K103077 · Meridian Dynamics Biomedical, Inc. · DPS · Jul 13, 2011 · Cardiovascular
Device Facts
Record ID
K103077
Device Name
ECG PEN
Applicant
Meridian Dynamics Biomedical, Inc.
Product Code
DPS · Cardiovascular
Decision Date
Jul 13, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2340
Device Class
Class 2
Indications for Use
The Vion.ECG Pen (Model no.: 800E) is intended for recording ECG data and displaying general information by adult patients who are concerned their hearter hythm and waveform. This Vion.ECG Pen allows the patients to record transient cardiac events voluntarily for providing to the healthcare professionals as references during office visits. In addition, the patient can also record their ECG data and transmit the recorded data to personal computer. The Vion.ECG Pen is not intended for recording and transmission of user's ECG signal simultaneously. Users with implanted pacemakers are not recommended to use this device. This device is not intended to substitute a conventional diagnostic tool.
Device Story
Small, portable, leadless electrocardiograph; records 30-second ECG measurements; stores up to 99 readings with date/time stamps. Operated by patient to capture transient cardiac events; data transferred via USB to personal computer for physician review during office visits. Assists healthcare professionals in monitoring patient heart rhythm; not a substitute for conventional diagnostic tools. Benefits include non-invasive, convenient documentation of cardiac symptoms.
Clinical Evidence
Clinical investigation compared Vion.800E to 12-lead standard ECG (PHILIPS PageWriter Trim II) in 206 patients. Evaluated performance metrics including heart rate, QRS interval, QT interval, PR interval, R-wave, S-wave, and T-wave. Results indicated performance is comparable to 12-lead ECG for recording and displaying general ECG information.
Technological Characteristics
Portable, leadless ECG recorder; USB interface for data transfer; internal memory for 99 measurements. Conforms to IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-2-25, and IEC 60601-2-47 (performance).
Indications for Use
Indicated for adult patients concerned about heart rhythm and waveform to record transient cardiac events for physician review. Contraindicated for users with implanted pacemakers.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
KI 03077
## 510(K) SUMMARY
JUL 1 3 2011
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR §807.92.
1.0 Submitter's Name: Meridian Dynamics Biomedical, Inc. (MD Biomedical, Inc.) 3E No 100.Sec 3. Mingsheng E. Rd., Taipei, Taiwan, 105 Address:
| Address: | 3F,NO 100,Sec 3, Mingsheng E. Rd., Taipei, Taiwan, 105 |
|----------|--------------------------------------------------------|
| Tel: | +886-2-27183000 Ext.103 |
| Fax: | +886-2-27180123 |
| Contact: | Carl Yu/Vice President |
| E-Mail: | carl@mdbiomedical.com |
- 2.0 Device Name Vion® ECG Pen(Model no .: 800E) Trade Name: Handheld ECG monitor Common Name: Electrocardiograph Classification name:
- 3.0 Classification: Class II DPS Product Code: 21CFR870.2340 Regulation No.:
- The predicate device is the Omron HCG-801 portable ECG 4.0 Predicate Device: Monitor (K060766) marketed by Omron Healthcare, Inc. & MD100 Handheld ECG Monitor (K080933) marketed by Beijing Choice Electronic Technology Co., Ltd..
- The Vion.ECG Pen (Model no.: 800E) is intended for 5.0 Intended Use: recording ECG data and displaying general information by adult patients who are concerned their heart rhythm and waveform. This Vion,ECG Pen allows the patients to record transient cardiac events voluntarily for providing to the healthcare professionals as references during office visits.
In addition, the patient can also record their ECG data and transmit the recorded data to personal computer.
The Vion.ECG Pen is not intended for recording and transmission of user's ECG signal simultaneously. Users with implanted pacemakers are not recommended to use this
Product: Vion. 800E ECG Pen by MD Biomedical, Inc. Section 4 - 510(k) Summary REV. 【C】 Page 1 of 3
{1}------------------------------------------------
device. This device is not intended to substitute a conventional diagnostic tool.
