Corsano CardioWatch 287-2 System

K232548 · Corsano Health B.V. · MSX · Mar 22, 2024 · Cardiovascular

Device Facts

Record IDK232548
Device NameCorsano CardioWatch 287-2 System
ApplicantCorsano Health B.V.
Product CodeMSX · Cardiovascular
Decision DateMar 22, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2300
Device ClassClass 2

Indications for Use

The CardioWatch 287-2 System is intended for reusable bedside, mobile and central multi-parameter, physiologic patient monitoring of adult patients in professional healthcare facilities, such as hospitals or skilled nursing facilities, or their own home. It is intended for monitoring of non-acutely ill patients by trained healthcare professionals. The CardioWatch 287-2 System is intended to provide visual and audible physiologic multi-parameter alarms. The CardioWatch 287-2 System is intended for monitoring of skin temperature at wrist of the patient or axillary temperature with connected thermometer device. The CardioWatch 287-2 System is intended for continuous monitoring of the following physiological indices in adults (over 22years old): • Pulse rate • Oxygen saturation • Temperature • Movement The CardioWatch 287-2 System is intended for intermittent monitoring with the CardioWatch Bracelet of the following physiological indices in adults (over 22years old): • Respiration rate. The CardioWatch 287-2 System is intended for intermittent or spot-check monitoring, in adults, of: • Non-invasive blood pressure • Lung function & spirometry • Weight The CardioWatch 287-2 System is not intended for use in high-acuity environments, such as ICU or operating rooms. The CardioWatch 287-2 System is not intended for use on acutely ill cardiac patients with the potential to develop life threatening arrhythmias e.g. very fast atrial fibrillation. For these patients, they should be monitored using a device with continuous ECG. The CardioWatch 287-2 system is not a substitute for an ECG monitor. The CardioWatch 287-2 System is not intended for SpO2 monitoring in conditions of high motion or low perfusion.

Device Story

The Corsano CardioWatch 287-2 System is a remote patient monitoring system comprising a wrist-worn bracelet, a web-based platform, and a mobile application. The bracelet uses PPG sensors (green, red, and IR LEDs), an accelerometer, and a thermopile to continuously monitor pulse rate, SpO2, skin temperature, and movement, and intermittently monitor respiration rate. It integrates with third-party devices for spot-check blood pressure, spirometry, and weight. Data is encrypted and transmitted via the patient app to a secure server. Healthcare professionals (HCPs) access data via the mobile app or web browser, which provide visual and audible alarms for out-of-limit parameters. The system supports clinical decision-making by enabling remote monitoring of non-acutely ill patients, allowing for timely intervention. It benefits patients by facilitating continuous monitoring outside of high-acuity settings, reducing the need for constant in-person observation while maintaining oversight by trained professionals.

Clinical Evidence

No clinical studies were conducted to prove substantial equivalence. Clinical testing was performed to support compliance with ISO 80601-2-61 and FDA pulse oximeter guidance. Respiratory rate validation was conducted against a gold standard (blinded clinician, manually counted end-tidal CO2 with FDA-cleared capnogram) to ensure accuracy in the intended population.

Technological Characteristics

Wearable wrist-worn device using PPG (525nm, 660nm, 880nm LEDs), accelerometer, and thermopile. Powered by rechargeable Lithium-Ion battery. Wireless communication. Non-sterile. Software developed per IEC 62304. Complies with IEC 60601-1, IEC 60601-1-2, ISO 80601-2-61, ISO 80601-2-56, and ISO 10993-1.

Indications for Use

Indicated for non-acutely ill adult patients (22+ years) in professional healthcare facilities or home settings for multi-parameter physiologic monitoring. Contraindicated for high-acuity environments (ICU/OR), acutely ill cardiac patients prone to life-threatening arrhythmias, and SpO2 monitoring during high motion or low perfusion.

