ORQIS EXTRACORPOREAL BLOOD PUMPING SYSTEM, MODEL BPS-100

K073631 · Orqis Medical Corporation · KFM · Jun 27, 2008 · Cardiovascular

Device Facts

Record IDK073631
Device NameORQIS EXTRACORPOREAL BLOOD PUMPING SYSTEM, MODEL BPS-100
ApplicantOrqis Medical Corporation
Product CodeKFM · Cardiovascular
Decision DateJun 27, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4360
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Orqis Extracorporeal Blood Pumping System is intended to be used as a blood pumping unit in extracorporeal circulation for cardiac support that may, or may not, require cardiopulmonary bypass (for up to six hours). The Orqis Pump Tubing is intended for use in extracorporeal circuits designed for cardiac support.

Device Story

The Orqis Extracorporeal Blood Pumping System is a non-pulsatile, non-roller centrifugal blood pump used for extracorporeal circulation. The system comprises a sterile, single-use disposable blood pump, a reusable motor unit, a microprocessor-based controller, an ultrasonic flow sensor, and medical-grade PVC pump tubing. The motor drives a magnet sealed within the pump's rotor to generate centrifugal force, moving blood through the circuit. The controller manages motor speed and monitors system parameters, including flow and pressure, providing visual and audible alarms. The device is intended for clinical use during cardiac support procedures lasting up to six hours. By providing mechanical circulatory support, the system assists in maintaining blood flow during complex surgeries, potentially improving patient outcomes during procedures where cardiopulmonary bypass is required or indicated.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Centrifugal blood pump; non-pulsatile; non-roller. Materials: Aluminum and polycarbonate (motor shell), medical-grade PVC (tubing). Sensing: Ultrasonic flow meter. Energy: Electric motor. Connectivity: Standalone controller with alphanumeric display and alarm system. Sterilization: Pump tubing provided sterile; motor provided non-sterile.

Indications for Use

Indicated for patients requiring extracorporeal circulatory support for up to six hours, including procedures requiring cardiopulmonary bypass or other circulatory support (e.g., valvuloplasty, mitral valve reoperation, surgery of the vena cava or aorta, liver transplants).

Regulatory Classification

Identification

A nonroller-type cardiopulmonary and circulatory bypass blood pump is a prescription device that uses a method other than revolving rollers to pump the blood through an extracorporeal circuit for periods lasting less than 6 hours for the purpose of providing either:(i) Full or partial cardiopulmonary bypass (i.e., circuit includes an oxygenator) during open surgical procedures on the heart or great vessels; or(ii) Temporary circulatory bypass for diversion of flow around a planned disruption of the circulatory pathway necessary for open surgical procedures on the aorta or vena cava.

