VITALCONNECT PLATFORM BY VITAL CONNECT INC.

K132447 · Vitalconnect, Inc. · DRG · Apr 3, 2014 · Cardiovascular

Device Facts

Record IDK132447
Device NameVITALCONNECT PLATFORM BY VITAL CONNECT INC.
ApplicantVitalconnect, Inc.
Product CodeDRG · Cardiovascular
Decision DateApr 3, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2910
Device ClassClass 2

Indications for Use

The VitalConnect Platform is a wireless monitoring system intended for use by healthcare professionals for unattended surveillance of physiological data within healthcare settings. This includes heart rate, electrocardiography (ECG), heart rate variability, respiratory rate, skin temperature, activity (including step count), and posture (body position relative to gravity including fall). Data is transmitted wirelessly to a central location where it is stored for analysis. The VitalConnect Platform can be configured by Authorized Persons to notify healthcare professionals when physiological data falls outside selected parameters. The device is not intended to be used on critical care patients and is intended to supplement vital signs monitoring by healthcare professionals, not to replace current standards of care. The device is intended for use on general care patients and on patients who are 18 years of age or older. It is not intended for home use.

Device Story

Wireless data collection system for physiological monitoring; consists of disposable adhesive patch with integrated sensor module, relay software library, and secure server software library. Sensor module captures ECG, HR, HRV, respiratory rate, skin temperature, activity, and posture; transmits encrypted data via relay to central server. Used in healthcare settings by healthcare professionals for unattended surveillance. System provides real-time configuration and notifications when physiological data exceeds user-defined parameters. Healthcare providers access and review stored data at central location. Supplements vital signs monitoring to assist clinical decision-making; intended to improve patient surveillance in general care environments.

Clinical Evidence

Clinical study performed on human subjects. Bench testing included ECG, HR, HRV, respiration rate, body impedance, activity/posture, Bluetooth verification, notification accuracy, data transmission/storage, reliability, electromagnetic compatibility, electrical safety, co-existence, software V&V, and biocompatibility. Results demonstrate the platform is safe and effective for its intended use.

Technological Characteristics

Wireless ambulatory monitoring system. Components: disposable adhesive patch, sensor module, relay software, server software. Sensors measure ECG, HR, HRV, respiratory rate, skin temperature, activity, and posture. Connectivity: Bluetooth/wireless transmission to relay and central server. Battery-operated. Software-based processing and notification. Biocompatible materials.

Indications for Use

Indicated for unattended surveillance of physiological data (HR, ECG, HRV, respiratory rate, skin temperature, activity, posture) in general care patients 18 years or older within healthcare settings. Not for critical care patients or home use.

