ScioCardio ECG transmitter is a physiological ECG acquisition module. It records and transmits the data acquired, to a compatible mobile phone via Bluetooth low energy link, for visualization without making any analysis or filtering to the data acquired. ScioCardio ECG transmitter acquires Lead III ECG signals meeting the AAMI/ANSI/IEC standards for ambulatory ECG acquisition, via three electrodes attached to the patient's body. ScioCardio ECG transmitter is a battery powered wearable device which allows the patient to be free to moving while wearing the device. ScioCardio ECG transmitter transmits the acquired physiological signals in real-time to a compatible mobile phone where ScioCardio Mobile App is installed. ECG data is stored in the mobile phone verbatim without any sampling adjustment or any other modifications and also is displayed in the Mobile App. ScioCardio ECG transmitter is a prescription only device which is intended for use under the supervision of a physician or those knowledgeable in all aspects of ECG morphology, rhythm and arrhythmia. ScioCardio ECG transmitter and ScioCardio Mobile App are intended to be used on adult patients only.
Device Story
Battery-powered wearable ECG transmitter; captures 3-lead ECG signals and motion data via IMU sensor; transmits raw data via Bluetooth Low Energy (BLE) to compatible mobile phone running ScioCardio Mobile App. Device is always-on; requires no patient intervention after initialization. Data stored verbatim on mobile phone without filtering or analysis; visualized via app. Used in ambulatory settings; allows patient mobility. Healthcare providers use visualized output for ECG monitoring, heart rate monitoring, and arrhythmia detection/analysis. Benefits include continuous, comfortable long-term monitoring.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, including electrical safety (AAMI/ANSI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), and ambulatory ECG system performance (AAMI/ANSI/IEC 60601-2-47). Additional bench testing validated 120-hour battery life, transmitter range, and patient simulator operability. Software/firmware validated per ISO 14971.
Technological Characteristics
Wearable 3-lead ECG transmitter; 3.7V rechargeable Li-ion battery; Bluetooth Low Energy (BLE) connectivity; 250Hz sampling rate with 17-bit ADC; 0.05-40 Hz bandwidth; 100 dB CMRR. Form factor: small, wearable. Standards: AAMI/ANSI ES60601-1, IEC 60601-1-2, AAMI/ANSI/IEC 60601-2-47.
Indications for Use
Indicated for adult patients requiring ambulatory ECG monitoring. Prescription only; intended for use under supervision of a physician or qualified personnel knowledgeable in ECG morphology, rhythm, and arrhythmia.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
{0}------------------------------------------------
May 30, 2018
SYNERGEN Technology Labs LLC % Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct Naples, Florida 34114
Re: K171019
Trade/Device Name: ScioCardio ECG Transmitter Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency Physiological Signal Transmitter And Receiver Regulatory Class: Class II Product Code: DRG Dated: April 30, 2018 Received: April 30, 2018
Dear Daniel Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
## Page 2 - Daniel Kamm
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
M.G. Hillemann
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K171019
Device Name ScioCardio ECG Transmitter
#### Indications for Use (Describe)
ScioCardio ECG transmitter is a physiological ECG acquisition module. It records and transmits the a compatible mobile phone via Bluetooth low energy link, for visualization without making any analysis or filtering to the data acquired.
ScioCardio ECG transmitter acquires Lead III ECG signals meeting the AAMI/ANSI/IEC standards for ambulatory ECG acquisition, via three electrodes attached to the patient's body.
ScioCardio ECG transmitter is a battery powered wearable device which allows the patient to be free to moving while wearing the device.
ScioCardio ECG transmitter transmits the acquired physiological signals in real-time to a compatible mobile phone where ScioCardio Mobile App is installed. ECG data is stored in the mobile phone verbatim without any sampling adjustment or any other modifications and also is displayed in the Mobile App.
ScioCardio ECG transmitter is a prescription only device which is intended for use under the supervision of a physician or those knowledgeable in all aspects of ECG morphology, rhythm and arrhythmia.
ScioCardio ECG transmitter and ScioCardio Mobile App are intended to be used on adult patients only.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Synergen Technology Labs. The logo features the word "SYNERGEN" in large, bold, blue letters. Below that is the text "Technology Labs" in a smaller, gray font. To the left of the word "SYNERGEN" is a semi-circular design made up of blue and gray dots.
# 510(k) Summary, SYNERGEN TECHNOLOGY LABS LLC
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92.
