K013533 · General Devices · DRG · Jan 18, 2002 · Cardiovascular
Device Facts
Record ID
K013533
Device Name
EMS TELEMEDICINE OPTION (FOR GEMS SERIES 4000)
Applicant
General Devices
Product Code
DRG · Cardiovascular
Decision Date
Jan 18, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2910
Device Class
Class 2
Indications for Use
The intended use of the GEMS Series 4000 EMS Telemetry Option is to provide a means to communicate with distant medical devices in an EMS pre-hospital setting and to acquire and present information, such as single/12-lead ECG and physiologic/non-physiologic data which has been gathered by these devices.
Device Story
EMS Telemedicine Option is an add-on hardware/software module for GEMS Series 4000 communications console; enables PC-based reception of physiologic data from remote medical devices. Inputs: single/12-lead ECG, physiologic/non-physiologic data via radio, landline, or wireless telephone. Signal processing: FM, DTMF, FSK, and modem demodulation via plug-in PCB. Output: visual presentation on CRT monitor, strip-chart, or laser printer. Used in EMS pre-hospital environments by emergency personnel. Healthcare providers use output to view patient status; facilitates clinical decision-making by providing remote access to patient data. Benefits: enables remote monitoring and data acquisition in pre-hospital settings.
Clinical Evidence
Bench testing only. Testing verified ability to communicate, convert, and present physiologic information under simulated use conditions. Confirmed performance of demodulation and display elements per ANSI/AAMI EC11-1991. No clinical data.
Technological Characteristics
PC option card (PCB) for GEMS Series 4000 console. Signal processing: FM, DTMF, FSK, modem. Connectivity: RS232, PC modem, radio, landline, wireless telephone. Power: +3.3, +/- 12 VDC from PC supply. Conforms to ANSI/AAMI EC11-1991 for heart rate meter and demodulator performance.
Indications for Use
Indicated for use in EMS pre-hospital settings to communicate with distant medical devices and acquire/present physiologic data (single/12-lead ECG) and non-physiologic data gathered by those devices.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
Predicate Devices
GEMS Series 2000 (Modified DR-2C/PMC 100) (K896153)
Rosetta-Rx
Related Devices
K201746 — Tempus Pro Patient Monitor · Remote Diagnostic Technologies Limited · Sep 18, 2020
K973318 — MOBIMED SYSTEM, (PEGASUS AND POLARIS) · Ortivus Us, Inc. · Feb 19, 1998
K170567 — Tempus Pro Patient Monitor · Remote Diagnostic Technologies , Ltd. · Jul 6, 2017
K091176 — PHILIPS TELEMEDICINE SERVER/SYSTEM, MODEL 861441 · Philips Medical Systems · Nov 13, 2009
K040663 — TELENABLE EMS SYSTEM · Molex, Inc. · Jun 30, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
'JAN 1 8 2002
K013533
p. 1/4
# 510(k) SUMMARY
| Applicant: | Michael Smith<br>President, General Devices<br>1000 River St.<br>Ridgefield, NJ 07657 |
|----------------------|------------------------------------------------------------------------------------------------------------------|
| Contact Person(s): | Applicant and/or<br>Curtis Bashford<br>Vice-President, General Devices<br>1000 River St.<br>Ridgefield, NJ 07657 |
| Telephone & Fax No.: | Tel. (201) 313-7075<br>Fax (201) 313-5671<br>E-Mail cbashford@general-devices.com |
| Date Prepared: | October 19, 2001 |
## Device Information:
| | Trade/Proprietary Name: EMS Telemedicine Option (for GEMS Series 4000 |
|-----------------|-----------------------------------------------------------------------|
| Common Name: | ECG Demodulator/Display |
| Classification: | Class II |
| Panel: | Cardiovascular Devices Classification |
| Product Code: | Not Applicable |
## Predicate Device Information:
| Trade/Proprietary Name: | GEMS Series 2000 (Modified DR-2C/PMC 100) |
|-------------------------|-------------------------------------------|
| Common Name: | EMS Communications Console |
| Manufacturer | General Devices (same as applicant) |
| Classification: | Class II |
| 510(k) No. | K896153 |
| Trade/Proprietary Name: | Rosetta-Rx |
| Common Name: | Demodulator, Data Translator |
| Manufacturer | General Devices (same as applicant) |
| Classification: | Class II |
{1}------------------------------------------------
#### DESCRIPTION OF DEVICE
The EMS Telemedicine Option is an optional element of the GEMS Series 4000 communications console which provides the means for performing telemedicine functions, namely, communicating with and presenting information acquired from distant medical devices.
