Comply with AAMI/ANSI EC57:2012 (Testing And Reporting Performance Results Of Cardiac Rhythm And ST-Segment Measurement Algorithms)
Complies with AAMI/ANSI EC57:2012
The AI ECG algorithm was developed using a machine learning based process (convolutional neural network or CNN).
Bench testing conducted to verify design specifications and compliance with AAMI/ANSI EC57:2012 and IEC 60601-2-47:2012.
Indications for Use
The AI-ECG Tracker is intended to be used by qualified healthcare professionals in hospitals and healthcare facilities for the assessment of arrhythmias using ECG data acquired from adults (age 22 and older) without pacemakers. The product supports downloading and analyzing data recorded from electrodes with conductive paste/gel placed on standard location in compatible formats from any device used for the arrhythmia diagnostics such as Holter, event recorder, 12-lead ambulatory ECG devices when assessment of the rhythm is necessary. The AI-ECG Tracker provides ECG signal processing and analysis on a beat by beat basis, QRS detection, Supraventricular Ectopic Beat detection, QRS feature extraction, interval measurement, and rhythm analysis. The AI-ECG Tracker is not for use in life supporting or sustaining systems or ECG monitoring and Alarm devices. The AI-ECG Tracker interpretation results are not intended to be the sole means of diagnosis for any abnormal ECG. It is offered to physicians and clinicians on an advisory basis only in conjunction with the physician's knowledge of ECG patterns, patient background, clinical history, symptoms, and other diagnostic information.
Device Story
Distributed ECG auto-analysis system; assists physicians in measuring/interpreting ambulatory ECG data. Inputs: ECG waveform data (leads I, II, III, V1, V5) uploaded by user. Processing: Server-side analysis using proprietary algorithm; includes beat-by-beat processing, QRS detection, Supraventricular/Ventricular Ectopic Beat detection, feature extraction, interval/heart rate measurement, rhythm analysis. Output: Analyzed ECG data downloaded to Physician Diagnostic Client for review, modification, confirmation, and report generation. Used in hospitals/healthcare facilities by qualified professionals. Clinical decision-making: Advisory tool; results not sole diagnostic means; requires physician integration with clinical history/symptoms. Benefits: Automated analysis support for arrhythmia assessment.
Clinical Evidence
Bench testing only. Compliance verified with AAMI/ANSI EC57:2012 (cardiac rhythm/ST-segment measurement algorithms), IEC 60601-2-47:2012 (ambulatory ECG systems), and IEC 60601-2-25:2011 (electrocardiograph safety). Software lifecycle managed per IEC 62304; risk management per ISO 14971.
Technological Characteristics
Distributed software system; C# and C/C++ language; Microsoft .Net framework and .Net Core runtime. Connectivity: Local wired or wireless network. Algorithm: Convolutional Neural Network (CNN) for beat classification (locked model). Standards: AAMI/ANSI EC57, IEC 60601-2-47, IEC 60601-2-25, IEC 62304, ISO 14971.
Indications for Use
Indicated for assessment of arrhythmias in adults (age 22+) without pacemakers. Contraindicated for patients with pacemakers, patients age 21 and under, non-cardiac abnormalities, and use as physiological monitoring of vital signs.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
Predicate Devices
Monebo Automated ECG Analysis and Interpretation Software Library, Version 3.0 (K062282)
K250569 — Cardiologs Holter Platform · Philips France Commercial · Aug 6, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 20, 2020
Shenzhen Carewell Electronics Co., Ltd. % Arthur Goddard President FDA Regulatory and Quality Systems Consultant 31853 Cedar Road Mayfield Heights, Ohio 44124-4445
Re: K200036
Trade/Device Name: AI-ECG Tracker Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK, DPS Dated: January 3, 2020 Received: January 8, 2020
Dear Arthur Goddard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Jessica Paulsen Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200036
Device Name AI-ECG Tracker
### Indications for Use (Describe)
The AI-ECG Tracker is intended to be used by qualified healthcare professionals in hospitals and healthcare facilities for the assessment of arrhythmias using ECG data acquired from adults (age 22 and older) without pacemakers. The product supports downloading and analyzing data recorded from electrodes with conductive paste/gel placed on standard location in compatible formats from any device used for the arrhythmia diagnostics such as Holter, event recorder, 12-lead ambulatory ECG devices when assessment of the rhythm is necessary. The AI-ECG Tracker provides ECG signal processing and analysis on a beat by beat basis, QRS detection, Supraventricular Ectopic Beat detection, QRS feature extraction, interval measurement, and rhythm analysis. The AI-ECG Tracker is not for use in life supporting or sustaining systems or ECG monitoring and Alarm devices.