Date Summary Prepared: January 26, 2017
### Submitter's Identification:
SYNERGEN Technology Labs LLC 3131 McKinney Avenue, Suite 602 Dallas, Texas 75204
### Summary Prepared by:
Sunil Konda, Vice President, Product Direct: 214-643-6002 Mobile: 214-435-0633 sunil.k@synergentl.com www.Synergentl.com
#### Device Name:
Trade /Proprietary Name: ScioCardio ECG Transmitter Regulation Number: 870.2910 Regulation Name: Transmitters and receivers, physiological signal, radiofrequency Regulatory Class: II Product Code: DRG
### Legally Marketed Predicate Devices Information:
Trade/Device Name: Cardioline HD+ made by Cardioline S.p.A. Premarket Notification K150289 Regulation Number: 870.2910 Regulation Name: Transmitters and Receivers, Physiological Signal, Radiofrequency Regulatory Class: II Product Code: DRG
Trade/Device Name: PocketECG v2 made by Medicalgorithmics LLC Premarket Notification K112921 Regulation Number: 870.1025 Regulation Name: Arrhythmia Detector and Alarm Regulatory Class: II Product Code: DSI
K171019
{4}------------------------------------------------
### Device Description:
ScioCardio ECG transmitter is a battery powered wearable device, capable of continuously capturing a body surface ECG from the patient, storing and transmitting the data to an external device or a system wirelessly. ECG signals are captured from the conventional solid gel electrodes worn on the body using a 3 lead ECG system. Raw ECG data is stored and transmitted wirelessly to any compatible device with Bluetooth Low Energy capability. This allows visualization, analysis and interpretation which can be used for a variety of applications such as ECG monitoring, heart rate monitoring, arrhythmia detection and analysis, out-patient monitoring etc. Other than the ECG data, ScioCardio ECG transmitter also captures, stores and transmits patient's motion data from an IMU sensor providing information about the patient's activities which can also be used in above mentioned analysis and interpretations. ScioCardio ECG transmitter is an always-on device. Data acquisition, storing and transmission are all done automatically and require no patient intervention other than initializing the device. ScioCardio ECG transmitter is small in form factor and has a higher battery life, making it possible for the user to wear the device comfortably for longer durations as per specific requirements.
Image /page/4/Figure/2 description: The image shows a diagram of a ScioCardio ECG Transmitter Device connected to a mobile phone, which is then connected to a remote system. The ScioCardio ECG Transmitter Device is on the left side of the image, and it is connected to the mobile phone by a blue arrow labeled "Data". The mobile phone is in the center of the image, and it is connected to the remote system by a blue arrow labeled "Data". The remote system is on the right side of the image, and it is represented by a cloud with a database icon.
### Intended Use/Indications for use:
ScioCardio ECG transmitter is a physiological ECG acquisition module. It records and transmits the data acquired, to a
compatible mobile phone via Bluetooth low energy link, for visualization without making any analysis or filtering to the data acquired.
ScioCardio ECG transmitter acquires Lead III ECG signals meeting the AAMI/ANS/JEC standards for ambulatory ECG acquisition, via three electrodes attached to the patient's body.
ScioCardio ECG transmitter is a battery powered wearable device which allows the patient to be free to moving while wearing the device.
ScioCardio ECG transmitter transmits the acquired physiological signals in real-time to a compatible mobile phone where ScioCardio Mobile App is installed. ECG data is stored in the mobile phone verbatim without any sampling adjustment or any other modifications and also is displayed in the Mobile App.
ScioCardio ECG transmitter is a prescription only device which is intended for use under the supervision of a physician or those knowledgeable in all aspects of ECG morphology, rhythm and arrhythmia.
ScioCardio ECG transmitter and ScioCardio Mobile App are intended to be used on adult patients only.
