K251189 · Inquis Medical · QEW · Jun 13, 2025 · Cardiovascular
Device Facts
Record ID
K251189
Device Name
Aventus Thrombectomy System
Applicant
Inquis Medical
Product Code
QEW · Cardiovascular
Decision Date
Jun 13, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Aventus Thrombectomy System is indicated for: • The non-surgical removal of emboli and thrombi from blood vessels. • Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel. The Aventus Thrombectomy System is intended for use in the peripheral vasculature and for the treatment of pulmonary embolism.
Device Story
Catheter-based manual aspiration system for minimally invasive removal of emboli/thrombi from peripheral vasculature and pulmonary arteries. System includes Aventus Thrombectomy Catheter, aspiration syringe, and clot canister. Catheter features atraumatic radiopaque distal tip with embedded sensors, angled aspiration orifice, and stainless-steel reinforced polymeric shaft. Clinician navigates catheter under fluoroscopy; manual 60-cc syringe provides aspiration force to draw thrombi into canister. Sensing electronics in handle assist navigation. Used in clinical settings by physicians. Enables rapid clot removal, potentially reducing pulmonary vascular resistance and improving right ventricular function in pulmonary embolism patients.
Clinical Evidence
Prospective, multicenter, single-arm, non-blinded AVENTUS trial (N=130; 120 ITT). Primary safety endpoint: composite of device-related major adverse events (MAEs) at 48 hours; 0% observed (p<0.0001 vs 25% goal). Primary effectiveness endpoint: change in RV/LV ratio from baseline to 48 hours; mean reduction of 0.47 (1.57 to 1.10, p<0.0001). No thrombolytics used in 99% of subjects.
Technological Characteristics
Catheter-based manual aspiration system. Materials: stainless steel reinforcement, polymeric jacket. Features: radiopaque distal tip, embedded sensors, angled aspiration orifice. Energy: manual aspiration (60-cc syringe). Connectivity: standalone. Sterilization: validated per ISO 14937. Software: embedded electronics for sensing. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, and IEC 62366-1.
Indications for Use
Indicated for non-surgical removal of emboli and thrombi from blood vessels, and injection, infusion, or aspiration of contrast media/fluids in peripheral vasculature and for treatment of pulmonary embolism.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
Predicate Devices
Inari FlowTriever Retrieval/Aspiration System (K191710)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 13, 2025
Inquis Medical
Zachary Woodson
VP of Regulatory Affairs & Quality Assurance
1530 O'Brien Drive
Suite A
Menlo Park, California 94025
Re: K251189
Trade/Device Name: Aventus Thrombectomy System
Regulation Number: 21 CFR 870.5150
Regulation Name: Embolectomy Catheter
Regulatory Class: Class II
Product Code: QEW, KRA
Dated: April 16, 2025
Received: April 17, 2025
Dear Zachary Woodson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251189 - Zachary Woodson
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K251189 - Zachary Woodson
Page 3
Sincerely,
Digitally signed by ARIEL G. ASH-SHAKOOR -S
Date: 2025.06.13
10:57:33 -04'00'
For
Gregory O'Connell
Assistant Director
DHT2C: Division of Coronary and Peripheral Interventional Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251189 | |
| Device Name Aventus Thrombectomy System | |
| Indications for Use (Describe) The Aventus Thrombectomy System is indicated for: • The non-surgical removal of emboli and thrombi from blood vessels. • Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel. The Aventus Thrombectomy System is intended for use in the peripheral vasculature and for the treatment of pulmonary embolism. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
INQUIS
MEDICAL
K251189
Aventus Thrombectomy System
(IM-2001/IM-0002 – PE Indication)
510k Pre-Market Notification
510(k) Summary
I. SUBMITTER
Inquis Medical
1530 O'Brien Drive, Ste. A
Menlo Park, CA 94025
Contact Person:
Zachary Woodson, VP of Regulatory Affairs & Quality Assurance
(707) 328-6522
Date Prepared: 03 June 2025
II. DEVICE
Name of Device: Aventus Thrombectomy System
Common or Usual Name: Aspiration Thrombectomy Catheter
Classification Name: Embolectomy Catheter
Regulatory Class: Class II
Product Code: QEW, KRA
Regulation Number: 21 CFR 870.5150
III. PREDICATE DEVICES
Predicate Device: Inari FlowTriever Retrieval/Aspiration System (K191710)
Reference Device: Aventus Thrombectomy System (K240117)
IV. DEVICE DESCRIPTION
The Aventus Thrombectomy System is a catheter-based manual aspiration system designed for the minimally invasive removal of emboli and thrombi from blood vessels, infusing fluids, and is intended to be used in the peripheral vasculature and for the treatment of pulmonary embolism. The System is comprised of the following major components:
- Aventus Thrombectomy Catheter
- Clot Management System consisting of:
- Aspiration Syringe
- Clot Canister
The System is compatible with a standard 24-French (24F) introducer sheath, 4F Navigation Catheters and is to be used with 0.035" guidewires. The System is provided sterile and is intended for single use only.
