Affinity Pixie Oxygenator with Balance Biosurface (Model BBP211), Affinity Pixie Oxygenator and Cardiotomy/Venous Reservoir with Balance Biosurface (Model BBP241), Affinity Pixie Oxygenator with Cortiva BioActive Surface (Model CBP211), Affinity Pixie Oxygenator and Cardiotomy/Venous Reservoir with Cortiva BioActive Surface (Model CBP241)
K183511 · Medtronic, Inc. · DTZ · Mar 18, 2019 · Cardiovascular
Device Facts
Record ID
K183511
Device Name
Affinity Pixie Oxygenator with Balance Biosurface (Model BBP211), Affinity Pixie Oxygenator and Cardiotomy/Venous Reservoir with Balance Biosurface (Model BBP241), Affinity Pixie Oxygenator with Cortiva BioActive Surface (Model CBP211), Affinity Pixie Oxygenator and Cardiotomy/Venous Reservoir with Cortiva BioActive Surface (Model CBP241)
Applicant
Medtronic, Inc.
Product Code
DTZ · Cardiovascular
Decision Date
Mar 18, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4350
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Intended Use
The Affinity Pixie™ Oxygenator with Balance™ Biosurface is indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass (CPB) procedures requiring a blood flow rate up to 2.0 L/min. The Affinity Pixie™ Oxygenator with Balance™ Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine CPB procedures up to 6 hours in duration. The Affinity Pixie™ Cardiotomy/Venous Reservoir with Balance™ Biosurface is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum-assisted venous drainage (VAVD) procedures.
Device Story
Affinity Pixie™ Oxygenator and Cardiotomy/Venous Reservoir (CVR) systems are extracorporeal perfusion components; used in CPB circuits for neonate, infant, and small pediatric patients. Devices oxygenate/remove CO2 from blood; provide thermal regulation (cooling/warming); collect venous/cardiotomy suctioned blood. CVR supports vacuum-assisted venous drainage (VAVD). Systems feature Balance™ Biosurface or Cortiva™ BioActive Surface coatings to improve hemocompatibility. Operated by perfusionists in clinical settings (e.g., OR) during procedures up to 6 hours. Output is processed blood returned to patient; assists clinicians in maintaining physiological support during cardiac surgery.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Extracorporeal oxygenator and reservoir system; includes Balance™ Biosurface or Cortiva™ BioActive Surface coatings. Designed for pediatric/neonate flow rates up to 2.0 L/min. Supports VAVD. Sterile, single-use device.
Indications for Use
Indicated for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass (CPB) procedures requiring blood flow rates up to 2.0 L/min.
Regulatory Classification
Identification
A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”
Related Devices
K141233 — AFFINITY PIXIE OXYGENATION SYSTEM WITH; CARMEDA BIOACTIVE SURFACE, BALANCE BIOSUFACE · Medtronic, Inc. · Jun 12, 2014
K172984 — Affinity Pixie Oxygenator with Balance Biosurface, Affinity Pixie Oxygenator with Cardiotomy/Venous Reservoir and Balance Biosurface, Affinity Pixie Oxygenator with Cortiva BioActive Surface, Affinity Fusion Oxygenator with Cardiotomy/Venous Reservoir and Cortiva BioActive Surface · Medtronic, Inc. · Nov 20, 2017
K100645 — AFFINITY PIXIE OXYGENATION SYSTEM WITH CARMEDA BIOACTIVE SURFACE OR BALANCE BIOSURFACE · Medtronic, Inc. · Sep 14, 2012
K191029 — Affinity NTTM Oxygenator (Model 511), Affinity® NT Hollow Fiber Oxygenator with Plasma Resistant Fiber (PRF) with TrilliumTM Biosurface (Model 511T), Affinity® NT Integrated CVR/Membrane Oxygenator with Plasma Resistant Fiber (Model 541), Affinity® NT Integrated CVR/Membrane Oxygenator with Plasma Resistant Fiber (Model 541-R), Affinity NT® Integrated CVR/Membrane Oxygenator with Trillium® Biosurface (Model 541T), Affinity NT® Integrated CVR/Membrane Oxygenator with Trillium® Biosurfac · Medtronic, Inc. · May 17, 2019
K973760 — AFFINITY HOLLOW FIBER OXYGENATOR WITH TRILLIUM BIOPASSIVE SURFACE · Avecor Cardiovascular, Inc. · Feb 18, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
March 18, 2019
Medtronic Inc Tyler Senjem Regulatory Affairs Specialist 7611 Northland Dr Minneapolis, Minnesota 55428
