The clue medical and clue medical BASIC are handheld, battery operated single channel electrocardiographs intended for recording and transmitting ECG data by patients who are concerned about their heart rhythm. The clue medical and clue medical BASIC allow the patient to record their ECG data into the device memory and transmit it for display by healthcare professionals. Specifically, the clue medical and clue medical BASIC are intended for patients that are concerned about their heart rhythm or have experienced the following symptoms that are suggestive of abnormal heart rhythm: Skipped beats Pounding heart (palpitations) History of arrhythmia The clue medical and clue medical BASIC ECG data is intended to be used by a licensed health care practitioner. These measurements are not intended for any specific clinical diagnosis. The clinical significance must be determined by the physician.
Device Story
Handheld, battery-operated single-channel ECG recorder; features four fixed electrodes and ECG lead connector. Patient records ECG (30s or 2min duration) into internal memory; stores up to ten recordings. Data transmitted via IrDA interface to local PC/printer or via IrDA-capable cell phone to Telovital server. Output is a printable PDF file. Used by patients at home; reviewed by physicians to assess heart rhythm. Device does not provide clinical diagnosis; physician determines clinical significance. Clue medical BASIC lacks cell phone transmission capability.
Clinical Evidence
Bench testing only. Device tested for compliance with IEC 60601-1 (General Safety), IEC 60601-1-1 (Medical Electrical Systems), IEC 60601-1-2 (EMC), IEC 60601-2-25 (Safety of Electrocardiographs), and IEC 60601-2-47 (Ambulatory ECG systems). Electrode leadwires compliant with IEC 60601-1 subclause 56.3(c). No clinical data provided.
Technological Characteristics
Handheld single-channel ECG recorder; battery-operated. Four fixed electrodes. IrDA interface for data transmission. Output: PDF file. Standards: IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-2-25, IEC 60601-2-47. Connectivity: IrDA to PC/printer or IrDA-capable cell phone.
Indications for Use
Indicated for patients concerned about heart rhythm or experiencing symptoms suggestive of arrhythmia (skipped beats, palpitations, history of arrhythmia). Intended for use by licensed healthcare practitioners to review ECG data recorded by patients.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
K083630 — CLUE MEDICAL AND CLUE MEDICAL BASIC · Telovital GmbH Telemedizin · Apr 23, 2009
K042814 — READMYHEART · Dailycare Biomedical, Inc. · Jan 6, 2005
K111438 — REKA E100 · Reka Pte, Ltd. · Sep 12, 2011
K060766 — PORTABLE, ELECTROCARDIOGRAPH (ECG) MONITOR, MODEL HCG-801 · Omron Healthcare, Inc. · Jun 9, 2006
K973267 — TTM5000 TELEPHONIC EKG MONITOR · Hds Medical, Inc. · Jan 29, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
NOV 10 2008
K082013
Pg 1 of 3
Class II
# Section 5: 510(k) Summary
and the same
. . . . . .
Monitor, Omron Healthcare
The following information is provided as required by 21 CFR § 807.87 for Telovital's 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the safety and effectiveness information upon which the substantial equivalence determination is based.
The safety and effectiveness of the clue medical and clue medical BASIC is based upon a determination of the substantial equivalence as well as the safety and effectiveness of its predicate device(s).
| Sponsor: | | Telovital GmbH Telemedizin<br>Office Park I / Top 4<br>A-1300 - Vienna - Airport | | |
|----------------------|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|----------|
| Contact: | | Calley Herzog<br>M Squared Associates<br>901 King St. Suite 200<br>Alexandria, VA 22314<br>Ph: 703-562-9800 x261<br>Fax: 703-562-9797<br>Email: cherzog@msquaredassociates.com | | |
| Date Prepared: | | October 10, 2008 | | |
| Proposed Class: | | II | | |
| Proprietary Name: | | clue medical and clue medical BASIC Telemedical<br>Systems | | |
| Common Name: | | Electrocardiograph | | |
| Classification Name: | | Electrocardiograph | | |
| Regulation Number: | | 21 CFR 870.2340, 21 CFR 870.2920 | | |
| Product Codes: | | DPS, DXH | | |
| Predicate Device(s): | | | | |
| | Device Name | Procode | Common Name | Class |
| K072353 | Portable ECScope,<br>DyAnsis Corp. | DPS | Electrocardiograph | Class II |
| K060766 | HCG Portable ECG | DPS | Electrocardiograph | Class II |
{1}------------------------------------------------
K082213
pg 2 of 3
#### K050670 Sensor mobile SM 100 DXH VITAPHONE
transmitters and receivers, electrocardiograph, telephone
Class II
#### Indication for Use
The clue medical and clue medical BASIC are handheld battery operated single channel electrocardiographs intended for recording and transmitting ECG data by patients who are concerned about their heart rhythm. The clue medical and clue medical BASIC allow the patient to record their ECG data into the device memory and transmit it for display by healthcare professionals.
