INSPIRE 8F DUAL HOLLOW FIBER OXYGENATOR WITH INTEGRATED ARTERIAL FILTER AND HARDSHELL RESERVOIR

K122844 · Sorin Group Italia S.R.L. · DTZ · Oct 17, 2012 · Cardiovascular

Device Facts

Record IDK122844
Device NameINSPIRE 8F DUAL HOLLOW FIBER OXYGENATOR WITH INTEGRATED ARTERIAL FILTER AND HARDSHELL RESERVOIR
ApplicantSorin Group Italia S.R.L.
Product CodeDTZ · Cardiovascular
Decision DateOct 17, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.4350
Device ClassClass 2
AttributesTherapeutic

Intended Use

INSPIRE 8F M: Hollow Fiber Oxygenator INSPIRE 8F M is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It provides gas exchange support and blood temperature control. INSPIRE 8F M integrated arterial filter provides additional protection against air and solid emboli. INSPIRE 8F M is intended to be used for 6 hours or less. INSPIRE HVR DUAL: Hardshell Venous/Cardiotomy Reservoir INSPIRE HVR DUAL is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It collects, defoams and filters venous blood and suction blood. INSPIRE HVR DUAL can be used post-operatively for chest drainage. INSPIRE HVR DUAL is intended to be used for 6 hours or less.

Device Story

The INSPIRE 8F DUAL is an integrated cardiopulmonary bypass system comprising a microporous hollow fiber membrane oxygenator, heat exchanger, arterial filter, and a dual-chamber hardshell venous/cardiotomy reservoir. It processes venous and suction blood by removing carbon dioxide, adding oxygen, controlling temperature, and filtering emboli. The device is operated by perfusionists in surgical settings. The output is oxygenated, temperature-regulated blood returned to the patient's arterial circulation. The system supports clinical decision-making by maintaining physiological gas exchange and hemodynamic stability during bypass. Benefits include integrated air/solid emboli protection and post-operative chest drainage capability. The device is a modification of the INSPIRE 8F, featuring a dual-chamber reservoir instead of a single-chamber design.

Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including structural integrity, blood/gas pathway integrity, air handling, filtration efficiency, hemolysis, and blood compatibility testing conducted in accordance with ISO 10993-1, ISO 15674, and relevant FDA guidance.

Technological Characteristics

Microporous hollow fiber membrane oxygenator; integrated heat exchanger; dual-chamber hardshell venous/cardiotomy reservoir; arterial filter. Ethylene oxide sterilized; non-pyrogenic fluid path; single-use. Flow rate capacity up to 8 L/min. Materials and construction identical to predicate (K121536).

Indications for Use

Indicated for adult and small adult patients undergoing surgical procedures requiring cardiopulmonary bypass for up to 6 hours. Provides gas exchange, blood temperature control, filtration of air/solid emboli, and collection/defoaming of venous and suction blood.