### Re: K183511
Trade/Device Name: Affinity Pixie™ Oxygenator with Balance™ Biosurface (Model BBP211), Affinity Pixie™ Oxygenator and Cardiotomy/Venous Reservoir with Balance™ Biosurface (Model BBP241), Affinity Pixie™ Oxygenator with Cortiva™ BioActive Surface (Model CBP211), Affinity Pixie™ Oxygenator and Cardiotomy/Venous Reservoir with Cortiva™ BioActive Surface (Model CBP241) Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II
Product Code: DTZ Dated: January 16, 2019 Received: January 17, 2019
Dear Tyler Seniem:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Fernando Aguel Fernando Aguel - S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known) K183511
Device Name
Affinity Pixie™ Oxygenator with Balance™ Biosurface (Model BBP211)
Indications for Use (Describe)
The Affinity Pixie™ Oxygenator with Balance™ Biosurface is indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass (CPB) procedures requiring a blood flow rate up to 2.0 L/min.
The Affinity Pixie™ Oxygenator with Balance™ Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine CPB procedures up to 6 hours in duration.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">✖</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{3}------------------------------------------------
510(k) Number (if known) K183511
#### Device Name
Affinity Pixie™ Oxygenator and Cardiotomy/Venous Reservoir with Balance™ Biosurface (Model BBP241)
### Indications for Use (Describe)
The Affinity Pixie™ Oxygenator and Cardiotomy/Venous Reservoir with Balance™ Biosurface are indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass (CPB) procedures requiring a blood flow rate up to 2.0 L/min.
The Affinity Pixie™ Oxygenator with Balance™ Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine CPB procedures up to 6 hours in duration.
The Affinity Pixie™ Cardiotomy/Venous Reservoir with Balance™ Biosurface is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum-assisted venous drainage (VAVD) procedures.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;"> <svg height="15" width="15"> <rect fill="white" height="15" stroke="black" width="15" x="0" y="0"></rect> <path d="M2,2 L13,13 M13,2 L2,13" stroke="black"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: sans-serif;"> <svg height="15" width="15"> <rect fill="white" height="15" stroke="black" width="15" x="0" y="0"></rect> </svg> </span> Over-The-Counter Use (21 CFR 807 Subpart C) |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| X | Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{4}------------------------------------------------
510(k) Number (if known) K183511
Device Name
Affinity Pixie™ Oxygenator with Cortiva™ BioActive Surface (Model CBP211)
Indications for Use (Describe)
The Affinity Pixie™ Oxygenator with Cortiva™ BioActive Surface is indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass (CPB) procedures requiring a blood flow rate up to 2.0 L/min.
The Affinity Pixie™ Oxygenator with Cortiva™ BioActive Surface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine CPB procedures up to 6 hours in duration.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
| <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{5}------------------------------------------------
510(k) Number (if known) K183511
#### Device Name
Affinity Pixie™ Oxygenator and Cardiotomy/Venous Reservoir with Cortiva™ BioActive Surface (Model CBP241)
### Indications for Use (Describe)
The Affinity Pixie™ Oxygenator and Cardiotomy/Venous Reservoir with Cortiva™ BioActive Surface are indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass (CPB) procedures requiring a blood flow rate up to 2.0 L/min.
The Affinity Pixie™ Oxygenator with Cortiva™ BioActive Surface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine CPB procedures up to 6 hours in duration.
The Affinity Pixie™ Cardiotomy/Venous Reservoir with Cortive Surface is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum-assisted venous drainage (VAVD) procedures.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|--------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.