The Aventus Thrombectomy Catheter includes an atraumatic radiopaque Distal Tip with embedded sensors, and an angled aspiration orifice for directional aspiration and navigation without a dilator. The Catheter Handle assists the clinician in navigating within the vasculature and includes a Sensing Indicator and houses the Sensing electronics. The
Page 1 of 5
{5}
INQUIS
MEDICAL
K251189
Aventus Thrombectomy System
(IM-2001/IM-0002 – PE Indication)
510k Pre-Market Notification
Aspiration Catheter shaft incorporates a metallic reinforcement layer made of stainless steel, an inner liner and polymeric outer jacket having variable stiffness.
The System is provided with a 60-cc dual action manual syringe which allows for directional flow control and directs aspirated blood and clot into the Clot Canister.
## V. INDICATIONS FOR USE
The Aventus Thrombectomy System is indicated for:
- The non-surgical removal of emboli and thrombi from blood vessels.
- Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.
The Aventus Thrombectomy System is intended for use in the peripheral vasculature and for the treatment of pulmonary embolism.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The subject device, Aventus Thrombectomy System, is substantially equivalent to the predicate device, the Inari FlowTriever Retrieval/Aspiration System cleared under K191710. The intended/indications for use of the subject device is the same as the predicate, namely removal of thrombi or emboli from blood vessels, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel and is intended to be used in the peripheral vasculature and for the treatment of pulmonary embolism.
The subject and reference device (Aventus Thrombectomy System cleared under K240117) are identical and share the same technological characteristics in that both devices are single patient use, large bore aspiration catheters which utilize a 60-cc manual syringe as the aspiration source. Both the subject and predicate device incorporate disposable powered electronics and embedded software. From a manufacturing standpoint, both devices utilize shafts made with metallic (stainless steel) reinforced polymeric jackets with variable stiffness, radiopaque markings at the distal tip for fluoroscopic visualization, and use of stopcocks to direct the flow of fluids. Neither the subject or reference device includes any coating or surface materials.
## VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
| Performance Testing | Data provided |
| --- | --- |
| Biocompatibility Testing | Biocompatibility testing was leveraged from the reference device which was successfully completed in accordance with ISO 10993-1:2018 and the FDA Guidance re: Use of ISO-10993. Testing included:
• Cytotoxicity
• Sensitization
• Intracutaneous Reactivity
• Acute Systemic Toxicity
• Material Mediated Pyrogenicity
• Hemocompatibility (Hemolysis, Complement Activation, Thrombogenicity) |
Page 2 of 5
{6}
INQUIS
MEDICAL
K251189
Aventus Thrombectomy System
(IM-2001/IM-0002 – PE Indication)
510k Pre-Market Notification
| Performance Testing | Data provided |
| --- | --- |
| | This testing demonstrated the materials of the Aventus Thrombectomy System do not pose a risk of negative interaction with patients. |
| Sterilization | Sterilization testing was leveraged from the reference device which was successfully completed in accordance with ISO 14937:2009 - Sterilization of health care products — General requirements for characterization of a sterilizing agent and the development, validation and routine control of a sterilization process for medical devices and demonstrated an SAL of 10^{-6}.
Bacterial endotoxins test (BET), a.k.a. Limulus amebocyte lysate (LAL) testing was conducted on the predicate device per current test guidelines: USP <85> Bacterial Endotoxin Test and AAMI ST72 Bacterial endotoxins-test methodologies, routine monitoring and alternatives to batch testing and confirmed that the System meets established pyrogen limit specifications. |
| Distribution, Packaging and Shelf-Life Testing | Distribution and packaging testing was leveraged from the reference device which successfully demonstrated the integrity of the sterile barrier and preservation of the System's properties.
Shelf-life testing was leveraged from the reference device which has demonstrated preservation of the System's properties for the labeled twelve-month shelf-life. |
| Software Testing | Software documentation was leveraged from the reference device as recommended by FDA Guidance: Content of Premarket Submissions for Device Software Functions, issued June 14, 2023. |
| Electrical Safety / EMC Testing | Electrical Safety and EMC testing were leveraged from the reference device to ensure the subject device complies with the requirements of IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6 and IEC 62366-1. |
| Performance Testing – Bench | Design verification testing was leveraged from the reference device and demonstrated that the physical and functional requirements were met.