Regulatory Classification

Identification

A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K122844 INSPIRE 8F DUAL Sorin Group Italia S.r.I. Special 510(k) September 14, 2012 OCT 1 7 2012 # 510(k) SUMMARY SUBMITTER: CONTACT PERSON: DATE PREPARED: COMMON NAMES: DEVICE TRADE NAME: Sorin Group Italia 86. Via Statale 12 Nord 41037 Mirandola (MO) Italy Luigi Vecchi Phone: 39 0535 29811 Fax: 39 0535 25229 September 14, 2012 INSPIRE 8F DUAL hollow fiber oxygenator with integrated arterial filter and hardshell venous/cardiotomy reservoir Hollow Fiber Oxygenator with integrated arterial filter and hardshell venous/cardiotomy reservoir Hollow Fiber Oxygenator with integrated arterial filter Hardshell Venous/Cardiotomy Reservoir Cardiopulmonary Bypass Oxygenator/ Cardiopulmonary Bypass Heat Exchanger/ Cardiopulmonary Bypass Blood Reservoir/ Cardiopulmonary Bypass Defoamer Cardiopulmonary Bypass Arterial Line Blood Filter UNMODIFIED DEVICES: CLASSIFICATION NAME: INSPIRE 8F hollow fiber oxygenator with integrated arterial filter and hardshell venous/cardiotomy reservoir (K121536) # DEVICE DESCRIPTION: The INSPIRE 8F DUAL is a high efficiency microporous hollow fiber membrane oxygenator integrated with an arterial filter and a heat exchanger (INSPIRE 8F M) and connected to a hardshell venous/cardiotomy reservoir (INSPIRE HVR DUAL). A molded fitting joint connects the oxygenator to the reservoir. The device can be operated at flow rates up to 8 liters per minute (I/min). The hollow fiber membrane oxygenator provides oxygenation and carbon dioxide removal from venous blood or suction blood. The integrated heat exchanger controls blood temperature and allows the use of hypothermia or aids in the maintenance of normothermia during surgery. The integrated arterial filter provides additional protection against air and solid emboli and the integrated hardshell reservoir collects, defoams, filters venous and suction blood, and can be used post-operatively for chest drainage. The INSPIRE 8F DUAL is a modified version of the currently marketed INSPIRE 8F integrated oxygenator/hardshell venous cardiotomy reservoir system. {1}------------------------------------------------ # INDICATION FOR USE: The intended uses for the two elements that constitute the integrated device are: #### INSPIRE 8F M: Hollow Fiber Oxygenator INSPIRE 8F M is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It provides gas exchange support and blood temperature control. INSPIRE 8F M integrated arterial filter provides additional protection against air and solid emboli. INSPIRE 8F M is intended to be used for 6 hours or less. #### INSPIRE HVR DUAL: Hardshell Venous/Cardiotomy Reservoir INSPIRE HVR DUAL is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It collects, defoams and filters venous blood and suction blood. INSPIRE HVR DUAL can be used post-operatively for chest drainage. INSPIRE HVR DUAL is intended to be used for 6 hours or less. ### TECHNOLOGICAL CHARACTERISTICS: The INSPIRE 8F DUAL has the same fundamental technological characteristics, principles of operation and control mechanisms as the unmodified device. The INSPIRE 8F DUAL integrated device will be provided with a dual chamber venous/cardiotomy reservoir (INSPIRE HVR DUAL) rather than a single chamber venous/cardiotomy reservoir (INSPIRE HVR) as for the INSPIRE 8F unmodified device. No change to the oxygenating module has been made as a result of this modification. The INSPIRE 8F DUAL modified device is provided with oxygenating module identical with respect to the unmodified device. No change to the intended use has been made as a result of these modifications. The INSPIRE 8F DUAL and the INSPIRE 8F unmodified device share the same fundamental technological characteristics except for some modifications that do not affect the basic device function. These differences do not raise any new issues of safety and effectiveness. The INSPIRE 8F DUAL is substantially equivalent to the INSPIRE 8F on the basis of operating principles and basic function. There are no differences in packaging type and material between unmodified and modified device. . The INSPIRE 8F DUAL is ethylene oxide sterilized and has a non-pyrogenic fluid path. It is for single use only. {2}------------------------------------------------ # NON CLINICAL TEST RESULTS: Applicable tests were conducted in accordance with the requirements of ISO 10993-1 and the FDA May 1st, 1995 Memorandum on the use of the ISO 10993 standard for biocompatibility testing of materials. # IN VITRO TEST RESULTS: In vitro testing was conducted in accordance with the relevant requirements of "Guidance for Extracorporeal Blood Circuit Defoamer 510(k) Submissions; Final Guidance for Industry and FDA" issued on November 29, 2000; ISO 15674, "Cardiovascular implants and artificial organs — Hard-shell cardiotomy/venous reservoir systems (with/without filter) and soft