INSPIRE 6F HOLLOW FIBER OXYGENATOR WITH INTEGRATED ARTERIAL FILTER AND HARDSHELL VENOUS/CARDIOTOMY RESERVIOR

K120185 · Sorin Group Italia S.R.L. · DTZ · May 11, 2012 · Cardiovascular

Device Facts

Record IDK120185
Device NameINSPIRE 6F HOLLOW FIBER OXYGENATOR WITH INTEGRATED ARTERIAL FILTER AND HARDSHELL VENOUS/CARDIOTOMY RESERVIOR
ApplicantSorin Group Italia S.R.L.
Product CodeDTZ · Cardiovascular
Decision DateMay 11, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4350
Device ClassClass 2
AttributesTherapeutic

Indications for Use

INSPIRE 6F M: Hollow Fiber Oxygenator INSPIRE 6F M is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It provides gas exchange support and blood temperature control. INSPIRE 6F M integrated arterial filter provides additional protection against air and solid emboli. INSPIRE 6F M is intended to be used for 6 hours or less. INSPIRE HVR: Hardshell Venous/Cardiotomy Reservoir INSPIRE HVR is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It collects, defoams and filters venous blood and suction blood. INSPIRE HVR can be used post-operatively for chest drainage. INSPIRE HVR is intended to be used for 6 hours or less.

Device Story

INSPIRE 6F is an integrated cardiopulmonary bypass system comprising a microporous hollow fiber membrane oxygenator, heat exchanger, arterial filter, and hardshell venous/cardiotomy reservoir. Used in OR by perfusionists/surgeons during adult/small adult cardiac surgery. Device receives venous blood; oxygenator performs gas exchange (O2/CO2); heat exchanger regulates blood temperature; integrated arterial filter removes air/solid emboli; reservoir collects/defoams/filters venous and suction blood. Output is oxygenated, temperature-controlled, filtered blood returned to patient. System supports clinical decision-making by maintaining physiological homeostasis during bypass. Benefits include integrated design reducing circuit complexity and providing continuous protection against emboli.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing only. In vitro studies performed per ISO 15675, ISO 15674, and FDA guidance documents. Testing included structural integrity, gas transfer, heat exchange, air handling, filtration efficiency, hemolysis, and blood compatibility. All tests passed, confirming performance equivalent to predicate devices.

Technological Characteristics

Integrated cardiopulmonary bypass system. Components: microporous hollow fiber membrane oxygenator, heat exchanger, arterial filter, hardshell venous/cardiotomy reservoir. Flow rate up to 6 L/min. Sterilization: ethylene oxide. Fluid path: non-pyrogenic. Single-use. Materials and design consistent with predicate devices (D905 EOS, D733).

Indications for Use

Indicated for adult and small adult patients undergoing surgical procedures requiring cardiopulmonary bypass for up to 6 hours. Provides gas exchange, blood temperature control, filtration of air/solid emboli, and collection/defoaming of venous and suction blood. Can be used post-operatively for chest drainage.