Regulatory Classification

Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. March 22, 2024 Corsano Health B.V. Peter Stas CEO - Head of Firm Wilhelmina van Pruisenweg 35 2595AN 's-Gravenhage The Netherlands Re: K232548 Trade/Device Name: Corsano CardioWatch 287-2 System Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II Product Code: MSX, DQA, BZG, BZQ, FLL, DRG, DXN, FRI Dated: February 16, 2024 Received: February 20, 2024 Dear Peter Stas: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. ### Jennifer W. Shih -S Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ Enclosure {3}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K232548 Device Name Corsano CardioWatch 287-2 System #### Indications for Use (Describe) The CardioWatch 287-2 System is intended for reusable bedside, mobile and central multi-parameter, physiologic patient monitoring of adult patients in professional healthcare facilities, such as hospitals or their own home. It is intended for monitoring of non-acutely ill patients by trained healthcare professionals. The CardioWatch 287-2 System is intended to provide visual and audible physiologic multi-parameter alarms. The CardioWatch 287-2 System is intended for monitoring of skin temperature at wrist of axillary temperature with connected thermometer device. The CardioWatch 287-2 System is intended for continuous monitoring of the following physiological indices in adults (over 22years old): - Pulse rate - · Oxygen saturation - · Temperature - Movement The Cardio Watch 287-2 System is intermittent monitoring with the CardioWatch Bracelet of the following physiological indices in adults (over 22years old): · Respiration rate. The CardioWatch 287-2 System is intended for intermittent or spot-check monitoring, in adults, of: - · Non-invasive blood pressure - · Lung function & spirometry • Weight The CardioWatch 287-2 System is not in high-acuity environments, such as ICU or operating rooms. The CardioWatch 287-2 System is not intended for use on acutely ill cardiac patients with the potential to develop life threatening arrhythmias e.g. very fast atrial fibrillation. For these patients, they should be monitored using a device with continuous ECG. The CardioWatch 287-2 system is not a substitute for an ECG monitor. The CardioWatch 287-2 System is not intended for SpO2 monitoring of high motion or low perfusion. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. {4}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ ### 510(k) Summary #### GENERAL INFORMATION 1 | Submitter and 510(k) Owner | | |----------------------------|-------------------------------------------------------------------------| | Name | Corsano Health B.V. | | Address : | Wilhelmina Van Pruisenweg 35<br>'s-Gravenhage<br>2595 AN<br>Netherlands | | Official Correspondent | Mr Peter STAS, CEO – Head of firm | | Phone | +41793102442 | | Email : | pcstas@corsano.com | | Date of Summary : | 22nd March 2024 | | Device Information | | |------------------------|-----------------------------------------------------------| | Trade/Proprietary Name | Corsano CardioWatch 287-2 System | | Common/Usual Name | Patient Remote Monitoring System | | Medical Specialty | Cardiovascular | | Regulation | 870.2300 | | Class: | II | | Classification Name: | System, Network and Communication, Physiological Monitors | | Product Codes: | MSX, DQA, BZG, FLL, DRG, DXN, FRI | | Primary Product Code | | | | | |------------------------------|-----------------------------------------------------------------|-----------------|-----------------|----------------------| | Classification<br>Regulation | Classification Name | Device<br>Class | Product<br>Code | Classification Panel | | 21 CFR 870.2300 | System, Network and<br>Communication, Physiological<br>Monitors | Class II | MSX | Cardiovascular | | Secondary Product Codes | | | | | |------------------------------|-----------------------------|-----------------|-----------------|-------------------------| | Classification<br>Regulation | Classification Name | Device<br>Class | Product<br>Code | Classification<br>Panel | | 21 CFR 870.2700 | Oximeter | Class II | DQA | Cardiovascular | | 21 CFR 868.1840 | Spirometer | Class II | BZG | Anesthesiology | | 21 CFR 868.2375 | Breathing Frequency Monitor | Class II | BZQ | Anesthesiology | {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the word "Corsano" in a teal color, followed by a teal-colored plus sign inside of a circle. The word is written in a sans-serif font and is the main focus of the image. The plus sign is located to the right of the word and is slightly smaller in size. | Secondary Product Codes | | | | | |------------------------------|---------------------------------------------------------------|-----------------|-----------------|-------------------------| | Classification<br>Regulation | Classification Name | Device<br>Class | Product<br>Code | Classification<br>Panel | | 21 CFR 870.2910 | Clinical Electronic Thermometer | Class II | FLL | General Hospital | | 21 CFR 870.2910 | Radiofrequency physiological<br>signal transmitter & receiver | Class II | DRG | Cardiovascular | | 21 CFR 870.1130 | Non-invasive Blood Pressure<br>Monitor | Class II | DXN | Cardiovascular | | 21 CFR 880.2700 | Stand-on Weight Scale | Class I | FRI | General Hospital | | Predicate Device | | |------------------------|-----------------------------------------------------------| | Trade/Proprietary Name | Current Wearable Health Monitoring System | | Common/Usual Name | Remote Patient Monitor | | Medical Specialty | Cardiovascular | | Regulation | 870.2300 | | Class: | II | | Classification Name: | System, Network and Communication, Physiological Monitors | | Product Codes: | MSX, FLL, DQA, BZQ, DRG, BZG | ### 2 Device Description The Corsano CardioWatch 287-2 System is a Remote-Patient Monitoring System that consists of a monitoring bracelet device worn on the wrist by adult patients (aged 22 years old and over), a web-based browser platform and a user mobile application operable in either Patient Mode or HealthCare Professional (HCP) Mode. Vital signs data both on mobile devices and web-based dashboard are available to the HealthCare Provider only. The Corsano CardioWatch 287-2 System is also integrated with third-party devices for displaying and monitoring physiological signs (spot monitoring of : non-invasive blood pressure (NIBP), lung function & spirometry (SPIRO), weight (WEIGHT) as well as continuous monitoring of axillary temperature (aTEMP). The Corsano Bracelet is intended to continuously monitor physiological vital sign data : Pulse Rate (PR), oxygen saturation (SpO2), skin temperature (sTEMP) and activity (STEPS) and for intermittent monitoring of respiratory rate (RR) from the person being monitored and securely transmit the encrypted data via the Patient User App to the secure server. The bracelet is intended for use in professional healthcare facilities, such as hospitals or skilled nursing facilities, or the home by trained healthcare professionals. {7}------------------------------------------------ The healthcare professional can securely access the patient physiological signs remotely via the Corsano mobile application HCP Mode or via a browser web-interface which are also intended to provide visual and audible physiologic multi-parameter alarms. #### 3 Intended Use / Indications for Use The CardioWatch 287-2 System is intended for reusable bedside, mobile and central multi-parameter, physiologic patient monitoring of adult patients in professional healthcare facilities, such as hospitals or skilled nursing facilities, or their own home. It is intended for monitoring of non-acutely ill patients by trained healthcare professionals. The CardioWatch 287-2 System is intended to provide visual and audible physiologic multi-parameter alarms. The CardioWatch 287-2 System is intended for monitoring of skin temperature at wrist of the patient or axillary temperature with connected thermometer device. The CardioWatch 287-2 System is intended for continuous monitoring of the following physiological indices in adults (over 22years old): - · Pulse rate - Oxygen saturation - · Temperature - Movement The CardioWatch 287-2 System is intended for intermittent monitoring with the CardioWatch Bracelet of the following physiological indices in adults (over 22years old): - Respiration rate. ● The CardioWatch 287-2 System is intended for intermittent or spot-check monitoring, in adults, of: - · Non-invasive blood pressure - · Lung function & spirometry - Weight The CardioWatch 287-2 System is not intended for use in high-acuity environments, such as ICU or operating rooms. The