The AI-ECG Tracker interpretation results are not intended to be the sole means of diagnosis for any abnormal ECG. It is offered to physicians and clinicians on an advisory basis only in conjunction with the physician's knowledge of ECG patterns, patient background, clinical history, symptoms, and other diagnostic information.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size:16px">❌</span> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> |
|-----------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span>☐</span> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Carewell. The logo consists of a pink circular design on the left, followed by the word "carewell" in a dark blue font. Below "carewell" is the phrase "health care" in a smaller, light gray font. Underneath the logo is the text "Shenzhen Carewell Electronics Co., Ltd."
# Section 5: 510(K) Summary
This summary of 510(K) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1900 and 21 CFR 807.92.
The assigned 510(K) Number:K200036
#### ട. 510(K) Summary
#### 5.1. Date of Preparation: January 3rd, 2020
#### 5.2. Sponsor
Shenzhen Carewell Electronics Co., Ltd.
Floor 4, BLD 9, Baiwangxin High-Tech Industrial Park, Songbai Road, Xili Street, Nanshan District 518108, Shenzhen, P.R. China Establishment Registration Number: 3010089768 Contact Person: Chang Liu Position: General Manager Tel: +86-755-86170389 Fax: +86-755-86170478 Email: standard@carewell.com.cn
### 5.3. Submission Correspondent
Mr. Arthur Goddard 31853 Cedar Road, Ohio, 44124-4445, U.S.A. Tel: (216) 233-5722 Email: asjgoddard@aol.com
#### Subject Device Identification 5.4.
Subject Device Name: AI-ECG Tracker Edition: S Edition Common name: ECG Analysis Software Classification Name(s): Programmable Diagnostic Computer / Electrocardiograph Product Code: DOK, DPS Regulation Number: 21 CFR 870.1425 Review Panel: Cardiovascular Classification: II
### 5.5. Predicate Device
510(k) Number: K062282
Device Name: Monebo Automated ECG Analysis and Interpretation Software Library, Version 3.0 Manufacturer: Monebo Technologies, Inc.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Carewell health care. The logo consists of a pink circular design on the left, followed by the word "carewell" in dark blue. Below "carewell" is the phrase "health care" in a smaller, lighter font. Underneath the logo is the text "Shenzhen Carewell Electronics Co., Ltd."
#### Reference Device 5.6.
510(k) Number: K113485
Device Name: Electrocardiograph Manufacturer: Shenzhen Carewell Electronics Co., Ltd.
### 5.7. Indications for use
The AI-ECG Tracker is intended to be used by qualified healthcare professionals in hospitals and healthcare facilities for the assessment of arrhythmias using ECG data acquired from adults (age 22 and older) without pacemakers. The product supports downloading and analyzing data recorded from electrodes with conductive paste/gel placed on standard location in compatible formats from any device used for the arrhythmia diagnostics such as Holter, event recorder, 12-lead ambulatory ECG devices when assessment of the rhythm is necessary. The AI-ECG Tracker provides ECG signal processing and analysis on a beat by beat basis, QRS detection, Supraventricular and Ventricular Ectopic Beat detection, QRS feature extraction, interval measurement, heart rate measurement, and rhythm analysis. The AI-ECG Tracker is not for use in life supporting or sustaining systems or ECG monitoring and Alarm devices.
The AI-ECG Tracker interpretation results are not intended to be the sole means of diagnosis for any abnormal ECG. It is offered to physicians and clinicians on an advisory basis only in conjunction with the physician's knowledge of ECG patterns, patient background, clinical history, symptoms, and other diagnostic information.
#### 5.8. Contraindications
Not suitable for diagnosis of non-cardiac abnormalities. Not suitable for diagnosing patients with pacemakers. Not suitable for diagnosing patients age 21 and under. Not suitable for diagnosis other than cardiac arrhythmia. The device must not be used as a physiological monitoring of vital signs.
#### 5.9. Special Requirement
The input ECG data is required to contain at least one of leads I, II, III, V1, and V5.
### 5.10. Device Description
The AI-ECG Tracker is a distributed ECG auto analysis system designed to assist physicians and qualified healthcare professionals in measuring and interpreting ambulatory ECG data. The interpretations by the analysis program can be confirmed, edited, modified, or deleted by the physician and qualified healthcare professionals. The program is intended for use by qualified healthcare professionals in hospitals and other healthcare facilities for the assessments of common cardiac arrhythmias using ECG data acquired from adults (age 22 and older) without pacemakers.
