K173830 · Ge Medical Systems Information Technologies, Inc. · DPS · Sep 18, 2018 · Cardiovascular
Device Facts
Record ID
K173830
Device Name
MAC VU360 Resting ECG Analysis System
Applicant
Ge Medical Systems Information Technologies, Inc.
Product Code
DPS · Cardiovascular
Decision Date
Sep 18, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The MAC VU360 Resting ECG Analysis System is a non-invasive prescription device. -The device is indicated for use to acquire, analyze, display and print electrocardiograms. -The device is indicated for use to provide interpretation of the data for consideration by a physician. -The device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician. It is not intended as a sole means of diagnosis. The interpretations of ECG offered by the device are only significant when used in conjunction with a physician overread as well as consideration of all other relevant patient data. -The device is indicated for use on adult and pediatric (birth through 21 years of age) populations.
Device Story
Mobile resting electrocardiograph; acquires 3, 6, 12, or 15-lead ECG signals via surface electrodes; processes data using 12SL algorithm to provide interpretive analysis; displays and prints reports. Used in hospitals, clinics, and offices by physicians or trained personnel. Connects to wired LAN or WiFi to interface with cardiology information systems (e.g., GEHC MUSE) for workflow integration; supports PDF reports; displays external web/Citrix content. Output assists physicians in clinical decision-making; requires physician overread. Benefits include standardized ECG analysis and efficient data management.
Clinical Evidence
No clinical data; bench testing only. Substantial equivalence supported by non-clinical verification and validation, including risk analysis, design reviews, software lifecycle management, module verification, system integration, and safety testing.
Technological Characteristics
Mobile wheeled trolley form factor; 3, 6, 12, 15-lead ECG acquisition; wired LAN and WiFi connectivity. Employs 12SL algorithm for analysis. No sterilization required. Complies with voluntary standards for electrocardiographs.
Indications for Use
Indicated for adult and pediatric (birth through 21 years) patients to acquire, analyze, display, and print ECGs for physician interpretation. Not for use during patient transport, with high-frequency surgical units, as an intra-cardiac application, as a sole diagnostic means, or as a vital signs monitor.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
September 18, 2018
GE Medical Systems Information Technologies, Inc. Marcella Entwisle Regulatory Affairs Leader 9900 West Innovation Drive Wauwatosa. Wisconsin 53226
Re: K173830
Trade/Device Name: MAC VU360 Resting ECG Analysis System Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS, DQK, DXH Dated: August 17, 2018 Received: August 20, 2018
Dear Marcella Entwisle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica E. Paulsen -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K173830
Device Name
MAC VU360 Resting ECG Analysis System
#### Indications for Use (Describe)
The MAC VU360 Resting ECG Analysis System is a non-invasive prescription device.
-The device is indicated for use to acquire, analyze, display and print electrocardiograms.
-The device is indicated for use to provide interpretation of the data for consideration by a physician.
-The device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician. It is not intended as a sole means of diagnosis.
-The interpretations of ECG offered by the device are only significant when used in conjunction with a physician overread as well as consideration of all other relevant patient data.
-The device is indicated for use on adult and pediatric (birth through 21 years of age) populations.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|---------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. The circle is surrounded by a white border with a swirling pattern. The logo is simple and recognizable, and it is associated with a well-known and established company.
# 510(k) Summary
## Section 5: 510(k) Summary
In accordance with 21 CFR Part 807.92(a)(1) the following summary of information is provided:
| Date: | September 17, 2018 |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Medical Systems Information Technologies, Inc.<br>9900 Innovation Drive<br>Wauwatosa, WI 53226 |
| Primary Contact: | Marcella Entwisle<br>Senior Regulatory Affairs Leader<br>Phone: (414) 721-3809<br>Fax: (414) 721-3863<br>email: marcella.entwisle@ge.com |
| Secondary Contact: | Joel Kent<br>Senior Regulatory Affairs Manager<br>GE Medical Systems Information Technologies, Inc.<br>Phone: (617) 851-0943<br>email: joel.kent@ge.com |
| PRODUCT IDENTIFICATION | |
| Device Trade Name: | MAC VU360 Resting ECG Analysis System |
| Common / Usual Name: | Electrocardiograph |
| Classification Names | 21 CFR 870.2340 – Electrocardiograph<br>21 CFR 870.1425 – Programmable Diagnostic Computer<br>21 CFR 870.2920 – Telephone electrocardiograph Transmitter and<br>Receiver |
| Product Codes: | DPS, DQK, and DXH |
| Predicate Device(s): | ELITM 380 Resting Electrocardiograph (K142105) |
| Reference Device(s): | MAC™ 5500 HD Resting ECG Analysis System (K073625)<br>12SLTM (v23) ECG Analysis Program (K141963) |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo, which is a stylized monogram of the letters 'G' and 'E' in a circular shape. The logo is colored in blue, and the letters are interconnected in a flowing, cursive style. The circular border around the letters has small, stylized water droplet shapes evenly spaced around it.
