DATRIX CARDIO WIFI ELECTROCARDIOGRAPH

K053083 · Datrix · DPS · Nov 30, 2005 · Cardiovascular

Device Facts

Record IDK053083
Device NameDATRIX CARDIO WIFI ELECTROCARDIOGRAPH
ApplicantDatrix
Product CodeDPS · Cardiovascular
Decision DateNov 30, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2340
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The Cardio WiFi electrocardiograph is a small portable digital electrocardiograph intended for use by medical professionals to acquire 12-lead ECG's from single or multiple patients in a clinical or point of care setting. ECG data is first recorded to a secure digital flashcard and then transferred to an ECG Management System for review by a physician or other qualified professional.

Device Story

Portable, battery-powered 12-lead electrocardiograph; records ECG data to flashcard; displays ECG on integrated LCD for periodic checks by medical professionals. Data transferred via WLAN (802.11 b/g) to external ECG Management System for physician review. Facilitates clinical decision-making by providing high-resolution ECG data for symptom evaluation. Used in clinical or point-of-care settings.

Clinical Evidence

Bench testing only. Compliance with AAMI EC11, IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-25 standards demonstrated performance, functionality, and reliability.

Technological Characteristics

Portable, battery-powered digital electrocardiograph. Features LCD display and WLAN (802.11 b/g) connectivity for data transfer. Records to secure digital flashcard. Tested to AAMI EC11, IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-25 standards.

Indications for Use

Indicated for use by medical professionals to acquire 12-lead ECGs from single or multiple patients in clinical or point-of-care settings.

Regulatory Classification

Identification

An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K053083; page 1 of 2 ### Datrix, Inc NOV 3 0 2005 ### Appendix G ### 510(k) Summary of Safety and Effectiveness for Datrix Cardio WiFi Electrocardiograph | 1. DATE SUMMARY PREPARED: | SEPTEMBER 26, 2005 | |----------------------------------|-------------------------------------------------------| | 2. SUBMITTER'S NAME AND ADDRESS: | Datrix<br>340 State Place<br>Escondido, CA 92016-1120 | | | Phone: (760) 480-8874 | | | Fax: (760) 480-9474 | | 3. CONTACT PERSON: | Linda Gluckman, QA Manager | | 4. DEVICE NAME: | | | Proprietary (trade) Name: | Cardio WiFi Electrocardiograph | | Common Name: | Portable Electrocardiograph without analysis | | Classification Name: | Electrocardiograph (CFR 870:2340) | | Product Code: | DPS | | Class: | 2 | #### PREDICATE DEVICE: 5. The legally marketed device/s to which equivalence is being claimed is the CardioCollect Portable Electrocardiograph manufactured by Reynolds Medical, Ltd. (K013367). #### 6. DEVICE DESCRIPTION The Cardio WiFi electrocardiograph continuously records data to a flashcard and displays it at high resolution. It features a large, liquid crystal display ("LCD") that allows a medical professional to periodically check the test subject's full, 12-lead ECG, either at intervals or in the event of a patient symptom. The device is portable and is battery-powered. Data is stored for later review at an ECG review station. Data transfer is accomplished via WLAN 802.11 b or g. {1}------------------------------------------------ ## K053083; page 2 of 2 # Datrix, Inc ### Appendix G #### 7. INTENDED USE . . The Cardio WiFi electrocardiograph is a small portable digital electrocardiograph intended for use by medical professionals to acquire 12-lead ECG's from single or multiple patients in a clinical or point of care setting. ECG data is first recorded to a secure digital flashcard and then transferred to an ECG Management System for review by a physician or other qualified professional. ### 8. NON-CLINICAL TESTS USED IN DETERMINATION OF SUBSTANTIAL EQUIVALENCE The substantial equivalence of the Datrix electrocardiograph (Cardio WiFi) is demonstrated by the following non-clinical testing: - Testing to applicable standards: AAMI EC11, IEC 60601-1; IEC 60601-1-2, . IEC 60601-2-25 - Testing for the performance, functionality, and reliability characteristics of the . device followed established test procedures in a quality system. #### 9. CONCLUSIONS FROM NON-CLINICAL TESTING Prior to marketing in the US, the Datrix Cardio WiFi electrocardiograph will have completed the testing listed above with acceptable results, demonstrating substantial equivalence. #### 10. SUBSTANTIAL EQUIVALENCE CONCLUSION In summary: Comparison to the predicate device listed in item #5 shows nearly identical technical data, same indications for use, same safety standards tested to, and raises no new questions of safety or efficacy. Therefore, the Datrix Cardio WiFi electrocardiograph supports a claim of substantial equivalence. {2}------------------------------------------------ Image /page/2/Picture/2 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo consists of a stylized eagle with three lines representing its wings and head. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 NOV 3 0 2005 Datrix, Inc. c/o Mr. Mark Job Regulatory Technology Services LLC 1394 25th Street NW Buffalo, MN 55313 Re: K053083 Trade Name: Datrix Cardio WiFi Electrocardiograph Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II (two) Product Code: DPS Dated: November 14, 2005 Received: November 16, 2005 Dear Mr. Job: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of general volurely would manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Mark Job Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, B. Hummerhofer Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### Datrix, Inc ### Indications for Use Statement 510(k) Number (if known): Prescription Use Device Name: Datrix Cardio WiFi Electrocardiograph Datrix Cardio WiFi Model: Indications for Use: The Cardio WiFi Electrocardiograph is a small portable digital cardiograph intended for use by medical professionals to acquire 12-lead ECG's from single or multiple patients in a clinical or point of care setting. ECG data is first recorded to a secure digital flashcard and then transferred to an ECG management system for review by a physician or other qualified professional. (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) AND/OR Over-The-Counter Use Виттиsence of CDRH, Office of Device Evaluation (ODF) vision Sign-Oft) vision of Cardiovascular Devices 10(k) Number K053083 X . 【
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...