QRS CARD S-T SEGMENT ANALYSIS PATIENT MONITORING SYSTEM MODEL: QRS CARD USB
Applicant
Pulse Biomedical, Inc.
Product Code
DPS · Cardiovascular
Decision Date
Jul 23, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2340
Device Class
Class 2
Indications for Use
This device is intended for use as a patient monitor for any patient undergoing ECG S-T Segment Analysis and/or ECG stress analysis during rest or exercise. Exercise can be performed treadmill or bicycle ergometer use. The intended use locations are either in a physician's office, hospital or rehabilitation facilities, or similar use areas. It is intended to be used by or on the order of a physician or similarly qualified health care professional. This device is intended for use in a hospital environment, physician's office, or similar settings. This device is not intended for home use.
Device Story
QRS Card™ S-T Segment Analysis Patient Monitoring System converts standard PCs into ECG monitors with ST segment measurement capabilities. System components include QRS CARD™ Office ECG device, patient cable/lead wires, and software. Input consists of ECG signals acquired from patients via lead wires; data transmitted to PC via Bluetooth wireless technology. Software processes signals to perform ST segment analysis and stress testing. Output displayed on PC screen for physician review; results can be printed or saved for future analysis. Used in hospitals, physician offices, or rehabilitation facilities by physicians or qualified healthcare professionals. Facilitates clinical decision-making through real-time monitoring and diagnostic data generation during rest or exercise.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Physiological monitor (Electrocardiograph). Components: ECG hardware, patient cables, PC-based software. Connectivity: Bluetooth wireless transmission. Energy source: PC-powered/battery. Environment: Clinical/hospital. Software-based analysis of ECG signals for ST segment measurement.
Indications for Use
Indicated for patients undergoing ECG S-T segment analysis and/or ECG stress analysis (treadmill or bicycle ergometer) at rest or during exercise in clinical settings. Not for home use.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
Predicate Devices
QRS CARD™ S-T Segment Analysis Patient Monitoring System (K972255)
K972255 — PULSE BIOMEDICAL INC, QRS CARD S-T SEGMENT ANALYSIS MONITORING SYSTEM · Pulse Biomedical, Inc. · Jul 7, 1998
K152427 — PC ECG · Edan Instruments, Inc. · Feb 5, 2016
K935732 — CARDIO PERFECT ST 2001 · Cardio Control BV · Apr 4, 1996
K052898 — IQMARK EZ STRESS · Brentwood Medical Technology Corp. · Mar 28, 2006
K022618 — X-SCRIBE II · Mortara Instrument, Inc. · Sep 6, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
# Attachment A. 510(k) Summary
Image /page/0/Picture/20 description: The image shows a logo with two distinct elements. On the left, there is a hexagonal shape filled with a textured pattern, and within it, a stylized electrocardiogram (ECG) waveform is visible. To the right of the hexagonal shape, the letters 'PBI' are printed in a bold, sans-serif font. The overall design appears to be for a company or organization with the initials PBI, possibly related to healthcare or cardiology.
**JUL 28 2010**
Ise Biomedical, Inc
**Pulse Biomedical GmbH**
Filhete Bihler Str 4, 72474 Winterlingen, GERALANY
Tel: +49-7434-316038 Fax:+49-7434-316039
Special 510(k) - QRS Card™ S-T Segment Analysis Patient Monitoring System Modification
Section 6
510(k) Summary - 807.92
## Applicant Name and Address
Pulse Biomedical Inc. 1305 Catfish Lane Norristown, PA 19403
#### Manufacturer Name and Address
Pulse Biomedical Inc. 1305 Catfish Lane Norristown PA 19403
#### Contact Person
Saleem Hasan - President Tel. No. - 610-660-5510 Fax No. - 610-666-5630
Establishment Registration Number 2529508 Device Name:
| Device<br>Trade<br>Name | Common<br>Classification<br>/ Name | Regulation<br>Number | Classification<br>Code | Device<br>Class | Classification<br>Name |
|-------------------------|------------------------------------|----------------------|------------------------|-----------------|------------------------|
| QRS<br>Card BT | Physiological<br>Monitor | N/A | DPS | II | Electrocardiograph |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo with two distinct elements. On the left, there is a stylized hexagon shape with a jagged line running through the center, resembling a heartbeat or electrical signal. To the right of the hexagon, the letters 'PBI' are printed in a bold, sans-serif font. The overall design is simple and professional.