Regulatory Classification

Identification

A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K132447 ## 510(K) SUMMARY This summary of 510(k)-safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. | Submitter: | Vital Connect Inc.<br>900 East Hamilton Avenue<br>Suite 500<br>Campbell, CA 95008 | |--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Sam Mostafavi<br>Vice President of Quality Assurance and Regulatory Affairs<br>E-mail: smostafavi@vitalconnect.com<br>Phone: (408) 963-4620<br>Fax: (408) 963-2828 | | Date Prepared: | Vital Connect Inc.<br>900 East Hamilton Avenue<br>Suite 500<br>Campbell, CA 95008<br>March 27, 2014 | | Trade name: | VitalConnect Platform (consisting of VitalConnect Sensor,<br>Relay Software Library and Secure Server Software Library) | | Classification: | Class II | | Classification: | 21 CFR 870.2910, Product Code: DRG<br>• Transmitters and Receivers, Physiological Signal,<br>Radiofrequency<br>21 CFR 870.1025, Product Code: DSI<br>• Arrhythmia detection and alarm (including ST-segment<br>measurement and alarm) | | Predicate Device: | • CareFusion, Wireless Monitoring System, 510(k) #:<br>K110809<br>• Preventice, BodyGuardian System, 510(k) #: K121197<br>• Corventis, Mobile Patient Management System, 510(k) #:<br>K083287 | | Classification<br>Panel: | Cardiology | | Device<br>Description: | The VitalConnect Platform is a wireless data collection system<br>that monitors physiological data and consists of the following<br>sub systems:<br>• VitalConnect Sensor (includes adhesive Patch and Sensor<br>Module)<br>• Relay Software Library<br>• Server Software Library | {1}------------------------------------------------ VitalConnect Platform sub-system includes: VitalConnect Sensor . a) Patch The Patch is designed as a disposable self-adhesive interface to the body. b) Sensor Module Residing within the patch, the sensor module performs processing functions related to capture of physiological data and also performs bi-directional communication with the relay device. VitalConnect Relay Software Library ● The Relay Software Library manages bi-directional communication between the Sensor Module and the Server Software Library and is installed on a relay device. VitalConnect Secure Server Software Library . The Server Software Library is installed on a central server, manages the upload, processing and storage of sensor data, as well as real-time configuration of and notifications from the VitalConnect Platform. The VitalConnect Sensor continuously gathers physiological data from the person being monitored and then transmits encrypted data via a bi-directional relay to the central server. The VitalConnect Sensor is a battery-operated adhesive patch with integrated sensors and wireless transceiver module, worn on the torso to record heart rate, electrocardiography (ECG), heart rate variability, respiratory rate, skin temperature, activity (including step count), and posture (body position relative to gravity including fall) for in-patient use. VitalConnect Sensor consisting of Patch and the Sensor Physical Description: Module dimensions are: Patch: approximately 111 mm long x 47mm wide x 6 mm . high (without the release liner) Sensor Module: approximately 21 mm long x 12 mm wide x . 4 mm high The VitalConnect Platform is a wireless monitoring system Intended Use: ● intended for use by healthcare professionals for unattended surveillance of physiological data within healthcare settings. This includes heart rate, electrocardiography (ECG), heart rate variability, respiratory rate, skin temperature, activity (including step count), and posture (body position relative to gravity including fall). Data is transmitted wirelessly to a central location where it is stored for analysis. The VitalConnect Platform can be configured by Authorized Persons to notify healthcare professionals when {2}------------------------------------------------ | | • physiological data falls outside selected parameters.<br>The device is not intended to be used on critical care<br>patients and is intended to supplement vital signs<br>monitoring by healthcare professionals, not to replace<br>current standards of care. The device is intended for use on<br>general care patients and on patients who are 18 years of<br>age or older. It is not intended for home use. | |---------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Comparison of<br>Technological<br>Characteristics: | VitalConnect Platform and the primary predicate device<br>(CareFusion, Wireless Monitoring System, 510(k) #: K110809)<br>contain small ambulatory monitoring sensors that measure<br>ECG, heart rate variability, respiratory rate, skin temperature,<br>activity (including step count), and posture (body position<br>relative to gravity including fall). All transmit their data to an<br>external device, which in turn, transmits the data to a remote<br>computer server that allows healthcare professionals to access<br>and review the data. There are no fundamental differences<br>between their technological characteristics.<br>Additionally, we claim Preventice BodyGuardian System,<br>K121197, and Corventis, Mobile Patient Management System,<br>K083287 as secondary predicates. Preventice BodyGuardian<br>System for ECG data, and Corventis, Mobile Patient<br>Management System for Activity and Posture. | | Non-Clinical<br>Testing: | The following bench testing was conducted on the<br>VitalConnect Platform system:<br>• Electrocardiograph (ECG)<br>• Heart rate<br>• Heart rate variability.<br>• Respiration rate<br>• Body impedance<br>• activity (including step count),<br>• Posture (body position relative to gravity including fall)<br>• Bluetooth verification<br>• Notification<br>• Measurement accuracy<br>• Communication, data transmission and storage<br>• Reliability<br>• Electromagnetic compatibility<br>• Electrical safety testing<br>• Co-existence testing<br>• Software verification and validation testing<br>• Biocompatibility verification and testing | | Clinical Testing: | Clinical study was performed on human subjects. | | Conclusion: | The results of the non-clinical and clinical studies demonstrate<br>that the VitalConnect Platform is safe and effective for its<br>intended use. | | | | | Based on the intended use and product performance results.<br>the subject device has been shown to be substantially | | | equivalent to the currently marketed predicate devices. | | : . . ・ · {3}------------------------------------------------ . . . . . . . . : : : {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with its wings spread. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 2, 2014 VITAL CONNECT, Inc. Sam Mostafavi VP of QA & RA 900 E. Hamilton Ave. Ste # 500 Campbell, CA 95008 US Re: K132447 Trade/Device Name: Vitalconnect platform by VITAL CONNECT Inc. Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (includes ST-segment measurement,alarm) Regulatory Class: Class II Product Code: DSI, DRG Dated: February 12, 2014 Received: February 20, 2014 Dear Sam Mostafavi: This letter corrects our substantially equivalent letter of April 3, 2014. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply {5}------------------------------------------------ Page 2 - Mr. Sam Mostafavi with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address: http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to: http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. FDA forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure This page intentionally left blank {6}------------------------------------------------ ## INDICATIONS FOR USE STATEMENT 510(k) Number (if known): K132447 Device Name: VitalConnect Platform Indication for use: - The VitalConnect Platform is a wireless monitoring system intended for use by . healthcare professionals for unattended surveillance of physiological data within healthcare settings. This includes heart rate, electrocardiography (ECG), heart rate variability, respiratory rate, skin temperature, activity (including step count), and posture (body position relative to gravity including fall). Data is transmitted wirelessly to a central location where it is stored for analysis. The VitalConnect Platform can be configured by Authorized Persons to notify healthcare professionals when physiological data falls outside selected parameters. - The device is not intended to be used on critical care patients and is intended to . supplement vital signs monitoring by healthcare professionals, not to replace current standards of care. The device is intended for use on general care patients and on patients who are 18 years of age or older. It is not intended for home use. Prescription Use x (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ## Kenneth J. Cavanaugh -S Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...