The Handheld ECG Monitor MD100 is intended for use in non-invasive recording and displaying ECG waveform by adult patients. In addition, it also can provide to the treating physician with relevant data on the cardiac rhythm in hospital patients. It is immediately available at any time to manually record transient cardiac events, suitable for patient and professional use, helpful in determining the cardiac rhythm at the time of symptoms. This ECG monitor allows the patient to record their ECG data into the device memory for display by healthcare professional during office visits. The product is not a conventional diagnostic tool.
Device Story
Handheld, battery-operated ECG monitor; records real-time cardiac event data via palm or chest contact; no external electrodes required. Device components include signal input/amplify units, CPU, display, and storage chip on a single PCB. Operated by patients or professionals to capture transient symptoms; data stored in device memory. Includes USB connectivity for data upload to a PC; "Keep-it-easy" software used for storage and playback. Healthcare providers review recorded waveforms during office visits to assist in cardiac rhythm assessment. Benefits include portability and ability to capture cardiac events at the time of symptoms.
Clinical Evidence
Bench testing only. Testing included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), ambulatory electrocardiograph performance (AAMI EC38), and biological evaluation (ISO 10993). No clinical data provided.
Technological Characteristics
Handheld device; ABS and metal construction. Sensing via direct skin contact (palm/chest). Powered by 2 AAA batteries. Features real-time ECG waveform display, internal storage, and USB connectivity. Software-driven. Standards: IEC 60601-1, IEC 60601-1-2, AAMI EC38, ISO 10993. Dimensions: 136x84x21mm; Weight: 100g.
Indications for Use
Indicated for adult patients for non-invasive recording and display of ECG waveforms to capture transient cardiac events and provide cardiac rhythm data to healthcare professionals.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
Kofo933
SEP 2 6 2008
# Appendix II 510(K) Summary
### Non-Confidential Summary of Safety and Effectiveness
This summary of 510(k) safety and effectiveness information is being submitted in accordance the requirements of 21 CFR 807.92
Beijing Choice Electronic Technology Co.,LTd. East Building 3F, No.5 Shixing Street, Badachu Hi-tech Zone Shijingshan District, Beijing Tel: 0086-10-88790480 China 100041 Fax:0086-10-88798860 Official Contact: Lei Chen Manager Email: chenlei@choicemmed.com Proprietary or Trade Name: MD100 Handheld ECG Monitor Common/Usual Name: Handheld, ECG Monitor Classification: Class II Device: MD100 Handheld ECG Monitor Predicate Devices: Omron HCG-801 Portable ECG Monitor K060766
{1}------------------------------------------------
#### Device Description:
The applicant device of Handheld ECG Monitor MD100 is a handheld device, which can records cardiac event data and displays the data in a clear and precise waveform.
The ECG Monitor is made up of signal input unit, signal amplify unit, CPU, Display unit, power unit and storage chip. They are all in one PCB that is designed and made by our company.
The MD100 Handheld ECG Monitor is activated by the user whenever symptoms are experienced. The recorded data serves as reliable evidence and are later shown to physicians or other health care professionals for confirmation of these symptoms. When a user feels that a cardiac event is occurring, the utilization of MD100 has the feature of recording this real time data that is normally difficult to capture.
The applicant device has "data upload" function which controlled by hardware, it can transmit the data measured by the device to computer via the USB port. The software attached with the device named " Keep-it-easy" in CD ROM, which is use for store and playback the data collected by MD100 ECG monitor. The "Keep-it-easy" software needs to install in the computer by user. The "Keep-it-Easy" software CD ROM is the accessory of the applicant device.
The applicant device has low battery voltage indication function. The power of the monitor works supplied by 2*AAA batteries.
### Non-Confidential Summary of Safety and Effectiveness
{2}------------------------------------------------
#### Indications for Use:
The Handheld ECG Monitor MD100 is intended for use in non-invasive recording and displaying ECG waveform by adult patients. It addition, it also can provide to the treating physician with relevant data on the cardiac rhythm in hospital patients. It is immediately available at any time to manually record transient cardiac events. suitable for patient and professional use, helpful in determining the cardiac rhythm at the time of symptoms. This ECG monitor allows the patient to record their ECG data into the device memory for display by healtheare professional during office visits.
The product is not a conventional diagnostic tool.
## Non-Confidential Summary of Safety and Effectiveness
Environment of Use - Prescription
{3}------------------------------------------------
4/7
.....
### Device Attributes:
:
| Features | MD100 |
|----------------------------------|----------------------------------------------------------------------------------------|
| Environment of use | Prescription use for self testing anywhere and anytime. |
| Patient Population | Adult |
| Type of waveform | Real time ECG Waveform display. |
| Heart Rate Range | 30-240 beats/min |
| Software driven | Yes |
| Materials in patient<br>contact | ABS, Metal |
| Standard met | IEC 60601-1, IEC 60601-1-2, AAMI EC38 |
| Measurement Rate | 30 Seconds |
| Components | 2 AAA Batteries, Hanging rope, Dustproof cover, Operation manual, USB Cable(Optional ) |
| Operating<br>conditions | 5°C~40 °C |
| Storage conditions | -20℃~40℃<br>≤RH80% |
| Dimensions(mm) | 136mm(W)×84mm(H)×21mm(D) |
| Weight(kg)<br>without<br>battery | 100 g |
## Non-Confidential Summary of Safety and Effectiveness
Differences Between Other Legally Marketed Predicate Devices
. 11:
{4}------------------------------------------------
The MD100 Handheld ECG Monitor is viewed as substantially equivalent to predicate device: Omron HCG-801 portable ECG Monitor. (K060766)
The OnIron HCG-801 portable ECG Monitor features real-time display of waveform and ECG recording.
