AlphaVac Multipurpose Mechanical Aspiration (MMA) F1885 System
Applicant
AngioDynamics, Inc.
Product Code
QEZ · Cardiovascular
Decision Date
Apr 4, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cannula is indicated for · the non-surgical removal of thrombi or emboli from venous vasculature. · aspiration of contrast media and other fluids from venous vasculature. The Cannula is intended for use in the venous system. The Handle is indicated as a vacuum source for the AlphaVac MMA System.
Device Story
Single-use, over-the-wire, catheter-based system for percutaneous thrombus/embolus removal. Components: 18F nitinol-reinforced cannula with 85° angled funnel tip, 22F sheath, obturator, handle, waste bag/tubing. Physician inserts cannula via Seldinger technique; nitinol basket expands into funnel at target site; handle creates manual vacuum suction to aspirate material into catheter for containment in waste bag. Used in minimally invasive procedures; provides mechanical aspiration to clear venous vasculature. Benefits patient by enabling non-surgical clot removal.
Clinical Evidence
Bench testing only. No clinical data. Testing included dimensional, tensile, column strength, cannula/funnel actuation, kink resistance, radiopacity, flushability, push/pull/retraction force, leak, siphoning, fluid volume removal, handle lock/pull force, pressure, and torque testing. Human factors/usability evaluation and simulated use testing performed.
Technological Characteristics
Large-bore catheter system; nitinol-reinforced distal funnel tip; 22F sheath with hemostatic valve; 85° angled configuration. Sterilized via ethylene oxide (EO). Biocompatibility per ISO 10993. Manual vacuum aspiration source.
Indications for Use
Indicated for non-surgical removal of thrombi or emboli from venous vasculature and aspiration of contrast media/fluids in venous system. Intended for use in venous vasculature including iliofemoral vein, IVC, SVC, and right atrium.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
Predicate Devices
AlphaVac Multipurpose Mechanical Aspiration (MMA) System (K211081)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
April 4, 2022
AngioDynamics, Inc. Kasey Newcomb Manager, Global Regulatory Affairs 26 Forest Street Marlborough, MA 01742
Re: K213388
Trade/Device Name: AlphaVac Multipurpose Mechanical Aspiration (MMA) F188 System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: QEZ Dated: February 23, 2022 Received: February 23, 2022
Dear Kasey Newcomb:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213388
Device Name
AlphaVac Multipurpose Mechanical Aspiration (MMA) F1885 System
Indications for Use (Describe)
- The Cannula is indicated for
- · the non-surgical removal of thrombi or emboli from venous vasculature.
- · aspiration of contrast media and other fluids from venous vasculature.
The Cannula is intended for use in the venous system.
The Handle is indicated as a vacuum source for the AlphaVac MMA System.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY FOR THE ALHAVAC MULTIPURPOSE MECHANICAL ASPIRATION (MMA) F1885 SYSTEM
### A. SPONSOR AngioDynamics, Inc. 603 Queensbury Ave Queensbury, NY 12804 USA B. CONTACT Kasey E Newcomb Troy Roberts Specialist II, Global Regulatory Affairs VP, Regulatory Affairs Tel: 508.658.7813 Tel: 508.658.7934 Email: knewcomb@angiodynamics.com Email: troberts@angiodynamics.com C. DEVICE NAME Trade Name: AlphaVac Multipurpose Mechanical Aspiration (MMA) F1885 System Common/Usual Name: Aspiration Thrombectomy Catheter Classification Name: Aspiration Thrombectomy Catheter (21 CFR § 878.5150, Class II, Pro-Code QEZ) Classification Panel: Cardiovascular D. PREDICATE DEVICE 510(k): K211081 Trade Name: AlphaVac Multipurpose Mechanical Aspiration (MMA) System Aspiration Thrombectomy Catheter Common/Usual Name: Classification Name: Aspiration Thrombectomy Catheter (21 CFR § 878.5150, Class II, Pro-Code QEZ) Classification Panel: Cardiovascular E. REFERENCE DEVICE 510(k): K212386 AngioVac F1885 Trade Name: Common/Usual Name: Catheter, Cannula and Tubing, Vascular, Cardiopulmonary Bypass Cardiopulmonary bypass vascular catheter, cannula, or tubing Classification Name: (21 CFR § 878.4210, Class II, Pro-Code DWF) Classification Panel: Cardiovascular
{4}------------------------------------------------
#### F. DEVICE DESCRIPTION
The AlphaVac Multipurpose Mechanical Aspiration (MMA) F1885 System is a single use-overwire catheter-based system that facilitates the removal of thrombus, embolus, or clot during minimally invasive percutaneous procedures. The AlphaVac MMA F186 System is comprised of six main components packaged together:
- . a flexible AlphaVac Cannula with self-expandable, nitinol reinforced, angled funnel shaped distal tip (18F)
- AlphaVac Sheath (22F)
- AlphaVac Obturator (17F) ●
- AlphaVac Handle .