6.0 Device Description: The MD's Vion.800E ECG Pen is a small, portable and easy-to-use electrocardiograph unit that can record and store electrocardiogram (ECG) measurements of your heart rhythm. Each ECG reading records 30 seconds measurement and these ECG readings can help your doctor monitor your condition.
> The unit included build-in memory that can store up to 99 measurements, including the ECG data along with date and time of measurement. With USB interface, it allows you or your doctor to view detail information of measurement.
> Furthermore, Vion® ECG Pen (Model no.: 800E) has similar general design with the Omron HCG-801 portable ECG Monitor (K060766 ) marketed by Omron Healthcare, Inc.. & MD100 Handheld ECG Monitor (K080933) marketed by Beijing Choice Electronic Technology Co., Ltd.
In terms of operating specification, Safety & EMC 7.0 Non-Clinical Performance Tests requirements, the device conforms to applicable standards include Summary: All Safety test: according to IEC 60601-2-25 & IEC
| All Burst, test: | 60601-1, |
|-------------------|-----------------------------|
| EMC test: | according to IEC 60601-1-2 |
| Performance test: | according to IEC 60601-2-47 |
A Clinical Investigation study was performed in such a way 8.0 Clinical test that compared the performance , including Heart rate. QRS Summary: interval, OT interval, PR interval, R-wave, S-wave, T-wave --------etc., between single channel patient-activated ECG device "Vion.800E" VS. 12-lead standard ECG "PHILIPS, PageWriter Trim II Cardiograph. 206 patients are involved in this study.
> According to the clinical investigation results and discussion, the performance of Vion.ECG Pen (Model: Vion.800E) is as
Product: Vion. 800E ECG Pen by MD Biomedical, Inc. Section 4 - 510(k) Summary REV. (C) Page 2 of 3
{2}------------------------------------------------
good as the 12-lead ECG (PHILIPS, PageWriter Trim II Cardiograph) in recording ECG data and displaying general information.
The patient-activated ECG system used in this study is a leadless recorder, which records 30 seconds ECG signal. The time it costs for use is much shorter than other conventional ECG recorders, and requires no external wires and electrodes. Therefore, the patient-activated ECG recorder, Vion.800E, is ideal to be used to document the behavior of rhythm and signal waveform during symptoms in order to replace a more expensive and invasive diagnostic procedures.
## Conclusions:
Vion® ECG Pen (Model no.: 800E) has the same intended use, principles of operation, and similar technological characteristics as predicate devices. Moreover, bench testing contained in this submission demonstrates that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. Thus, the Vion® ECG Pen (Model no.: 800E) is substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Meridian Dynamics Biomedical, Inc. C/O Jennifer Reich Harvest Consulting Corp. 2904 N. Boldt Drive Flagstaff, AZ 86001
JUL 1 3 2011
Re: K103077
Trade/Device Name: ECG Pen Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: June 14, 2011 Received: June 24, 2011
Dear Ms. Reich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it ' may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{4}------------------------------------------------
Page 2 – Ms. Jennifer Reich
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.html for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bryan D. Zuckerman, M.D.
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): ___K103077
Device Name: Vion.® 800E ECG Pen MD Biomedical, Inc.
Indications for Use:
The Vion.ECG Pen (Model no.: 800E) is intended for recording ECG data and displaying general information by adult patients who are concerned their hearter hythm and waveform. This Vion.ECG Pen allows the patients to record transient cardiac events voluntarily for providing to the healthcare professionals as references during office visits.
In addition, the patient can also record their ECG data and transmit the recorded data to personal computer.
The Vion.ECG Pen is not intended for recording and transmission of user's ECG signal simultaneously. Users with implanted pacemakers are not recommended to use this device. This device is not intended to substitute a conventional diagnostic tool.
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | V | AND/OR | Over-The-Counter Use<br>(21 CFR 807 Subpart C) | |
|-------------------------------------------------|---|--------|------------------------------------------------|--|
|-------------------------------------------------|---|--------|------------------------------------------------|--|
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
| 510(k) Number | K103077 |
|---------------|---------|
|---------------|---------|
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.