Device Description: The MAC VU360 Resting ECG Analysis System is a mobile electrocardiograph designed to acquire, analyze, display, and record ECG signals from surface ECG electrodes. The MAC VU360 can capture 3, 6, 12, or 15 lead electrocardiograms, provide interpretive analysis, and print reports. The MAC VU360 is indicated for use on adult and pediatric (birth through 21 years of age) populations. The device can connect to a network, either through a wired LAN connection or via wireless WiFi access points. Once on the network, the device can optionally interface with the cardiology information systems such as the GEHC MUSE® system to participate in a complete electrocardiology workflow. The device can also optionally display data from external systems which provide web or Citrix content. The device provides state-of-the-art information technology security features and a contemporary user interface. Mobility is provided via a wheeled trolley. The MAC VU360 is intended to be a high-performance, Intended Use / multichannel resting electrocardiograph. As a resting electrocardiograph, the MAC VU360 simultaneously acquires data from each lead. Once the data is acquired, it can be analyzed, reviewed, stored, printed or transmitted. It is a device primarily intended for use in hospitals, but may be used in medical clinics and offices of any size. The MAC VU360 is intended to be used under the direct supervision of a licensed healthcare practitioner, by trained operators in a hospital or medical professional's facility. Indications for Use / The MAC VU360 Resting ECG Analysis System is a non-invasive prescription device. -The device is indicated for use to acquire, analyze, display and print electrocardiograms. -The device is indicated for use to provide interpretation of the data for consideration by a physician. -The device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician. It is not intended as a sole means of diagnosis. The interpretations of ECG offered by the device are only significant when used in conjunction with a physician overread as well as consideration of all other relevant patient data. -The device is indicated for use on adult and pediatric (birth through 21 years of age) populations.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a blue circle. There are three white water droplet shapes surrounding the circle.
| Contraindications: | The MAC VU360 Resting ECG Analysis System is not intended for<br>use in the following manner:<br>-During patient transport<br>-With high-frequency surgical units<br>-As an intra-cardiac application<br>-As a sole means of diagnosis<br>-As a vital signs physiological monitor |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technology: | The MAC VU360 employs the same fundamental scientific<br>technology, basic design, construction, materials, energy source,<br>control mechanism, operating principles as the predicate device ELI<br>380 and the MAC 550 HD in acquiring, analyzing, recording,<br>displaying and printing ECG data for both adult and pediatric<br>populations. The MAC VU360 utilizes the 12SL Algorithm as does<br>the MAC 5500 HD.<br><br>The basic system prints 6 or 12 leads of ECG and provides optional<br>transmission and reception of ECG data to and from a central ECG<br>cardiovascular information system.<br><br>The MAC VU360 is similar to the ELI 380 Resting ECG Analysis<br>System, K142105, in the technology of downloading orders and<br>patient demographics from a central ECG cardiovascular information<br>system (e.g. MUSE) as well as supporting ECG reports in PDF. Both<br>are able to use WiFi communication.<br><br>For more on the predicate device comparison refer to Section 12 of<br>this submission. |
| Performance Standards: | The MAC VU360 Resting ECG Analysis System complies with the<br>voluntary standards as detailed in Section 9 of this submission. |
| Sterilization: | The MAC VU360 Resting ECG Analysis System does not require<br>sterilization. |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circle. The circle has small, wave-like elements around its perimeter, giving it a dynamic appearance. The logo is presented in a blue color scheme.
| Determination of<br>Substantial Equivalence: | Summary of Non-Clinical Tests: |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The MAC VU360 Resting ECG Analysis System and its applications<br>comply with voluntary standards. The following quality assurance<br>measures were applied to the development of the system: |
| | ● Risk Analysis<br>● Required Reviews<br>● Design Reviews<br>● Software Development Lifecycle<br>● Testing on unit level (Module verification)<br>● Integration testing (System verification)<br>● Performance testing (Verification)<br>● Safety testing (Verification) |
| Summary of Clinical<br>Tests: | The subject of this premarket submission, MAC VU360 Resting<br>ECG Analysis System, did not require clinical studies to support<br>substantial equivalence. |
| Conclusion: | GE Healthcare considers the MAC VU360 Resting ECG Analysis<br>System to be as safe, and as effective, and performance is<br>substantially equivalent to the predicate device, ELI 380 Resting<br>Electrocardiograph. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.