Regulatory Classification

Identification

A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ # K120185 #### 510(k) SUMMARY May 08, 2012 arterial filter # MAY 1 1 2012 Sorin Group Italia S.r.I. SUBMITTER: 86, Via Statale 12 Nord 41037 Mirandola (MO) Italy CONTACT PERSON: Luigi Vecchi Phone: 011 39 0535 29811 Fax: 011 39 0535 25229 DATE PREPARED: DEVICE TRADE NAME: COMMON NAMES: CLASSIFICATION NAMES: Cardiopulmonary Bypass Oxygenator/ Cardiopulmonary Bypass Heat Exchanger/ Cardiopulmonary Bypass Blood Reservoir/ Cardiopulmonary Bypass Defoamer/ Cardiopulmonary Bypass Arterial Line Blood Filter INSPIRE 6F hollow fiber oxygenator with Hollow Fiber Oxygenator with integrated Hollow Fiber Oxygenator with integrated Hardshell Venous/Cardiotomy Reservoir integrated arterial filter and hardshell venous/cardiotomy reservoir arterial filter and hardshell venous/cardiotomy reservoir D905 EOS: hollow fiber oxygenator with integrated hardshell venous/cardiotomy reservoir (K043323) D733 MICRO 40 Ph.I.S.I.O. Arterial Filter: Sorin D733 MICRO 40 Arterial Filter with 40 micron screen with phosphoryIcholine coating (K112525) #### DEVICE DESCRIPTION: PREDICATE DEVICE: The INSPIRE 6F is consisting of an oxygenator, integrated with an arterial filter and a heat exchanger (INSPIRE 6F M), and a hardshell venous/cardiotomy reservoir (INSPIRE HVR). The reservoir is connected to the gas exchange module by means of a molded fitting joint. The INSPIRE 6F is a high efficiency microporous hollow fiber membrane oxygenator, integrated with an arterial filter and a heat exchanger, connected to a hardshell venous/cardiotomy reservoir. {2}------------------------------------------------ The device can be operated at flow rates up to 6 liters per minute (I/min). The hollow fiber membrane oxygenator provides oxygenation and carbon dioxide removal from venous blood. The integrated heat exchanger controls blood temperature and allows the use of hypothermia or aids in the maintenance of normothermia during surgery. The integrated arterial filter provides additional protection against air and solid emboli and the integrated hardshell reservoir collects, defoams, filters venous and suction blood, and can be used post-operatively for chest drainage. The INSPIRE 6F is a modified version of the currently marketed integrated oxygenator/hardshell venous cardiotomy reservoir system (D905 EOS) and of the arterial filter (D733 MICRO 40 Ph.I.S.I.O., hereinafter referred to as D733) respectively in commercial distribution as separate units. #### INDICATION FOR USE: The intended use for the two elements that constitute the integrated device are: #### INSPIRE 6F M: Hollow Fiber Oxygenator INSPIRE 6F M is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It provides gas exchange support and blood temperature control. INSPIRE 6F M integrated arterial filter provides additional protection against air and solid emboli. INSPIRE 6F M is intended to be used for 6 hours or less. #### INSPIRE HVR: Hardshell Venous/Cardiotomy Reservoir INSPIRE HVR is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It collects, defoams and filters venous blood and suction blood. INSPIRE HVR can be used post-operatively for chest drainage. INSPIRE HVR is intended to be used for 6 hours or less. #### TECHNOLOGICAL CHARACTERISTICS: The INSPIRE 6F has the same fundamental technological characteristics, principles of operation and control mechanisms as the predicate devices. Sorin believes that the INSPIRE 6F is substantially equivalent to the D905 EOS on the basis of operating principles and basic function. The integrated arterial filter of INSPIRE 6F is also substantially equivalent to the D733 predicate device with respect to the expected main function of an arterial filter. The INSPIRE 6F and the predicate devices share the same fundamental technological characteristics except for some modifications that do not affect the basic device function. Any differences do not raise any new issues of safety and effectiveness. The INSPIRE 6F is ethylene oxide sterilized and has a non-pyrogenic fluid path. It is for single use only. #### NON CLINICAL TEST RESULTS: Applicable tests were carried out in accordance with the requirements of ISO 10993-1 and the FDA May 1st, 1995 Memorandum on the use of the ISO 10993 standard for biocompatibility testing of materials. {3}------------------------------------------------ #### IN VITRO TEST RESULTS: In vitro testing was carried out in accordance with the relevant requirements of "Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions; Final Guidance for Industry and FDA Staff" issued on November 13, 2000, "Guidance for Extracorporeal Blood Circuit Defoamer 510(k) Submissions; Final Guidance for Industry and FDA" issued on November 29, 2000; "Guidance for Cardiopulmonary Bypass Arterial Line Blood Filter 510(k) Submissions; Final Guidance for Industry and FDA" issued on November 29, 2000; ISO 15675 "Cardiovascular implants and artificial organs - Cardiopulmonary bypass systems - Arterial blood line filters" and ISO 15674, "Cardiovascular implants and artificial organs - Hard-shell cardiotomy/venous reservoir systems (with/without filter) and soft venous reservoir bags". In vitro testing was carried out to demonstrate both the substantial equivalence