Special Controls

*Classification* —Class II (special controls). The special controls for this device are:(i) Non-clinical performance testing must perform as intended over the intended duration of use and demonstrate the following: Operating parameters, dynamic blood damage, heat generation, air entrapment, mechanical integrity, and durability/reliability; (ii) The patient-contacting components of the device must be demonstrated to be biocompatible; (iii) Sterility and shelf life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components; and (iv) Labeling must include information regarding the duration of use, and a detailed summary of the device- and procedure-related complications pertinent to use of the device. (b) *Nonroller-type temporary ventricular support blood pump* —(1)*Identification.* A nonroller-type temporary ventricular support blood pump is a prescription device that uses any method resulting in blood propulsion to provide the temporary ventricular assistance required for support of the systemic and/or pulmonary circulations during periods when there is ongoing or anticipated hemodynamic instability due to immediately reversible alterations in ventricular myocardial function resulting from mechanical or physiologic causes. Duration of use would be less than 6 hours.(2) *Classification.* Class III (premarket approval).(c) *Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with FDA on or before September 8, 2015, for any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976, or that has, on or before September 8, 2015, been found to be substantially equivalent to any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976. Any other nonroller-type temporary ventricular support blood pump shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 2 510(k) Summary | Submitter: | Orqis Medical Corporation<br>14 Orchard Road, Suite 100<br>Lake Forest, CA 92630 | |------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Ryan Kelly<br>Senior Regulatory Affairs Specialist | | Date Prepared: | December 21, 2007 | | Trade Name: | Orqis Extracorporeal Blood Pumping System<br>Orqis Blood Pump<br>Orqis Motor<br>Orqis Controller<br>Orqis Flow Sensor<br>Orqis Pump Tubing | | Classification Name: | Pump, Blood, Cardiopulmonary Bypass, Non-roller type<br>(21 CFR 870.4360)<br>Control, Pump Speed, Cardiopulmonary Bypass (21 CFR 870.4380)<br>Tubing, Pump, Cardiopulmonary Bypass (21 CFR 870.4390) | | Predicate Devices: | Levitronix CentriMag® Extracorporeal Blood Pumping System, K020271<br>Cardiac Assist TandemHeart Escort Controller, K061369<br>Olsen Medical Sales Plastron Pump Tubing, K013578 | | Device Description: | The Orqis Extracorporeal Blood Pumping System consists of a pump, independent motor unit, a controller, a flow sensor, and some pump tubing. The system also has two accessories: a holster for the pump and motor, and a controller stand.<br>The Orqis Blood Pump is a sterile, single-use, disposable, non-pulsatile, non-roller pump that utilizes a rotor to impart energy to the blood in an extracorporeal circuit through centrifugal forces. The inlet of the pump runs concentric with the axis of the rotor. | | The Orqis Motor is a small, compact, reusable unit that holds the disposable blood pump and drives the rotor inside the blood pump. The Motor turns a magnet that is hermetically sealed in the pump's rotor at a speed that is set by the controller. The outer shell of the motor is built from aluminum and polycarbonate. The motor is provided non-sterile. | | | The Orqis Controller is a microprocessor-based device, which drives and controls the blood pump, through the motor. The Controller also provides monitoring of motor speed, fluid flow, and pressure through the use of alphanumeric display and audible and visual alarms. | | | The Orqis Flow Sensor is an ultrasonic flow meter that measures fluid flow through standard 3/8" tubing. | | | The Orqis Pump Tubing is medical grade PVC tubing used to connect the Blood Pump to external devices. The Pump Tubing is provided sterile. | | | Intended Use: | The Orqis Extracorporeal Blood Pumping System is intended to be used as a blood pumping unit in extracorporeal circulation for cardiac support that may, or may not, require cardiopulmonary bypass (for up to six hours). | | The Orqis Pump Tubing is intended for use in extracorporeal circuits designed for cardiac support. | | | Comparative Analysis: | The Orqis Extracorporeal Blood Pumping System has been demonstrated to be as safe and effective as the predicate device for its intended use. | | Functional Testing: | The Orqis Extracorporeal Blood Pumping System has successfully undergone functional testing demonstrating equivalence to the predicate device | | Conclusion: | The Orqis Extracorporeal Blood Pumping System is substantially equivalent to the predicate devices. | {1}------------------------------------------------ : : . {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The eagle is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" in capital letters. The seal represents the department's mission to protect the health of all Americans and provide essential human services. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 2 7 2008 Mr. Ryan Kelly Senior Regulatory Affairs Specialist Orqis Medical Corporation 14 Orchard Road, Suite 100 Lake Forest, CA 92630 K073631 Re: Orqis® Extracorporeal Blood Pumping System Regulation Number: 21 CFR 870.4360 Regulation Name: Pump, Blood, Cardiopulmonary Bypass, Non-roller type Regulatory Class: Class III Product Code: KFM Dated: June 11, 2008 Received: June 13, 2008 Dear Mr. Kelly: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ ## Page 2 - Mr. Ryan Kelly Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a dctermination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. R. Kochner A Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## 1 Indications for Use Statement 510(k) Number (if known): K073631 Device Name: Orqis® Extracorporeal Blood Pumping System Indications for Use: The Orqis Extracorporeal Blood Pumping system is indicated to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants, etc). The Orqis Pump Tubing is indicated for use in extracorporeal circuits for circulatory support lasting six-hours or less. Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR (Part 21 CFR 801 Subpart D) Over-The-Counter Use _ (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) R. Jachner (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number. Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...