CardioWatch 287-2 System is not intended for use on acutely ill cardiac patients with the potential to develop life threatening arrhythmias e.g. very fast atrial fibrillation. For these patients, they should be monitored using a device with continuous ECG. The CardioWatch 287-2 system is not a substitute for an ECG monitor. The CardioWatch 287-2 System is not intended for SpO2 monitoring in conditions of high motion or low perfusion. {8}------------------------------------------------ #### Substantial Equivalence ব The Corsano CardioWatch 287-2 System has an intended use statement identical to the predicate device with the exception of several word changes specific to the CardioWatch 287-2 System, both the devices also share same conditions of use in terms measurement technology, measurements, transmission, data visualization and alarms whilst using third party devices, as well as in the same healthcare environments and the same intended patients. #### Technological characteristics Corsano CardioWatch 287-2 System share the same technological characteristics as the predicate device, providing patient physiological parameters measurements using PPG technology on wearable sensors, transmit and provide data visualization and alarms (visual & audible) to HCP's. Both systems provide the possibility of using validated third-party devices that provide additional physiological parameters & data visualization to HCP's. | Features | Proposed Device:<br>CardioWatch 287-2 System | Predicate Device:<br>Current Wearable Health<br>Monitoring System | Comment | |-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device<br>Manufacturer | Corsano Health B.V. | Current Health Ltd | - | | Device<br>Classification | II | II | Identical | | 510(k)<br>Number | N/A | K210133 | - | | Primary<br>Product Code | MSX | MSX | Identical | | Secondary<br>Product Code | DQA | DQA | Identical | | | BZQ | BZQ | Identical | | | FLL | FLL | Identical (Clinical Electronic<br>Thermometer) both for wearable<br>sensor & third-party device | | | DRG | DRG | Identical (Radiofrequency<br>physiological signal transmitter &<br>receiver) | | | BZG | BZG | Identical (Spirometer - third-party<br>device) | | | DXN | - | Both devices use the NIPB (third-<br>party device), however the<br>Predicate device has not<br>included the Product Code. | | Features | Proposed Device:<br>CardioWatch 287-2 System | Predicate Device:<br>Current Wearable Health<br>Monitoring System | Comment | | | FRI | - | Both devices use the Scales<br>(third-party device), however the<br>Predicate device has not<br>included the Product Code. | | Device<br>Classification<br>Name | System, Network and<br>Communication,<br>Physiological Monitors | System, Network and<br>Communication,<br>Physiological Monitors | Identical | | Intended Use/<br>Indication for Use | The CardioWatch 287-2<br>System is intended for<br>reusable bedside, mobile<br>and central multi-<br>parameter, physiologic<br>patient monitoring of adult<br>patients in professional<br>healthcare facilities, such<br>as hospitals or skilled<br>nursing facilities, or their<br>own home. It is intended for<br>monitoring of non-acutely ill<br>patients by trained<br>healthcare professionals.<br>The CardioWatch 287-2<br>System is intended to<br>provide visual and audible<br>physiologic multi-parameter<br>alarms.<br>The CardioWatch 287-2<br>System is intended for<br>monitoring of skin<br>temperature at wrist of the<br>patient or axillary<br>temperature with<br>connected thermometer<br>device.<br>The CardioWatch 287-2<br>System is intended for<br>continuous monitoring of<br>the following physiological<br>indices in adults (over<br>22years old):<br>• Pulse rate<br>• Oxygen saturation<br>• Temperature | The Current Wearable<br>Health Monitoring System<br>is intended for reusable<br>bedside, mobile and central<br>multi-parameter,<br>physiologic patient<br>monitoring of adult patients<br>in professional healthcare<br>facilities, such as hospitals<br>or skilled nursing facilities, or<br>their own home. It is<br>intended for monitoring of<br>patients by trained<br>healthcare professionals.<br>The Current Wearable<br>Health Monitoring System<br>is intended to provide<br>visual and audible<br>physiologic multi-parameter<br>alarms.