venous reservoir bags". In vitro testing was conducted on hardshell venous/cardiotomy reservoir to demonstrate unmodified device substantial equivalency and compliance to safety and effectiveness requirements. The modified INSPIRE 8F DUAL and unmodified INSPIRE 8F utilize identical oxygenating modules integrated with an arterial filter and a heat exchanger. This 510(k) cross references the following oxygenating module performance data previously submitted in the INSPIRE 8F (K121536): structural integrity, blood, water, gas pathway integrity, blood volume capacity, gas transfer performance/blood side pressure drop, heat exchange performance/water side pressure drop, air handling capability, filtration efficiency, hemolysis, blood compatibility, leaching of coating, flaking of coating, uniformity of coating, The following table lists the performance and physical/mechanical integrity tests conducted to demonstrate compliance to the product's performance specifications. The INSPIRE 8F DUAL successfully met all acceptance criteria for each test. | TEST | TEST CLASSIFICATION | TEST TITLE | |------|------------------------|--------------------------------------------| | 1 | Physical/Mechanical | Structural integrity | | 2 | Physical/Mechanical | Blood pathway integrity | | 3 | Functional/Performance | Blood rest volume | | 4 | Functional/Performance | Air handling | | 5 | Functional/Performance | Break-through time and volume | | 6 | Functional/Performance | Defoaming efficiency | | 7 | Functional/Performance | Dynamic priming volume / Hold-up | | 8 | Functional/Performance | Filtration efficiency - venous section | | 9 | Functional/Performance | Filtration efficiency - cardiotomy section | | 10 | Functional/Performance | Flow rate capacity | {3}------------------------------------------------ | TEST | TEST CLASSIFICATION | TEST TITLE | |------|------------------------|-----------------------| | 11 | Functional/Performance | Pressure drop | | 12 | Functional/Performance | Hemolysis | | 13 | Functional/Performance | Blood compatibility | | 14 | Functional/Performance | Leaching of coating | | 15 | Functional/Performance | Flaking of coating | | 16 | Functional/Performance | Uniformity of coating | ### CONCLUSIONS: The results of in vitro studies demonstrate that the modified device performs in a manner substantially equivalent to the unmodified device with respect to the relevant functional parameters. Test results of this study demonstrate the INSPIRE 8F DUAL is equivalent to the INSPIRE 8F unmodified device with respect to device function. Additional testing has also demonstrated the effectiveness of production techniques to assure that the device is sterile and non-pyrogenic. {4}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image contains the text "Public Health Service". The text is in a simple, sans-serif font and is horizontally aligned. The text is likely part of a document or sign, indicating an affiliation with a public health organization. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 OCT 1 7 200 Sorin Group USA, Inc. c/o Mr. Scott Light Regulatory Affairs Manager 14401 W. 65th Way Arvada, CO 80004 Re: K122844 Trade/Device Names: INSPIRE 8F DUAL Hollow Fiber Oxygenator with Integrated Arterial Filter and Dual Chamber Hardshell Venous/Cardiotomy Reservoir Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II Product Code: DTZ, DTN Dated: September 14, 2012 Received: September 17, 2012 Dear Mr. Light: We have reviewed your Section 510(k) premarket notification of intent to market the devices referenced above and have determined the devices are substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ # Page 2 - Mr. Scott Light Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, G.A. Kleiber Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ 510(k) Number (if known): Device Name: INSPIRE 8F DUAL hollow fiber oxygenator with integrated arterial filter and hardshell venous/cardiotomy reservoir K122844 Indication for Use: #### INSPIRE 8F M: Hollow Fiber Oxygenator INSPIRE 8F M is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It provides gas exchange support and blood temperature control. INSPIRE 8F M integrated arterial filter provides additional protection against air and solid emboli. INSPIRE 8F M is intended to be used for 6 hours or less. ### INSPIRE HVR DUAL: Hardshell Venous/Cardiotomy Reservoir INSPIRE HVR DUAL is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It collects, defoams and filters venous blood and suction blood. INSPIRE HVR DUAL can be used post-operatively for chest drainage. INSPIRE HVR DUAL is intended to be used for 6 hours or less. Prescription Use × (Part 21CFR 801 Subpart D) Over-the- Counter Use AND/OR (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) 177844 M.A. Killelea on Signician Devices 510(k) Number_
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...