BASIC COMPARISONS BETWEEN MD100 and HCG-801,
| | MD100 | HCG-801 |
|-------------------------------------------|-----------------------------------|-----------------------------------|
| SUBSTANTIAL<br>EQUIVALENCE<br>COMPARISONS | K060766 | |
| Intended Use | Same | Same |
| Prescription/Over<br>the Counter | Prescription | Prescription |
| Display of waveform | Display real-time ECG<br>waveform | Display real-time ECG<br>waveform |
| Type of Transmission | USB Transmission | Non-USB Transmission |
| Transmission Tool | USB | Standard compact memory<br>card |
| Lead placement on body | Palm and Chest Placement | Chest Placement |
| Multiple Event Recording | Yes | Yes |
| Base-line stabilization | Yes | Yes |
| Battery Life Indicator | Yes | Yes |
| Lead Cable | No | No |
| Pacemaker Detection | No | No |
## Non-Confidential Summary of Safety and Effectiveness
In summary, the MD100 Handheld ECG Monitor is substantially equivalent to Omron HCG-801 portable ECG Monitor in the following ways:
{5}------------------------------------------------
MD100 Handheld ECG Monitor is schematically similar to Omron HCG-801 portable ECG Monitor.
Both devices are handheld, portable, personal type ECG monitor. MD100 Handheld ECG Monitor does provide the data transmission option which is not an option the Omron HCG-801 ECG offers. Both devices are prescription devices intended for self-testing by patients under doctors' supervision. In both devices, user can place device on his/her chest and hold it steadily to test although the MD100 Handheld ECG Monitor provides another method that test by the centre of the palm. In both devices the user is not required to apply external electrodes to the body. Both devices have the capability to record real time heart rhythm waveform and heart beat and store data that can be displayed and downloaded.
The MD100 Handheld ECG Monitor constitutes a safe, accurate, and reliable means for recording of ECG data. When this device is used as intended it is as safe and effective as the predicate device. As shown, MD100 device has generally the same technological characteristics and intended use as Omron HCG-801 portable ECG Monitor but more advantageous and practical in terms of ease of use and reliability.
Validation testing contained in the submission demonstrates that any differences in their technological characteristics do not raise any new questions of safety and effectiveness. When the device is used as it is intended it poses no adverse health effects of safety risks to users.
Performance Testing:
We performed the following bench testing to demonstrate safety and effectiveness and equivalency to the predicate device:
IEC 60601-1 Medical Electrical Equipment Part 1: General Requirements for Safety; Amendment 1, 1991-11,Amendment 2,1995-03. Version 1995
EMC tests according to IEC 60601-1-2 Medical Electrical Equipment Part 1-2: General Requirements for Safety-Collateral Standard: Electromagnetic Compatibility Requirements and Tests. Version 2001
AAMI EC38 Ambulatory Electrocardiographs ISO 10993, Biological Evaluation of Medical Devices
## Non-Confidential Summary of Safety and Effectiveness
Conclusion
{6}------------------------------------------------
Based upon the performance testing and comparison to legally marketed predicate device ( for indications for use, technology, and performance ) we have demonstrated that the MD100 Handheld ECG Monitor is substantially equivalent in safety and effectiveness to the predicate device.
There are no significant differences that affect the safety or effectiveness of the intended device as compared to the predicate device.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an abstract eagle-like symbol with three stylized wing segments. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
SEP 2 6 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Beijing Choice Electronic Technology Co., Ltd. c/o Ms. Diana Hong Shanghai Midlink Business Consulting Co., Ltd. Suite 8D, Zhongxin Zhongshan Mansion, No. 19, Lanc 999, Zhong Shan Nan Er Rd. Shanghai 200030, CHINA
Re: K080933
Trade/Device Name: Handheld ECG Monitor MD 100 Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II (Two) Product Code: DPS Dated: September 18, 2008 Received: September 19, 2008
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the individe for use stated in the enclosure) to legally marketed predicate devices marked in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drig and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, inte Act. devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Province
{8}------------------------------------------------
Page 2 - Mr. Mark Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as seting forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Duna R. Vachner
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
# Appendix I Indications for Use Form
510(k) Number: K080993
#### Device Name: Handheld ECG Monitor MD100
Indications for Use:
The Handheld ECG Monitor MD100 is intended for use in non-invasive recording and displaying ECG waveform by adult patients. In addition, it also can provide to the treating physician with relevant data on the cardiac rhythm in hospital patients. It is immediately available at any time to manually record transient cardiac events, suitable for patient and professional use, helpful in determining the cardiac rhythm at the time of symptoms. This ECG monitor allows the patient to record their ECG data into the device memory for display by healthcare professional during office visits.
The product is not a conventional diagnostic tool.
Prescription Use
Over-The-Counter Use
AND/OR
( Part 21 CFR 801 Subpart D)
( Part 21 CFR 801 Subpart C)
( PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
Duma D. Vahine
Division Sign-Off)
Division of Cardiovascular Devices
Page _1_ of _1
510(k) Number K800733
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.