K250202 · Inquis Medical · QEW · Feb 26, 2025 · Cardiovascular
Device Facts
Record ID
K250202
Device Name
Aventus Thrombectomy System
Applicant
Inquis Medical
Product Code
QEW · Cardiovascular
Decision Date
Feb 26, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Aventus Thrombectomy System is indicated for: · The non-surgical removal of emboli and thrombi from blood vessels. · Iniection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel. The Aventus Thrombectomy System is intended for use in the peripheral vasculature.
Device Story
Aventus Thrombectomy System is a catheter-based manual aspiration system for minimally invasive removal of emboli/thrombi from peripheral vasculature. System components include an aspiration catheter, aspiration syringe, and clot canister. The catheter features an atraumatic radiopaque distal tip with embedded sensors, an angled aspiration orifice, and a handle with a sensing indicator. Clinicians navigate the catheter through the vasculature; manual 60-cc syringe provides aspiration force to draw clots into the canister. The device is used in clinical settings by physicians. Embedded sensors and electronics assist in navigation and sensing. The system is compatible with 24F introducer sheaths, 4F/5F navigation catheters, and 0.035" guidewires. It is provided sterile for single use. The device benefits patients by enabling mechanical clot removal without open surgery.
Clinical Evidence
Bench testing only. Performance data included visual/dimensional inspection, tensile strength, pressure/leak integrity, clot burden removal validation, vacuum/leak tests, compatibility, torque transmission, kink testing, simulated use, and verification/validation of sensing. Biocompatibility (ISO 10993-1), sterilization (ISO 14937), and electrical safety/EMC (IEC 60601-1 series) were leveraged from the predicate. GLP animal study data from the predicate was also leveraged to demonstrate safety and performance.
Technological Characteristics
Catheter-based aspiration system; stainless steel reinforced polymeric shaft with variable stiffness; radiopaque distal tip; 60-cc manual syringe; disposable powered electronics with embedded sensors; compatible with 24F introducer sheaths and 0.035" guidewires. Sterilized to SAL 10^-6. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, and IEC 62366-1.
Indications for Use
Indicated for non-surgical removal of emboli and thrombi from peripheral blood vessels and for injection, infusion, or aspiration of contrast media and other fluids in the peripheral vasculature.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
February 26, 2025
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue and "ADMINISTRATION" in a smaller font size.
Inquis Medical Zachary Woodson VP of Regulatory Affairs & Ouality Assurance 1530 O'Brien Drive Suite A Menlo Park, California 94025
Re: K250202
Trade/Device Name: Aventus Thrombectomy System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: OEW. KRA Dated: January 23, 2025 Received: January 24, 2025
Dear Zachary Woodson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Digitally signed by Gregory W. Gregory W. O'connell -S O'connell -S Date: 2025.02.26
14:46:42 -05'00'
Gregory O'Connell Assistant Director DHT2C: Division of Coronary And Peripheral Interventional Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K250202
Device Name
Aventus Thrombectomy System
Indications for Use (Describe)
The Aventus Thrombectomy System is indicated for:
· The non-surgical removal of emboli and thrombi from blood vessels.
· Iniection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.
The Aventus Thrombectomy System is intended for use in the peripheral vasculature.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Inquis Medical. The word "INQUIS" is in a dark blue sans-serif font, with a green circle in the middle of the "Q". Below that, the word "MEDICAL" is in a lighter blue sans-serif font.
## 510(k) Summary
#### 1. SUBMITTER
Inquis Medical 1530 O'Brien Drive, Ste. A Menlo Park, CA 94025
Contact Person: Zachary Woodson, VP of Regulatory Affairs & Quality Assurance
Date Prepared: 10 February 2025
#### II. DEVICE
Name of Device: Common or Usual Name: Classification Name: Regulatory Class: Product Code: Requlation Number:
Aventus Thrombectomy System Aspiration Thrombectomy Catheter Embolectomy Catheter Class II QEW, KRA 21 CFR 870.5150
#### = PREDICATE DEVICES
| Predicate Device: | Aventus Thrombectomy System (K240117) |
|-------------------|---------------------------------------|
| Reference Device: | Flowtriever - Triever 24 (K213402) |
#### IV. DEVICE DESCRIPTION
The Aventus Thrombectomy System is a catheter-based manual aspiration svstem designed for minimally invasive removal of emboli and thrombi from the peripheral vasculature and/or infusing fluids into the peripheral vasculature. The System is comprised of the following major components:
- Aventus Thrombectomy Catheter .
- Clot Management System consisting of:
- . Aspiration Syringe
- . Clot Canister
The System is compatible with a standard 24-French (24F) introducer sheath, 4F or 5F Navigation Catheters (based on Aventus Thrombectomy Catheter model) and is to be used with 0.035" guidewires. The System is provided sterile and is intended for single use only.
The Aventus Thrombectomy Catheter includes an atraumatic radiopaque Distal Tip with embedded sensors, and an angled aspiration orifice for directional aspiration and navigation without a dilator. The Catheter Handle assists the clinician in navigating within the vasculature and includes a Sensing Indicator and houses the Sensing electronics.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Inquis Medical. The word "INQUIS" is in a bold, dark blue font, with a yellow circle in the middle of the "Q". Below that, the word "MEDICAL" is in a thinner, light blue font. The logo is simple and modern.
The Aspiration Catheter shaft incorporates a metallic reinforcement layer made of stainless steel, an inner liner and polymeric outer jacket having variable stiffness.
