K193296 · Ecordum, Inc. · DXH · Oct 30, 2020 · Cardiovascular
Device Facts
Record ID
K193296
Device Name
eCordum Cardiac Monitor (eCordum CM)
Applicant
Ecordum, Inc.
Product Code
DXH · Cardiovascular
Decision Date
Oct 30, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2920
Device Class
Class 2
Indications for Use
The eCordum™ Cardiac Monitor is intended to record, transfer and store single-channel electrocardiogram (ECG) data via Bluetooth transmission to Bluetooth enabled devices. The monitor is intended for use by healthcare professionals, patients with known or suspected heart conditions and health conscious individuals. The ECG data is intended to supplement other patient data and is not intended for automated analysis. The device has not is not intended for pediatric use.
Device Story
Device records single-channel ECG data via chest-attached disposable OTS electrodes; powered by replaceable battery; features interchangeable covers. Collects and encrypts ECG data for unidirectional Bluetooth transmission to Bluetooth-enabled devices. Used in ambulatory settings by healthcare professionals, patients, or health-conscious individuals. No automated analysis or arrhythmia detection performed. Output viewed by healthcare provider to supplement patient data; aids in cardiac monitoring/assessment. Benefits include long-term ECG recording capability for clinical evaluation.
Clinical Evidence
Bench testing only. Device performance validated against ANSI/AAMI/IEC 60601-2-47:2012, ANSI/AAMI ES 60601-1:2005, and IEC 60601-1-2:2014. Biocompatibility assessed per ISO 10993-1:2009. Software verification and validation conducted per FDA guidance.
Technological Characteristics
Single-channel ECG monitor; Bluetooth Class II connectivity; replaceable battery power; uses standard disposable OTS electrodes with conductive gel. Complies with ANSI/AAMI ES 60601-1, IEC 60601-1-2, and IEC 60601-2-47. Biocompatibility per ISO 10993-1. Software-based data encryption for transmission.
Indications for Use
Indicated for healthcare professionals, patients with known or suspected heart conditions, and health-conscious individuals for recording, transferring, and storing single-channel ECG data. Not intended for pediatric use.
Regulatory Classification
Identification
A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.
K063044 — CARDIOBELT ELECTRODE BELT · Monebo Technologies, Inc. · Mar 23, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized image of a human figure, and the text on the right is in blue. The text is arranged in three lines, with "FDA" on the first line, "U.S. FOOD & DRUG" on the second line, and "ADMINISTRATION" on the third line.
October 30, 2020
eCordum, Inc. % Becky Ditty Consultant Biologics Consulting 1555 King St. Suite 300 Alexandria, Virginia 22314
Re: K193296
Trade/Device Name: eCordum Cardiac Monitor (eCordum CM) Regulation Number: 21 CFR 870.2920 Regulation Name: Telephone electrocardiograph transmitter and receiver Regulatory Class: Class II Product Code: DXH Dated: November 26, 2019 Received: November 27, 2019
Dear Becky Ditty:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193296
Device Name eCordum™ Cardiac Monitor (eCordum™ CM)
Indications for Use (Describe)
The eCordum™ Cardiac Monitor is intended to record, transfer and store single-channel electrocardiogram (ECG) data via Bluetooth transmission to Bluetooth enabled devices. The monitor is intended for use by healthcare professionals, patients with known or suspected heart conditions and health conscious individuals. The ECG data is intended to supplement other patient data and is not intended for automated analysis. The device has not is not intended for pediatric use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K193296 510(k) Summary
#### 1. Submission Sponsor
eCordum, Inc. 419 1st Street East Northport, NY 11731 USA Vladislav Bukhman President Email: vlad.bukhman@ecordum.com Tel number: 631.223.6447
#### 2. Submission Correspondent
Biologics Consulting 1555 King Street, Suite 300 Alexandria, VA 22314 USA Becky Ditty Consultant Email: bditty@biologicsconsulting.com Tel number: 269.888.2516
#### 3. Date Prepared
November 26, 2019
## 4. Device Identification
| Trade/Proprietary Name: | eCordum™ Cardiac Monitor (eCordum™ CM) |
|----------------------------|-----------------------------------------------------------|
| Common/Usual Name: | Transmitters and receivers, electrocardiograph, telephone |
| Regulation Name: | Telephone electrocardiograph transmitter and receiver |
| Classification Regulation: | 21 CFR§870.2920 |
| Product Codes: | DXH |
| Device Class: | II |
| Classification Panel: | Cardiovascular |
#### 5. Legally Marketed Predicate Device(s)
K121197 BodyGuardian System manufactured by Preventice Solutions
#### 6. Device Description
The eCordum™ Cardiac Monitor (eCordum™ CM) is intended to record, transfer and store single-channel electrocardiogram (ECG) data via Bluetooth transmission to Bluetooth enabled devices. The monitor is intended for use by healthcare professionals, patients with known or suspected heart conditions and health conscious individuals. The device has not been tested and it is not intended for pediatric use.
