catheter lumen with integrated Navigation Catheter, a dedicated infusion lumen for injection of fluids, and
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Inquis Medical. The word "INQUIS" is in a dark blue sans-serif font, with a green circle in the middle of the "Q". Below that, the word "MEDICAL" is in a smaller, lighter blue sans-serif font.
a Handle designed to navigate within the vasculature. The Aspiration Catheter shaft incorporates a metallic reinforcement made of stainless steel, an inner liner and polymeric jacket material with variable stiffness.
The System is provided with a 60-cc dual action manual syringe which allows for directional flow control and directs aspirated blood and clot into the Clot Canister.
#### V. INDICATIONS FOR USE
The Aventus Thrombectomy System is indicated for:
- . The non-surgical removal of emboli and thrombi from blood vessels.
- Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood . vessel.
The Aventus Thrombectomy System is intended for use in the peripheral vasculature.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The subject device, Aventus Thrombectomy System, is substantially equivalent to the predicate device: Malibu Aspiration Catheter cleared under K223929. The intended use of the subject device is the same as the predicate, namely removal of thrombi or emboli from the peripheral vasculature system and injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.
The subject and predicate device share similar technological characteristics in that both devices are large bore aspiration catheters which utilize a 60-cc manual syringe as the aspiration source. Both devices utilize a dedicated infusion port for injection and infusion of contrast media and other fluids into or from the blood vessel. Both devices utilize similar construction such as metallic (stainless steel) reinforcement and a polvmeric jacket with variable stiffness for the shaft, radiopaque markings at the distal tip for fluoroscopic visualization, and use of flow-control valves to direct the flow of fluids.
The subject device differs from the predicate device in that it incorporates additional features that increase the convenience of use and does not include any coating or surface materials like the predicate device.
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for Inquis Medical. The word "INQUIS" is in a bold, sans-serif font, with a green circle inside the "Q". Below that, the word "MEDICAL" is in a thinner, sans-serif font. The overall design is clean and modern.
| Performance Testing | Data provided |
|-------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility<br>Testing | Biocompatibility testing was successfully completed in accordance with<br>ISO 10993-1:2018 and the FDA Guidance re: Use of ISO-10993.<br>Testing included:<br>Cytotoxicity Sensitization Intracutaneous Reactivity Acute Systemic Toxicity Material Mediated Pyrogenicity Hemocompatibility (Hemolysis, Complement<br>Activation, Thrombogenicity) |
| | This testing demonstrated the materials of the Aventus Thrombectomy |
| | System do not pose a risk of negative interaction with patients. |
| Sterilization | Sterilization testing was successfully completed in accordance with ISO<br>14937:2009 - Sterilization of health care products – General<br>requirements for characterization of a sterilizing agent and the<br>development, validation and routine control of a sterilization process for<br>medical devices and demonstrated an SAL of 10-6. |
| | Bacterial endotoxins test (BET), a.k.a. Limulus amebocyte lysate |
| | (LAL) testing was conducted per current test guidelines: USP <85> |
| | Bacterial Endotoxin Test and AAMI ST72 Bacterial endotoxins- test |
| | methodologies, routine monitoring and alternatives to batch testing |
| | and confirmed that the System meets established pyrogen limit<br>specifications. |
| Distribution,<br>Packaging and Shelf-<br>Life Testing | Distribution testing and Accelerated Aging simulating six-month shelf<br>life was successfully completed. |
| | Final packaging and device performance were successfully tested |
| | demonstrating integrity of the sterile barrier and preservation of the |
| | System's properties for the labeled six-month shelf-life. |
| Performance Testing -<br>Bench | Design verification testing was performed demonstrated that the physical<br>and functional requirements were met.<br>Specifically, the following was tested:<br>Visual and Dimensional Inspection Tensile Strength Pressure/ Leak Integrity Clot Burden Removal Validation Vacuum Test Leak Test Pouch Seal Visual Inspection and Bubble Test Pouch Seal Strength Compatibility Testing Torque Transmission Testing Kink Testing Kink and Torque to Failure Simulated-Use |
| Performance Testing | Data provided |
| Performance Testing –<br>Non-Clinical | A GLP animal study was completed in compliance with GLP<br>regulation (21 CFR Part 58) and in accordance with FDA Guidance:<br><i>General Considerations for Animal Studies for Cardiovascular</i><br><i>Devices</i> (July 2010) and FDA Guidance: <i>General Considerations for<br/>Animal Studies Intended to Evaluate Medical devices</i> (March 2023),<br>demonstrated that the System was able to be used safely in a chronic<br>large animal GLP study and met all pre-defined study endpoints. |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Inquis Medical. The word "INQUIS" is in a dark blue sans-serif font, with a yellow circle in the "Q". Below that, the word "MEDICAL" is in a lighter blue sans-serif font.
#### VIII. CONCLUSIONS
In conclusion, the intended use, indications for use, and technological characteristics of the Aventus Thrombectomy System are the same or equivalent to the predicate device. Performance testing has demonstrated that the Aventus Thrombectomy System is substantially equivalent to the predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.