The clue medical and clue medical BASIC are handheld, battery operated single channel electrocardiographs intended for recording and transmitting ECG data by patients who are concerned about their heart rhythm. The clue medical and clue medical BASIC allow the patient to record their ECG data into the device memory and transmit it for display by healthcare professionals. Specifically, the clue medical and clue medical BASIC are intended for patients that are concerned about their heart rhythm or have experienced the following symptoms that are suggestive of abnormal heart rhythm: Skipped beats Pounding heart (palpitations) History of arrhythmia The clue medical and cluc medical BASIC ECG data is intended to be used by a licensed health care practitioner. These measurements are not intended for any specific clinical diagnosis. The clinical significance must be determined by the physician.
Device Story
Handheld, battery-operated single-channel ECG recorder; features four fixed electrodes on device back; records up to ten ECG segments (30 seconds or 2 minutes). Patient-operated; requires holding device to chest. Transforms ECG signals into PDF files. Transmits data via IrDA or Bluetooth to local PC/printer; clue medical model also supports transmission via Bluetooth-enabled cell phone to Telovital server. Healthcare providers receive and display PDF files to assess heart rhythm; device does not provide automated diagnosis. Benefits patients by enabling remote monitoring of symptomatic heart rhythm concerns.
Clinical Evidence
Bench testing only. The device was tested for compliance with IEC 60601-1 (safety), IEC 60601-1-1 (systems), IEC 60601-1-2 (EMC), IEC 60601-2-25 (ECG safety), and IEC 60601-2-47 (ambulatory ECG performance). No clinical data was required or provided.
Technological Characteristics
Handheld single-channel ECG recorder; battery-operated. Features four fixed electrodes. Connectivity: IrDA and Bluetooth. Data output: PDF files. Standards: IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-2-25, IEC 60601-2-47. No software algorithm class specified.
Indications for Use
Indicated for patients concerned about heart rhythm or experiencing symptoms suggestive of arrhythmia, including skipped beats, palpitations, or history of arrhythmia. Intended for use by patients capable of operating the device and by licensed healthcare practitioners for review.
Regulatory Classification
Identification
A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.
K082013 — CLUE MEDICAL AND CLUE MEDICAL BASIC · Telovital GmbH Telemedizin · Nov 10, 2008
K050670 — SENSOR MOBILE SM 100 VITAPHONE 1001R REMOS · Tms Telemedizinische Systeme GmbH · Mar 30, 2005
K103077 — ECG PEN · Meridian Dynamics Biomedical, Inc. · Jul 13, 2011
K042814 — READMYHEART · Dailycare Biomedical, Inc. · Jan 6, 2005
K111438 — REKA E100 · Reka Pte, Ltd. · Sep 12, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
K0f3630
# Section 6: 510(k) Summary
:
## APR 2 3 2009
### 510(k) Summary
The following information is provided as required by 21 CFR § 807.87 for Telovital's 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the safety and effectiveness information upon which the substantial equivalence determination is based.
The safety and effectiveness of the clue medical and clue medical BASIC is based upon a determination of the substantial equivalence as well as the safety and effectiveness of its predicate device(s).
| Sponsor / Manufacturer: | Telovital GmbH Telemedizin<br>Office Park I / Top 4<br>A-1300 - Vienna - Airport<br>Austria | | | | |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|----------|--------------------|----------|
| Contact: | Cherita James<br>M Squared Associates<br>901 King St. Suite 200<br>Alexandria, VA 22314<br>Ph: 703-562-9800 x257<br>Fax: 703-562-9797<br>Email: CJames@MSquaredAssociates.com | | | | |
| Date Prepared: | December 8, 2008 | | | | |
| Proposed Class: | II | | | | |
| Proprietary Name: | clue medical and clue medical BASIC | | | | |
| Common Name: | Electrocardiograph | | | | |
| Classification Name: | Electrocardiograph | | | | |
| Regulation Number: | 21 CFR 870.2340, 21 CFR 870.2920 | | | | |
| Product Codes: | DPS, DXH | | | | |
| Predicate Device: | | | | | |
| | Device Name | | Procode | Common Name | Class |
| K082013 | clue medical and clue<br>medical BASIC | | DPS, DXH | Electrocardiograph | Class II |
{1}------------------------------------------------
#### Indication for Use
The clue medical and clue medical BASIC are handheld, battery operated single channel electrocardiographs intended for recording and transmitting ECG data by patients who are concerned about their heart rhythm. The clue medical BASIC allow the patient to record their ECG data into the device memory and transmit it for display by healthcare professionals.
Specifically, the clue medical and clue medical BASIC are intended for patients that are concerned about their heart rhythm or have experienced the following symptoms that are suggestive of abnormal heart rhythm:
Skipped beats
Pounding heart (palpitations)
History of arrhythmia
The clue medical and cluc medical BASIC ECG data is intended to be used by a licensed health care practitioner. These measurements are not intended for any specific clinical diagnosis. The clinical significance must be determined by the physician.