{4}------------------------------------------------
The eCordum™ CM device is powered by a replaceable battery and has two interchangeable attachable covers with respective ECG electrodes for long-term ECG recording. The Wearable Unit for long-term recording is used with two standard disposable Off-The-Shelf (OTS) ECG electrodes. For long term recording ECG data is obtained by attaching the device to user's chest using OTS ECG electrodes. Long-term recordings provide a single channel ECG signal. During recording the device collects and encrypts ECG data in the storage media for unidirectional transmission to a Bluetooth enabled device.
## 7. Indication for Use Statement
The eCordum™ Cardiac Monitor is intended to record, transfer and store single-channel electrocardiogram (ECG) data via Bluetooth transmission to Bluetooth enabled devices. The monitor is intended for use by healthcare professionals, patients with known or suspected heart conditions and health conscious individuals. The ECG data is intended to supplement other patient data and is not intended for automated analysis. The device has not been tested and it is not intended for pediatric use.
#### 8. Substantial Equivalence Discussion
#### Comparison of Indications
Table 1 compares the indications for use of the subject device with the indications for use of the two predicate devices.
| eCordum CM<br>(K193296) | Preventice<br>(K121197) |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The eCordum™ Cardiac Monitor is<br>intended to record, transfer and<br>store single-channel<br>electrocardiogram (ECG) data via<br>Bluetooth transmission to Bluetooth<br>enabled devices. The monitor is<br>intended for use by healthcare<br>professionals, patients with known<br>or suspected heart conditions and<br>health conscious individuals. The<br>ECG data is intended to supplement<br>other patient data and is not<br>intended for automated analysis.<br>The device has not been tested and it<br>is not intended for pediatric use. | The BodyGuardian System detects and<br>monitors non-lethal cardiac<br>arrhythmias in ambulatory patients,<br>when prescribed by a physician or<br>other qualified healthcare professional.<br>The BodyGuardian System<br>continuously records, stores and<br>periodically transmits the following<br>physiological data to a remote<br>computer server for up to 30 days at a<br>time:<br>• ECG<br>• Heart rate (including HR variability<br>and HR reliability)<br>• Respiration rate<br>• Activity |
Comparison of Indications for Use Table 1:
{5}------------------------------------------------
The subject device and the predicate devices have tundamental intended use, namely to record, store and transfer ECGs. The subject eCordum device does NOT perform any arrhythmia detection, and does not detect heart rate, respiration rate nor activity.
In regard to the eCordum wearable, the subject device and the predicate Preventice device are both intended for prescription use only.
These differences do not raise different questions of safety and effectiveness, and thus do not constitute a new intended use.