The AI-ECG Tracker receives ECG waveform data uploaded by a user, analyzes ECG data and automatically interprets on a computer server. The ECG measurement, interpretation and waveform
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Shenzhen Carewell Electronics Co., Ltd. The logo includes a circular graphic with pink lines and the company name in dark blue, with "health care" in a smaller, lighter font underneath. The full company name, "Shenzhen Carewell Electronics Co., Ltd.", is printed in black below the logo.
data are then downloaded to the Physician Diagnostic Client for a physician to review, modify, confirm the analysis statements, and print the report. The original ECG waveform data is stored permanently in the user's server computer securely.
The system uses a machine learning based process (convolutional neural network or CNN) only for development of the AI ECG algorithm. The AI ECG algorithm is only used for beat classification. After the AI ECG algorithm is developed, the AI-based beat classification model is locked in the released product which means the marketed device doesn't have active machine learning or selflearning features.
| Item | Subject Device<br>AI-ECG Tracker | Predicate Device<br>K062282 Monebo<br>Automated ECG Analysis<br>and Interpretation Software<br>Library, Version 3.0 | Remark |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for use | The AI-ECG Tracker is intended to be used by qualified healthcare professionals in hospitals and healthcare facilities for the assessment of arrhythmias using ECG data acquired from adults (age 22 and older) without pacemakers. The product supports downloading and analyzing data recorded from electrodes with conductive paste/gel placed on standard location in compatible formats from any device used for the arrhythmia diagnostics such as Holter, event recorder, 12-lead ambulatory ECG devices when assessment of the rhythm is necessary. The AI-ECG Tracker provides ECG signal processing and analysis on a | The Automatic Analysis and Interpretation Software Library is intended for use by qualified medical professionals for the assessment of arrhythmias using historic ambulatory EGG data. The product supports downloading and analyzing data recorded in compatible formats from any device used for the arrhythmia diagnostics such as Holter, Event Monitor, 12 lead ambulatory or resting EGG devices, or other similar devices when assessment of the rhythm is necessary. The Automatic Analysis and Interpretation Software Library can also be electronically interfaced, and perform analysis with data | No substantial difference |
| Item | Subject Device<br>AI-ECG Tracker | Predicate Device<br>K062282 Monebo<br>Automated ECG Analysis<br>and Interpretation Software<br>Library, Version 3.0 | Remark |
| | beat by beat basis, QRS detection, Supraventricular and Ventricular Ectopic Beat detection, QRS feature extraction, interval measurement, heart rate measurement, and rhythm analysis. The AI-ECG Tracker is not for use in life supporting or sustaining systems or ECG monitoring and Alarm devices. | transferred from other computer based EGG systems, such as an EGG management system. The Automatic Analysis and Interpretation Software Library provides EGG signal processing and analysis on a beat by beat basis, QRS and Ventricular Ectopic Beat detection, QRS feature extraction, interval measurement, heart rate measurement, and rhythm analysis for up to sixteen(16) leads of captured data. The library is not for use in life supporting or sustaining systems or EGG monitoring and Alarm devices<br>The product can be integrated into computerized EGG monitoring devices. In this case the medical device manufacturer will identify the indication for use depending on the application of their device. | |
| | The AI-ECG Tracker interpretation results are not intended to be the sole means of diagnosis for any abnormal ECG. It is offered to physicians and clinicians on an advisory basis only in conjunction with the physician's knowledge of ECG patterns, patient background, clinical history, symptoms, and other diagnostic information. | | |
| Algorithm | Proprietary Algorithm | Proprietary Algorithm | Different |
| Level of<br>Concern of<br>the software | Major | Moderate | Different |
| Diagnostic<br>Statement | Heart rate determination for<br>non-paced adult<br>QRS Detection<br>Non-paced arrhythmia interpretation<br>for adult patients | Heart rate determination for<br>non-paced adult<br>QRS Detection<br>Non-paced arrhythmia<br>interpretation for adult patients | Same |
| Item | Subject Device<br>AI-ECG Tracker | Predicate Device<br>K062282 Monebo<br>Automated ECG Analysis<br>and Interpretation Software<br>Library, Version 3.0 | Remark |
| | Non-paced ventricular arrhythmia<br>calls for adult patients<br>Intervals measurement<br>Ventricular ectopic beat detection | Non-paced ventricular arrhythmia<br>calls for adult patients<br>Intervals measurement<br>Ventricular ectopic beat detection | |