<br>The Current Wearable<br>Health Monitoring System<br>is intended for temperature<br>monitoring where<br>monitoring temperature at<br>the upper arm is clinically<br>indicated.<br>The Current Wearable<br>Health Monitoring System<br>is intended for continuous<br>monitoring of the following<br>parameters in adults:<br>• Pulse rate<br>• Oxygen saturation<br>• Temperature<br>• Movement | Identical both devices can be<br>considered similar in terms of<br>intended use (measurements,<br>transmission, data visualization,<br>alarms and third party devices) in<br>the same healthcare<br>environments and the same<br>intended patients. | | Features | Proposed Device:<br>CardioWatch 287-2 System | Predicate Device:<br>Current Wearable Health<br>Monitoring System | Comment | | | The CardioWatch 287-2<br>System is intended for<br>intermittent monitoring with<br>the CardioWatch Bracelet<br>of the following<br>physiological indices in<br>adults (over 22years old):<br>• Respiration rate. | The Current Wearable<br>Health Monitoring System<br>is intended for intermittent<br>or spot-check monitoring, in<br>adults, of:<br>• Respiration rate<br>• Non-invasive blood<br>pressure<br>• Lung function &<br>spirometry<br>• Weight | | | | The CardioWatch 287-2<br>System is intended for<br>intermittent or spot-check<br>monitoring, in adults, of:<br>• Non-invasive blood<br>pressure<br>• Lung function &<br>spirometry<br>• Weight | The Current Wearable<br>Health Monitoring System<br>is not intended for use in<br>high-acuity environments,<br>such as ICU or operating<br>rooms. | | | | The CardioWatch 287-2<br>System is not intended for<br>use in high-acuity<br>environments, such as ICU<br>or operating rooms. | The Current Wearable<br>Health Monitoring System<br>is not intended for use on<br>acutely ill cardiac patients<br>with the potential to<br>develop life threatening<br>arrhythmias e.g. very fast<br>atrial fibrillation. For these<br>patients, they should be<br>monitored using a device<br>with continuous ECG. The<br>Current Wearable Health<br>Monitoring System is not a<br>substitute for an ECG<br>monitor. | | | | The CardioWatch 287-2<br>System is not intended for<br>use on acutely ill cardiac<br>patients with the potential<br>to develop life threatening<br>arrhythmias e.g. very fast<br>atrial fibrillation. For these<br>patients, they should be<br>monitored using a device<br>with continuous ECG. The<br>CardioWatch 287-2 system<br>is not a substitute for an<br>ECG monitor. | The Current Wearable<br>Health Monitoring System<br>is not intended for SpO2<br>monitoring in conditions of<br>high motion or low<br>perfusion. | | | | The CardioWatch 287-2<br>System is not intended for<br>SpO2 monitoring in<br>conditions of high motion or<br>low perfusion. | | | | Features | Proposed Device: CardioWatch 287-2 System | Predicate Device: Current Wearable Health Monitoring System | Comment | | Target Population | Adult (22yrs & older) | Adult (18yrs & older) | Similar ; both devices are intended for use on adults | | Patient target category | Non acutely ill patients | Non acutely ill patients | Identical | | Wearing location | Wrist | Upper arm | The difference in wearing location on the body does not raise new questions of safety or efficacy.<br>Both locations are valid clinical locations allowing for monitoring peripheral arteries. | | Duration of Use | Upto 24 hours per day as required by the practician | Upto 24 hours per day as required by the practician | Identical | | Sterile | Non-sterile | No-sterile | Identical | | Mechanism of Action / Principle of Operation | The Corsano CardioWatch 287-2 System continuously monitors Pulse Rate (PR), oxygen saturation (SpO2), temperature (sTEMP) and Activity (STEPS) and intermittently the respiration rate (RESP) through sensors in the wrist bracelet.<br><br>The Corsano System monitors skin temperature (sTEMP) at wrist or axillary temperature (aTEMP) with connected thermometer.<br><br>The device has wireless communication capabilities.<br>Intermittent or spot-checking monitoring of blood pressure (BP), spirometry & lung function (SPIRO) and weight (WEIGHT) through third-party product integration. | The Current Health System continuously monitors pulse rate (PR), oxygen saturation (SpO2), skin temperature (TEMP) and movement (MOVEMENT) through sensors in a device designed to be worn around the upper ar…
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...