# Comparison of Technological Comparison
Table 2 provides a detailed comparison between the subject device and the predicate devices.
| | Proposed eCordum CM Device | Predicate Preventice Wearable<br>ECG Monitor |
|------------------------------|--------------------------------------------------------------------------|--------------------------------------------------------------------------|
| 510(k) Number | K193296 | K121197 |
| Applicant | eCordum, Inc. | Preventice |
| Device Name | eCordum™ CM | Preventice BodyGuardian Device |
| Classification<br>Regulation | 21 CFR§870.2920 | 21 CFR§870.1025 |
| Product Code | DXH | DSI |
| Rx or OTC | RX for wearable option | RX |
| Intended Use | Ambulatory | Ambulatory |
| Alarm | No | No |
| Adult/Pediatric | Adult | Adult |
| Electrodes | Attachable standard<br>ambulatory OTS electrodes<br>with conductive gel. | Attachable standard ambulatory<br>OTS electrodes with conductive<br>gel. |
#### Table 2: Device Comparison Table
{6}------------------------------------------------
| | Proposed eCordum CM Device | Predicate Preventice Wearable<br>ECG Monitor |
|---------------------------|------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Single<br>Use/Reusable | Disposable ambulatory OTS<br>electrodes. | Disposable ambulatory OTS<br>electrodes. |
| ECG and Events<br>Storage | Transmit ECG data to a<br>Bluetooth enabled device. | Stores and periodically transmits<br>events and ECG data to remote<br>server via a mobile device when<br>internet is available. |
| Real time ECG<br>View | No | No |
| Arrhythmia<br>Detection | No | Yes |
| Sampling Rate | 250 Hz | 250 Hz |
| Application/Wear | Chest for ambulatory OTS<br>electrodes. | Chest for ambulatory OTS<br>electrodes. |
| User Interface | Automatic turn on upon skin<br>contact<br>Mobile app | Same |
| Connection | Class II Bluetooth | Class II Bluetooth |
| Power Supply | Replaceable battery | Single use rechargeable battery |
The eCordum™ CM has a same intended use and similar technological characteristics performance as the predicate devices. The subject device has been tested and shown to comply with IEC 60601-2-47. The results of firmware testing showed that eCordum CM device meet all requirements and specifications as described in the Software Validation and Verification protocol and reports.
# 9. Performance Data
The eCordum™ Cardiac Monitor meets all the requirements for overall design, biocompatibility, electrical, EMC safety and cybersecurity protection. The eCordum™ Cardiac Monitor passed all testing and supports the claims of substantial equivalence and safe operation.
{7}------------------------------------------------
## Biocompatibility
A biocompatibility assessment for patient contacting materials was performed in accordance with ISO 10993-1:2009. All biocompatibility testing passed.
#### Software Documentation
Software verification and validation testing was conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". Verification of the eCordum™ Cardiac Monitor was conducted to ensure that the product works as designed. Validation was conducted to check the design and performance of the product.
## EMC and Electrical Safety
The eCordum™ CM was assessed for conformity with the relevant requirements of the following standards and found to comply:
- ANSI/AAMI ES 60601-1:2005/(R) 2012 and A1:2012, C1:2009/(R) 2012 and A2:2010/(R) 2012
- IEC 60601-1-2:2014 (Fourth Edition)
# Bench Testing
Bench testing was done to support device performance and substantial equivalence. Testing demonstrating that the eCordum™ Cardiac Monitor complies with ANSI/AAMI/IEC 60601-2-47:2012 (60601-2-47): Medical electrical equipment – Part 2-47. The eCordum™ Cardiac Monitor met all the requirements for overall design confirming that design output meets design inputs and specifications.
#### 10. Statement of Substantial Equivalence
The eCordum™ Cardiac Monitor is substantially equivalent to predicate device. The eCordum™ Cardiac Monitor has the same intended use and the same technological characteristics as the previously cleared predicate device, BodyGuardian monitor by Preventice. The new device (eCordum™ Cardiac Monitor) does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.