The GEMS Series 4000 communications console is a standard radio/telephony communications console used for voice and data communications. The GEMS Series 4000 communications console, without the EMS Telemetry Option is marketed for non-medical applications. Personal Computer (PC) based, it contains various means to interface to and control communications elements (i.e.; 2-way radio equipment, landline and wireless telephone) as which are commonly used for commercial and public safety (i.e.: police, fire and EMS) communications purposes.
With the EMS Telemedicine Option, the console can perform medical functions, such as communicating physiologic information (i.e .: single/12-Lead ECG and physiologic/non-physiologic data) with distant medical devices. The signal processing means employed for this includes: frequency modulation (FM), which is commonly used in EMS ECG telemetry applications; DTMF (touch-tone) signaling, which is commonly used for telecommunications signaling: Frequency Shift Keying (FSK), which is commonly used for data communications; as well as standard digital communications schemes such as modulation/demodulation (modem) communications. It also provides for the display and printing of physiologic information, such as single and 12-Lead ECG, and non-physiologic data, such as EMS Run ID, patient name, time of day, etc.
The EMS Telemedicine Option consists of hardware and software elements that are installed into the GEMS Series 4000 communications console. These hardware & software elements convert received information into a digital form compatible with standard PC functions, such as information presentation by a CRT monitor and hardcopy printouts. Signal processing is performed by a printed circuit board (PCB) that plugs into the console's PC Mother Board. The EMS Telemedicine Option's features and functions are similar to those found in this type of equipment, including the predicate devices (GEMS Series 2000 and Rosetta-Rx). The digital and analog technologies employed by the above stated purposes are commonly employed for this purpose and are well understood. The functioning of the heart rate meter and the performance of the demodulator/signal processing and display elements conform to ANSI/AAMI EC11-1991.
#### INTENDED USE
The intended use of the GEMS Series 4000 EMS Telemetry Option is to provide a means to communicate with distant medical devices in an EMS pre-hospital setting and to acquire and present information, such as single/12-lead ECG and physiologic/non-physiologic data which has been gathered by these devices.
{2}------------------------------------------------
K013533 2314
## COMPARISON WITH PREDICATE DEVICES
| Function or<br>Specification | New Device | Predicate Devices | |
|----------------------------------------------------|-------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| | ECG Telemetry Option | GEMS Series 2000 | ROSETTA-Rx |
| Form Factor | PC Option Card (for PC cabinet) | Desktop/rack mount cabinet | Option Card (for GEMS<br>Series 2000) or Stand-Alone<br>device |
| ECG Presentation Means | CRT Display (PC monitor), strip-<br>chart recorder, laser printer | CRT Display (Vectored Beam<br>CRT), strip-chart recorder, laser<br>printer (w/Rosetta Option) | Stand-Alone: Laser printer<br>Option Card: GEMS 2000<br>display and record features |
| Communication Means | Radio, landline telephone,<br>wireless telephone | Radio, landline telephone,<br>wireless telephone | Radio, landline telephone,<br>wireless telephone |
| Digital Communication<br>Interface | Serial (RS232), PC Modem | Serial (RS232) | Serial (RS232) |
| ECG & physiologic/non-<br>physiologic data Formats | FM, DTMF, FSK, Modem | FM, DTMF | FM, DTMF, & FSK |
| Demodulation Means | Hardware/Software | Hardware | Hardware/Software |
| ECG Demodulation<br>Format | 1400 Hz Center frequency, 50 &<br>250 mV/Hz deviation | 1400 Hz Center frequency, 50<br>mV/Hz deviation | 1400 Hz Center frequency,<br>50 & 250 mV/Hz deviation |
| Data Demodulation<br>Format | FSK 1.2kHz/1.6kHz | FSK 1.2kHz/1.6kHz(w/Rosetta<br>Option) | FSK 1.2kHz/1.6kHz |
