MOBILEECG SYSTEM

K103427 · Memtec Corp. · DPS · Mar 24, 2011 · Cardiovascular

Device Facts

Record IDK103427
Device NameMOBILEECG SYSTEM
ApplicantMemtec Corp.
Product CodeDPS · Cardiovascular
Decision DateMar 24, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2340
Device ClassClass 2
AttributesPediatric

Indications for Use

The MobileECG System is intended to provide an analysis of rhythm and of detailed morphology of complex cardiac complexes for the resting 12 lead ECG in all situations, whether in a hospital or primary care setting. It is adjunct to the diagnosis of all commonly recognized ECG abnormalities such as myocardial infarction (MI), including acute MI, ventricular hypertrophy, abnormal ST-T changes and common abnormalities of rhythm. The MobileECG System is intended for use in adults and children of any age from birth upwards.

Device Story

MobileECG System consists of Model 950-12LR HHC wireless module and CardioVu interpretation software. Module captures 12-lead electrical heart activity; transmits data wirelessly to laptop, netbook, tablet, or desktop computer (up to 100 feet). Wireless transmission reduces cable-related artifacts. CardioVu software receives, displays, and stores ECG data in real-time; applies ECG algorithms to provide advisory interpretation. Used in hospitals or primary care settings by physicians or qualified medical professionals. Output serves as advisory tool for clinical decision-making; not intended as sole diagnostic means. Benefits include patient comfort and reduced motion artifacts by eliminating tethered cables.

Clinical Evidence

Bench testing only. No clinical data provided. Equivalence established through comparative non-clinical testing against standards including ANS/AAMI EC11, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4, ANSI/AAMI EC53, and FCC Part 15.247.

Technological Characteristics

System comprises Model 950-12LR HHC wireless module and CardioVu software. Wireless communication via dedicated secure link. Standards: ANS/AAMI EC11 (ECG performance), IEC 60601-1-1 (medical electrical systems), IEC 60601-1-2 (EMC), IEC 60601-1-4 (programmable electrical medical systems), ANSI/AAMI EC53 (ECG cables/leads), FCC Part 15.247 (radio frequency).

Indications for Use

Indicated for adults and pediatric patients (birth upwards) requiring resting 12-lead ECG analysis for rhythm and morphology, including detection of myocardial infarction, ventricular hypertrophy, ST-T changes, and rhythm abnormalities. Intended for use in hospital or primary care settings as an adjunct to clinical diagnosis.

Regulatory Classification

Identification

An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 2011 510(k) K103427 # SUMMARY OF SAFETY AND EFFECTIVENE ## 510(k) SUMMARY of Safety and Effectiveness ## 1. General Information | Submitter's Name:<br>Address: | Memtec Corporation<br>68 Stiles Road Unit D<br>Salem, NH 03079 | |-------------------------------|----------------------------------------------------------------| | Telephone: | 603 893-8080 Ext. 204 | | Contact Person: | Dennis Garboski | | Trade Name: | MobileECG System | | Common Name: | Electrocardiograph (per 21 CFR 870.2340) | | Class: | II | #### 2. Predicated Devices The legally marketed predicated devices to which equivalence is being claimed is: | Micromedical, Inc. | Biolog 3000 | K974351 | |--------------------|-----------------|---------| | Datrix, Inc. | Cardio WiFi and | K053083 | | Datrix, Inc. | CardioServer | K052883 | | QRS Diagnostic | EKGCARD SYSTEM | K030535 | | QRS Diagnostic | Cardioview 32 | K083749 | #### 3. Description of Device · The MobileECG System is an electrocardiograph consisting of Model 950-12LR HHC module and CardioVu interpretation software. The Model 950-12LR HHC module is a small, light weight, single battery, wireless module designed to transmit 12Lead electrical heart activity of a patient to a computer. This module will transmit a patient's ECG wirelessly to a laptop, netbook, tablet, or desktop computer containing the CardioVu interpretation software from up to 100 feet away. Elimination of the cable reduces a source of artifact resulting in high-quality ECG tracings. Wireless dedicated and secure communication links the Model 950-12LR HHC module to CardioVu interpretation software. The CardioVu software receives, displays, and stores the patients ECG data wirelessly on a laptop, netbook, tablet, or desktop computer in "real time". The software also has the data using ECG algorithms for review by a physician or other qualified medical professional as an advisory basis only in conjunction with the physician's knowledge of ECG. The wireless feature allows the patient not to be tethered to the computer or a bulky box thus reducing motion artifacts caused by movement of the connecting cables and providing comfort to the patient. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the text "K103427 P2/2" in a handwritten style. The text appears to be a code or identifier, possibly related to a document or item. The handwriting is clear and legible, with distinct characters. #### 4. Indications For Use The MobileEG System is intended to provide an interpretation of the resting 12 lead ECG in all situations, whether in a hospital or primary care setting. It is capable of diagnosing all commonly recognized ECG abnormalities such as myocardial infarction (MI), including acute MI, ventricular hypertrophy, abnormal ST-T changes and common abnormalities of rhythm. The MobileECG System is intended for use in adults and children of any age from birth upwards. The MobileECG System's interpretation software is not intended as sole means of diagnosis and is offered to physicians and clinicians on an advisory basis only in conjunction with the physician's knowledge of ECG. #### 5. Non-clinical Tests Used in Determination of Substantial Equivalence Non-clinical tests were performed to compare the MobileECG System to the predicated devices. The following applicable standards were used to compare the MobileECG System to the predicated devices: ANS/AAMI EC11, IEC 60601-1-2, IEC 60601-1-1, IEC 60601-1-4, 21 CFR 898, ANSI/AAMI EC53, AND FCC Part 15.247. ### 6. Conclusions From Non-clinical Testing After comparing predicated devices to Memtec's MobileECG System, results show that with the intended use, the Model 950-12LR HHC module with CardioVu interpretation software is equivalent in safety and effectiveness. Therefore Memtec supports a claim of substantial equivalence for the MobileECG System. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is an image of an eagle with its wings spread, with three lines representing the feathers. The eagle is facing to the right. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Memtec Corporation c/o Mr. Dennis Garboski President 68 Stiles Road Unit D Salem, NH 03079 MAR 2 4 2511 K103427 Re: > Trade/Device Name: MobileECG Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II (two) Product Code: DPS Dated: March 14, 2011 Received: March 16, 2011 Dear Mr. Garboski: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 –Mr. Dennis Garboski Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, 17 Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## INDICATIONS FOR USE STATEMENT Indications for Use 510(k) Number(if known) KIO 3427 Device Name: MobileECG System Indications for Use The MobileECG System is intended to provide an analysis of rhythm and of detailed morphology of complex cardiac complexes for the resting 12 lead ECG in all situations, whether in a hospital or primary care setting. It is adjunct to the diagnosis of all commonly recognized ECG abnormalities such as myocardial infarction (MI), including acute MI, ventricular hypertrophy, abnormal ST-T changes and common abnormalities of rhythm. The MobileECG System is intended for use in adults and children of any age from birth upwards. Prescription Use Over-The-Counter Use_ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (Please Do Not Write Below This Line) Concurrence of CDRH, Office of Device Evaluation (ODE) A.D.O. for B.Zuckerman (Division Sign-Off) ular Device 510(k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...