K032736 · Aerotel Medical Systems (1998) , Ltd. · DXH · Oct 3, 2003 · Cardiovascular
Device Facts
Record ID
K032736
Device Name
HEART 2006
Applicant
Aerotel Medical Systems (1998) , Ltd.
Product Code
DXH · Cardiovascular
Decision Date
Oct 3, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2920
Device Class
Class 2
Indications for Use
The Heart 2006 is a long-term portable electrocardiogram monitor intended to be used for long-term cardiac out-patient management. The unit records a period of electrocardiogram whenever the patient feels symptoms, as indicated to him by a physician, and presses a button on the unit. The electrocardiograms are then sent to the physician by telephone. The unit is indicated whenever it is desired to have electrocardiograms of a symptomatic patient at the time of the symptoms. There are no known contraindications.
Device Story
Battery-powered ECG event recorder/transmitter; records ECG segments before and after patient-activated button press; transmits stored ECG data via telephone to central receiving station (e.g., Heartline Receiving Station). Used by patients in home/out-patient settings for long-term cardiac management. Physician reviews transmitted ECGs to aid clinical decision-making regarding symptomatic cardiac events. Benefits include remote capture of transient symptomatic ECG data.
Clinical Evidence
No clinical tests were performed. Equivalence established via non-clinical bench testing (electrical safety and electromagnetic compatibility).
Technological Characteristics
Battery-powered ECG event recorder/transmitter. Electrical safety and performance validated per IEC 601-1, IEC 601-1-2, and EC38 standards. Connectivity via telephone transmission to central receiving station.
Indications for Use
Indicated for long-term cardiac out-patient management in symptomatic patients requiring ECG monitoring at the time of symptoms. No known contraindications.
Regulatory Classification
Identification
A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.
K033451 — HEARTRAK SMART AT/HEARTRAK SMART2 · Universal Medical, Inc. · Feb 5, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
, P
# OCT 03 2003
Image /page/0/Picture/2 description: The image shows the text "K032736 page 1 of 2 Page A 2". The text appears to be handwritten, with the exception of "Page A 2". The text is black and is set against a white background. The image is likely a page from a document.
## Aerotel Medical Systems (1998) Ltd. Special 510(k) Heart 2006 Cardiac Event Recorder
### 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements Regulation par. 807.92, effective March 14, 1995.
#### 1. Submitter
Name: Aerotel Medical Systems (1998) Ltd Address: 5 Hazoref St., Holon 58856, Israel Telephone Number: 972-3-5596111 Contact person: Dr. George Myers, 210-787- 1703 Date prepared: March 7, 2000
#### 2. Device
Proprietary name: Heart 2006 Common Name: Electrocardiograph Event Recorder and Telephonic Transmitter Classification Name: Transmitters and Receivers, electrocardiograph, telephone
The Heart 2006 is a battery-powered ECG event recorder and transmitter that is capable of storing an electrocardiogram and transmitting it by means of a telephone to a central receiving station. It records a portion of the ECG both before and after a "Record" button is depressed.
3. Predicate Device
Heart 2005. K000775. manufactured by Aerotel Medical Systems (1998) Ltd..
4. Description
The H2006 is a battery powered ECG Event Recorder and Transmitter which is intended to be used by the patient to record portions of a patient's electrocardiogram (ECG) and to send it to a receiving center such as Aerotel's Heartline Receiving Station (FDA # K022073) or equivalent. It records portions of the ECG both before and after the Record button is depressed.
5. Intended Use
The Heart 2006 is a long-term portable electrocardiogram monitor intended to be used for long-term cardiac out-patient management. The unit records a period of electrocardiogram whenever the patient feels symptoms, as indicated to him by a
{1}------------------------------------------------
#### Aerotel H2006
physician, and presses a button on the unit. The electrocardiograms are then sent to the physician by telephone.
The unit is indicated whenever it is desired to have electrocardiograms of a symptomatic patient at the time of the symptoms. There are no known contraindications.
6. Comparison
The Heart 2006 is a modified version of the predicate device. It has basically the same electrical and mechanical characteristics, and the use of the two devices by the patient is equivalent.
7. Performance Data
(1) Non-clinical tests
The modifications were validated by repeating IEC 601-1 and IEC 601-1-2 tests (electrical safety and electromagnetic compatibility) and the electrical characteristic tests of standard EC38.
(2) Clinical Tests
No clinical tests were performed.
#### 8. Conclusion
The conclusion drawn from these tests is that the Heart 2006 is equivalent in safety and efficacy to its predicate device.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines that suggest feathers or wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 03 2003
Aerotel Medical Systems (1998) Ltd. c/o George H. Myers, Sc.D. President Medsys Inc. 377 Route 17 South Hasbrouck Heights, NJ 07604
Re: K032736
Trade Name: Heart 2006 Regulation Number: 21 CFR 870.2920 Regulation Name: Telephone Electrocardiograph Transmitter and Receiver Regulatory Class: Class II (two) Product Code: DXH Dated: September 2, 2003 Received: September 4, 2003
Dear Dr. Myers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Dr. George Myers
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
V. Valenti
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Aerotel H2006 Recorder
Page 3
K032736
Page _________________________________________________________________________________________________________________________________________________________________________
510(k) Number (if known):
Indications for Use Form
Device Name: Heart 2005
Indications for Use:
The Heart 2006 is a long-term portable electrocardiogram monitor intended to be used for long-term cardiac out-patient management. The unit records a period of electrocardiogram whenever the patient feels symptoms, as indicated to him by a physician, and presses a button on the unit. The electrocardiograms are then sent to the physician by telephone.
The unit is indicated whenever it is desired to have electrocardiograms of a symptomatic patient at the time of the symptoms. There are no known contraindications.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | X |
|----------------------|---|
| Use | |
| (Per 21 CFR 810.109) | |
OR
Over-the-Counter
(Optional Format 1-2-96)
OaheTm
510(k) Number
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.