The System is provided with a 60-cc dual action manual syringe which allows for directional flow control and directs aspirated blood and clot into the Clot Canister.
#### V. INDICATIONS FOR USE
The Aventus Thrombectomy System is indicated for:
- The non-surgical removal of emboli and thrombi from blood vessels. ●
- . Iniection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.
The Aventus Thrombectomy System is intended for use in the peripheral vasculature.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The subject device, Aventus Thrombectomy System, is substantially equivalent to the predicate device: Aventus Thrombectomy System cleared under K240117. The intended use of the subject device is the same as the predicate, namely removal of thrombi or emboli from the peripheral vasculature system and infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.
The subject and predicate devices share the same technological characteristics in that both devices are single patient use, large bore aspiration catheters which utilize a 60-cc manual syringe as the aspiration source. Both the subject and predicate device incorporate disposable powered electronics and embedded software. From a manufacturing standpoint, both devices utilize shafts made with metallic (stainless steel) reinforced polymeric jackets with variable stiffness, radiopaque markings at the distal tip for fluoroscopic visualization, and use of stopcocks to direct the flow of fluids. Neither the subject nor predicate device includes any coating or surface materials.
The subject device (with model IM-2200 Aventus Thrombectomy catheter), is modified from the predicate device (with model IM-2001 Aventus Thrombectomy Catheter) as follows:
- Simplification of lumen construction from three to one, re-routing of applicable ports, 1. and addition of Navigation Port valve.
- 2. Repositioning of electrical wires to within the catheter wall.
- 3. Removal of distal infusion holes.
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Inquis Medical. The word "INQUIS" is in a bold, sans-serif font, with a green circle in the middle of the "Q". Below that, the word "MEDICAL" is in a smaller, thinner, sans-serif font. The overall design is clean and modern.
| Performance Testing | Data provided |
|-------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility<br>Testing | Biocompatibility testing was leveraged from the predicate device which was successfully completed in<br>accordance with ISO 10993-1:2018 and the FDA Guidance re: Use of ISO-10993. Testing included:<br>• Cytotoxicity<br>• Sensitization<br>• Intracutaneous Reactivity<br>• Acute Systemic Toxicity<br>• Material Mediated Pyrogenicity<br>• Hemocompatibility (Hemolysis, Complement Activation, Thrombogenicity)<br>This testing demonstrated the materials of the Aventus Thrombectomy System do not pose a risk of negative interaction with patients. |
| Sterilization | Sterilization testing was leveraged from the predicate device which was successfully completed in<br>accordance with ISO 14937:2009 - Sterilization of health care products - General requirements for<br>characterization of a sterilizing agent and the development, validation and routine control of a sterilization<br>process for medical devices and demonstrated an SAL of 10-6.<br>Bacterial endotoxins test (BET), a.k.a. Limulus amebocyte lysate (LAL) testing was conducted on the<br>predicate device per current test guidelines: USP <85> Bacterial Endotoxin Test and AAMI ST72<br>Bacterial endotoxins- test methodologies, routine monitoring and alternatives to batch testing and<br>confirmed that the System meets established pyrogen limit specifications. |
| Distribution,<br>Packaging and Shelf-<br>Life Testing | Distribution and packaging testing was leveraged from the predicate device which successfully<br>demonstrated the integrity of the sterile barrier and preservation of the System's properties.<br>Shelf-life testing was leveraged from the predicate device which has demonstrated preservation of the<br>System's properties for the labeled six-month shelf-life. |
| Software Testing | Software documentation was leveraged from the predicate device as recommended by FDA Guidance:<br>Content of Premarket Submissions for Device Software Functions, issued June 14, 2023. |
| Electrical Safety / EMC<br>Testing | Electrical Safety and EMC testing were leveraged from the predicate device to ensure the subject device<br>complies with the requirements of IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6 and IEC 62366-1.<br>Electrical Leak Testing per IEC 60601-1 was repeated on the subject device. |
| Performance Testing –<br>Bench | Design verification testing was performed and demonstrated that the physical and functional requirements<br>were met.<br>Specifically, the following was tested:<br>• Visual & Dimensional Inspection<br>• Tensile Strength<br>• Pressure/Leak Integrity<br>• Clot Burden Removal Validation<br>• Vacuum Test<br>• Leak Test<br>• Compatibility Testing<br>• Torque Transmission Testing<br>• Kink Testing<br>• Simulated Use<br>• Verification and Validation of Sensing |
| Performance Testing –<br>Non-Clinical | GLP animal testing completed in compliance with GLP regulation (21 CFR Part 58) and in accordance<br>with FDA Guidance: General Considerations for Animal Studies for Cardiovascular Devices (July<br>2010) and FDA Guidance: General Considerations for Animal Studies Intended to Evaluate Medical<br>devices (March 2023) provided for the predicate device, was leveraged for the subject device and<br>demonstrated that the System was able to be used safely in a chronic large animal GLP study and met<br>all pre-defined study endpoints. |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Inquis Medical. The word "INQUIS" is in a bold, sans-serif font, with a green lemon shape in the middle of the "Q". Below that, the word "MEDICAL" is in a thinner, sans-serif font.
#### VIII. CONCLUSIONS
In conclusion, the intended use, indications for use, and technological characteristics of the Aventus Thrombectomy System are the same or equivalent to the predicate devices. Performance testing has demonstrated that the Aventus Thrombectomy System is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.