| User Controls | Pointing Device (mouse) | Mechanical front panel<br>switches | Mechanical front panel<br>switches |
| Power | +3.3, +/- 12 VDC from PC power<br>supply | 115Volt AC, 60Hz | Stand alone: External 9VDC<br>adapter<br>Option Card: +5, +/- 12<br>VDC from GEMS Series<br>2000 power supply |
#### SUMMARY OF DIFFERENCES
The differences between the predicate and the new device relate primarily to: the form of system integration, in that the EMS Telemetry Option is an "add-on" to a nonmedical device; the use of a modem to communicate information; the use of a pointing device, rather than mechanical control keys, an lastly, minor differences, such as formfactor, size, appearance and minor operating features.
## SUMMARY OF NON-CLINICAL TESTS
The EMS Telemetry Option was subjected to non-clinical testing to insure proper performance. The testing consisted of the following parts:
{3}------------------------------------------------
- · A determination of the ability of the EMS Telemetry Option to correctly and accurately communicate desired information from the acquisition device under simulated use conditions
- · A determination to the ability of the EMS Telemetry Option to correctly convert the gathered information to the desired communications scheme under simulated use conditions
- · A determination of the efficacy of the intended transmission schemes under simulated use conditions
- · A determination of the ability of the EMS Telemetry Option to correctly and accurately acquire the information from the transmission means under simulated use conditions
- · A determination of the ability of the EMS Telemetry Option to correctly convert the gathered information for presentation by the desired presentation means and accurately present this information under simulated use conditions
### DISCUSSION OF HOW TEST RESULTS SUPPORT SUBSTANTIAL EQUIVALENCY
The testing that was performed on the EMS Telemetry Option demonstrates that the system is substantially equivalent to the predicate devices in that physiologic information acquired by an external source was reliably and accurately received and reconstructed for presentation at the far end.
#### CONCLUSION DRAW BY NON-CLINICAL TESTING
The conclusions draw by the non-clinical testing indicate that the EMS Telemetry Option/EMS Telemetry Option successfully perform the intended task under normal use conditions. The system behaved as expected and demonstrates the application of well-understood technology as well as the absence of any electrical risk factors to the patient.
END OF SUMMARY
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. In the center of the seal is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect and promote the health and well-being of the nation.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
'JAN 1 8 2002
Mr. Michael Smith President General Devices 1000 River St. Ridgefield, NJ 07657
Re: K013533
Trade Name: EMS Telemedicine Option (for GEMS Series 4000) Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency Physiological Signal Transmitter and Receiver Regulatory Class: Class II (two) Product Code: DRG Dated: October 19, 2001 Received: October 23, 2001
#### Dear Mr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
Page 2 - Mr. Michael Smith
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
h. D. Zeller, M.D.
Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
6.2
#### INDICATIONS FOR USE STATEMENT
PMN 510(k) Number: _ KO1333
Device Name: ECG Telemedicine Option (for GEMS Series 4000)
Indications For Use:
.
The ECG Telemedicine Option for the GEMS Series 4000 Communications Console is indicated for use when it is desired to communicate with distant medical devices in an EMS pre-hospital setting and to acquire and present information, such as single/12lead ECG and physiologic/non-physiologic data, or other physiologic information which has been gathered by these devices.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
OR
Over-The-Counter Use **__**
(Per 21 CFR 801.109)
(Optional Format 1-2-96)
Division of Cardiovascular & Respiratory Devices
510(k) Number: KA3533
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.