- . Waste Bag and Tubing
The AlphaVac Cannula is placed within target vasculature using standard percutaneous vascular access techniques (i.e., Seldinger) and commonly available vascular access tools (e.g., guidewire, vascular introducers, etc.). Once the cannula is in place, the AlphaVac Handle and waste bag is connected and primed. The AlphaVac Cannula is advanced out of the sheath and the nitinol basket automatically expands into a funnel, aiding in the removal of thrombus, emboli, and clot. The aspiration handle is pulled back creating suction and pulling the material into the catheter, removing it from the vasculature. The aspirated material is captured and contained within the waste bag for disposal. Target vessels include, but are not limited to, the iliofemoral vein, Inferior Vena Cava (IVC), Superior Vena Cava (SVC), and Right Heart (Atrium (RA)). The device is provided in ~85° (AlphaVac F1885) angled configuration.
#### G. INDICATION FOR USE
The Cannula is indicated for
- . the non-surgical removal of thrombi or emboli from venous vasculature.
- . aspiration of contrast media and other fluids from venous vasculature.
The Cannula is intended for use in the venous system.
The Handle is indicated as a vacuum source for the AlphaVac MMA System.
#### H. STERILIZATION/SHELF LIFE
The AlphaVac MMA F1885 System is sterilized via ethylene oxide (EO). A series of tests, performed by AngioDynamics and independent test houses, have been conducted to assess the suitability of the sterile packaging to protect the proposed AlphaVac MMA F1865 System and ensure sterility within its stated shelf life at point of use. These tests confirm the packaging integrity, sterility and distribution cycle. Testing demonstrated that the packaging is robust enough to withstand extreme distribution conditions at the most extreme environmental conditions while maintaining packaging integrity and sterility.
#### I. BIOCOMPATIBILITY
The AlphaVac MMA F1865 System is a sterile single-use disposable instrument. The AlphaVac MMA F1885 System has met the biocompatibility testing requirements identified in ISO 10993: Biological Evaluation of Medical Devices Part 1: Evaluation and testing within a risk management process. Specifically, the following tests were performed with acceptable results; cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, and hemocompatibility.
{5}------------------------------------------------
#### J. TECHNOLOGY CHARACTERISTICS
Predicate device, AlphaVac Multipurpose Mechanical Aspiration (MMA) System cleared via K211081, was used to support and evaluate substantial equivalence of the subject device. Both the subject device and specified predicate device include the following technological characteristics:
- . Both the proposed and predicate devices are designed for the non-surgical removal of thrombi or emboli from vasculature.
- Both devices are intended to be used with commonly available vascular access tools (e.g., . guidewire, vascular introducer, etc.) to facilitate removal of thromboemboli during minimally invasive percutaneous procedures.
- . Both devices are large bore catheters with a syringe-like aspiration source.
- Both devices have similar operating principles by being advanced through a sheath over a ● guidewire to target location with a mechanical aspiration source that is used to aspirate thrombus/emboli from vasculature.
Technological characteristics that are different in the subject device are as follows:
- The cannula is available with an 85°angle distal tip
- The cannula outside diameter is 22F and working length of 44.4 inches ●
- . Hemostatic Valve on sheath proximal end allows sheath and cannula to be separated
The technological characteristics of the proposed AlphaVac MMA F1885 System are substantially equivalent with respect to the basic system design and function to that of the specified predicate device.
Additionally, a reference predicate (K212386 - AngioVac F1885) has been identified to provide additional support as it is identical in materials, dimensions, manufacture, and used within the same vasculature to the proposed AlphaVac F1885 System
#### K. PERFORMANCE DATA
Comprehensive bench testing (integrity and functional performance) was performed to support substantial equivalence of the specified predicate device. The AlphaVac MMA F1885 System met all specified design and performance requirements below:
- Dimensional Testing .
- Visual Inspection
- Tensile Testing ●
- Column Strength
- Cannula and Funnel Actuation ●
- Distal Cannula Shape ● Manipulation
- Hub Rotation
- Distal Tip Functionality .
- Kink Resistance ●
- Radiopacity ●
- Flushability
- Product Interface (Compatibility) Testing
- Push/Pull/Retraction Force ●
- Leak Testing ●
- Siphoning Testing ●
- Fluid Volume Removal
- Handle Lock Testing ●
- Pressure Testing
- Handle Pull Force ●
- Human Factors Evaluation/Usability Evaluation
- Simulated Use ●
- Torque
#### L. CONCLUSIONS
The results of the non-clinical testing and a comparison of similarities and differences demonstrates that the proposed and predicate devices are substantially equivalent.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.