1305 Catfish Lane, Norristown, PA 19403, USA
Tel: 610-666-5510 Fax: 610-666-5630
Wuneim Binler Str 4, 72474 Winterlingen, GERMANY
Tel: +49-7434-316038 Fax: +49-7434-316039
# Special 510(k) - QRS Card™ S-T Segment Analysis Patient Monitoring System Modification
## Legally Marketed Devices to Which Substantial Equivalence is Claimed
| 510(k) Number | Trade Name | Company |
|---------------|---------------------------------------------------------------------|------------------------|
| K972255 | QRS CARD™ S-T Segment<br>Analysis Patient Monitoring<br>System | Pulse Bio Medical Inc. |
| K080141 | PC ECG 1200W System | Norav Medical LTD |
| K061977 | Plex 04 Wireless<br>Electrocardiograph with<br>Software Accessories | PinMed, Inc. |
| K082077 | BR3/6, BR12 | Coscience GmbH & Co KG |
### Device Description:
The QRS CARD™ S-T Segment Analysis Patient Monitoring System converts a commercially available personal computer (PC) into an ECG monitor with ST segment measurement capability. Physicians can initiate ST evaluations on patients and the data are transmitted, using Bluetooth wireless technology, to a computer screen. Test results can also be printed at any time during the testing, or saved in computer memory for future review or additional report generation.
The QRS CARD™ S-T Segment Analysis Patient Monitoring System consists of the following components:
- The QRS CARD™ Office ECG Device ●
- ECG Patient Cable and Lead Wires ●
- Product Software on CD .
- User Manual .
#### Intended Use:
This device is intended for use as a patient monitor for any patient undergoing ECG S-T Segment Analysis and/or ECG stress analysis during rest or exercise. Exercise can be performed treadmill or bicycle ergometer use. The intended use locations are either in a physician's office, hospital or rehabilitation facilities, or similar use areas. It is intended to be used by or on the order of a physician or similarly qualified health care professional. This device is intended for use in a hospital environment, physician's office, or similar settings. This device is not intended for home use.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo with two distinct elements. On the left, there is a stylized, geometric shape resembling a hexagon with a jagged line running through its center, possibly representing a heartbeat or signal. To the right of this shape, the letters 'PBI' are printed in a bold, sans-serif font. The overall design is simple and professional, suggesting a company or organization with a focus on precision or technology.
#### Pulse Biomedica
1/3
-666-5630
# Special 510(k) - QRS Card™ S-T Segment Analysis Patient Monitoring System Modification
## Conclusion:
The technological characteristics of the QRS CARD™ S-T Segment Analysis Patient Monitoring System are safe and effective and substantially equivalent to the legally marketed predicate devices.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Pulse Biomedical, Inc. c/o Mr. Saleem Hasan President 1305 Catfish Lane Norristown, PA 19403
# JUL 2 3 2010
K100813 Re:
Trade/Device Name: QRS Card™ ST-Segment Analysis Patient Monitoring System Regulatory Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II (Two) Product Code: DPS Dated: June 22, 2010 Received: June 23, 2010
Dear Mr. Hasan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for assession to May 28, 1976, the enactment date of the Medical Device Amendments, or to conninered processified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and coosmeter fore, market the device, subject to the general controls provisions of the Act. The I ca mayy nevels provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Mr. Saleem Hasan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
buna R. Vohrer
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo with the letters "PBI" to the right of a hexagonal shape. Inside the hexagon is a line graph that resembles a heartbeat. The hexagon is textured with a rough, possibly stone-like pattern. The letters "PBI" are in a simple, bold font.
Pulse Biomedical, Inc.
1305 Catfish Lane, Norristown, PA 19403, USA el: 610-666-5510
Pulse Biomedical GmbH Wilhelm Bihler Str 4; 72474 Winterlingen, GERMANY 49-7434-316038
Special 510(k) - QRS Card™ S-T Segment Analysis Patient Monitoring System Modification
Section 4
Indications for Use Statement
510(k) Number: _______________________________________________________________________________________________________________________________________________________________ (if known)
Device Name:
QRS Card™ S-T Segment Analysis Patient Monitoring System
Indications for Use:
This device is intended for use as a patient monitor for any patient undergoing ECG S-T Segment Analysis and/or ECG stress analysis during rest or exercise. Exercise can be performed treadmill or bicycle ergometer use. The intended use locations are either in a physician's office, hospital or rehabilitation facilities, or similar use areas. It is intended to be used by or on the order of a physician or similarly qualified health care professional. This device is intended for use in a hospital environment, physician's office, or similar settings. This device is not intended for home use.
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of the CDRH, Office of Device Evaluation (ODE)
Prescription Use X OR Over the Counter Use
Dina P. Weiner
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number_ K100813
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.