#### Device Description
The clue medical is an handheld battery operated single-channel ECG recorder featuring four fixed electrodes on the back of the device, an ECG lead connector, an IrDA and a Bluetooth interface and one single button to record and store up to ten ECG recordings with 30 seconds or two minutes duration. The stored ECG tests are processed into a printable Portable Document Format (PDF) that can be transmitted two ways. The PDF can be transmitted via IrDA or Bluetooth to an IrDa or Bluetooth capable local PC or printer. If the receiving device is not IrDa or Bluetooth capable, an optional IrDA USB stick is provided with the clue medical. The PDF file can then be saved, sent, displayed or printed. Alternatively, the PDF can be transmitted via a Bluetooth capable cell phone directly to the Telovital server for storage.
The device requires that a user is mentally and physically capable of reading and understanding the Patient Operating Instructions. The patient must also be physically capable of pressing the record button and holding the device to his/her chest for an adequate reading.
The clue medical BASIC is identical to the clue medical in design. The only difference is that the clue medical BASIC does not provide the added capability of sending the ECG data via cell
{2}------------------------------------------------
phone directly to the Telovital server, it is only capable of transmission to a local PC or printer. The purpose of this Special 510(k) is to modify the method of wireless data transmission for both the clue medical and clue medical BASIC to include Bluetooth technology.
#### Performance Testing
The clue medical and clue medical BASIC were tested to meet the following standards for its previous clearance in K082013:
- . IEC 60601-1, "Medical Electrical Equipment. General Requirements for Safety"
- IEC 60601-1-1, "Safety Requirements for Medical Electrical Systems" .
- . IEC 60601-1-2, " Part 1-2: General requirements for safety -Collateral standard: Electromagnetic compatibility -Requirements and tests "
- . IEC 60601-2-25, "Medical Electrical Equipment Part 2: Particular Requirements for the Safety of Electrocardiographs"
- . IEC 60601-2-47, "Medical Electrical Equipment, Particular requirements for safety, including performance, for ambulatory electrocardiographic systems"
Electrode leadwires were found to be compliant to IEC 60601-1, subclause 56.3(c).
For the purpose of this device modification the device was again tested to meet:
- . IEC 60601-1, "Medical Electrical Equipment, General Requirements for Safety
- . IEC 60601-1-2, " Part 1-2: General requirements for safety -Collateral standard: Electromagnetic compatibility -Requirements and tests "
- . IEC 60601-2-25, "Medical Electrical Equipment Part 2: Particular Requirements for the Safety of Electrocardiographs"
- . IEC 60601-2-47, "Medical Electrical Equipment, Particular requirements for safety, including performance, for ambulatory electrocardiographic systems"
#### Summary of Technological Characteristics:
The only modification to the clue medical and clue medical BASIC since its previous clearance in K082013 is to enhance the method of wireless data transmission.
#### Substantial Equivalence Discussion:
The change to the method of wireless data transfer to include Bluetooth does not change the intended use or performance characteristics of the clue medical and clue medical BASIC as
{3}------------------------------------------------
previously cleared in K082013.
#### Conclusion:
The modified clue medical and clue medical BASIC have the following similarities to the devices previously cleared in K082013:
4/4
- has the exact same indication for use, .
- uses the same operating principle, .
- incorporates the same basic device design, ●
- incorporates the same materials. .
Therefore the modification to the clue medical and clue medical BASIC can be found substantially equivalent to the same devices as cleared in K082013.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and tail feathers. The eagle faces to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### APR 2 8 2009
Telovital GmbH Telemedizin c/o Ms. Cherita James M Squared Associates 901 King St., Suite 200 Alexandria, VA 22314
Re: K083630
> Trade/Device Name: clue medical and clue medical BASIC Regulation Number: 21 CFR 870.2920 Regulation Name: Telephone Electrocardiograph Transmitter and Receiver Regulatory Class: Class II (Two) Product Code: DXH and DPS Dated: March 26, 2009 Received: March 27, 2009
Dear Ms. James:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Ms. Cherita James
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Section 5: Indications for Use Statement
## Indications for Use
510(k) Number:
KO 83630
Device Name: clue medical and clue medical BASIC
Indications for Use:
The clue medical and clue medical BASIC are handheld, battery operated single channel electrocardiographs intended for recording and transmitting ECG data by patients who are concerned about their heart rhythm. The clue medical and clue medical BASIC allow the patient to record their ECG data into the device memory and transmit it for display by healthcare professionals.
Specifically, the clue medical and clue medical BASIC are intended for patients that are concerned about their heart rhythm or have experienced the following symptoms that are suggestive of abnormal heart rhythm: Skipped beats
Pounding heart (palpitations) History of arrhythmia
The clue medical and clue medical BASIC ECG data is intended to be used by a licensed health care practitioner. These measurements are not intended for any specific clinical diagnosis. The clinical significance must be determined by the physician.
Prescription Use
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|-------------------------|
| | AtthBte for B Zuckerman |
1122/09 (Division Sign-Off) Division of Cardiovascular Devices
Page 1 of 1
510(k) Number K082630
Telovital GmbH December 2008
Page 15 of 33
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.