Specifically, the following was tested:
• Visual & Dimensional Inspection
• Tensile Strength
• Pressure/Leak Integrity
• Clot Burden Removal Validation
• Vacuum Test
• Leak Test
• Compatibility Testing
• Torque Transmission Testing
• Kink Testing
• Simulated Use
• Verification and Validation of Sensing |
| Performance Testing – Non-Clinical | GLP animal testing completed in compliance with GLP regulation (21 CFR Part 58) and in accordance with FDA Guidance: General Considerations for Animal Studies for Cardiovascular Devices (July 2010) and FDA Guidance: General Considerations for Animal Studies Intended to Evaluate Medical devices (March 2023) provided for the reference device, was leveraged for the subject device and demonstrated that the System was able to be used safely in a chronic large animal GLP study and met all pre-defined study endpoints. |
| Performance Testing – Clinical | Inquis Medical completed the AVENTUS trial, a prospective, multicenter, single-arm, non-blinded clinical trial to evaluate the safety and effectiveness of the Aventus Thrombectomy System. A total of 130 subjects were enrolled (120 intent-to-treat and 10 roll-in) at 22 sites in the United States. Review of computed tomography angiography was conducted by a centralized core laboratory to determine change in RV/LV ratio from baseline to 48 hours. Major adverse events were adjudicated for clinical endpoint analysis by an independent clinical events committee (CEC). The primary safety endpoint was the composite of device related major adverse events (MAEs) within 48 hours of the index procedure. The primary effectiveness endpoint was the change in RV/LV ratio from baseline to 48 hours.
Of the 120 subjects, the majority were men (58.8%), and the cohort had a mean BMI of 35.6. White non-Hispanic or Latino subjects were in the majority, with 17.2% of subjects being Black/African American. Key |
Page 3 of 5
{7}
INQUIS
MEDICAL
K251189
Aventus Thrombectomy System
(IM-2001/IM-0002 – PE Indication)
510k Pre-Market Notification
| Performance Testing | Data provided |
| --- | --- |
| | medical history included hypertension (54.2%), concomitant deep vein thrombosis (84.3%), prior PE (8.3%) and history of cancer (9.2%).
The primary safety endpoint was met as there were no device-related MAEs in the study. |
| | ITT (N=120) |
| 48-Hr* Event Type | # AEs | # (%) Subjects | p-value** |
| Death | 0 | 0 (0.0%) | |
| Major Bleeding | 0 | 0 (0.0%) | |
| Clinical Deterioration | 0 | 0 (0.0%) | |
| Pulmonary Vascular Injury | 0 | 0 (0.0%) | |
| Cardiac Injury | 0 | 0 (0.0%) | |
| Total | 0 | 0 (0.0%) | <0.0001 |
| * Window for follow-up was 48 hours ± 8 hours. Data table includes device-related MAEs through 56 hours.
**p-value comparing the proportion of subjects with a MAE within 56 hours of the procedure to the performance goal of 25% |
| Information on non-device-related major adverse events at 48 hours is presented in the table below: |
| | ITT (N=120) |
| 48-Hr* Event Type | # AEs | # (%) Subjects |
| Death | 0 | 0 (0.0%) |
| Major Bleeding | 1 | 1 (0.8%) |
| Clinical Deterioration | 2 | 2 (1.7%) |
| Pulmonary Vascular Injury | 0 | 0 (0.0%) |
| Cardiac Injury | 0 | 0 (0.0%) |
| Total | 3 | 3 (2.5%) |
| * Window for follow-up was 48 hours ± 8 hours. Data table includes all MAEs through 56 hours.
The primary efficacy endpoint was met as the change in RV/LV from baseline to 48 hours was significantly better than the performance goal. |
| RV/LV Ratio | Baseline | 48 Hour/Discharge | Change from Baseline | Percent Change from Baseline |
| N | 119 | 116 | 116 | 116 |
| Mean ± SD | 1.57 ± 0.40 | 1.10 ± 0.20 | 0.47 ± 0.36 | 27.0% ± 17.1% |
| Median | 1.50 | 1.08 | 0.40 | 27.5% |
| Min, Max | 0.85, 3.01 | 0.70, 1.93 | -0.45, 1.76 | -52.9%, 58.5% |
| p-value | | | <0.0001 | |
| The total number of days in the hospital following the index procedure was 2.6 ± 2.2 days and there were no thrombolytics used within 48 hours of the procedure, N=119.
As part of the AVENTUS trial, the cleared Blood Return Filter component of the Aventus Thrombectomy System was utilized to return blood to patients. Additional information may be found in the Instructions for Use. |
Page 4 of 5
{8}
INQUIS
MEDICAL
K251189
Aventus Thrombectomy System
(IM-2001/IM-0002 – PE Indication)
510k Pre-Market Notification
## VIII. CONCLUSIONS
In conclusion, the intended use, indications for use, and technological characteristics of the Aventus Thrombectomy System are substantially equivalent to the predicate and reference devices. Clinical performance testing has demonstrated that the Aventus Thrombectomy System is substantially equivalent to the predicate device.
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.