| Fundamental<br>scientific<br>technology | The AI-ECG Tracker consists of:<br>- Interfaces which provide tools to measure, analyze, interpret, review ECGs, and to generate and print reports.<br>- Automated ECG interpretation algorithms that measure and analyze ECGs to provide supplementary information for ECG diagnosis.<br>The device is developed in C# and C/C++ language, supported by Microsoft .Net framework and .Net Core runtime.<br>Components and libraries can be accessed through the Application Programming Interface (API). Carewell<br>AI-ECG Tracker requires local wired or wireless network | The predicate device is a collection of callable functions that have been complied into machine code or IDL code of the computer on which they execute.<br>The predicate device consists of a basic application for viewing, analyzing and annotating ECG data, and a callable object library built on the Microsoft .Net framework.<br>An application software can be written to invoke some or all the functions in an object library.<br>The library can be accessed through an Application program Interface (API) as a callable function. This allows the library to be used as an accessory to an ECG management application or as a stand-alone product | No substantial difference.<br>The subject device AI-ECG Tracker and the predicate device K062282 both built with Microsoft .Net framework with similar application architect, both have user's interface for viewing, analyzing and interpreting ECG data, and both allow the core algorithm library to be accessed via APIs. |
## 5.11. Predicate Devices and Subject Device Comparison
# Table 3-1 Feature Comparison with Predicate Devices
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Carewell Health Care logo. The logo consists of a circular graphic on the left and the text "carewell" in a bold, dark blue font on the right. Below "carewell" is the text "health care" in a smaller, lighter gray font. The circular graphic is made up of several pink, curved lines that create a spherical shape.
Shenzhen Carewell Electronics Co., Ltd.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Carewell Health Care. The logo consists of a circular graphic on the left, made up of pink curved lines. To the right of the graphic is the text "carewell" in a bold, dark blue font. Below "carewell" is the text "health care" in a smaller, lighter font.
Shenzhen Carewell Electronics Co., Ltd.
## Table 3-2 Performance Comparison
| Item | Subject Device<br>AI-ECG Tracker | Predicate Device<br>K062282 Monebo Automated ECG<br>Analysis and Interpretation<br>Software Library, Version 3.0 | Remark |
|-------------------------------|----------------------------------|------------------------------------------------------------------------------------------------------------------|--------|
| Basic safety<br>and essential | Comply with IEC 60601-2-25 | Comply with IEC 60601-2-25 | Same |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Carewell health care. The logo includes a stylized globe-like image with curved lines, followed by the word "carewell" in a bold, sans-serif font. Below "carewell" is the phrase "health care" in a smaller, lighter font. The text "Shenzhen Carewell Electronics Co., Ltd." is located below the logo.
Traditional 510(k) Premarket Notification AI-ECG Tracker
| Item | Subject Device | Predicate Device | Remark |
|----------------------------|---------------------------------------------------|----------------------------------------------------------------------------------------------|--------|
| | AI-ECG Tracker | K062282 Monebo Automated ECG<br>Analysis and Interpretation<br>Software Library, Version 3.0 | |
| performance | | | |
| Measurement<br>performance | Comply with AAMI/ANSI EC57 and<br>IEC 60601-2-47. | Comply with AAMI/ANSI<br>EC57 and IEC 60601-2-47. | Same |
### 5.1. Non-Clinical Test Conclusion
Bench test were conducted to verify that the subject device met all design specifications, as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device complies with the following standards.
AAMI ANSI EC57:2012 Testing And Reporting Performance Results Of Cardiac Rhythm And ST-Segment Measurement Algorithms;
IEC 60601-2-47:2012 Medical Electrical Equipment - Part 2-47: Particular Requirements For The Basic Safety And Essential Performance of Ambulatory Electrocardiographic Systems;
IEC 60601-2-25:2011, Medical Electrical Equipment -- Part 2-25: Particular requirements for the safety of electrocardiographs;
IEC 62304 Edition 1.1 2015-06, Medical device software - Software life-cycle;
ISO 14971:2007, Medical devices-Application of risk management to medical device.
### 5.2. Substantially Equivalent Conclusion
The subject device, AI-ECG Tracker, is determined to be Substantially Equivalent (SE) to the predicate device, Monebo Automated ECG Analysis and Interpretation Software Library, Monebo Technologies, Inc. K062282.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.