Specifically, the clue medical and clue medical BASIC are intended for patients that are concerned about their heart rhythm or have experienced the following symptoms that are suggestive of abnormal heart rhythm:
Skipped beats Pounding heart (palpitations) History of arrhythmia
The clue medical and clue medical BASIC ECG data is intended to be used by a licensed health care practitioner. These measurements are not intended for any specific clinical diagnosis. The clinical significance must be determined by the physician.
#### Device Description
The clue medical is an handheld battery operated single-channel ECG recorder featuring four fixed electrodes on the back of the device, an ECG lead connector, an IrDA1 interface and one single button to record and store up to ten ECG recordings with 30 seconds or two minutes duration. The stored ECG tests are processed into a printable Portable Document Format (pdf) that can be transmitted two ways. The pdf can be transmitted via IrDA to an IrDA capable local PC or printer. If the receiving device is not IrDA capable, an optional IrDA USB stick is provided with the clue medical. The pdf file can then be saved, sent, displayed or printed. Alternatively, the pdf can be transmitted via an IrDA capable cell phone directly to the Telovital server for storage.
IIDA refers to Infrared Data Association, a standard for communication between devices (such as computers, PDAs and mobile phones) over short distances using infrared Data Association (IrDA) defines physical specifications communications protocol standards for the short-range exchange of data over infrared light.
{2}------------------------------------------------
K082013
pg 3 of 3
The clue medical BASIC is identical to the clue medical in design. The only difference is that the clue medical BASIC does not provide the added capability of sending the ECG data via cell phone directly to the Telovital server, it is only capable of IrDA transmission to a local PC or printer.
#### Performance Testing
The clue medical and clue medical BASIC have been tested to mcct all of the following standards:
- IEC 60601-1, "Medical Electrical Equipment, General Requirements for Safety" �
- IEC 60601-1-1, "Safety Requirements for Medical Electrical Systems" .
- IEC 60601-1-2, "Part 1, General Requirements for Safety" .
- IEC 60601-2-25, "Medical Electrical Equipment Part 2: Particular Requirements for . the Safety of Electrocardiographs"
- IEC 60601-2-47, "Medical Electrical Equipment, Particular requirements for safety, . including performance, for ambulatory electrocardiographic systems"
Electrode leadwires were found to be compliant to IEC 60601-1, subclause 56.3(c),
### Technological Characteristics and Substantial Equivalence
The clue medical and clue medical BASIC are substantially equivalent to one or more of the predicate devices in design, measurement capability intended use and output format.
Their design, measurement capability and output are equivalent to the previously cleared Sensor Mobile device (K050670) and Portable EC Scope (K072353). Their intended use is equivalent to the Omron HCG-801 Portable ECG Monitor (K060766). This premarket notification has described the characteristics of the clue medical and clue medical BASIC in sufficient detail to assure substantial equivalence.
{3}------------------------------------------------
Public Health Service
Image /page/3/Picture/2 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with outstretched wings, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle. The seal is simple and monochromatic.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Telovital GmbII Telemedizin c/o Ms. Calley Herzog M Squared Associates 901 King St. Suite 200 Alexandria, VA 22314
NOV 1 0 2008
Re: K082013
Trade Name: clue medical and clue medical BASIC Telemedical Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Codes: DPS, DXH Dated: October 9, 2008 Received: October 10, 2008
Dear Ms. Herzog:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Ms. Calley Herzog
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Onna R. bo dines
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{5}------------------------------------------------
## Section 4: Indications for Use Statement
K082013 510(k) Number:
Device Name: clue medical and clue medical BASIC
#### Indications for Use:
The clue medical and clue medical BASIC are handheld, battery operated single channel electrocardiographs intended for recording and transmitting ECG data by patients who are concerned about their heart rhythm. The clue medical BASIC allow the patient to record their ECG data into the device memory and transmit it for display by healthcare professionals.
Specifically, the clue medical and clue medical BASIC are intended for patients that are concerned about their heart rhythm or have experienced the following symptoms that are suggestive of abnormal heart rhythm:
Skipped beats Pounding heart (palpitations)
History of arrhythmia
The clue medical and clue medical BASIC ECG data is intended to be used by a licensed health care practitioner. These measurements are not intended for any specific clinical diagnosis. The clinical significance must be determined by the physician.
Prescription Use
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
er R. vi hines
of Cardiovascular Devices
KOBZOIB 510(k) Number ...
Page 1 of
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.