Safety and Effectiveness: As shown in the table below, the design presents no issues regarding safety and effectiveness as compared to the predicate devices. K171019 Page 2 of 7
{5}------------------------------------------------
### Table of comparison to Legally Marketed Devices:
| Comparison<br>Criteria | K150289 Cardioline HD+<br>Product Code DRG | Synergen ScioCardio ECG<br>Transmitter<br>Product Code DRG | SE Discussion |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Photo | Image: K150289 Cardioline HD+ | Image: Synergen ScioCardio ECG Transmitter | Equivalent technology employed:<br>Battery operated Bluetooth ECG<br>recorder/ transmitter |
| Basic<br>Comparison | 12 lead ECG without arrhythmia<br>detection software, Bluetooth<br>communication. Analysis program on<br>the host is a separate product not<br>marketed with the HD+. | 3 Lead ECG without arrhythmia<br>detection software. Uses Bluetooth<br>communication. Compatible with 3rd<br>party arrhythmia detection not<br>marketed with the ScioCardio. | Like K150289, i.e. without arrhythmia<br>but 3 leads instead of 12, same<br>product code. |
| Mechanical<br>Design | Portable plastic case with internal<br>batteries. | SAME | Substantially equivalent. |
| Comparison<br>Criteria | K150289 Cardioline HD+<br>Product Code DRG | Synergen ScioCardio ECG<br>Transmitter<br>Product Code DRG | SE Discussion |
| Indications for<br>Use | HD+ is a physiological ECG acquisition<br>module. HD+ transmits wireless, via<br>Bluetooth to a PC or Tablet, the data<br>acquired, without making any analysis or<br>filtering on the data acquired.<br>HD+ acquires 12-lead ECG waveforms<br>meeting the standards for clinical and<br>diagnostic applications (AAMI, ANSI, AHA,<br>ACC) and offers full ECG acquisition. HD+ is<br>designed to acquire and transmit a high<br>quality ECG data allowing the patient to be<br>free to moving (without cable connected to<br>the processing unit). The HD+ transmits the<br>acquired physiological signals in real-time to<br>a computer/device where a compatible<br>application is installed. All data acquired are<br>sent via Bluetooth to a receiver that it can be<br>a PC, tablet or device capable of receiving BT<br>data. The ECG is transmitted verbatim to the<br>receiving system, without LSB or sampling<br>adjustment. It is up to the receiving<br>system/application to perform the necessary<br>processing such as (but not limited to) LSB<br>scaling, signal filtering, Resting ECG analysis<br>etc... The device HD+ is intended to be used<br>on adult and on all pediatric patients. The<br>device is intended for use by qualified,<br>trained nurses and physicians operating in<br>hospitals, clinics and medical practices. | ScioCardio ECG transmitter is a physiological<br>ECG acquisition module. It records and<br>transmits the data acquired, to a<br>compatible mobile phone via Bluetooth low<br>energy link, for visualization without making<br>any analysis or filtering to the data acquired.<br>ScioCardio ECG transmitter acquires Lead II<br>and Lead III ECG signals meeting the<br>AAMI/ANSI/IEC standards for ambulatory<br>ECG acquisition, via three electrodes<br>attached to the patient's body.<br>ScioCardio ECG transmitter is a battery<br>powered wearable device which allows the<br>patient to be free to moving while wearing<br>the device.<br>ScioCardio ECG transmitter transmits the<br>acquired physiological signals in real-time to<br>a compatible mobile phone where<br>ScioCardio Mobile App is installed. ECG data<br>is stored in the mobile phone verbatim<br>without any sampling adjustment or any<br>other modifications and also is displayed in<br>the Mobile App.<br>ScioCardio ECG transmitter is a prescription<br>only device which is intended for use under<br>the supervision of a physician or those<br>knowledgeable in all aspects of ECG<br>morphology, rhythm and arrhythmia.<br>ScioCardio ECG transmitter and ScioCardio<br>Mobile App are intended to be used on adult<br>patients only. | Like K112921 PocketECG v2 in that it records<br>3 leads (but we don't perform the analysis,<br>that is done by a 3rd party program.<br>Like K150289 Cardioline HD+ in that it the<br>analysis is done by a 3rd party program, but<br>we record 3 leads instead of 12.<br>The compatible 3rd party analysis program<br>(Telemed, K142349) works with either 12 or<br>3 leads.<br>Substantially equivalent. |
| Comparison | K150289 Cardioline HD+ | Synergen ScioCardio ECG | SE Discussion |
| Criteria | Product Code DRG | Transmitter | |
| | | Product Code DRG | |
| ECG Leads | Up to 12 leads | 3 leads | Substantially equivalent. Analysis |
| | | | programs can utilize either 3 or 12 |
| | | | lead ECG formats. |
| Communications | Bluetooth, Full ECG | Bluetooth Low Energy (BLE) Full ECG | SAME technology |
| | | to Mobile Phone | |
| Analysis | 3rd party program | 3rd party program | Includes aspects of both predicates |
| Power Source | 2x AAA Batteries | Rechargeable 3.7V Li-ion battery | Battery operated: Substantially |
| | | | equivalent |
| Battery Life | 10 Hours | 120 Hours | The subject device has the longest |
| | | | battery life. |
| Powering the | Device is always on | Device is always on | SAME |
| Device | Powered on when the battery is | Powered on when the battery is | |
| | inserted to the device, no power | inserted to the device, no power…
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.