with the predicate device and also to comply with safety and effectiveness requirements. Testing supplied in the 510(k) premarket notification includes performance tests and mechanical integrity tests that demonstrate compliance with performance specifications. The tests that were performed are listed in the following summary table. The · INSPIRE 6F passed each test mentioned in the table below. | TEST | TEST<br>CLASSIFICATION | TEST TITLE | |------|------------------------|-------------------------------------------------------| | 1 | Physical/Mechanical | Structural Integrity | | 2 | Physical/Mechanical | Blood, water, gas pathway integrity | | 3 | Physical/Mechanical | Blood pathway integrity | | 4 | Functional/Performance | Blood volume capacity | | 5 | Functional/Performance | Blood rest volume | | 6 | Functional/Performance | Gas transfer performance/blood side<br>pressure drop | | 7 | Functional/Performance | Heat exchange performance/water side<br>pressure drop | | 8 | Functional/Performance | Air handling capability | | 9 | Functional/Performance | Air handling | | 10 | Functional/Performance | Break-through time and volume | | 11 | Functional/Performance | Defoaming efficiency | | 12 | Functional/Performance | Dynamic priming volume / Hold-up | | 13 | Functional/Performance | Filtration Efficiency | | 14 | Functional/Performance | Filtration Efficiency - venous section | | 15 | Functional/Performance | Filtration Efficiency - cardiotomy section | | 16 | Functional/Performance | Flow rate capacity | | 17 | Functional/Performance | Pressure drop | | 18 | Functional/Performance | Hemolysis | {4}------------------------------------------------ | TEST | TEST<br>CLASSIFICATION | TEST TITLE | |------|------------------------|-------------------------------| | 19 | Functional/Performance | Reservoir Hemolysis | | 20 | Functional/Performance | Blood compatibility | | 21 | Functional/Performance | Reservoir Blood compatibility | | 22 | Functional/Performance | Leaching of coating | | 23 | Functional/Performance | Reservoir Leaching of coating | | 24 | Functional/Performance | Flaking of coating | | 25 | Functional/Performance | Reservoir Flaking of coating | | 26 | Functional/Performance | Uniformity of coating | #### CONCLUSIONS: The results of in vitro studies demonstrate that the INSPIRE 6F performs in a manner substantially equivalent to the D905 EOS predicate device with respect to the relevant functional parameters. Also, the INSPIRE 6F performs in a manner substantially equivalent to the D733 predicate device, with respect to the filtering and air handling performances. Test results of this study suggest the INSPIRE 6F is equivalent to the predicate devices with respect to device function. Additional testing has also demonstrated the effectiveness of production techniques to assure that the device is sterile and non-pyrogenic. {5}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 MAY 1 1 2012 Sorin Group Italia s.r.l. c/o Mr. Barry Sall Principal Consultant Parexel International Consulting 195 West Street Waltham, WA 02451 #### Re: K120185 Trade/Device Name: Inspire 6F Hollow Fiber Oxygenator with Integrated Arterial Filter and Hardshell Venous Cardiotomy/Reservoir Regulation Number: 21 CFR 870.4350 · Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II Product Code: DTZ Dated: May 8, 2012 Received: May 9, 2012 Dear Mr. Sall: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). Tournete Frev (100) 2007 the device, subject to the general controls provisions of the Act. The I va may, allers, which of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {6}------------------------------------------------ Page 2 - Mr. Barry Sall Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must of any I outlar state and the equirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you desire specific advisor your your your contrices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please the Comer Devices and National by reference to premarket notification" (21CFR Part note the regulation onlined, The reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the rou may of amall Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address (000) 050 x 6da.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ # Indications for Use #### 510(k) Number (if known): K120185 INSPIRE 6F Hollow Fiber Oxygenator with integrated Arterial Filter Device Name: and Hardshell Venous/Cardiotomy Reservoir: #### Indications for Use: INSPIRE 6F M: Hollow Fiber Oxygenator INSPIRE 6F M is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It provides gas exchange support and blood temperature control. INSPIRE 6F M integrated arterial filter provides additional protection against air and solid emboli. INSPIRE 6F M is intended to be used for 6 hours or less. ### INSPIRE HVR: Hardshell Venous/Cardiotomy Reservoir INSPIRE HVR is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It collects, defoams and filters venous blood and requiring blood. INSPIRE HVR can be used post-operatively for chest drainage. INSPIRE HVR is intended to be used for 6 hours or less. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ## Concurrence of CDRH, Office of Device Evaluation (ODE) Division Sign-Off
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...